A review on low-dimensional physics-based models of systemic arteries: application to estimation of central aortic pressure

Abstract

The physiological processes and mechanisms of an arterial system are complex and subtle. Physics-based models have been proven to be a very useful tool to simulate actual physiological behavior of the arteries. The current physics-based models include high-dimensional models (2D and 3D models) and low-dimensional models (0D, 1D and tube-load models). High-dimensional models can describe the local hemodynamic information of arteries in detail. With regard to an exact model of the whole arterial system, a high-dimensional model is computationally impracticable since the complex geometry, viscosity or elastic properties and complex vectorial output need to be provided. For low-dimensional models, the structure, centerline and viscosity or elastic properties only need to be provided. Therefore, low-dimensional modeling with lower computational costs might be a more applicable approach to represent hemodynamic properties of the entire arterial system and these three types of low-dimensional models have been extensively used in the study of cardiovascular dynamics. In recent decades, application of physics-based models to estimate central aortic pressure has attracted increasing interest. However, to our best knowledge, there has been few review paper about reconstruction of central aortic pressure using these physics-based models. In this paper, three types of low-dimensional physical models (0D, 1D and tube-load models) of systemic arteries are reviewed, the application of three types of models on estimation of central aortic pressure is taken as an example to discuss their advantages and disadvantages, and the proper choice of models for specific researches and applications are advised.

Introduction

Cardiovascular diseases have become a dominant factor of mortality all over the world [1]. Nearly 17.5 million people die of cardiovascular disease [2] and billions of dollars are spent every year on related healthcare [3]. Nowadays, cardiovascular research has become an important topic and been paid significant attention by researchers. Cardiovascular system is a complex circulatory system consisting of the heart, arteries and veins [4]. In recent years, due to the significant improvements in computer technology, modeling based on physical principles has become a powerful tool to simulate the hemodynamic properties of cardiovascular system and has been playing an increasingly important role in the diagnosis of cardiovascular diseases and the development of medical devices [5,6,7].

Current physics-based models can be divided into two categories, high-dimensional models and low-dimensional models as shown in Fig. 1. High-dimensional models including 2D models and 3D models can give detailed descriptions of the local flow field of the blood. These models describe the complex hemodynamic phenomenon of a specific region in the cardiovascular system. 2D models are generally used to describe changes of the radial blood flow velocity in an axisymmetric tube [8, 9]. 3D models are usually applied to simulate the fluid-structure interaction between the vascular walls and blood [10, 11]. To establish a 3D model of the entire arterial tree, the complex geometrical and mechanical information needs to be provided, which results in the enormous computational complexity, so that it cannot be readily implemented in practice. Consequently, high dimensional models can generally be used to simulate local hemodynamics of specific arterial sites, instead of the whole arterial tree.

Fig. 1
figure 1

The structure diagram of physics-based models in the cardiovascular system

In contrast to high-dimensional models, low-dimensional models with small computational costs can readily reproduce the pulse wave propagation phenomenon and realize patient-specific modeling. Thus, the low-dimensional modeling can be an effective way to describe the hemodynamic properties of the entire arterial tree in practical applications. At present, the available low-dimensional models mainly consist of 0D models, 1D models and tube-load models. 0D models, also called lumped parameter models, can describe global properties of the arterial system. The lumped parameter models are characterized by their pulse waveforms as a function of time only. The most well-known lumped parameter model is the Windkessel model [12], which includes mono-compartment models and multi-compartment models [13, 14]. 1D models and tube-load models are distributed parameter models, which can represent distributed properties of the arterial system. In the latter two types of models, their pulse waveforms depend on both time and space. In the distributed parameter models, 1D model based on the simplified Navier–Stokes equation is commonly used to reproduce pressure and flow at any position in the entire arterial tree [15,16,17,18,19]. The Windkessel model is computationally simple but less accurate. On the other hand, 1D model can represent the wave propagation phenomenon accurately but need a relatively large amount of computation. Taking advantage of both Windkessel models (simplicity) and 1D models (accuracy), some researchers developed tube-load models [20, 21]. Tube-load models can monitor multiple arterial hemodynamic parameters such as pulse transit time, arterial compliance, pulse wave velocity, and so on.

So far, these three types of low-dimensional models have been extensively used in the study of cardiovascular dynamics as shown in Table 1. Especially, applying physics-based models to estimate central aortic pressure has been paid much attention in recent decades [22,23,24,25,26,27]. However, to our best knowledge, there has been few review paper about the reconstruction of central aortic pressure using these physics-based models. Additionally, estimating central aortic pressure is a common application of these three types of models, which contributes to compare their advantages and disadvantages fully. Thus, this paper is to review three types of low-dimensional physics-based models (0D models, 1D models and tube-load models) of the arterial system and take the application of estimating central aortic pressure as an example to compare their advantages and disadvantages. To begin with, the theories and applications of Windkessel models including mono-compartment and multi-compartment models are described. Then, the theories and applications of two types of distributed parameter models, namely 1D models and tube-load models, are elaborated. Next, the advantages and disadvantages of these three models are discussed. Finally, future challenges and final conclusion are presented.

Table 1 Main applications of Windkessel models, 1D models and tube-load models

0D models

In 0D models (lumped parameter models), the Windkessel theory is applied to the modeling of the arterial system [12, 28, 29]. Windkessel models are divided into two classes: mono-compartment models and multi-compartment models. Theories and applications of two categories of models are elaborated in this part, respectively. Furthermore, the comparison of different Windkessel models is made in Table 2.

Table 2 Comparison of different Windkessel models

Model descriptions

Mono-compartment models

The mono-compartment model is a combination of inductance, compliance and resistance. According to the number of elements included, current mono-compartment models are classified into four main types: two-element, three-element, four-element and complex mono-compartment Windkessel models.

a. Two-element Windkessel model

The two-element Windkessel model is the simplest mono-compartment model presented by Frank [12], which is made up of a resistor (R) and a capacitor (C) as shown in Fig. 2a. In this model, the resistor describes the resistance of small peripheral vessels and the capacitor describes the distensibility of large arteries. The two-element Windkessel model simply describes the pressure decay of the aorta in diastole. This model cannot signify the high frequency effects because there is merely a time constant in the model. Owing to its simplicity, this model can be used in clinical practice readily such as total arterial compliance estimation [30] and blood pressure estimation [31].

Fig. 2
figure 2

The mono-compartment models. a Two-element Windkessel model; b three-element Windkessel model; c four-element Windkessel model

b. Three-element Windkessel model

Adding a characteristic impedance (\(Z_c\)) to the two-element Windkessel model, the three-element Windkessel model is formed as shown in Fig. 2b [15]. The characteristic impedance is equal to oscillatory pressure divided by oscillatory flow. Although it is found that a resistance numerically equals approximately a characteristic impedance, the characteristic impedance is different from the resistance. The characteristic impedance is merely used to signify oscillatory phenomena. Owing to the inclusion of the characteristic impedance, this model can simulate high frequency effects. Simultaneously, the introduction of the characteristic impedance also results in some errors at the low frequency. In contrast with the two-element Windkessel model, the three-element Windkessel model can have a higher accuracy. Therefore, the three-element Windkessel model has been extensively used in theoretical research [32,33,34].

c. Four-element Windkessel model

Taking the inertance of blood flow into consideration on the basis of the three-element Windkessel model, Stergiopulos et al. [35] proposed a four-element Windkessel model as shown in Fig. 2c. Due to the addition of the inertance, this model can represent middle frequency effects. In other words, the four-element Windkessel model can simulate all frequency effects. It has been validated that the four-element model can give a better description of the impedance characteristics [36]. Some nonlinear regression analysis methods are applied to the estimation of the four-element model parameters. Compared with the two-element and three-element Windkessel models, it is more difficult to identify the model parameters of the four-element Windkessel model. Consequently, only a few researchers make use of this model [23, 37, 38].

d. Complex mono-compartment models

For the sake of the further improvement of the arterial model, a few researchers developed more complex Windkessel models in which more resistive, inductive and capacitive components were introduced [39, 40]. By including more resistive and inductive elements, the laminar oscillatory flow impedance can be simulated. Owing to the high complexity of these models, there has been no further development by other investigators.

These models described above focus on simulating the pressure and flow characteristics of the arterial vessels without considering the effect of the venous vessels. In fact, with regard to the coronary and pulmonary circulation, the pressure and flow of the veins have a significant impact on the global hemodynamics [41]. Under this circumstance, the venous side cannot be ignored. In order to describe the characteristics of the veins, extra resistance, inertance and compliance are added to form more complex (five, six and seven element) arterial models. In contrast with the two, three and four element Windkessel model, the five-element model simulates the characteristics of microcirculation hemodynamics more accurately, and the six-element model accounts for the hemodynamic contribution of the venous vessels in the cardiovascular system more precisely, and the seven-element model further gives the representation of the systemic circulation through the improvement on the description of the venous system.

Multi-compartment models

Regardless of spatial information, the mono-compartment model regards all arteries as a single block. In order to represent the distribution of flow and pressure, some multi-compartment models which were composed of a series of mono-compartment models were established. Figure 3 is an example of a simple multi-compartment model of the systemic arteries [42]. Every mono-compartment model is a combination of resistance (R), inertance (L) and compliance (C). At present, there are four typical compartmental configurations in the multi-compartment model: T, \(\Pi\) and inverted L element, respectively [13]. The corresponding compartmental configuration should be chosen appropriately according to the characteristics of the particular arteries. Since the multi-compartment model represents position information roughly but in general does not signify the nonlinear convective acceleration term of 1D model, this model is usually seen as the first order discretization of the one-dimensional linear model [43].

