Table 3 Lasting alterations in baseline pressure of Raumedic sensors following ESD

From: The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges

ICP Sensor

Baseline pressure (mmHg)

Raumedic

SensorID

Before

After

Maximum change 1

NeuroVent P-C

10

18.4

-9.3

-27.7

"

11

-0.3

-11.5

-11.2

"

12

9.5

3.6

-5.9

"

13

0.6

6.9

6.3

"

14

-5.4

11.4

16.8

"

15

-0.7

-14.1

-13.4

"

16

0

18.3

18.3

"

18

-1.4

1.1

2.5

"

29

-5.7

17.3

23

"

46

-12.1

-0.1

12

"

57

1.3

-8.3

-9.6

NeuroVent P

17

1.6

0.1

-1.5

"

30

-0.2

-0.5

-0.3

"

31

-0.05

-0.9

-0.85

"

32

-0.3

4.7

5

"

33

0

9.6

9.6

"

34

0.2

0.7

0.5

"

35

0.1

0.5

0.4

"

47

2.5

-0.1

-2.6

"

48

0.4

-0.1

-0.5

"

54

0.6

-0.9

-1.5

"

55

0.3

0.5

0.2

"

56

0

-0.1

-0.1

NeuroDur

19

0.1

-0.6

-0.7

"

21

-0.05

-0.1

-0.05

"

22

0

0.1

0.1

"

23

52.4

0.8

-51.6

"

24

0.8

0.9

0.1

"

25

0.4

10.7

10.3

"

26

0.8

0.3

-0.5

"

27

0.1

0.2

0.1

"

28

0

39

39

  1. 1Change in baseline pressure ≥ 2 mmHg (absolute value) is highlighted.