Skip to main content

The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges

Abstract

Background

The monitoring of intracranial pressure (ICP) has a crucial role in the surveillance of patients with brain injury. During long-term monitoring of ICP, we have seen spontaneous shifts in baseline pressure (ICP sensor zero point), which are of technical and not physiological origin. The aim of the present study was to explore whether or not baseline pressures of ICP sensors can be affected by electrostatics discharges (ESD's), when ESD's are delivered at clinically relevant magnitudes.

Methods

We performed bench-testing of a set of commercial ICP sensors. In our experimental setup, the ICP sensor was placed in a container with 0.9% NaCl solution. A test person was charged 0.5 - 10 kV, and then delivered ESD's to the sensor by touching a metal rod that was located in the container. The continuous pressure signals were recorded continuously before/after the ESD's, and the pressure readings were stored digitally using a computerized system

Results

A total of 57 sensors were tested, including 25 Codman ICP sensors and 32 Raumedic sensors. When charging the test person in the range 0.5-10 kV, typically ESD's in the range 0.5 - 5 kV peak pulse were delivered to the ICP sensor. Alterations in baseline pressure ≥ 2 mmHg was seen in 24 of 25 (96%) Codman sensors and in 17 of 32 (53%) Raumedic sensors. Lasting changes in baseline pressure > 10 mmHg that in the clinical setting would affect patient management, were seen frequently for both sensor types. The changes in baseline pressure were either characterized by sudden shifts or gradual drifts in baseline pressure.

Conclusions

The baseline pressures of commercial solid ICP sensors can be altered by ESD's at discharge magnitudes that are clinically relevant. Shifts in baseline pressure change the ICP levels visualised to the physician on the monitor screen, and thereby reveal wrong ICP values, which likely represent a severe risk to the patient.

Background

In patients with brain injury due to traumatic brain injury, stroke, or complications to neurosurgery, the continuous monitoring of intracranial pressure (ICP) is crucial for surveillance [13], even though no randomized trials have confirmed the benefit of ICP monitoring in patients with brain injury [4].

Modern ICP monitoring was first introduced by Janny in 1950 [5] and Lundberg in 1960 [6]. While ICP initially was mostly measured from fluid-filled catheters in connection with the ventricular cerebrospinal fluid (CSF), the first ICP micro transducers were introduced in the 1980's [7, 8]. The ICP micro transducers most commonly used today include Camino [9] and Codman [10] ICP sensors which were introduced in the 1980's, the Spiegelberg ICP sensor [11] introduced in the 1990's, the Raumedic ICP sensor introduced in the beginning of 2000 [12], and the Pressio ICP sensor [13] introduced more recently. There is an extensive literature on the assessment of these ICP sensors, including bench testing [10, 11, 1319] and clinical evaluation [9, 12, 2030].

In our hospital, we have particularly addressed the problem of spontaneous shifts in baseline pressure (zero point) that occur during continuous ICP monitoring. Simultaneous monitoring from two ICP sensors placed nearby in the brain demonstrated spontaneous shifts in baseline pressure, which produced differences in ICP even > 20 mmHg [31]. Since the differences in ICP were accompanied by close to identical ICP waveforms, the differences in ICP could be explained by shifts in baseline pressure of technical, not physiological, origin. Similar observations of marked differences in ICP even > 10-20 mmHg despite identical ICP waveform were done when the ICP sensors were placed in different intracranial locations [30, 32]. Moreover, during long-term ICP monitoring, sudden shifts in baseline ICP occurred with few hours interval [33]. The reasons for spontaneous shifts in baseline pressure have not been identified.

The present study was undertaken to explore whether or not commercial ICP sensors are affected by electrostatic discharges (ESD's). To our knowledge this topic has previously not been addressed. The issue of electrical safety in hospitals has received much attention since many years [3437]. The need for increased awareness of electromagnetic interference with medical equipment also was addressed more recently [38]. In the hospital environment, ESD's can be evoked during patient care such as bedding of hospital beds [39]. The frequency and severity of ESD's are affected by numerous factors such as humidity and temperature, the in-house environment (textiles used in clothing, antistatic floor and washing).

In this study, we made an experimental set-up to deliver ESD's to ICP sensors at magnitudes that are clinically relevant. We tested different types of solid ICP sensors that have previously been extensively tested and are still widely used, namely the Codman ICP sensor, and the Raumedic NeuroVent and NeuroDur ICP sensors.