Fig. 3
figure 3

A multi-compartment model of systemic circulation. ao aortic root, at artery, ar arteriole, cp capillary, vn vein

Applications

Mono-compartment models are simplified descriptions of an arterial system, simulating physiological properties of the arterial vessels with a few parameters. Multi-compartment models construct a full arterial network by connecting several mono-compartment models, describing particular information of different vessel compartments. Due to the simplicity of mono-compartment and multi-compartment models, merely a few researchers used them to reconstruct central aortic pressure.

In Windkessel models, proximal flow or peripheral pressure measurements were frequently used as inflow condition. Many researchers chose aortic flow as model input [22, 44, 45]. Flow at other positions was selected as model input, such as carotid flow [46], left ventricle flow [23] and mitral valve mean flow [47]. Few researchers chose brachial and radial pressure as model input [48]. For model parameter acquisition, the majority of researchers calculated model parameters by population averaging [22, 23, 44,45,46, 48] and only a small number of researchers adopted partially individualized model parameters [47].

Mono-compartment models are more commonly applied to the estimation of central aortic pressure than multi-compartment models. For example, the pressure waveform in the aorta was reconstructed by Stergiopulos et al. [44] and Struijk et al. [45] using a two-element Windkessel model. Cai et al. [22], Zala et al. [46] and Vennin et al. [22] employed a three-element Windkessel model to calculate central aortic pressure. A four-element Windkessel model was adopted by Her et al. [23] to reproduce the patient’s aortic pressure waveform undergoing counter-pulsation control by ventricular assist devices. Revie et al. [47] used a six-chamber lumped parameter model consisting of left ventricle and right ventricle, aorta, pulmonary artery, pulmonary vena cava and pulmonary vein to monitor aortic pressure changes. It was verified in clinical data [22, 23, 47] that Windkessel models can describe the general shape of pressure waveform in the ascending aorta, however, it is difficult to show the details of pressure waveform such as dicrotic notch features. Therefore, Windkessel models have limited accuracy for the estimation of central aortic pressure.

1D models

1D models are distributed parameter models. Theory and application of 1D models are described in this part, respectively. 1D models mainly focus on methods that solve one-dimensional equations and boundary conditions (including inflow, bifurcation and outflow conditions).

Model descriptions

Model derivations

The one-dimensional arterial flow theory was proposed by some researchers [16, 17]. Euler [49] first established a 1D model using the one-dimensional theory. Although assumptions of the model are simple, it laid foundations for further studies. Reymond et al. [50] extended the existing 1D model to a more detailed model consisting of the foot and hand circulation. Moreover, the ventricular-arterial coupling model was developed and 1D model of the circulation was validated. 1D models are commonly used to represent pulse wave propagation phenomena of large arteries [29]. In the 1D model, the blood is assumed to be an incompressible Newton fluid and the vessel is an axisymmetric cylindrical tube. The 1D model is governed by two equations [51]. A continuity equation (see Eq. 1) and a momentum equation (see Eq. 2) both together describe the motion of the blood flow and the vessel wall. The formulas are as follows

$$\begin{aligned} \frac{\partial q}{\partial x}+\frac{\partial A}{\partial t}= & {} 0. \end{aligned}$$
(1)
$$\begin{aligned} \frac{\partial q}{\partial t}+\frac{4}{3}\frac{\partial \frac{q^2}{A}}{\partial x}= & {} -\frac{A}{\rho }\frac{\partial p}{\partial x}-\frac{8\mu }{\rho r^2}q. \end{aligned}$$
(2)

where x is the distance along the vessel, t is time, q is the blood flow rate, p is the blood pressure, A is the cross-sectional area, r is the vessel radius, \(\rho\) is the blood density and \(\mu\) is the viscosity.

Solving methods

For solving 1D Navier–Stokes equations, there are two types of methods including time domain and frequency domain methods. The time domain method can solve linear or nonlinear equations and frequency domain method can only solve linear equations.

a. Time domain method

Generally, Navier–Stokes equations of 1D models are nonlinear, which are solved in the time domain using numerical methods. At present, there are many numerical methods for solving the partial differential equations. Each method has its scope of application. The method of characteristics, finite difference method, finite volume method, finite element method and spectral method are frequently used to solve 1D pulse propagation equations.

The method of characteristics is a basic method of solving the partial differential equations. The essence of this method is the integral along the characteristic line of the partial differential equations to simplify the form of equations. The characteristic method has a clear physical meaning and a wide application scope. As for solving the differential equations of three independent variables, the method of characteristic can be very complicated, and there are still some problems to be solved. The governing equations can be solved by taking use of the method of characteristics [52,53,54].

The finite difference method is a numerical method for solving complex partial differential equations by approximating the derivatives with finite differences. While the principle of the finite difference method is simple, it can give the corresponding difference equation for any complex partial differential equation. Difference equations can only be considered as mathematical approximations of differential equations. This method has been used by a number of researchers. For instance, Olufsen et al. applied the two-step Lax–Wendroff method to solving the continuity and momentum equations [18, 19].

A finite volume method is developed on the basis of the finite difference method. To begin with, the calculated region is divided into a series of control volumes and there is a control volume at the surrounding of each grid point. Then each control volume is integrated and a set of discrete equations are obtained. Finally, the discrete equations need to be solved. The finite volume method is suitable for the computation of fluid. This method has a high computing speed and low requirements for the grid, but its precision is limited. To solve the differential equations, the finite volume method is often used [55, 56].

The finite element method uses the variational principle to minimize the error function. The advantage of the finite element method is that this method can simulate complex curve or surface boundary accurately. Furthermore, the division of the grid is arbitrary and it can design the general program easily. Nevertheless, the finite element method cannot give a reasonable physical explanation and some errors in the calculation are still difficult to improve. Recently, some investigators used the finite element method to solve the differential equations [57,58,59].

The spectral method is a class of computing techniques of using an orthogonal function or intrinsic function as an approximate function to solve certain differential equations. The superiorities of the spectral method are to obtain a higher precision using fewer grid points. The poor stability and high complexity in the treatment of boundary conditions are the major weaknesses of this method. The spectral method has been utilized to solve the one-dimensional pulse wave propagation equations by a few researchers [60, 61].

b. Frequency domain method

In order to reduce the computational complexity of the nonlinear model, a transmission line method is used to solve the Navier–Stokes equations in the frequency domain. The method requires that the 1D Navier–Stokes equations are linearized. According to the similarity of electromagnetic propagation theory and pulse wave propagation theory, linear 1D Navier–Stokes equations (see Eqs. 3, 4) in hemodynamics are converted into electrical transmission line equations (see Eqs. 5, 6) in the circuit [62]. The subsequent work is that we can employ methods of solving electric circuit to solve 1D Navier–Stokes equations. A transmission line equivalent circuit of an arterial segment is represented as shown in Fig. 4.

$$\begin{aligned} -\frac{\partial q}{\partial x}= \, {} \frac{dA}{dp}\frac{\partial p}{\partial t}=0. \end{aligned}$$
(3)
$$\begin{aligned} -\frac{\partial p}{\partial x}= \, {} \frac{\rho }{A}\frac{\partial q}{\partial t}+\frac{8\mu }{\pi r^2}q. \end{aligned}$$
(4)
$$\begin{aligned} -\frac{\partial I}{\partial x}= \, {} UG+C\frac{\partial U}{\partial t}. \end{aligned}$$
(5)
$$\begin{aligned} -\frac{\partial U}{\partial x}= \, {} IR+L\frac{\partial I}{\partial t}. \end{aligned}$$
(6)

where U is the voltage, I is the current, \(R=\frac{8\mu }{\pi r^4}\) is the resistance, \(L=\frac{\rho }{A}=\frac{\rho }{\pi r^2}\) is the inductance, \(C=\frac{dA}{dp}=\frac{3 \pi r^2}{2Eh}\) is the capacitance, E is the Young’s modulus, and h is the arterial wall thickness. G is the conductance, describing blood flow leakage, which is usually neglected. The electrical circuit is comprised of resistive, inductive and capacitive elements. The values of these elements are calculated from mechanical and geometric parameters in the arterial tree.

Fig. 4
figure 4

Transmission line equivalent circuit. a Arterial segment of unit length; b transmission line segment. Where \(Z_{input}\) is the input impedance, \(Z_L\) is the terminal impedance, \(\gamma = \sqrt{(R+jwL)(G+jwC)}\) is the propagation constant and \(Z_c= \sqrt{(R+jwL)/(G+jwC)}\) is the characteristic impedance. When the transmission line is lossless (\(R=G=0\)), \(\gamma = w \sqrt{LC}\) and \(Z_c= \sqrt{L/C}\)

Boundary conditions

For 1D pulse wave propagation equations, boundary conditions, commonly including inflow, bifurcation and outflow boundary conditions, need to be determined.

a. Inflow conditions

The flow waveform measured in vivo at ascending aorta or aortic root serves as the inflow condition using a magnetic resonance imaging or ultrasound equipment. Alternatively, a flow function derived from a simple model of the heart can serve as the inflow condition. The function is periodic which is mainly determined by the cardiac period and the cardiac output parameters [18].

$$\begin{aligned} q(0,t)= \, {} CO\frac{t}{\tau ^2}e^{\frac{-t^2}{2\tau ^2}}, \quad 0\le t<T. \end{aligned}$$
(7)
$$\begin{aligned} q(0,t+jT)= \, {} q(0,t), \quad j=1,2,3,\ldots \end{aligned}$$
(8)

where CO denotes the cardiac output, \(\tau\) denotes the time for cardiac output to reach its maximum and T denotes the cardiac period.

The other inflow condition is that the left ventricle of the heart is coupled to 1D arterial tree model. Two main heart models are developed to represent the relationship between the ventricular pressure and volume. The time-varying elastance model is that the heart is seen as an elastance varying with time [63]. This left ventricular model indicates the instantaneous change of pressure and volume in left ventricle.

$$\begin{aligned} E(t)=\frac{P(t)}{V(t)-V_0}. \end{aligned}$$
(9)

where E is the elastance, P is the instantaneous pressure of left ventricle, V is the instantaneous volume of left ventricle and \(V_0\) is the volume intercept of the end-systolic line.

