Stability of different fixation methods after reduction malarplasty under average and maximum masticatory forces: a finite element analysis

Abstract

Background

Although titanium plates/screws are effective fixation methods (FM) after L-shaped osteotomy reduction malarplasty (LORM), the ideal FM remains controversial. This first finite element analysis (FEA) aimed to study the effect of various zygomatic body/zygomatic arch FM combinations and their placement vectors on the zygoma complex stability after virtual LORM under the effect of both average (150 N/mm2) and maximum (750 N/mm2) forces and three-dimensional (3D) mapping of stress and strain parameters distribution over the zygomatic bone, fixation methods, and total model.

Results

The fixation methods about the short-arm of the L-shaped osteotomy showed lower stress, strain, and displacement values than those across the long-arm osteotomy site. Combined with any zygomatic arch fixation methods (ZAFm), the two bicortical screws group (2LS) on the zygomatic body osteotomy site resulted in smaller displacements and the lowest zygoma bone stress and displacement when combined with Mortice–Tenon structure (MT) as zygomatic arch fixation method. Applied forces caused statistically significant differences in zygomatic bone stress (P < 0.001 and P = 0.001) and displacement (P = 0.001 and P = 0.002).

Conclusion

All FMs both on the zygomatic body and zygomatic arch provide adequate zygomatic complex stability after LORM. The 2LS group showed better resistance than rectangular plate (RP) and square plate (SP) with lower stress concentrations. The L-shaped plate with short-wing on the maxilla (LPwM) is more stable than having the short-wing on the zygoma bone (LPwZ). Future prospective clinical studies are required to validate the current findings.

Background

The zygomatic bone is a prominent midface structure that is crucial for three-dimensional facial aesthetic appearance. It shows racial characteristics by determining the lateral face's height, width, and projection [1]. Orientals are characterised by high and overly protruding zygomas that upset the facial units’ harmony with rough, aggressive, and masculine appearance [1,2,3,4,5]. Therefore, reduction malarplasty (RM) is commonly practised in East Asia [6]. Over the past three decades, several surgical techniques have been introduced, with progressive advances in simplicity, safety, and aesthetic outcomes [2, 3, 7,8,9,10,11,12,13,14]. Most of these studies were focused on the osteotomy design on both the zygomatic body (ZB) and the zygomatic arch (ZA). Currently, L-shaped osteotomy RM (LORM) consisting of intraoral L-shaped osteotomy (LO) with simultaneous short-sideburn incision is most popular because of its advantages: short surgical time, small scar, low risk of facial nerve injury, and fast recovery [6, 12, 14,15,16,17,18]. The optimal RM procedure should fulfil the following goals: midfacial width reduction, flattening of the zygomatic prominence, preservation of the natural malar curvature, nonvisible scar, and zygomatic complex sustained height as proposed by Nakanishi et al. [10]. Of equal importance is maintaining intimate bone-to-bone contact on the zygomatic body’s osteotomy line and zygomatic complex (body and arch) fixation without bony dehiscence [19]. The fundamental problem of LORM is the instability due to poor fixation, which is exaggerated by powerful masseteric inferior pulling action [19, 20]. Not surprisingly, the literature revealed that postoperative zygomatic complex mobility is conductive to complications such as malunion or nonunion with resultant unfavourable outcomes such as sagging of the cheek, facial asymmetry, malar depression, and restricted jaw movement [16, 21,22,23,24,25,26,27]. All these outcomes mandate corrective surgical procedures through aggressive approaches such as the bicoronal flap approach that could result in major morbidities like big facial scar, hair loss, and facial nerve injury [16, 22, 24]. Therefore, prevention of these unfoavourable outcomes should be kept in surgeons mind. This prevention could be achieved by choosing robust internal fixation methods after establishing an intimate bone-to-bone contact in the L-shaped osteotomy region. Of the same importance, is the fixation methods’ placement vector about the L-shaped osteotomy line. To the best of our knowledge, there are no comparative studies showed the effect of combined zygomatic body fixation methods (ZBFm) with zygomatic arch fixation methods (ZAFm) on the total stability of the zygomatic complex after L-shaped osteotomy reduction malarplasty. Most previous studies focused on either the fixation methods on the zygomatic body or on the zygomatic arch separately. Kim et al. [15] demonstrated in their experiment on skull replicate samples that it is better to locate the fixation point at a higher level in the medial part of the zygomatic body, closer to the lateral orbital rim. According to studies by Baek et al. [16, 23, 24] and Lee and Lee [22], the intraoral approach does not allow fixation of the lateral orbital rim side, it cannot withstand the strong action of the masseter muscle, which may cause the unfavourable outcomes mentioned above. On the other hand, Hwang et al. [19] and Wang et al. [21] explained the effect of zygomatic arch fixation method on the stability of RM outcomes despite the fixation methods used on ZB region.