Methods

ICP sensors

The following types of commercially available ICP sensors were tested: Codman ICP MicroSensor (Codman, Johnson & Johnson, Raynham, MA, USA; Figure 1a), Raumedic NeuroVent P-C (Raumedic AG, Münchberg, GE; Figure 1b), Raumedic NeuroVent P (Raumedic AG, Münchberg, GE; Figure 1c), and Raumedic NeuroDur sensor (Raumedic AG, Münchberg, GE; Figure 1d).

Figure 1
figure 1

The different sensors/sensor transducers applied in this study. In this study we tested the (a) Codman Microsensor, (b) Raumedic NeuroVent P-C, (c) Raumedic NeuroVent P, and (d) Raumedic NeuroDur solid sensors.

Those ICP sensors that had previously been used in patients were stored in closed plastic bags at room temperature.

This study did not include research on humans or animals; ethical approval for the study was not applicable.

Experimental setup

The experimental setup is illustrated in Figure 2. The sensor was placed in a container filled with 0.9% NaCl solution. Also a metal rod was placed in the container, whereby a test-person could deliver ESD's to the sensor. The sensor cable was wrapped around the container in order to compare with the clinical situation, wherein the cable is placed on the patient, resulting in increased capacitance. The wrapping procedure was not done for SensorID's 1-3. The Codman sensor was connected to a Codman ICP Express (Codman, Johnson & Johnson, Raynham, MA, USA), which is a pressure transducer. The Raumedic sensors were connected to a MPR1 Raumedic pressure transducer (Raumedic AG, Münchberg, GE). The continuous ICP signals from the Codman ICP express were sampled at 200 Hz using the Sensometrics pressure logger, which is an analogue to digital converter, and stored on a computer using the Sensometrics® software (dPCom A/S, Oslo). The continuous signals provided by the Raumedic MPR-1 were transferred directly to the computer and stored using Sensometrics® software.

Figure 2
figure 2

Experimental set-up. The ICP sensor (a) was placed in a container (b) containing 0.9% NaCl solution. A metal rod (c) was placed in the 0.9% NaCl solution; thereby the test person (d) could deliver ESD to the sensor (a). The ICP sensor was connected via a cable (e) to the pressure transducer (f), which was the Codman ICP Express for Codman sensors, and Raumedic MPR-1 for the various Raumedic sensors. The pressure transducer (f) was further connected with a laptop computer (g) with Sensometrics software for sampling and storage of the continuous pressure signals. In order to record the ESD delivered to the sensor, a meter (h) was connected with the container to read the magnitude of ESD.

In order to record the electrostatic levels reached, we used the Stat Arc II model 265 (Monroe Electronics Inc., 100 Housel Ave., Lyndonville, N.Y.).

The testing was done in standardized room temperature of 22-23°C.

Testing of effects of electrostatic discharges

After initiating the pressure recording, the test person was charged using a Metriso 5000 insulation tester (Metrawatt GmbH, Germany) and thereafter touching the metal rod. The test-person was charged in the following sequence: 0.5 kV, 1.0 kV, 2.5 kV, and 5.0 kV. For deliverance of 10 kV, the test person first charged the container to 5kV and then swiftly charged himself to 5 kV opposite polarity before touching the metal rod. The test-person noted whether or not he detected the current impulse being delivered.

The testing of a Codman sensor is shown in Additional file 1, and the testing of a Raumedic sensor is shown in Additional file 2.

Additional file 1: The testing of a Codman sensor. The video shows the testing of a Codman sensor. A test person is being charged, and then touches the metal rod, leading the electrostatic discharge to the container filled with 0.9% NaCl Ringer solution, wherein the sensor is placed. A schematic illustration is shown in Figure 2. (WMV 5 MB)

Additional file 2: The testing of a Raumedic sensor. The video shows the testing of a Raumedic sensor. A test person is being charged, and then touches the metal rod, leading the electrostatic discharge to the container filled with 0.9% NaCl Ringer solution, wherein the sensor is placed. A schematic illustration is shown in Figure 2. (WMV 5 MB)

Assessment of leakage current

All sensors were tested for leakage current using the Metriso 5000 insulation tester and a Fluke 87 III True RMS Multimeter (in the voltage range). The procedure of testing for leakage current is shown in Figure 3. All leakage current measurements were done at 500V unless otherwise stated; resistance was calculated.