The one-fiber model is another heart model in which the heart is described as a rotationally-symmetric cylindrical or spherical cavity [64]. The left-ventricular pump function is reflected by wall tissue function. It is assumed that fiber stress and strain are homogeneously distributed in the thick wall. The ratio of the fiber stress (\(\tau _f\)) to the pressure of the left ventricle (\(P_{lv}\)) is closely related to the ratio of the wall volume (\(V_w\)) and the cavity volume (\(V_{lv}\)).

$$\begin{aligned} \frac{P_{lv}}{\tau _f}=\frac{1}{3}ln\left( 1+\frac{V_w}{V_{lv}}\right) . \end{aligned}$$
(10)
b. Bifurcation conditions

In arterial networks, vessel branching is another important sort of boundary condition. At the bifurcation, the principle of pressure and flow continuity was applied [32, 55, 59, 65]. It is assumed that all bifurcations are situated at a point and the effect of the bifurcation angles is ignored. Without any blood leakage, according to the conservation of mass, the outlet flow of a parent vessel is equal to the sum of the inlet flow of two daughter vessels at the bifurcation (see Eq. 11). Considering the continuity of pressure, the pressure of a parent vessel and the pressure of each daughter vessel are identical at the bifurcation (see Eq. 12). In reality, there exist energy losses at the bifurcations. Generally, the energy loss at a bifurcation is quite small and they are often neglected. For the bifurcation at the aortic arch, however, large energy losses are brought about. Because it has a nearly right-angled turn and a high velocity of blood at this bifurcation, remarkable vortices are produced. In order to represent the loss at the aortic arch bifurcation, a loss coefficient K is introduced [19] (see Eq. 13).

$$\begin{aligned} q_{pa}= \, {} q_{d1}+q_{d2}. \end{aligned}$$
(11)
$$\begin{aligned} p_{pa}= \, {} p_{d1}=p_{d2}. \end{aligned}$$
(12)
$$\begin{aligned} p_{di}= \, {} p_{pa}+\frac{\rho }{2}((\bar{u_x})^2_{pa}- (\bar{u_x})^2_{di})-\frac{K_{di}}{2}(\bar{u_x})^2_{pa}, \quad i=1,2. \end{aligned}$$
(13)

where \(\bar{u_x}\) denotes the average axial velocity and \(\rho\) denotes the density. The subscripts pa and d indicate the parent and the daughter vessel, respectively.

c. Outflow conditions

The simplest outflow boundary condition is that the terminal of vessel is seen as a pure resistive load [18]. Nevertheless, a precise peripheral resistance value is not easy to be determined. Assuming a constant relation between pressure and flow, pressure and flow are in phase, actually, which is not physiologically reasonable for large arteries. The pure resistance model is merely suitable for small arteries. To overcome the weaknesses, a phase-shift between pressure and flow should be applied to the downstream boundary. The terminal impedance for the pure resistance is as follows.

$$\begin{aligned} Z_L(w)=R_T \end{aligned}$$
(14)

where \(Z_L(w)\) denotes the terminal impedance of large arteries and \(R_T\) denotes the peripheral resistance.

The distal network of truncated vessel is represented as the terminal impedance which is modeled by a three-element Windkessel model [66]. The three-element Windkessel model is made up of a resistance \(R_1\) in series with a parallel combination of a capacitor \(C_T\) and another resistance \(R_2\). This model cannot represent the wave propagation effects. The frequency dependent impedance of Windkessel model is given by

$$\begin{aligned} Z_L(w)=\frac{R_1+R_2+iwC_TR_1R_2}{1+iwC_TR_2}. \end{aligned}$$
(15)

The relation between pressure and flow at the truncated arteries is given by the following differential equation.

$$\begin{aligned} \frac{\partial p}{\partial t}=R_1\frac{\partial q}{\partial t}-\frac{p}{R_2C_T}+\frac{q(R_1+R_2)}{R_2C_T}. \end{aligned}$$
(16)

In recent years, the structured-tree model presented by Olufsen [67] has become a popular outflow boundary condition. Compared with the resistance and Windkessel model, the structured-tree model can simulate the impedances of small arteries more accurately. At the terminal branches of the truncated arterial tree, a structured-tree model, which is based on linear one-dimensional Navier–Stokes equations, provides a dynamic boundary condition for large arteries. The model can describe the phase lag between pressure and flow and the high frequency oscillations. Meanwhile, it can also represent the wave propagation effects of arterial system. According to the convolution theorem, the outflo w boundary condition is obtained by

$$\begin{aligned} p(x,t)=\frac{1}{T}\int _{-T/2}^{T/2}z(x,t-\tau )q(x,\tau )d\tau . \end{aligned}$$
(17)

The root impedance is computed from the relationship between pressure and flow as follow

$$\begin{aligned} Z_L(w)=\frac{ig^{-1}sin(wL/c)+Z(L,w)cos(wL/c)}{cos(wL/c)+igZ(L,w)sin(wL/c)}. \end{aligned}$$
(18)

where \(g=cC\), c denotes the wave propagation velocity, \(Z_L(w)\) denotes the root impedance, namely, the terminal impedance of large arteries, Z(Lw) denotes the terminal impedance of small arteries, L denotes the vessel length and w denotes the angular frequency.

Applications

1D models can simulate pressure and flow waveforms at any point of the arterial network according to their distributed properties. Since 1D models include too many vascular parameters, they haven’t been extensively used to reconstruct central aortic pressure up to now. By limiting the number of personalized parameters, 1D models may have a great potential for estimation of central aortic pressure in clinical practice.

For 1D models, aortic flow waveform is the most common inflow condition [24, 68]. Few researchers used peripheral pressure measurement (e.g. brachial pressure or radial pressure) as model input [25, 69]. The aortic flow waveform can be measured by ultrasound equipment, however, it is not accurate. Meanwhile, it is very expensive to obtain aortic flow waveforms by MRI equipment. The pressure waveforms of good stability can readily be recorded using peripheral pressure sensors such as applanation tonometry. Generally, vascular geometric parameters of 1D models are measured by MRI or CT equipment, which is costly and complex. Many researchers used population averages for these geometric parameters. Nevertheless, Harana et al. [24] measured aortic geometry parameters including ascending aorta, descending aorta and three supra-aortic branches using MRI equipment. Meanwhile, pulse wave velocity (PWV) and vascular resistance and compliance parameters for each subject were calculated from measured data. Remaining blood flow parameters such as density and viscosity were assumed to be constants.

In recent years, 1D models with different degrees of complexity have been utilized by several researchers to reconstruct central aortic pressure. For example, Bárdossy et al. [69] presented a “backward” calculation method to derive central aortic pressure waveform in a 1D model comprising 50-segment arteries. A personalized transfer function between aorta and radial was established by Jiang et al. [24] to estimate central aortic pressure based on 1D model including 55 large arteries and 28 small arteries. Khalifé et al. [68] estimated absolute pressure in the aorta by combining a reduced 1D model including an ascending aorta branch and a descending aorta branch with MRI. A non-invasive personalized estimation method of central aortic pressure was developed by Harana et al. [24] using a 1D aortic blood flow model. Owing to the complexity of 1D models, the details of pressure waveform can be described easily [24]. If all vascular geometric parameters are measured by noninvasive equipments, 1D models can provide accurate estimation of central aortic pressure.

Tube-load models

Tube-load models are distributed parameter models. Theory and application of tube-load models are described in this part, respectively. Tube-load models mainly focus on various tube models based on different assumptions.

Model descriptions

Tube-load models are a kind of highly simplified transmission line models, which are made up of multiple parallel tubes with loads [70]. The simplest tube-load model whose tube is taken as lossless, linear and uniform, only consists of a tube and a load as shown in Fig. 5. The tube signifies the wave transmission pathway of large arteries and the load signifies the wave refection site in the arterial terminal. The formulas are as follows.

$$\begin{aligned} T_d = \, {} \sqrt{LC} \end{aligned}$$
(19)
$$\begin{aligned} Z_c = \, {} \sqrt{L/C} \end{aligned}$$
(20)
$$\begin{aligned} \Gamma (jw) = \, {} \frac{Z_L(jw)-Z_c}{Z_L(jw)+Z_c} \end{aligned}$$
(21)
$$\begin{aligned} P(x,jw) = \, {} P_f(0,jw)e^{jwT_d x/d}+P_b(0,jw)e^{-jwT_d x/d} \end{aligned}$$
(22)
$$\begin{aligned} Q(x,jw) = \, {} \frac{1}{Z_c}(P_f(0,jw)e^{jwT_d x/d}+P_b(0,jw)e^{-jwT_d x/d}) \end{aligned}$$
(23)
$$\begin{aligned} P_p(jw) = \, {} \frac{\left( jw+\frac{1}{RC}+\frac{1}{2Z_c C}\right) e^{jwT_d} +\frac{1}{2Z_c C}e^{-jwT_d}}{jw+\frac{1}{RC}+\frac{1}{Z_c C}}P_c(jw) \end{aligned}$$
(24)

where L denotes the large artery inertance, C denotes the large artery compliance, d denotes length of tube, \(T_d\) denotes the time delay, \(Z_c\) denotes the characteristic impedance, \(Z_L\) denotes the terminal impedance, \(\Gamma\) denotes the wave reflection coefficient, P denotes the pressure, Q denotes the flow. In addition, the subscripts f, b, c and p are for the forward and backward wave, aorta and peripheral artery, respectively.