Hence, there is no consensus regarding the optimum FM combinations and best placement vector that will provide long-term stable outcomes. Therefore, choosing ZB/ZA fixation method combinations and their placement vector is still controversial among surgeons. One reason such controversies persist is that the basic biomechanics of the zygomatic complex region are not well understood [28,29,30].

FEA is a numerical approach simulating the dynamics of physical objects with confirmed benefits in evaluating facial fracture plating techniques [31, 32]. It has the following advantages: graphic visualisation of unseen regions, precise model simulation of perioperative behaviour, time savings, cost-effectiveness, repeatability, and various clinical scenario simulations through variations in force application point, magnitude, and direction [31].

However, the fixation method efficiency after LORM has never been studied using 3D-FEA. Therefore, this first time FEA study aimed to investigate the stability of commonly used zygomatic body and zygomatic arch fixation method combinations under normal (150 N/mm2) and maximum (750 N/mm2) masticatory forces and 3D-mapping of the stress and strain distribution over zygoma bone, FM, and the overall model after virtual LORM. In addition, we studied the effect of placement vector of L-shaped plate through comparing the stability and the stress distribution pattern when the short wing fixated on the zygoma (LPwZ) versus fixated on the maxilla (LPwM). Therefore, the results of this study can serve as an evidence-based guide for the selection of stable ZB/ZA fixation method combinations and the optimal placement vectors for these combinations. With such evidence, disputes among surgeons in this area will be resolved for good clinical practice, which will positively impact the health of patients as well as the profession and reputation of surgeons. Furthermore, the present results can be applied to industrial fixation methods (titanium plates and screws). Therefore, stronger titanium plates can be produced by strengthening the weak regions that appeared in the stress–strain 3D-mapping under loading conditions.

Results

Stress analysis

Under 150 and 750 N/mm2, the highest stress values of the overall model (373.58 and 1817.5, respectively) and zygoma bone (604.1 and 2934.6, respectively) belonged to the SP*3HP combination, as shown in the colour-coded diagram (Fig. 1). The LPwZ*MT combination recorded the lowest values of the overall model under both forces (58.714 and 157.39, respectively), while the 2LS*MT combination showed zygoma bone’s lowest values (149.73 and 748.66, respectively). Regarding the stress distribution over the fixation methods, the highest values were demonstrated by the LPwM*MT combination under both forces (265.53 and 1325.2, respectively). However, the lowest value under 150 N/mm2 was shown by the 2LS*MT combination (42.566), which was increased almost four times under 750 N/mm2 with the LPwZ*MT combination (163.03) (Table 1).

Fig. 1
figure 1

Colour-coded spectrum demonstrating the highest and lowest Von Mises stress distribution over the overall model, zygoma bone, and fixation methods under average (150 N/mm2) and maximum (750 N/mm2) masticatory forces. SP square plate, 3HP 3-hole-plate, LPwM L-shaped plate with a short wing on the maxilla, MT Mortice–Tenon, LPwZ L-shaped plate with a short wing on the zygoma, 2LS 2 long bicortical screws

Table 1 The average and maximum values of stress concentration associated with tested fixation methods under normal (150 N/mm2) and maximum (750 N/mm2) masticatory forces for zygoma bone, fixation methods, and overall model

The colour-coded diagram (Fig. 2) showed that LPwM outperformed LPwZ under both forces. The ANOVA test revealed that all ZBFm groups under both forces showed statistically significant differences in stress distribution over zygoma bone (P < 0.001 and P = 0.001, respectively) and the overall model (P < 0.001 and P = 0.001, respectively) with nonsignificant differences over fixation methods (P = 0.071 and P = 0.173, respectively). ZAFm had nonsignificant differences in all measuring parameters.