Figure 3
figure 3

Set-up for testing of leakage current. The ICP sensor (a) was placed in a container (b) containing 0.9% NaCl solution. A metal rod (c) was placed in the 0.9% NaCl solution and connected to one side of a 500V source (h). All pins of the ICP sensor connector (f) were connected to the Volts input of the digital multimeter (g) and the multimeter Common input was connected to the other side of the 500V source. The multimeter's input impedance of 10 MΩ thus acted as a shunt resistor providing a current scale of 100pA/mV.

Results

ICP sensors

We tested a total of 57 sensors (25 Codman and 32 Raumedic sensors). Six Codman sensors (SensorID's 39 and 49-53) and seven Raumedic sensors (SensorID's 46-48 and 54-57) were new while the other sensors had previously been used in patients. The proportion of ICP sensors with changes in baseline pressure ≥ 2 mmHg is indicated in Table 1.

Table 1 ICP sensors tested for altered baseline pressure following ESD

The Codman sensors typically responded with sudden shifts in baseline pressure, though gradual drifts in baseline pressure were seen in two sensors (SensorID's 7 and 37). The maximum lasting changes in baseline pressure of Codman sensors are shown in Table 2. Charging the test person to 5 kV caused baseline shift in five sensors (SensorID's 37, 38, 40, 41 and 50), with a measured potential change of 2-5 kV in the 0.9% NaCl solution with the sensor. In the other Codman sensors, changes in baseline pressure occurred when the test person was charged to a 10 kV differential, in which potential changes to the sensor were comparable to that evoked when charging the person to 5 kV. For only one sensor (SensorID 38), we managed to deliver a 10 kV potential change directly to the sensor (taking place after the sensor already had responded to 5 kV). Figure 4 illustrates shifts in baseline pressure in two sensors, and gradual drift of baseline pressure in another. An animation of baseline shift subsequent to ESD's of SensorID 4 is shown in Additional File 3.

Table 2 Lasting alterations in baseline pressure of Codman sensors following ESD
Figure 4
figure 4

Continuous pressure signal from Codman sensors before/after electrostatic discharge. The continuous pressure signals retrieved from Codman Microsensor are shown before and after electrostatic discharges (ESD's) for SensorID 4 (a), SensorID 38 (b), and SensorID 7 (c). Note that sudden changes in baseline pressure occurred for SensorID's 4 and 38, while SensorID 7 showed gradual drift of baseline pressure. The baseline pressure level (mmHg) is indicated on the y axis, and the time line on the × axis levels; the ESD is indicated by an arrow.

The Raumedic sensors responded differently to ESD's depending on the design of the sensors. While Raumedic NeuroVent P-C was highly unstable to ESD's, the Raumedic NeuroVent P and NeuroDur sensors differed. The maximum lasting changes in baseline pressure for the individual Raumedic sensors are presented in Table 3.

Table 3 Lasting alterations in baseline pressure of Raumedic sensors following ESD

All the Raumedic NeuroVent P-C sensors responded with gradual drifts in their baseline pressure; drifts occurred when the test person was charged to 0.5 kV in 3 sensors (SensorID's 16, 29 and 46), 2.5 kV in 6 sensors (SensorID's 11-15, and 18), and 5 kV in two sensors (SensorID's 10 and 57). The gradual drift of NeuroVent P-C in SensorID 14 is shown in Figure 5a (see also Additional file 4).

Figure 5
figure 5

Continuous pressure signal from Raumedic sensors before/after electrostatic discharge. The continuous pressure signals retrieved from Raumedic NeuroVent sensors are shown before and after electrostatic discharges (ESD's) for SensorID 14 (NeuroVent P-C; a), and for SensorID 33 (NeuroVent P; b). The baseline pressure level (mmHg) is indicated on the y axis, and the time line on the × axis levels; the ESD is indicated by an arrow. Note that gradual drifts in baseline pressure occurred. For SensorID 33, repeated low ESD's of 0.5 kV (not being sensed by the test person) causes gradual build-up of baseline pressures.

The shifts in baseline pressure of NeuroVent P sensors were seen after charging the test person to 0.5 kV in two (SensorID's 33 and 47), and to 2.5 kV in another (SensorID 32). In Figure 5b is illustrated how SensorID 33 gradually changed its baseline pressure following repetitive ESD's of 0.5 kV (an animation of the pressure signal is shown in Additional file 5).