Fig. 5
figure 5

Single tube model with a load. R the peripheral resistance, C the load compliance, \(Z_c\) the characteristic impedance and \(T_d\) the time delay

For tube-load models, there are three main types of loads, including pure resistance, generic pole-zero, three-element Windkessel models. The pure resistance load is a high simplification of small arterial vessels, accounting for the peripheral resistance [26, 71]. The greatest advantage of this kind of load is simplicity and its disadvantage is that there is a big difference from the real vascular structure. The generic pole-zero models as a terminal impedance can change the order of system flexibly [72]. However, the weakness of the model is that model parameters do not have physiological significance. The most frequently used load is the three-element Windkessel model, which consists of a characteristic impedance, a resistance and a compliance [73, 74]. Although the Windkessel model fails to provide the detailed anatomical and mechanical information of arterial network, it can describe the lumped properties of terminal arterial vessels well.

Based on different assumptions, tube-load models have been developed into T-tube, lossy tube-load, nonlinear tube-load and non-uniform tube-load models, which are summarized in Table 3. In comparison with the simplest tube-load model, these models have a higher accuracy in hemodynamic simulation.

Table 3 Summary of tube-load models based on different assumptions

T-tube model

The T-tube model is a frequently-used tube-load model, in which there are two tubes with two terminal loads as shown in Fig. 6 [75, 76]. While the tubes signify head-end and body-end travel paths, the loads represent head-end and body-end reflection sites, respectively. The advantages of the T-tube model are that it is a simple model and that it can indicate main features of pressure and flow waveforms in the large blood vessels. However, it cannot represent the wave reflection intensely depending on frequency.

Fig. 6
figure 6

The T-tube model. R the peripheral resistance, C the load compliance, \(Z_c\) the characteristic impedance; subscripts b and h are for the body load and head load, respectively

Lossy tube-load model

Tube-load models mentioned above are based on lossless tube-load models. The lossless tube-load model is a kind of ideal and simple model. In some situations, ignoring the loss of tube can bring large errors [77,78,79,80]. For example, for reconstructing the central aortic pressure waveform from peripheral pressure waveforms, it is generally assumed that the mean pressure is same at any position. In fact, the blood pressure loss is large along the arterial tree in pathophysiologic conditions or postoperative period. In order to improve the accuracy of tube-load models, Abdollahzade et al. [20] proposed the lossy tube-load model of arterial tree in humans as shown in Fig. 7. Compared with lossless tube-load models, lossy tube-load models have smaller errors and larger efficacy.

Fig. 7
figure 7

The lossy tube-load model. P the blood pressure, \(\gamma\) the wave propagation constant, \(l_0\) the length of the tube, R the peripheral resistance, C the load compliance, \(Z_c\) the characteristic impedance; subscripts i and o are for the inlet and outlet of the tube, respectively

Nonlinear tube-load model

Previously, tube-load models were regarded as linear models. In recent years, Gao et al. [81, 82] has developed a nonlinear tube-load model based on the exponential relationship between blood pressure and compliance as shown in Fig. 8. In the nonlinear tube-load model, arterial compliance is no longer a constant but a function of blood pressure. In contrast with the linear tube-load model, the nonlinear tube-load model has a higher accuracy in estimating pulse transit time [21].

Fig. 8
figure 8

The nonlinear tube-load model. P the blood pressure, \(\alpha\) the constant, L the large artery inertance, R the peripheral resistance, \(C_0\) the large artery compliance, C the load compliance, \(Z_c\) the characteristic impedance; subscripts p and d are for the inlet and outlet of the tube, respectively

Non-uniform tube-load model

The electrical transmission line theory is applied to mathematical modeling of arterial vessels. It is generally assumed that the transmission line model is a uniform tube with a terminal load. Taylor [83] explored the wave propagation properties of a non-uniform transmission line, since the uniform tube is too simple to reflect the real characteristics of the arteries. Subsequently, Einav et al. [84] proposed an exponentially tapered transmission line model of the arterial system, in which the geometrical properties and wall elasticity of the tube exponentially tapered along the length of arterial vessels. At present, there are two methods to describe the taper effects of a non-uniform tube. One method is that the inductance and capacitance of the tube change with the position exponentially as shown in Fig. 9a [85,86,87]. Another method is that an artery is separated into several smaller segments and each segment is viewed as a uniform tube as shown in Fig. 9b [88, 89].

Fig. 9
figure 9

The non-uniform tube-load model. a A non-uniform tube tapering with the position exponentially; b a non-uniform tube consisting of multiple uniform tubes with a successive decrease in diameter dimension

Applications

Tube-load models include tubes and loads, describing wave propagation and refection phenomenon with only a few parameters. Combining the advantages of Windkessel models (simplicity) and 1D models (accuracy), tube-load models have become an attractive tool for the reconstruction of central aortic pressure waveform.

For tube-load models, inflow conditions are obtained from one or two measured peripheral pressure waveforms. A radial [71, 73, 74], brachial [26] or femoral [72] pressure measurement is commonly used as model input. Moreover, some researchers [27, 90,91,92] chose two measured peripheral pressure waveforms as inflow conditions such as radial and femoral arteries. Because intravascular and extravascular pressure at radial artery is very close, the radial pressure waveform can be accurately recorded using an applanation tonometry. The brachial cuff-based measurement is a very convenient approach for obtaining brachial pressure, especially suitable for longer term outpatient monitoring of central aortic pressure or potential general hospital ward usage.

Single tube or T-tube models with different loads are used to described the relation between central and peripheral pressures. For example, Swamy et al. [72] employed a single tube with a generic pole-zero load to establish an adaptive transfer function between aortic and femoral pressures. A single tube model with a resistance load was applied by Gao et al. [71] and Natarajan et al. [26] to the estimation of central aortic pressure. Individualized transfer functions were built by Sugimachi et al. [73] and Hahn et al. [74] using a single tube model with a three element Windkessel load. Ghasemi et al. [27, 90], Lee [91] and Kim et al. [92] utilized a T-tube model with a three-element Windkessel load, to reconstruct central aortic pressure from two measured peripheral pressure.

In order to acquire an adaptive or individualized transfer function, most researchers measured the pulse transit time for each subject through some noninvasive approaches and calculated the remaining parameters by population averaging [26, 27, 71,72,73,74, 92, 93]. Blind system identification is another method of estimating model parameters, which can reconstruct the central aortic pressure waveform from two distinct peripheral pressure waveforms [90, 91]. The blind system identification method can obtain fully individualized parameters. The weakness of this method is that it is inconvenient to measure multiple distinct peripheral pressure waveforms in clinical practice. To examine the performance of individualization, Hahn [93] made a comparative study on the estimation of central pressure among a fully individualized, two partially individualized and a fully generalized transfer functions based on tube-load models. The 9 swine experiment results showed that the fully individualized transfer function had higher accuracy than two partially individualized functions and the fully nonindividualized transfer function. Since in the tube-load model, only one parameter, the pulse transit time, could be readily individualized, tube-load models had moderate accuracy for estimation of central aortic pressure.

Discussions and conclusions

Comparisons of three types of models

In this paper, recent progresses of Windkessel, 1D and tube-load models in the arterial system are reviewed. Windkessel models are developed into increasingly complicated and detailed structures and a variety of Windkessel models are established [39, 94]. 1D models including more arterial segments and coupling the heart have been set up in recent years [29, 50]. Tube-load models with various types of tubes based on different assumptions are investigated [77, 81, 83].

To select an appropriate model, the comparisons among three types of low-dimensional models are made. The Windkessel model can give a global description of the arterial system and every model element has a particular physiological meaning [15, 28]. Windkessel models only include a few parameters and the parameters are usually estimated from measured aortic pressure and flow waveforms. Due to its simplicity, the Windkessel models have a low accuracy. The iterative and system identification techniques are adopted by most researchers, such as linear least-squares method [95] and subspace model identification method [96]. Since the Windkessel model is established by a set of ordinary differential equations, this model is simpler than the distributed parameter model built by a set of partial differential equations. The model is fit for calculating hemodynamic parameters and simulating the whole circulatory system.

1D models and tube-load models are distributed parameter models which signify the distributed properties along the arterial vessels. 1D models can accurately predict flow and pressure in the entire arterial tree and the model parameters can truly reflect the physiological properties of arterial vessels [49]. A 1D model is determined by a set of partial differential equations with a lot of parameters. It is very difficult to determine a large number of parameters using system identification methods. The majority of geometrical and mechanical parameters can be directly measured by MRI, CT or Doppler ultrasound equipment. The rest of the parameters are approximated or seen as constants such as the thickness of arterial wall and width of boundary layer. The 1D model is an appropriate approach to study pulse wave propagation phenomenon in arterial system.

Tube-load models are a parallel connection of multiple tubes with parametric loads, which is a simplified 1D model. The tube indicates the path of pulse wave propagation and the load indicates the effective reflection point [75, 76]. The transit time of pressure and flow waves is described by the time delay constant parameter of the model. Tube-load models can represent the relationship between central and peripheral arteries with a few parameters. Once proximal and distal waveforms of the arterial tree are obtained, the parameters of tube-load models can be determined by system identification techniques. In comparison with 1D model, tube-load model has a lower computation cost. Meanwhile, the shortage of the tube-load model is that it is less accurate than 1D model. Tube-load models are suitable for investigating wave propagation and reflection.

In general, computational time of three types of models is short and they have the potential for real-time and general use clinical monitoring in intensive care. In specific applications, however, in order to obtain more accurate clinical parameters, these models may need to be optimized such as personalization of the model which will be made a detailed discussion later. The personalized model requires that model parameters can be obtained individually. The real-time character depends on required patient-specific model parameters, which should be discussed in the following two cases. If the patient-specific model parameters cannot be real-time obtained, the central aortic pressure monitoring is impossible. For example, the pulse transit time parameter in tube-load models can be obtained from various combinations of physiological signals such as two pulse waveforms (e.g. carotid and radial sites) and a combination of ECG and pulse waveforms (e.g. radial site) methods. If we choose the former method (two pulse waveforms), the pulse transit time parameter cannot be real-time obtained since the carotid pulse waveform is difficult to measure for long time. In this case, we cannot monitor central aortic pressure in real-time. The pulse transit time parameter can be real-time estimated if we choose the latter method (ECG and pulse waveforms) because ECG and radial pulse waveforms can be measured simultaneously for long time, real-time central aortic pressure can be monitored. Windkessel models and tube-load models can be used in long-term monitoring of central aortic pressure/potential general hospital ward usage, however, 1D models cannot. This is because geometric parameters of 1D model need to be measured by CT or MRI equipment, which is very complex and costly. Detailed comparisons of three types of low-dimensional models are summarized in Table 4.