Fig. 2
figure 2

Colour-coded spectrum demonstrating the zygoma bone, fixation methods, and overall model stress concentration associated with LPwM compared to LPwZ as zygomatic body fixation methods combined with 3-hole plate, Mortice–Tenon, and short screw as zygomatic arch fixation methods under normal (150 N/mm2) and maximum (750 N/mm2) masticatory forces

Strain analysis

The SP*3HP combination recorded the highest strain values for the overall model and zygoma bone under 150 N/mm2 (0.00283 and 0.00311, respectively) and 750 N/mm2 (0.01382 and 0.01512, respectively). The lowest values for the overall model were shown by the LPLS*SS combination (150 N/mm2 = 0.00068 and 750 N/mm2 = 0.00340), while zygoma bone values were revealed by 2LS*MT (150 N/mm2 = 0.00076 and 750 N/mm2 = 0.00378). The highest values for fixation methods under both forces were shown by the LPwM*MT combination (150 N/mm2 = 0.00287 and 750 N/mm2 = 0.01453, respectively), while the lowest values were recorded by 2LS*MT (150 N/mm2 = 0.00044) and LPwZ*MT (750 N/mm2 = 0.00170). The ANOVA test among ZBFm at 150 N/mm2 revealed a statistically significant difference for zygoma bone (P = 0.000) and fixation methods (P = 0.039), while the overall model had a statistically nonsignificant difference (P = 0.394). In contrast, under 750 N/mm2, the differences were statistically significant for both the overall model (P = 0.032) and zygoma bone (P = 0.013), but for the fixation methods, they were statistically nonsignificant (P = 0.241). Regarding the ZAF methods, both forces yielded statistically nonsignificant differences for all parameters (Table 2).

Table 2 The average and maximum values of strain distribution associated with tested fixation methods under normal (150 N/mm2) and maximum (750 N/mm2) masticatory forces for zygoma bone, fixation methods, and overall model

Displacement analysis

The highest displacement values for the overall model under both forces were associated with the 2LS*SS combination (150 N/mm2 = 0.049 mm and 750 N/mm2 = 0.245 mm) (Table 3).

Table 3 The average and maximum values of displacement values associated with tested fixation methods under normal (150 N/mm2) and maximum (750 N/mm2) masticatory forces for zygoma bone, fixation methods, and overall model

However, LPLS*SS recorded the lowest value (0.011 mm) under 150 N/mm2, while under 750 N/mm2, almost double the value (0.020 mm) was associated with the LPwZ*MT combination. The magnitude of the displacement related to the RP*MT combination, as illustrated by a colour-coded spectrum, was the highest value under both forces for the zygoma bone (150 N/mm2 = 0.244 mm and 750 N/mm2 = 1.135 mm) and the fixation methods’ displacement under 750 N/mm2 (0.504 mm) (Fig. 3). In addition, the highest fixation method displacement value under 150 N/mm2 was recorded by the SP*3HP combination (0.113 mm). On the other hand, the lowest displacement values under both forces for zygoma bone were 0.037 mm under 150 N/mm2 and 0.187 mm under 750 N/mm2, and the lowest values for fixation methods (150 N/mm2 = 0.011 mm and 750 N/mm2 = 0.057 mm) were associated with the 2LS*MT combination. In addition, the colour-coded diagram (Fig. 4) showed the superiority of LPwM over LPwZ in holding the zygoma complex against the masseteric simulation forces. The zygomatic bone and fixation methods showed statistically significant differences under 150 N/mm2 (P = 0.001 and P = 0.004, respectively) and 750 N/mm2 (P = 0.020 and P = 0.040, respectively). At the same time, the overall model failed to show statistically significant differences (P = 0.491 and P = 0.403), as revealed by ANOVA. At the same time, all ZAFM groups’ displacements were statistically nonsignificant differences.

Fig. 3
figure 3

Colour-coded spectrum demonstrating the highest and lowest displacement values of the overall model, zygoma bone, and fixation methods under average (150 N/mm2) and maximum (750 N/mm2) masticatory forces

Fig. 4
figure 4

Colour-coded spectrum demonstrating the zygoma bone, fixation methods, and overall model displacement values associated with LPwM compared to LPwZ as zygomatic body fixation methods combined with 3-hole plate, Mortice–Tenon, and short screw as zygomatic arch fixation methods under normal (150 N/mm2) and maximum (750 N/mm2) masticatory forces

Discussion

The FEA is a powerful tool that has been used extensively in structural analysis of facial skeletons that allows detailed visualisation of where structures bend or twist and indicates the distribution of stresses and displacements [31, 32]. As a result of the computing advancement, we are increasingly turning to virtual analysis and tools tailored to our needs to test the mechanical properties of facial soft and hard tissues and osteosynthesis materials, repetitively, safely, and cost-effectively, which are all met by FEA [33, 34]. Herein, we innovatively implemented the 3D-FEA to map the stress distribution over the most commonly used FM after LORM, which enabled the prediction of areas more susceptible to fracture under simulated masticatory forces.