All the three NeuroDur sensors responding to ESD (SensorID's 23, 25 and 28) responded when the test person was charged to 0.5 kV, with a corresponding 0.5 kV potential change in the solution around the sensor. The responses of SensorID's 25 and 28 are shown in Figure 6 (see also Additional file 6).

Figure 6
figure 6

Continuous pressure signal from Raumedic sensors before/after electrostatic discharge. The continuous pressure signals retrieved from Raumedic NeuroDur sensors are shown before and after electrostatic discharges (ESD's) of SensorID 28 (a), and for SensorID 25 (b). The baseline pressure level (mmHg) is indicated on the y axis, and the time line on the × axis levels; the ESD is indicated by an arrow. Note that gradual drifts in baseline pressure occurred.

Experimental setup

When the test-person was charged to 0.5 kV, the ESD delivered to the sensor was typically 0.5 kV pulse peak. Charging to 5 kV provided for a potential change of 2.5 kV on average (range 2-5 kV). When charging to 10 kV, typically 3.5-4.0 kV change was seen; in a few occasions we observed even 7 kV change (10 kV on one occasion). While ESD's < 3 kV hardly provided any unpleasant sensations, ESD's of about 5 kV gave weak unpleasant sensations, while ESD's ≥ 7 kV provided a sensation which is evident, though not painful, to the test person.

Leakage current

The results of testing leakage current are presented in Table 4. Leakage current was seen in 1 of the 25 Codman sensors (SensorID 45); this sensor revealed leakage current of up to100 nA with continuous breakthrough.

Table 4 Results of testing of current leakage

The Raumedic sensors differed depending on design. The NeuroVent P-C sensors showed a marginally higher leakage current. Two of the NeuroVent P sensors responding to ESD's (SensorID's 32 and 33) showed leakage current. The three NeuroDur sensors (SensorID's 23, 25 and 28) being affected by ESD's of 0.5 kV, all showed abnormal current leakage.

Discussion

This study shows that the baseline pressures (zero point) of Codman and Raumedic ICP sensors can be altered by ESD's at magnitudes that are clinically relevant. The observations indicate severe limitations with currently used ICP sensors.

ICP sensors used for clinical monitoring of ICP

The ICP sensors tested in this study are widely used ICP sensors. Both the Codman [10, 14, 16, 21, 22, 25, 27] and Raumedic [12, 19, 29, 30] sensors have undergone extensive bench and clinical testing. In general, the assessment of ICP sensors has previously focused on long-term-drift of the sensors, sensitivity to temperature changes, and inter-sensor accuracy comparisons [10, 11, 1319, 22, 25].

All the ICP sensors measure pressure relative to atmospheric pressure, which means that they have to be zeroed before measuring ICP. Hence, their zero point equates the atmospheric pressure, and the ICP level displayed on the monitor represents the difference between pressure level within the intracranial compartment and the sensor zero point. It should be noted that in daily clinical practice, various notations are used to refer to the zero point, such as set point, reference pressure, or baseline pressure. In this paper and previous publications [3033], we have preferred the term baseline pressure, when referring to the zero point of the ICP sensor.

Depending on clinical state, the upper normal threshold of ICP varies between 15 and 25 mmHg [14]. Obviously, if the baseline pressure (zero point) spontaneously shifts > 10-20 mmHg, the ICP presented to the physician becomes wrong. Since the continuous monitoring of ICP is done for surveillance of patients with brain injury, e.g. due to traumatic brain injury, stroke complications to brain surgery [13], false ICP values represent a likely hazard to the patient. For example, when ICP increases, efforts may be done to reduce the ICP; such efforts include medication, artificial ventilation and surgical procedures.

When the impact of ESD's on ICP sensors previously has not been considered, the reason may be that the issues of non-physiological changes in baseline pressure have not been regarded as a problem in ICP monitoring.

Electrostatic discharges in the hospital environment

There are different ways to test ESD's; it has also been addressed that there is a need for more standardized methods [40]. The rational for our experimental setup was to best possible test ESD's of clinically relevant magnitudes. Therefore, the ESD's were delivered from a test person, and caused pulse peak discharges to the sensor typically in the range 0.5 - 5 kV. Such ESD's may not be unpleasant to the test person, and are below the levels that can be seen clinically [39, 41]. ESD's of magnitudes < 2-3 kV may not even be appreciated by the personnel taking care of the patients. It was previously demonstrated that potentials > 30 kV could be induced on the bed framework when the bedding is pulled from the bed; the degree of charging being dependent on the material of hospital bedding [39]. In comparison, previous tests in our hospital showed that ESD's of 20-40 kV could be seen, depending on the textiles used in clothing (Jensen, Grimnes, unpublished data). Using the test approach described here, we avoided ESD's of magnitudes that are not clinically relevant. Only in a few instances, we managed to deliver 7 kV potential changes to the sensor (10 kV in one sensor that first responded markedly to 5 kV). Accordingly, the voltages referred to here are quite low.