Table 4 Comparison of Windkessel, 1D and tube-load models

Future challenges

Although a variety of physics-based models have been developed, there still exist a number of challenging problems to be solved. Multi-scale modeling, coupling of various systems and patient-specific modeling are very significant research subjects at present.

Multi-scale modeling

A model of each scale has its scope of application [29, 97, 98]. Low-dimensional models have low computational cost but poor accuracy, which is applicable to represent the global properties of the arterial networks. Nevertheless, high-dimensional models can offer high accuracy simulation but with greater complexity. They are commonly used to describe the local properties of arterial vessels in detail. Therefore, coupling models of various different scales can combine the advantages of different dimensional models [43, 99, 100]. Multi-scale modeling of arterial vessels can be a powerful tool for providing potential applications in clinical practice.

Coupling of various systems

There exist a variety of biological systems such as the nervous system and the respiratory system in human body which run simultaneously and interactively. The effects of other biological systems are usually ignored in physical modeling of cardiovascular system. As a matter of fact, the nervous system has a significant impact on cardiovascular system [101]. For example, as the blood pressure changes, to avoid dysfunctions, the sympathetic nerves and the parasympathetic nerves are usually motivated to regulate cardiovascular response [101]. Smith et al. [102] proposed a modified cardiovascular model by including a minimal autonomic nervous system activation model. This modified model can simulate various cardiovascular diseases such as hypovolemic shock septic shock, cardiogenic shock, pericardial tamponade and pulmonary embolism.

Due to the fact that both the respiratory system and the cardiovascular system located in the thoracic cavity, the cardiovascular system can be influenced by the respiratory system strongly [103]. For example, an integrated model of the cardiopulmonary system was present by Albanese et al. [104], which included cardiovascular circulation, respiratory mechanics, gas exchange and neural control mechanisms. The physiological parameters in normal and pathological conditions were simulated and the interactions between the cardiovascular and respiratory systems were explained. Trenhago et al. [105] proposed a refined coupled model of the cardiovascular and respiratory systems, consisting of 19 compartments, in which the respiratory system was extended to include a complex system for gas exchange and transport. The advantage of the refined model is that it enables to simulate situations in which existing models cannot predict mimic and it helps us to understand complex mechanisms better. For modeling of cardiovascular system, the respiratory effects should be considered. In comparison with previous cardiovascular model, a modified cardiovascular model integrating neurological and respiratory components has better accuracy. Hence, combining the nervous system and the respiratory system with the cardiovascular system might be a good way to improve the accuracy of models.

Patient-specific modeling

Patient-specific models can provide opportunities for improving accuracy [106,107,108]. The patient-specific parameters can be obtained from imaging techniques such as magnetic resonance imaging, computed tomography and ultrasound. A properly personalized model can predict physiological or pathological status more accurately [109, 110]. The personalized modeling in the arterial system can play an increasingly key role in the development of medical instruments.

This paper takes the estimation of central arterial pressure as an example to introduce personalization of physics-based models. Although several individualized estimation methods of central aortic pressure have been proposed, these methods haven’t been sufficiently verified in clinical practice yet. An accurate and convenient method of reconstructing central aortic pressure waveform with sufficient verification is a current hot topic. The physics-based models with clear physiological meaning may provide great potential for individualized estimation approach of central aortic pressure. In these three types of physics-based models, 1D models are the most accurate and complicated. Under the condition of guaranteeing a high accuracy, reducing the complexity of the model as much as possible is the prefered method. By examining the influence of the complexity of the arterial tree on the accuracy of the model, the complex model can be greatly simplified. For example, a 1D aortic model consisting of ascending aorta, aortic arch, thoracic aorta and abdominal aorta with Windkessel outflow conditions may be created to reconstruct aortic pressure waveform with a high accuracy. Another feasible way is to use body size parameters to replace complex blood vessel parameters. Since 1D models have too many parameters, it is actually impossible to measure all parameters, however, human body size parameters are available easily. A statistical analysis has shown that the blood vessel sizes have close correlation with sex, age, height, and weight of a subject [111]. Furthermore, Young’s modulus representing vascular stiffness has a strong correlation with age. It might be possible to build relationships between geometrical and mechanical parameters of 1D models and body size parameters of the subject. The geometrical and mechanical parameters can be firstly employed by population averages and then the average parameters can be calibrated with body size parameters. Combining with the two methods above, a modified 1D model may be the best choice for estimating central aortic pressure.

Conclusion

In conclusion, different physics-based models in cardiovascular system have different traits and the selection of models mainly depends on the aim of modeling including complexity and accuracy required. By discussing the advantages and disadvantages of various physics-based models, this review contributes to a better understanding of physiological mechanism in the arterial system and provides effective guidance on low-dimensional physics-based modeling.

References

  1. Nejad SE, Carey JP, Mcmurtry MS, Hahn J-O. Model-based cardiovascular disease diagnosis: a preliminary in-silico study. Biomech Model Mechanobiol. 2016;16(2):549–60.

    Article  Google Scholar 

  2. Mendis S, Puska P, Norrving B, Mendis S, Puska P, Norrving B. Global atlas on cardiovascular disease prevention and control. Geneva: World Health Organization; 2011.

    Google Scholar 

  3. Weintraub WS, Daniels SR, Burke LE, Franklin BA Jr, Hayman LL, Lloyd-Jones D, Pandey DK, Sanchez EJ, Schram AP. Value of primordial and primary prevention for cardiovascular disease: a policy statement from the american heart association. Circulation. 2011;124(8):967–90.

    Article  Google Scholar 

  4. Quarteroni A, Formaggia L. Mathematical modelling and numerical simulation of the cardiovascular system. Handb Numer Anal. 2004;12:7–9.

    Google Scholar 

  5. Quarteroni A, Manzoni A, Vergara C. The cardiovascular system: mathematical modelling, numerical algorithms and clinical applications. Acta Numer. 2017;26:365–590.

    Article  MathSciNet  MATH  Google Scholar 

  6. Capoccia M, Marconi S, Singh SA, Pisanelli DM, De CL. Simulation as a preoperative planning approach in advanced heart failure patients. a retrospective clinical analysis. Biomed Eng Online. 2018;17(1):52–72.

    Article  Google Scholar 

  7. Tang D, Li ZY, Gijsen F, Giddens DP. Cardiovascular diseases and vulnerable plaques: data, modeling, predictions and clinical applications. Biomed Eng Online. 2015;14(1):1–7.

    Article  Google Scholar 

  8. Ghigo AR, Fullana JM, Lagree PY, Ghigo AR, Fullana JM, Lagree PY, Ghigo AR, Fullana JM, Lagree PY. A 2D nonlinear multiring model for blood flow in large elastic arteries. J Comput Phys. 2016;350:136–65.

    Article  MathSciNet  MATH  Google Scholar 

  9. Boujena S, Kafi O, El Khatib N. A 2D mathematical model of blood flow and its interactions in an atherosclerotic artery. Math Model Nat Phenom. 2014;09(6):46–68.

    Article  MathSciNet  MATH  Google Scholar 

  10. Lopezperez A, Sebastian R, Ferrero JM. Three-dimensional cardiac computational modelling: methods, features and applications. Biomed Eng Online. 2015;14(1):35–65.

    Article  Google Scholar 

  11. Xie X, Zheng M, Wen D, Li Y, Xie S. A new CFD based non-invasive method for functional diagnosis of coronary stenosis. Biomed Eng Online. 2018;17(1):36–48.

    Article  Google Scholar 

  12. Frank O. Grundform des arteriellen pulses. Z Biol. 1899;37:483–526.

    Google Scholar 

  13. Shi Y, Lawford P, Hose R. Review of zero-D and 1-D models of blood flow in the cardiovascular system. Biomed Eng Online. 2011;10(1):33–70.

    Article  Google Scholar 

  14. Malatos S, Raptis A, Xenos M. Advances in low-dimensional mathematical modeling of the human cardiovascular system. J Hypertens Manag. 2016;2(2):1–10.

    Google Scholar 

  15. Westerhof N, Bosman F, De Vries CJ, Noordergraaf A. Analog studies of the human systemic arterial tree. J Biomech. 1969;2(2):121–43.

    Article  Google Scholar 

  16. Lambert Wallace J. Fluid flow in a nonrigid tube. PhD thesis, Purdue University, Mechanical Engineering Department. 1956.

  17. Hughes TJR, Lubliner J. On the one-dimensional theory of blood flow in the larger vessels. Math Biosci. 1973;18(1):161–70.

    Article  MATH  Google Scholar 

  18. Olufsen MS. Modeling the arterial system with reference to an anesthesia simulator. PhD thesis, Roskilde University, Mathematics Department. 1998.

  19. Olufsen MS, Peskin CS, Kim WY, Pedersen EM, Nadim A, Larsen J. Numerical simulation and experimental validation of blood flow in arteries with structured-tree outflow conditions. Ann Biomed Eng. 2000;28(11):1281–99.

    Article  Google Scholar 

  20. Abdollahzade M, Kim CS, Fazeli N, Finegan BA, Sean MM, Hahn J-O. Data-driven lossy tube-load modeling of arterial tree: in-human study. J Biomech Eng. 2014;136(10):101011–7.

    Article  Google Scholar 

  21. Gao M, Cheng H, Sung S, Chen C, Olivier NB, Mukkamala R. Estimation of pulse transit time as a function of blood pressure using a nonlinear arterial tube-load model. IEEE Trans Biomed Eng. 2017;64(7):1524–34.