Generally, the current findings showed that all ZBFm*ZAFm combinations were adequately resistant to zygoma bone displacement, as all records were below 2.5 mm, which from our clinical experience, is not recognised clinically and did not contribute to any complications that warrant reoperation [21].

The highest stress associated with RP and SP combined with any ZAFm over the overall model and zygoma bone could be explained by two points. First, their location was near the site of the maximum masseteric effect along the L-shaped osteotomy line’s long-arm. Second, the thick bone in this region tolerated longer screws, providing more bone surface contact and higher resistance to the masseter muscle. Interestingly, under masseteric action simulated forces, although, among the ZBFm groups, the maximum zygoma bone displacement did not reach above 2.4 mm, RP and SP showed the highest values. This may be due to the parallelism of the plate placement vector with the masseter muscle's vector, which may enhance the muscle's inferior displacement simulated forces. In addition, their highest stress concentration, as shown by the colour-coded diagram, could lead to either screw loosening or plate bridge deformities, as revealed by the high strain values. Such findings could not be explored without FEA help, as they cannot be understood or entirely revealed by in vitro biomechanical studies [35, 36]. Although it seems that RP and SP recorded parameters (stress, strain, and displacement) were higher than other ZBFm groups, all of them still showed good resistance against the inferior displacement. In line with Kim et al. [15] biomechanical study, placing the fixation tools at a higher level near the orbital rim provides more favourable outcomes. Baek et al. [16] also align with these findings. They proposed that placement of the fixation methods at a higher level across the osteotomy line would provide more support against the displacing forces.

Another interesting finding is the effect of changing the placement vector on the FMs’ performance. This is obvious in the amount of resistance the single LP provided in different placement vectors. Having the short-wing fixed to the maxilla (fixed bone), the stress and displacement values of the zygoma bone were almost 50% lower than when fixed to the zygoma (mobile bone). This could have two explanations: the anti-rotation provided by the short-wing vertically arranged screws over the maxilla bone offers less stress and better resistance. Second, the plate's perpendicular placement vector across the short-arm allows support of the greater force. In addition, the distance between the short-wing and the osteotomy line is short, resulting in a higher fixation level with a smaller external rotating force [15].

Our results could explain the outperformance of the 2LS and LPLS groups, holding the zygoma complex (body and arch) more stable due to their double bridges of fixation. These findings concord with Baek et al. [16] who emphasised that at least a two-bridge fixation method on the ZB should be placed to provide adequate support. Hence, the 2LS group is highly recommended when there is a good amount of bone, as it is strong enough to hold the freed zygoma against the rotational masseter muscle force. Furthermore, from our clinical experience, they require less surgical placement time; postoperatively, they are less palpable, no migration could occur compared to plates, and they are less costly.

On the other hand, the ZAF methods after RM also have no standard protocol. Nevertheless, from the literature, the most commonly used methods are either MT structure or titanium plates and screws to ensure good bony consolidation [7, 14, 19, 37]. The authors first proposed the MT structure in 2014, in which the Mortice is formed by the ZA free end and the Tenon by the gap between the zygomatic process laterally and temporal bone medially [7, 21].

The current study confirmed the importance of ZBFm and ZAFm combined stability on freed zygomatic complex (body and arch) stability after RM. Meanwhile, they showed the effect of each type of ZAFm on the stress, strain, and displacement values over the zygomatic complex (body and arch). Furthermore, the findings showed that all ZAF methods performed well against the inferior displacement exerted by the masseter muscle simulated by the two forces (150 N/mm2 and 750 N/mm2).