Different characteristics of Codman and Raumedic ICP sensors

There were some differences between the Codman and Raumedic sensors in their responses to ESD's. The Codman sensors consistently responded to electrostatic changes of 2-3 kV, with sudden shifts in baseline pressure. Gradual drift was only seen in 2 of 25 Codman sensors (8%). These findings compare with our clinical observations of spontaneous alterations in baseline ICP despite unchanged ICP waveform. The observation that baseline pressure was changed maximally > 10 mmHg in 13 of 25 (52%) sensors (and > 20 mmHg in 3 (12%) sensors), indicate that effects of ESD's are of a magnitude that likely would affect patient management.

The Raumedic sensors responded differently depending on their design. Two types of responses were seen, namely gradual drifts and sudden shifts in baseline pressure. While the NeuroVent P-C was completely unstable to ESD's, even at levels of 0.5 kV, the NeuroVent P was less affected. The P-C type incorporates a ceramic coating on the sensor tip while the P type uses titanium. Also the NeuroDur sensor using titanium was more stable, where the tip seemed connected to the sensor with a 5 MΩ resistance. Nevertheless, the observation of alterations in baseline pressure > 10 mmHg in 10 of 32 (31%) Raumedic sensors (> 20 mmHg in 4 of 32 (12.5%) indicate that the effects of ESD's would affect patient management also when using these sensors. In a recent study comparing simultaneous ICP signals from Raumedic NeuroVent and NeuroDur sensors, we encountered average differences between sensors during over-night monitoring > 10 mmHg in 4 of 12 (33%) patients [30].

While leakage current was seen in only one Codman sensor, and no Raumedic NeuroVent P-C sensors, current leakage was seen in 2 of 3 Raumedic NeuroVent P sensors that responded to ESD's, and in all three Raumedic NeuroDur sensors responding to ESD's. The testing of leakage current indicated that in Raumedic titanium sensors (NeuroVent P and NeuroDur) there is an internal 5 MΩ resistance between the metal shell and the connector. Hence, the sensors with resistance different from 5 MΩ (Table 4) might have a broken protection resistor. We found, however, no evidence of sensor damage using microscopy, though damage to ICP sensors may happen both during the implantation and explanation.

Control of risk associated with ESD's

A major issue with both the Codman and Raumedic ICP sensors is that the health care personnel get no warning about sudden shifts in baseline pressure (zero point) of ICP sensors, or even damage to the ICP sensor during/after implantation. Thereby it is impossible for the physician or nurse to know whether changes in ICP are related to ESD's or not. The Codman sensor cannot be re-zeroed because this is done within the operating room before sensor implantation. The Raumedic sensors, on the other hand, can be re-zeroed after implantation; however, this procedure is not necessarily done by the nurse/physician when ICP is changing.

While the present study focused on effects of ESD's on ICP sensors, the baseline pressure can also be affected by user-related wrong zeroing or even damage to the sensor during implantation, which may not be recognized. Therefore, it can be questioned why modern monitoring systems include no warning. Such warning should be incorporated as part of risk control.

We suggest that a robust way of incorporating risk control is by determining the ICP from the ICP waveform itself. Thereby quality control is accomplished and the issue of baseline pressure alterations is eliminated. The first author previously described a procedure for automatic identification of the cardiac-induced waves in the ICP waveform [42]. Using this approach, the ICP parameters such as the mean ICP wave amplitude (MWA), can be determined from the cardiac induced ICP waves [42]. Since such determination of single wave pressure parameters is done within the ICP signal itself, the analysis results is not affected by changes in baseline pressure. The automatic identification of verified cardiac induced ICP waves also recognizes other ICP sensor-related issues. For example, if an ICP sensor is placed wrong by mistake, artificial waves and no cardiac induced ICP waves will be identified, providing feedback to the user that the ICP signal is erroneous.