    Article  Google Scholar 

  22. Vennin S, Li Y, Willemet M, Fok H, Gu H, Charlton P, Alastruey J, Chowienczyk P. Identifying hemodynamic determinants of pulse pressure. Hypertension. 2017;70(5):1176–82.

    Article  Google Scholar 

  23. Her K, Kim JY, Lim KM, Choi SW. Windkessel model of hemodynamic state supported by a pulsatile ventricular assist device in premature ventricle contraction. Biomed Eng Online. 2018;17(1):18–30.

    Article  Google Scholar 

  24. Harana MJ. Non-invasive, MRI-based calculation of the aortic blood pressure waveform by 0-dimensional flow modelling: development and testing using in silico and in vivo data. Master’s thesis, King’s college London, Department of Biomedical Engineering. 2017.

  25. Jiang S, Zhiqiang, Wang F, Wu J-K. A personalized-model-based central aortic pressure estimation method. J Biomech. 2016;49(16):4098–106.

    Article  Google Scholar 

  26. Natarajan K, Cheng H-M, Liu J, Gao M, Sung S-H, Chen C-H, Hahn J-O, Mukkamala R. Central blood pressure monitoring via a standard automatic arm cuff. Sci Rep. 2017;7(1):14441–52.

    Article  Google Scholar 

  27. Ghasemi Z, Lee JC, Kim C-S, Cheng H-M, Sung S-H, Chen C-H, Mukkamala R, Hahn J-O. Estimation of cardiovascular risk predictors from non-invasively measured diametric pulse volume waveforms via multiple measurement information fusion. Sci Rep. 2018;8(1):10433–43.

    Article  Google Scholar 

  28. Westerhof N, Lankhaar JW, Westerhof BE. The arterial Windkessel. Med Biol Eng Comput. 2009;47(2):131–41.

    Article  Google Scholar 

  29. Liu H, Liang F, Wong J, Fujiwara T, Ye W, Tsubota KI, Sugawara M. Multi-scale modeling of hemodynamics in the cardiovascular system. Acta Mech Sin. 2015;31(4):446–64.

    Article  MathSciNet  MATH  Google Scholar 

  30. Liu Z, Brin KP, Yin FC. Estimation of total arterial compliance: an improved method and evaluation of current methods. Am J Physiol. 1986;251(2):588–600.

    Google Scholar 

  31. Lee J, Sohn JJ, Park J, Yang SM, Lee S, Kim HC. Novel blood pressure and pulse pressure estimation based on pulse transit time and stroke volume approximation. Biomed Eng Online. 2018;17(1):81–100.

    Article  Google Scholar 

  32. Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol. 1993;74(5):2566–73.

    Article  Google Scholar 

  33. Laskey WK, Parker HG, Ferrari VA, Kussmaul WG, Noordergraaf A. Estimation of total systemic arterial compliance in humans. J Appl Physiol. 1990;69(1):112–9.

    Article  Google Scholar 

  34. Stergiopulos N, Young DF, Rogge TR. Computer simulation of arterial flow with applications to arterial and aortic stenoses. J Biomech. 1992;25(12):1477–88.

    Article  Google Scholar 

  35. Stergiopulos N, Westerhof BE, Westerhof N. Total arterial inertance as the fourth element of the Windkessel model. Am J Physiol. 1999;276(1 Pt 2):81–8.

    Google Scholar 

  36. Deswysen B, Charlier AA, Gevers M. Quantitative evaluation of the systemic arterial bed by parameter estimation of a simple model. Med Biol Eng Comput. 1980;18(2):153–6.

    Article  Google Scholar 

  37. Burattini R, Di SP. Development of systemic arterial mechanical properties from infancy to adulthood interpreted by four-element windkessel models. J Appl Physiol. 2007;103(1):66–79.

    Article  Google Scholar 

  38. Segers P, Georgakopoulos D, Afanasyeva M, Champion HC, Judge DP, Millar HD, Verdonck P, Kass DA, Stergiopulos N, Westerhof N. Conductance catheter-based assessment of arterial input impedance, arterial function, and ventricular-vascular interaction in mice. Am J Physiol Heart Circ Physiol. 2005;288(3):1157–64.

    Article  Google Scholar 

  39. Jager GN, Westerhof N, Noordergraaf A. Oscillatory flow impedance in electrical analog of arterial system: representation of sleeve effect and non-newtonian properties of blood. Circ Res. 1965;16(2):121–33.

    Article  Google Scholar 

  40. Avanzolini G, Barbini P, Cappello A, Cevenini G, Moller D, Pohl V, Sikora T. Electrical analogs for monitoring vascular properties in artificial heart studies. IEEE Trans Biomed Eng. 1989;36(4):462–70.

    Article  Google Scholar 

  41. Frasch HF, Kresh JY, Noordergraaf A. Two-port analysis of microcirculation: an extension of Windkessel. Am J Physiol. 1996;270(2):376–85.

    Google Scholar 

  42. Shi Y, Lawford PV, Hose DR. Numerical modeling of hemodynamics with pulsatile impeller pump support. Ann Biomed Eng. 2010;38(8):2621–34.

    Article  Google Scholar 

  43. Quarteroni A, Veneziani A, Vergara C. Geometric multiscale modeling of the cardiovascular system, between theory and practice. Comput Meth Appl Mech Eng. 2016;302:193–252.

    Article  MathSciNet  Google Scholar 

  44. Stergiopulos N, Westerhof N. Role of total arterial compliance and peripheral resistance in the determination of systolic and diastolic aortic pressure. Pathologie Biologie. 1999;47(6):641–7.

    Google Scholar 

  45. Struijk PC, Mathews VJ, Loupas T, Stewart PA, Clark EB, Steegers EAP, Wladimiroff JW. Blood pressure estimation in the human fetal descending aorta. Ultrasound Obstet Gynecol. 2008;32(5):673–81.

    Article  Google Scholar 

  46. Zala D. Arterial flow based transfer function and ascending aorta pressure waveform estimation. Master’s thesis, The State University of New Jersey, The Graduate School of Biomedical Science. 2017.

  47. Revie JA, Stevenson D, Chase JG, Pretty CJ, Lambermont BC, Ghuysen A, Kolh P, Shaw Geoffrey M, Desaive T. Evaluation of a model-based hemodynamic monitoring method in a porcine study of septic shock. Comput Math Method Med. 2013;2013:1–17.

    Article  Google Scholar 

  48. Cai Y, Ma C, Zhang P, Liu J. A novel method to reconstruct central aortic pressure signal using dual-peripheral pressure waves. In: IEEE international conference on information science & technology; 2014. p. 565–8.

  49. Euler L. Principia pro motu sanguinis per arterias determinando. J Biol Phys. 1775;2(1):814–23.

    Google Scholar 

  50. Reymond P, Merenda F, Perren F, Rufenacht D, Stergiopulos N. Validation of a one-dimensional model of the systemic arterial tree. Am J Physiol Heart Circ Physiol. 2009;297(1):208–22.

    Article  Google Scholar 

  51. Saito M, Ikenaga Y, Matsukawa M, Watanabe Y, Asada T, Lagreee P-Y. One-dimensional model for propagation of a pressure wave in a model of the human arterial network: comparison of theoretical and experimental results. J Biomech Eng. 2011;133(12):121005–13.

    Article  Google Scholar 

  52. Parker KH, Jones CJH. Forward and backward running waves in the arteries: analysis using the method of characteristics. J Biol Phys. 1990;112(3):322–6.

    Google Scholar 

  53. Wang JJ, O’Brien AB, Shrive NG, Parker KH, Tyberg JV. Time-domain representation of ventricular-arterial coupling as a Windkessel and wave system. Am J Physiol Heart Circ Physiol. 2003;284(4):1358–68.

    Article  Google Scholar 

  54. Wang JPK. Wave propagation in a model of the arterial circulation. J Biomech. 2004;37(4):457–70.

    Article  Google Scholar 

  55. Brook BS, Falle SAEG, Pedley TJ. Numerical solutions for unsteady gravity-driven flows in collapsible tubes: evolution and roll-wave instability of a steady state. J Fluid Mech. 1999;396(396):223–56.

    Article  MathSciNet  MATH  Google Scholar 

  56. Brook BS, Pedley TJ. A model for time-dependent flow in (giraffe jugular) veins: uniform tube properties. J Biomech. 2002;35(1):95–107.

    Article  Google Scholar 

  57. Porenta G, Young DF, Rogge TR. A finite-element model of blood flow in arteries including taper, branches, and obstructions. J Biomech Eng. 1986;108(2):161–7.

    Article  Google Scholar 

  58. Formaggia L, Gerbeau J-F, Nobile F, Quarteroni A. On the coupling of 3D and 1D navier-stokes equations for flow problems in compliant vessels. Comput Meth Appl Mech Eng. 2001;191(6–7):561–82.

    Article  MathSciNet  MATH  Google Scholar 

  59. Wan J, Steele B, Spicer SA, Strohband S, Feijoo G, Hughes TJR, Taylor CA. A one-dimensional finite element method for simulation-based medical planning for cardiovascular disease. J Biomech Eng. 2002;5(3):195–206.

    Google Scholar 

  60. Sherwin SJ, Franke V, Peireo J, Parker K. One-dimensional modelling of a vascular network in space-time variables. J Eng Math. 2003;47(3–4):217–50.

    Article  MathSciNet  MATH  Google Scholar 

  61. Bessems D, Giannopapa CG, Rutten MCM, Vosse FNVD. Experimental validation of a time-domain-based wave propagation model of blood flow in viscoelastic vessels. J Biomech. 2008;41(2):284–91.

    Article  Google Scholar 

  62. John LR. Forward electrical transmission line model of the human arterial system. Med Biol Eng Comput. 2004;42(3):312–21.