From all these findings, we appreciate that FEA can vividly show the dynamic behaviour of the zygomatic complex under masseter muscle action and reveal the outcomes of this biomechanical interaction on the implanted plates and screws. Additionally, it is helpful to analyse the stress concentration on various parts of the zygomatic complex incorporating the fixation methods across the osteotomy lines. Therefore, these findings could guide surgeons in choosing the best type of fixation method and the optimum placement vector that provides better stability. In addition, industrial-wise, more robust plates could be produced by strengthening the weak areas that appeared on the stress–strain mapping under loading conditions. This is one of the strong points of the current study. Furthermore, we have studied, for the first time, the ability of 18 ZBFm* ZAFm combinations to withstand the inferiorly directed masseteric force, which was simulated by the loaded forces (the average (150 N/mm2) and voluntary maximum (750 N/mm2) masticatory forces after RM. However, FEA is just a method that simulates real physical conditions; thus, it cannot be completely accurate [32, 38]. Therefore, the findings of this FEA study need to be validated by conducting a prospective clinical study comparing the long-term effect of the same fixation methods studied here on the stability of the zygomatic complex (body and arch) after LORM.

Conclusion

The findings of this first FE simulation suggest that all ZBFm*ZAFm combinations after LORM can provide adequate stability. RP and SP combined with any form of ZAFm did not differ significantly from 2LS, LPLS, or LPwM concerning zygoma bone displacement. The 2LS group showed better resistance with less stress concentrations. The single L-shaped plate will be more stable if its short-wing is fixed on the maxilla. Future studies validating the current findings are recommended.

Materials and methods

FE model creation and virtual surgery

Preoperative CT of a 26-year-old healthy consented female who underwent LORM to correct prominent zygomas was used to produce the virtual 3D zygoma complex model. The research committees at West China Hospital of Stomatology and Sichuan University approved this study (WCHS-CRSE-2022-103).

The Digital Imaging and Communications in Medicine (DICOM) format of the CT with 0.5 mm thickness was imported to 3D-Slicer software to extract the CT images and select the midface bone using the Hounsfield threshold (785.99 to 3071.00) to create a standard tessellation language file (STL) format (https://www.slicer.org) [39]. The 3D model was constructed in all three spatial planes of axial, coronal, and sagittal by manual segregation for better accuracy with a slice thickness of 0.5 mm. The STL file was edited in PTC Creo 4.0 M010 2016 (PTC, Boston, Massachusetts) software (https://www.ptc.com) to create FEA solid model, maintaining only the left zygoma complex bone with part of lateral and inferior orbital rims and posterolateral maxilla anteriorly, articular tubercle posteriorly, and temporal bone medially to fasten the calculation process.

In PTC Creo 4.0, a virtual LORM procedure based on outlines drawn by a senior surgeon [21] was conducted, resulting in a freed zygomatic complex and stable surrounding borders (Fig. 5). The whole complex was then repositioned superomedially and posteriorly, guided by anterior bone-to-bone contact using the software's segmentation function.

Fig. 5
figure 5

L-shaped osteotomy reduction malarplasty was carried out virtually resulting in two separated segments: zygomatic complex (body and arch: red labelled) and boundaries bones (white labelled). Blue line: long arm, yellow line: short arm. A: 45° view, B superior view, C inferior view

Then, the fixation methods were reverse-engineered in PTC Creo 4.0. All plates were made of titanium, 2.0 mm in thickness, namely, L-shaped plate (LP) and square plate (SP). Each plate consisted of 4 monocortical screws of 5 mm length, the rectangular plate (RP) consisted of 6 monocortical screws of 5–6 mm length, and the 3-hole plate (3HP) with 1.7 mm thickness and consisted of 3 monocortical screws of 5 mm length was used for zygomatic arch osteotomy line fixation. Additionally, the short screw (SS) was developed at 7 mm in length. Bicortical screws (LS) (2.0 mm) were designed 15 mm in length. Maintaining the freed zygoma in the new position, the six different zygomatic body fixation methods (ZBFm), namely, across the L-shaped osteotomy line’s short-arm: 2LS, LPLS, LP with a short-wing on the maxilla (LPwM), and LP with a short-wing on the zygoma (LPwZ). SP and RP were implanted across the L-shaped osteotomy line’s long-arm. Each ZBFm group combined three different zygomatic arch fixation methods (ZAFm): MT, 3HP, and SS. A total of 18 virtual models were developed, incorporating the virtually operated and fixated zygoma bone and the internal fixation methods (IFM) (Fig. 6).