Conclusions

The baseline pressure (zero point) of the Codman and Raumedic ICP sensors can be altered by ESD's at discharge magnitudes that are clinically relevant levels. The shifts in baseline pressure will directly affect the ICP levels visualised to the health care personnel. The alterations in baseline pressure can be extensive (> 10-20 mmHg), thereby revealing wrong ICP values, which subsequently poses a high risk for erroneous treatment.

Abbreviations

ICP:

Intracranial pressure

ESD:

Electrical discharge

kV:

kilo Volt

CSF:

cerebrospinal fluid

MWA:

mean ICP wave amplitude

SW:

single wave.

References

  1. Marmarou A, Anderson RL, Ward JD, et al.: Impact of ICP instability and hypotension on outcome in patients with severe head trauma. J Neurosurg 1991, 75: S59-S66.

    Google Scholar 

  2. Ghajar J: Traumatic brain injury. The Lancet 2000, 356: 923–929. 10.1016/S0140-6736(00)02689-1

    Article  Google Scholar 

  3. Czosnyka M, Pickard JD: Monitoring and interpretation of intracranial pressure. J Neurol Neurosurg Psychiatry 2004, 75: 813–821. 10.1136/jnnp.2003.033126

    Article  Google Scholar 

  4. Padayachy LC, Figaji AA, Bullock MR: Intracranial pressure monitoring for traumatic brain injury in the modern era. Child's Nerv Syst 2010, 26: 441–452. 10.1007/s00381-009-1034-0

    Article  Google Scholar 

  5. Janny P: La pression intracrannienne Chez l'Homme. Paris, Thesis; 1950.

    Google Scholar 

  6. Lundberg N: Continuous recording and control of ventricular fluid pressure in neurosurgical practice. Acta Psychiatr Neurol Scand 1960, 36: 1–193.

    Google Scholar 

  7. Allen R: Intracranial pressure: A review of clinical problems, measurement techniques and monitoring methods. J Med Engin Tech 1986, 10: 299–320. 10.3109/03091908609022923

    Article  Google Scholar 

  8. Zhong J, Dujovny M, Park HK, et al.: Advances in ICP monitoring techniques. Neurol Res 2003, 25: 339–350. 10.1179/016164103101201661

    Article  Google Scholar 

  9. Gambardella G, d'Avella D, Tomasello F: Monitoring of brain tissue pressure with a fiberoptic device. Neurosurgery 1992, 31: 918–922. 10.1227/00006123-199211000-00014

    Article  Google Scholar 

  10. Piper IR, Miller JD: The evaluation of the wave-form analysis capability of a new strain-gauge intracranial pressure microsensor. Neurosurgery 1995, 36: 1142–1145. 10.1227/00006123-199506000-00012

    Article  Google Scholar 

  11. Yau YH, Piper I, Clutton RE Whittle IR: Experimental evaluation of the Spiegelberg intracranial pressure and intracranial compliance monitor. J Neurosurg 2000, 93: 1072–1077. 10.3171/jns.2000.93.6.1072

    Article  Google Scholar 

  12. Stendel R, Heidenreich J, Schilling A, et al.: Clinical evaluation of a new intracranial pressure monitoring device. Acta Neurochir 2003, 145: 185–193. 10.1007/s00701-002-1052-0

    Article  Google Scholar 

  13. Allin D, Czosnyka M, Czosnyka Z: Laboratory testing of the Pressio intracranial pressure monitor. Neurosurgery 2008, 62: 1158–1161. 10.1227/01.neu.0000325878.67752.eb

    Article  Google Scholar 

  14. Czosnyka M, Czosnyka Z, Pickard JD: Laboratory testing of three intracranial pressure microtransducers: Technical report. Neurosurgery 1996, 38: 219–224. 10.1097/00006123-199601000-00053

    Article  Google Scholar 

  15. Morgalla MH, Mettenleiter H, Bitzer M, et al.: ICP measurement control: Laboratory test of 7 types of intracranial pressure transducers. J Medical Engineer Techn 1999, 23: 144–151. 10.1080/030919099294195

    Article  Google Scholar 

  16. Morgalla MH, Krasznai L, Dietz K, et al.: Methods of experimental and clinical assessment of the relative measurement accuracy of an intracranial pressure transducer. Technical note. J Neurosurg 2001, 95: 529–532. 10.3171/jns.2001.95.3.0529

    Article  Google Scholar 

  17. Morgalla MH, Dietz K, Deininger M, et al.: The problem of long-term ICP drift assessment: Improvement by use of the ICP drift index. Acta Neurochir 2002, 144: 57–61. 10.1007/s701-002-8274-2

    Article  Google Scholar 

  18. Piper I, Barnes A, Smith D, et al.: The Camino intracranial pressure sensor: Is it optimal technology? An internal audit with a review of current intracranial pressure monitoring technologies. Neurosurgery 2001, 49: 1158–1165.

