Skip to main content

Table 1 ICP sensors tested for altered baseline pressure following ESD

From: The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges

Type of sensor

Number

Number (%) with change baseline pressure ≥ 2 mmHg

Codman ICP Microsensor

25

24 (96)

Raumedic

  

NeuroVent P-C

11

11 (100)

NeuroVent P

12

3 (25)

NeuroDur

9

3(33)