    Article  Google Scholar 

  63. Sunagawa K, Sagawa K. Models of ventricular contraction based on time-varying elastance. Crit Rev Biomed Eng. 1982;7(3):193–228.

    Google Scholar 

  64. Cox LG, Loerakker S, Rutten MC, de Mol BA, van de Vosse F. A mathematical model to evaluate control strategies for mechanical circulatory support. Artif Organs. 2009;33(8):593–603.

    Article  Google Scholar 

  65. Lighthill J. Mathematical biofluiddynamics. Comput Methods Appl Mech Eng. 1975;386(4):1–3.

    MATH  Google Scholar 

  66. Vosse FNVD, Stergiopulos N. Pulse wave propagation in the arterial tree. Annu Rev Fluid Mech. 2011;43(1):467–99.

    Article  MathSciNet  MATH  Google Scholar 

  67. Olufsen MS. Structured tree outflow condition for blood flow in larger systemic arteries. Am J Physiol. 1999;276(1 Pt 2):257–68.

    Google Scholar 

  68. Khalife M, Decoene A, Caetano F, Rochefort L, Durand E, Rodriguez D. Estimating absolute aortic pressure using mri and a one-dimensional model. J Biomech. 2014;47(13):3390–9.

    Article  Google Scholar 

  69. Bardossy G, Halesz G. A “backward” calculation method for the estimation of central aortic pressure wave in a 1D arterial model network. Comput Fluids. 2013;73:134–44.

    Article  MathSciNet  MATH  Google Scholar 

  70. Zhang G, Hahn J-O, Mukkamala R. Tube-load model parameter estimation for monitoring arterial hemodynamics. Front Physiol. 2011;2:72–89.

    Google Scholar 

  71. Gao M, Rose William C, Fetics B, Kass DA, Chen C-H, Mukkamala R. A simple adaptive transfer function for deriving the central blood pressure waveform from a radial blood pressure waveform. Sci Rep. 2016;6(1):33230–8.

    Article  Google Scholar 

  72. Swamy G, Xu D, Olivier NB, Mukkamala R. An adaptive transfer function for deriving the aortic pressure waveform from a peripheral artery pressure waveform. Am J Physiol Heart Circ Physiol. 2009;297(5):1956–63.

    Article  Google Scholar 

  73. Sugimachi M, Shishido T, Miyatake K, Sunagawa K. A new model-based method of reconstructing central aortic pressure from peripheral arterial pressure. Jpn J Physiol. 2001;51(2):217–22.

    Article  Google Scholar 

  74. Hahn J-O, Reisner AT, Jaffer FA, Harry H. Subject-specific estimation of central aortic blood pressure using an individualized transfer function: a preliminary feasibility study. IEEE Trans Inform Technol Biomed. 2011;16(2):212–20.

    Article  Google Scholar 

  75. Campbell KB, Burattini R, Bell DL, Kirkpatrick RD, Knowlen GG. Time-domain formulation of asymmetric T-tube model of arterial system. Am J Physiol. 1990;258(6 Pt 2):1761–74.

    Google Scholar 

  76. Burattini R, Campbell KB. Modified asymmetric T-tube model to infer arterial wave reflection at the aortic root. IEEE Trans Biomed Eng. 2002;36(8):805–14.

    Article  Google Scholar 

  77. Gravlee GP, Brauer SD, O’Rourke MF, Avolio AP. A comparison of brachial, femoral, and aortic intra-arterial pressures before and after cardiopulmonary bypass. Anaesth Intensive Care. 1989;17(3):305–11.

    Article  Google Scholar 

  78. Nakayama R, Goto T, Kukita I, Sakata R. Sustained effects of plasma norepinephrine levels on femoral–radial pressure gradient after cardiopulmonary bypass. J Anesth. 1993;7(1):8–15.

    Article  Google Scholar 

  79. Pauca AL, Wallenhaupt SL, Kon ND, Tucker WY. Does radial artery pressure accurately reflect aortic pressure? Chest. 1992;102(4):1193–8.

    Article  Google Scholar 

  80. Young DF, Cholvin NR, Roth AC. Pressure drop across artificially induced stenoses in the femoral arteries of dogs. Circ Res. 1975;36(6):735–43.

    Article  Google Scholar 

  81. Gao M, Mukkamala R. Perturbationless calibration of pulse transit time to blood pressure. Conf Proc IEEE Eng Med Biol Soc. 2012;2012(2012):232–5.

    Google Scholar 

  82. Hughes DJ, Babbs CF, Geddes LA, Bourland JD. Measurements of Young’s modulus of elasticity of the canine aorta with ultrasound. Ultrason Imaging. 1979;1(4):356–67.

    Article  Google Scholar 

  83. Taylor MG. Wave-travel in a non-uniform transmission line, in relation to pulses in arteries. Phys Med Biol. 1965;10(4):539–50.

    Article  Google Scholar 

  84. Einav S, Aharoni S, Manoach M. Exponentially tapered transmission line model of the arterial system. IEEE Trans Biomed Eng. 1988;35(5):333–9.

    Article  Google Scholar 

  85. Chang KC, Tseng YZ, Lin YJ, Kuo TS, Chen HI. Exponentially tapered T-tube model of systemic arterial system in dogs. Med Eng Phys. 1994;16(5):370–8.

    Article  Google Scholar 

  86. Chang KC, Tseng YZ, Kuo TS, Chen HI. Impedance and wave reflection in arterial system: simulation with geometrically tapered T-tubes. Med Biol Eng Comput. 1995;33(5):652–60.

    Article  Google Scholar 

  87. Fogliardi R, Burattini R, Campbell KB. Identification and physiological relevance of an exponentially tapered tube model of canine descending aortic circulation. Med Eng Phys. 1997;19(3):201–11.

    Article  Google Scholar 

  88. Segers P, Carlier S, Pasquet A, Rabben SI, Hellevik LR, Remme E, De BT, De SJ, Thomas JD, Verdonck P. Individualizing the aorto-radial pressure transfer function: feasibility of a model-based approach. Am J Physiol Heart Circ Physiol. 2000;279(2):542–9.

    Article  Google Scholar 

  89. Matonick JP, Li KJ. A new nonuniform piecewise linear viscoelastic model of the aorta with propagation characteristics. Cardiovasc Eng. 2001;1(1):37–47.

    Article  Google Scholar 

  90. Ghasemi Z, Kim C-S, Ginsberg E, Gupta A, Hahn J-O. Model-based blind system identification approach to estimation of central aortic blood pressure waveform from noninvasive diametric circulatory signals. J Dyn Syst Meas Control. 2016;139:1–10.

    Google Scholar 

  91. Lee J. Subject-specific multichannel blind system identification of human arterial tree via cuff oscillation measurements. Master’s thesis, University of Maryland, Department of Mechanical Engineering. 2016.

  92. Kim C-S, Fazeli N, McMurtry MS, Finegan BA, Hahn J-O. Quantification of wave reflection using peripheral blood pressure waveforms. IEEE J Biomed Health Inform. 2015;19(1):309–16.

    Article  Google Scholar 

  93. Hahn J-O, Reisner AT, Jaffer FA, Harry H. Individualized estimation of the central aortic blood pressure waveform: a comparative study. IEEE J Biomed Health Inform. 2014;18(1):215–21.

    Article  Google Scholar 

  94. Wilde RBPD, Schreuder JJ, Berg PCMVD, Jansen JRC. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery. Anaesthesia. 2007;62(8):760–8.

    Article  Google Scholar 

  95. Shim Y, Pasipoularides A, Straley CA, Hampton TG, Soto PF, Owen CH, Davis JW, Glower DD. Arterial windkessel parameter estimation: a new time-domain method. Ann Biomed Eng. 1994;22(1):66–77.

    Article  Google Scholar 

  96. Taco K, Faes TJC, Jan-Willem L, Anton VN, Michel V. Estimation of three- and four-element windkessel parameters using subspace model identification. IEEE Trans Biomed Eng. 2010;57(7):1531–8.

    Article  Google Scholar 

  97. Alastruey J, Parker KH, Peiro J, Sherwin SJ. Lumped parameter outflow models for 1-D blood flow simulations: effect on pulse waves and parameter estimation. Commun Comput Phys. 2008;4(2):317–36.

    MathSciNet  MATH  Google Scholar 

  98. Taelman L, Degroote J, Verdonck P, Vierendeels J, Segers P. Modeling hemodynamics in vascular networks using a geometrical multiscale approach: numerical aspects. Ann Biomed Eng. 2013;41(7):1445–58.

    Article  Google Scholar 

  99. Chen WW, Gao H, Luo XY, Hill NA. Study of cardiovascular function using a coupled left ventricle and systemic circulation model. J Biomech. 2016;49(12):2445–54.

    Article  Google Scholar 

  100. Zhang Y, Barocas VH, Berceli SA, Clancy CE, Eckmann DM, Garbey M, Kassab GS, Lochner DR, Mcculloch AD, Tran-Son-Tay R. Multi-scale modeling of the cardiovascular system: disease development, progression, and clinical intervention. Ann Biomed Eng. 2016;44(9):1–19.

    Article  Google Scholar 

  101. Liang F. An integrated computational study of multi-scale hemodynamics and multi-mechanism physiology in human cardiovascular system. PhD thesis, Chiba University, Artificial System Science Department. 2007.

  102. Smith BW, Andreassen S, Shaw GM, Jensen PL, Rees SE, Chase JG. Simulation of cardiovascular system diseases by including the autonomic nervous system into a minimal model. Comput Methods Programs Biomed. 2007;86(2):153–60.

    Article  Google Scholar 

  103. Gutta S, Cheng Q, Benjamin BA. Control mechanism modeling of human cardiovascular-respiratory system. In: Global S. I. P; 2016. p. 918–22.