    Google Scholar 

  19. Citerio G, Piper I, Cormio M, et al.: Bench test assessment of the new Raumedic Neurovent-P ICP sensor: a technical report by the Brain-IT group. Acta Neurochir (Wien) 2004, 146: 1221–1226. 10.1007/s00701-004-0351-z

    Article  Google Scholar 

  20. Pople IK, Muhlbauer MS, Sanford RA, et al.: Results and complications of intracranial pressure monitroing in 303 children. Pediatr Neurosurg 1995, 23: 64–67. 10.1159/000120938

    Article  Google Scholar 

  21. Gray WP, Palmer JD, Gill J, et al.: A clinical study of parenchymal and subdural miniature strain-gauge transducers for monitoring intracranial pressure. Neurosurgery 1996, 39: 927–932.

    Google Scholar 

  22. Fernandes HM, Bingham K, Chambers IR, Mendelow AD: Clinical evaluation of the Codman microsensor inracranial pressure monitoring system. Acta Neurochir 1998, 71(Suppl):44–46.

    Google Scholar 

  23. Guyot LL, Dowling C, Diaz FG, et al.: Cerebral monitoring devices: Analysis of complications. Acta Neurochir 1998, 71(Suppl):47–49.

    Google Scholar 

  24. Rossi S, Buzzi F, Paparella A, et al.: Complications and safety associated with ICP monitoring: A study of 542 patients. Acta Neurochir 1998, 71(Suppl):91–93.

    Google Scholar 

  25. Signorini DF, Shad A, Piper IR, et al.: A clinical evaluation of the Codman Microsensor for intracranial pressure monitoring. Br J Neurosurg 1998, 1: 223–227.

    Google Scholar 

  26. Lang J-M, Beck J, Zimmermann M: Clinical evaluation of intraparenchymal Spiegelberg pressure sensor. Neurosurgery 2003, 52: 1455–1459. 10.1227/01.NEU.0000065136.70455.6F

    Article  Google Scholar 

  27. Koskinen L-O, Olivecrona M: Clinical experience with the intraparenchymal intracranial pressure monitoring Codman Microsensor system. Neurosurgery 2005, 56: 693–698. 10.1227/01.NEU.0000156609.95596.24

    Article  Google Scholar 

  28. Gelabert-Gonzàlez M, Ginesta-Galan V, Sernamito-Garcia R, et al.: The Camino intracranial pressure device in clinical practice. Assessment in a 1000 cases. Acta Neurochir 2006, 148: 435–441. 10.1007/s00701-005-0683-3

    Article  Google Scholar 

  29. Poca M, Martinez-Ricarte F, Sahuquillo J, et al.: Intracranial pressure monitoring with the Neurodur-P epidural sensor: A prospective study in patients with adult hydrocephalus or idiopathic intracranial hypertension. J Neurosurg 2008, 108: 934–942. 10.3171/JNS/2008/108/5/0934

    Article  Google Scholar 

  30. Eide PK, Sorteberg W: Simultaneous measurements of intracranial pressure parameters in the epidural space and in brain parenchyma in patients with hydrocephalus. J Neurosurg 2010, 113: 1317–1325. 10.3171/2010.7.JNS10483

    Article  Google Scholar 

  31. Eide PK: Comparison of simultaneous continuous intracranial pressure (ICP) signals from a Codman and a Camino ICP sensor. Med Eng Physics 2006, 28: 542–549. 10.1016/j.medengphy.2005.09.003

    Article  Google Scholar 

  32. Eide PK: Comparison of simultaneous continuous intracranial pressure (ICP) signals from ICP sensors placed within the brain parenchyma and the epidural space. Med Eng Physics 2008, 30: 34–40. 10.1016/j.medengphy.2007.01.005

    Article  Google Scholar 

  33. Eide PK, Rapoport BI, Gormley WB, et al.: A dynamic nonlinear relationship between the static and pulsatile components of intracranial pressure in patients with subarachnoid hemorrhage. J Neurosurg 2010, 112: 616–625. 10.3171/2009.7.JNS081593

    Article  Google Scholar 

  34. Bruner JM: Hazards of electrical apparatus. Anesthesiology 1967, 28: 396–425. 10.1097/00000542-196703000-00023

    Article  Google Scholar 

  35. Bruner JM, Aronow S, Cavicchi RV: Electrical incidents in a largehospital: a 42 month register. J Assoc Adv Med Instrum 1972, 6: 222–230.

    Google Scholar 

  36. Leonard PF: Characteristics of electrical hazards. Anesth Analg 1972, 51: 797–809.

    Article  Google Scholar 

  37. Weibell FJ: Eletrical safety in the hospital - 1974. Ann Biomed Engineer 1974, 2: 126–148. 10.1007/BF02368486

    Article  Google Scholar 

  38. Abenstein JP: Safety While Swimming in a Sea of Energy. Mayo Clinic Proceedings 2007, 82: 276–278. 10.4065/82.3.276

    Article  Google Scholar 

  39. Holdstock P, Wilson N: The effect of static charge generated on hospital bedding. EOS/ESD Symposium 1996, 356–364.

    Google Scholar 

  40. Smallwood J: Standardisation of electrostatic test methods and electrostatic discharge prevention measures for the world market. J Electrostatics 2005, 63: 501–508. 10.1016/j.elstat.2005.03.009

    Article  Google Scholar 

  41. Kathirgamanathan P, Toohey MJ, Haase J, et al.: Measurements of incendivity of electrostatic discharges from textiles used in personal protective clothing. J Electrostatics 2000, 49: 51–70. 10.1016/S0304-3886(00)00003-6

    Article  Google Scholar 

  42. Eide PK: A new method for processing of continuous intracranial pressure signals. Med Eng Physics 2006, 28: 579–587. 10.1016/j.medengphy.2005.09.008

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Per K Eide.

Additional information

Competing interests

AB reports no conflicts of interest. PKE has financial interest in the software company (dPCom A/S) that manufactures the software (Sensometrics® Research software and Sensometrics® Software), which was used for digital recording of the continuous pressure signals in this study.

Authors' contributions

Both authors have made substantial contributions to conception and design, acquisition of data, analysis and interpretation of data; and have been involved in drafting the manuscript or revising it critically for important intellectual content. Both authors have read and approved the final manuscript.

Electronic supplementary material

12938_2011_489_MOESM3_ESM.mpg

Additional file 3: A continuous pressure signal before and after ESD - Codman sensor. The continuous pressure signal of a Codman sensor (SensorID 4) is shown at a higher velocity (about × 30) than normal. The test person was charged to 10 kV, which delivered an ESD of 5 kV to the sensor. At the time of the ESD, a sudden change in baseline pressure occurs. See also Figure 4a. (MPG 1 MB)

12938_2011_489_MOESM4_ESM.mpg

Additional file 4: A continuous pressure signal before and after ESD - Raumedic NeuroVent P-C sensor. The continuous pressure signal of a Raumedic P-C sensor (SensorID 14) is shown at a higher velocity (about × 30) than normal. The test person was charged to 2.5 kV three times, which delivered ESD's of 1-2.5 kV to the sensor. At the time of the first ESD, a slight reduction of baseline pressure occurred, while at the second and third ESD's, the baseline pressures drifted to higher levels. See also Figure 5a. (MPG 2 MB)

12938_2011_489_MOESM5_ESM.mpg

Additional file 5: A continuous pressure signal before and after ESD - Raumedic NeuroVent P sensor. The continuous pressure signal of a Raumedic P sensor (SensorID 33) is shown at a higher velocity (about × 30) than normal. The test person was repeatedly charged to 0.5 kV, which delivered ESD's of 0.5 kV to the sensor. At the time of each ESD, a slight increase of baseline pressure occurred. See also Figure 5b. (MPG 5 MB)

12938_2011_489_MOESM6_ESM.mpg

Additional file 6: A continuous pressure signal before and after ESD - Raumedic NeuroDur sensor. The continuous pressure signal of a Raumedic NeuroDur sensor (SensorID 28) is shown at a higher velocity (about × 30) than normal. The test person was charged to 0.5 kV, which delivered an ESD of 0.5 kV to the sensor. At the time of ESD, a marked downward drift of baseline pressure occurred. See also Figure 6a. (MPG 2 MB)

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Eide, P.K., Bakken, A. The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges. BioMed Eng OnLine 10, 75 (2011). https://doi.org/10.1186/1475-925X-10-75

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1475-925X-10-75

Keywords