  104. Albanese A, Cheng L, Ursino M, Chbat NW. An integrated mathematical model of the human cardiopulmonary system: model development. Am J Physiol Heart Circ Physiol. 2015;310(7):899–921.

    Article  Google Scholar 

  105. Trenhago PR, Fernandes LG, Mueller LO, Blanco PJ, Feijoo RA. An integrated mathematical model of the cardiovascular and respiratory systems. Int J Numer Methods Biomed. 2016;32(1):1–25.

    MathSciNet  Google Scholar 

  106. Kiselev IN, Kolpakov FA, Biberdorf EA, Baranov VI, Komlyagina TG, Suvorova IY, Melnikov VN, Krivoschekov SG. Patient-specific 1D model of the human cardiovascular system. In: Proceeding of international conference biomedical engineering computer technology; 2015. p. 76–81.

  107. Taylor CA, Figueroa CA. Patient-specific modeling of cardiovascular mechanics. Annu Rev Biomed Eng. 2009;11:109–34.

    Article  Google Scholar 

  108. Tay WB, Tseng YH, Lin LY, Tseng WY. Towards patient-specific cardiovascular modeling system using the immersed boundary technique. Biomed Eng Online. 2011;10(1):52–68.

    Article  Google Scholar 

  109. Poleszczuk J, Debowska M, Dabrowski W, Wojcik-Zaluska A, Zaluska W, Waniewski J. Patient-specific pulse wave propagation model identifies cardiovascular risk characteristics in hemodialysis patients. PLoS Comput Biol. 2018;14(9):1–15.

    Article  Google Scholar 

  110. Zheng D, Yin M, Fan X, Yang X, Luo X. A patient-specific lumped-parameter model of coronary circulation. Sci Rep. 2018;8(1):874–83.

    Article  Google Scholar 

  111. Mortensen JD, Talbot S, Burkart JA. Cross-sectional internal diameters of human cervical and femoral blood vessels: relationship to subject’s sex, age, body size. Anat Rec. 1990;226(1):115–24.

    Article  Google Scholar 

  112. Parlikar TA, Heldt T, Ranade GV, Verghese GC. Model-based estimation of cardiac output and total peripheral resistance. In: Computers in cardiology; 2007. p. 379–82.

  113. Stergiopulos NMJWN. Evaluation of methods for estimation of total arterial compliance. Am J Physiol. 1995;268(2):1540–8.

    Google Scholar 

  114. Burkhoff D, Alexander J, Schipke J. Assessment of Windkessel as a model of aortic input impedance. Am J Physiol. 1988;255(2):742–53.

    Google Scholar 

  115. Gnudi G. Analytical relationship between arterial input impedance and the three-element Windkessel series resistance. Med Biol Eng Comput. 1998;36(4):480–4.

    Article  Google Scholar 

  116. Kamoi S, Pretty C, Balmer J, Davidson S, Pironet A, Desaive T, Shaw GM, Chase JG. Improved pressure contour analysis for estimating cardiac stroke volume using pulse wave velocity measurement. Biomed Eng Online. 2017;16(1):16–51.

    Article  Google Scholar 

  117. Kamoi S, Pretty C, Chiew YS, Davidson S, Pironet A, Desaive T, Shaw GM, Chase JG. Relationship between stroke volume and pulse wave velocity. In: 9th IFAC symposium on biological and medical systems; 2015. p. 285–90.

  118. Myers TG, Ripoll VR, de Tejada Cuenca AS, Mitchell SL, McGuinness MJ. Modelling the cardiovascular system for assessing the blood pressure curve. Math Ind Case Stud. 2017;8(1):2–17.

    Article  Google Scholar 

  119. Huppert TJ, Allen MS, Benav H, Devor A, Jones P, Dale A, Boas DA. A multi-compartment vascular model for inferring arteriole dilation and cerebral metabolic changes during functional activation. J Cereb Blood Flow Metab. 2007;27(6):1262–79.

    Article  Google Scholar 

  120. Huppert TJ, Allen MS, Diamond SG, Boas DA. Estimating cerebral oxygen metabolism from fMRI with a dynamic multi-compartment Windkessel model. Hum Brain Mapp. 2009;30(5):1548–67.

    Article  Google Scholar 

  121. Huberts W, Bode AS, Kroon W, Planken RN, Tordoir JH, Fn vdV, Bosboom EM. A pulse wave propagation model to support decision-making in vascular access planning in the clinic. Med Eng Phys. 2012;34(2):233–48.

    Article  Google Scholar 

  122. Jalali A, Jones G, Licht D, Nataraj C. Application of mathematical modeling for simulation and analysis of hypoplastic left heart syndrome (HLHS) in pre- and postsurgery conditions. Biomed Res Int. 2015;2015:1–14.

    Article  Google Scholar 

  123. Frolov SV, Sindeev SV, Lischouk VA, Gazizova DS, Liepsch D, Balasso A. A lumped parameter model of cardiovascular system with pulsating heart for diagnostic studies. Biomed Res Int. 2016;17(3):1750056–76.

    Google Scholar 

  124. Bodley WE. The non-linearities of arterial blood flow. Phys Med Biol. 1971;16(4):663–72.

    Article  Google Scholar 

  125. Streeter VL, Keitzer WF, Bohr DF. Pulsatile pressure and flow through distensible vessels. Circ Res. 1963;13(1):3–20.

    Article  Google Scholar 

  126. Reymond P, Merenda F, Perren F, Rufenacht D, Stergiopulos N. Validation of a one-dimensional model of the systemic arterial tree. Am J Physiol Heart Cric Physiol. 2009;297(1):208–22.

    Article  Google Scholar 

  127. Pan Q, Wang R, Reglin B, Cai G, Yan J, Pries AR, Ning G. A one-dimensional mathematical model for studying the pulsatile flow in microvascular networks. J Biomech Eng. 2014;136(1):011009–19.

    Article  Google Scholar 

  128. Sun YH, Anderson TJ, Parker KH, Tyberg JV. Wave-intensity analysis: a new approach to coronary hemodynamics. J Appl Physiol. 2000;89(4):1636–44.

    Article  Google Scholar 

  129. Hollander EH, Dobson GM, Wang JJ, Parker KH, Tyberg JV. Direct and series transmission of left atrial pressure perturbations to the pulmonary artery: a study using wave-intensity analysis. Am J Physiol Heart Circ Physiol. 2004;286(1):267–75.

    Article  Google Scholar 

  130. Zambanini A, Cunningham SL, Parker KH, Khir AW, McG Thom SA, Hughes AD. Wave-energy patterns in carotid, brachial, and radial arteries: a noninvasive approach using wave-intensity analysis. Am J Physiol Heart Circul Physiol. 2005;289(1):270–6.

    Article  Google Scholar 

  131. Charlton P, Aresu M, Spear J, Chowienczyk P, Alastruey J. Indices to assess aortic stiffness from the finger photoplethysmogram: in silico and in vivo testing. Artery Res. 2018;24:128.

    Article  Google Scholar 

  132. Willemet M, Chowienczyk P, Alastruey J. A database of virtual healthy subjects to assess the accuracy of foot-to-foot pulse wave velocities for estimation of aortic stiffness. Am J Physiol Heart Circul Physiol. 2015;309(4):663–75.

    Article  Google Scholar 

  133. Vennin S, Mayer A, Li Y, Fok H, Clapp B, Alastruey J, Chowienczyk P. Noninvasive calculation of the aortic blood pressure waveform from the flow velocity waveform: a proof of concept. Am J Physiol Heart Circul Physiol. 2015;309(5):969–1000.

    Google Scholar 

  134. Mukkamala R, Hahn J-O, Inan OT, Mestha LK, Kim C-S, Toreyin H, Kyal S. Toward ubiquitous blood pressure monitoring via pulse transit time: theory and practice. IEEE Trans Biomed Eng. 2015;62(8):1879–901.

    Article  Google Scholar 

  135. Gao M, Zhang G, Olivier NB, Mukkamala R. Improved pulse wave velocity estimation using an arterial tube-load model. IEEE Trans Biomed Eng. 2014;61(3):848–58.

    Article  Google Scholar 

  136. Swamy G, Olivier B, Mukkamala R. Calculation of forward and backward arterial waves by analysis of two pressure waveforms. IEEE Trans Biomed Eng. 2010;57(12):2833–9.

    Article  Google Scholar 

  137. Kim C, Fazeli N, McMurtry MS, Finegan BA, Hahn J-O. Quantification of wave reflection using peripheral blood pressure waveforms. IEEE J Biomed Health Inform. 2015;19(1):309–16.

    Article  Google Scholar 

  138. Leung M, Dumont G, Sandor GGS, Potts JE. Estimating arterial stiffness using transmission line model. In: Conference proceeding on IEEE engineering medical biological society; 2006. p. 1375–8.

  139. Mo LY, Bascom PA, Ritchie K, McCowan LME. A transmission line modelling approach to the interpretation of uterine Doppler waveforms. Ultrasound Med Biol. 1988;14(5):365–76.

    Article  Google Scholar 

Download references

Authors’ contributions

SZ performed the literature review and drafted the manuscript; LX conceived the study and provided the theory of mathematical modeling; LH, HX and YaY checked the manuscript; LQ and YuY gave suggestions and revised the manuscript. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

Not applicable.

Consent for publication

All authors consent for the publication of this manuscript.

Ethics approval and consent to participate

Not applicable.

Funding

National Natural Science Foundation of China (Nos. 61773110, 61374015, 61202258 and 61701099); Fundamental Research Funds for the Central Universities (N161904002 and N172008008); Open Program of Neusoft Research of Intelligent Healthcare Technology (NRIHTOP1801).

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Lisheng Xu.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhou, S., Xu, L., Hao, L. et al. A review on low-dimensional physics-based models of systemic arteries: application to estimation of central aortic pressure. BioMed Eng OnLine 18, 41 (2019). https://doi.org/10.1186/s12938-019-0660-3