Fig. 6
figure 6

Models with fixation techniques across the short arm after virtual reduction malarplasty (A–D) and across the long arm (E, F): zygomatic body fixation methods (frontal view): A L-shaped plate with the short wing on the zygoma, B L-shaped plate with the short wing on the maxilla, C L-shaped plate with one bicortical screw D Two bicortical screws, E 6-hole rectangular plate F Square plate. (1–3): Zygomatic arch fixation methods (lateral view): 1: Mortice–Tenon structure, 2: 3-hole plate, 3: short screw

Material properties

The zygomatic complex virtual model was constructed of homogenous bone. The bone properties were derived from previous studies, with average values representing normal healthy adult bone [40, 41]. The IFMs had the properties of commercially pure titanium Ti-4AI-6V (Table 4).

Table 4 Materials properties used in FEA

Mesh creation and simulation analysis

ANSYS R19.2 (ANSYS, Inc., Canonsburg, Pennsylvania, U.S.) was used to independently mesh bones, screws, and plates of the assembled model, creating a model consisting of a maximum element size of 3 mm and a minimum of 0.5 mm. The surrounding bone meshed with software default mesh size; only the operated assembled zygomatic bone meshed finer for more accurate results (199,292 elements with 344,199 nodes). The freed zygoma and the surrounding bone contact surfaces were movable and nonpenetrating, while the screws were rigidly fixed to the bone (Fig. 7A). Frictional contact was set between the titanium plate surface and the fixated bone. Zygoma bone was the only movable part under the applied forces, while the surrounding bones were fully restrained, forming the first boundary condition. Then, we depicted the inferior surface of the zygomatic complex (body and arch), simulating the location and direction of the masseteric force representing the second boundary condition (Fig. 7B). The force magnitudes were determined based on the studies of Okiyama et al. [42] and Sarkarat et al. [43] to be 150 N/mm2 (average mastication force) and 750 N/mm2 (maximal mastication force) along the Z-axis. The remote displacement formed the third boundary condition. In this step, we guided ANSYS to allow the operated zygoma to rotate in all directions and be displaced in the Z and X directions, representing the masseteric inferolateral pulling force. Otherwise, the zygoma bone could not show the displacement and the generated stress.

Fig. 7
figure 7

A Zygoma bone was the only movable part of the model that could be displaced under the applied pressure, while the surrounding bones were fully restrained, forming the first boundary condition. B Force location and direction in the FEA model simulating masseter muscle

Finally, to obtain the required outcomes, we ran the solution after setting up all the models in terms of stress, strain, and displacement registration over the overall model, zygoma bone, and fixation methods, as shown in the flowchart (Fig. 8).

Fig. 8
figure 8

Finite element analysis processing flowchart

Statistical analysis

Descriptive statistics were used regarding the average and maximum values of the recorded stress, strain, and displacement. ANOVA was used to reveal the presence of statistically significant differences among the ZBFm and ZAFm groups under the loaded forces. If significant, a post hoc test with Bonferroni correction was used for multiple comparisons. SPSS Statistics version 25 (IBM Corp., Armonk, NY, USA) was used for all analyses. A P-value of less than 0.05 was considered statistically significant.

Availability of data and materials

The data used and/or analysed during the present study are available from the corresponding author on reasonable request.

Abbreviations

FM:

Fixation methods

LORM:

L-shaped osteotomy reduction malarplasty

FEA:

Finite element analysis

N/mm2 :

Newton/squared millimetre

3D:

Three dimensional

2LS:

Two bicortical screws

MT:

Mortice–tenon

RP:

Rectangular plate

SP:

Square plate

LPwM:

L-shaped plate with short-wing on the maxilla

LPwZ:

L-shaped plate with short-wing on the zygoma

LPLS:

L-shaped plate with one bicortical screw

RM:

Reduction malarplasty

LO:

L-shaped osteotomy

3HP:

Three-hole plate

SS:

Short screw

3D-FEA:

Three-dimensional finite element analysis

ZBFm:

Zygomatic body fixation methods

ZAFm:

Zygomatic arch fixation methods

CT:

Computed tomography

DICOM:

Digital Imaging and Communications in Medicine

STL:

Standard tessellation language

IFM:

Internal fixation methods

ANSYS:

A general-purpose, finite-element modeling package

ANOVA:

Analysis of variance

SPSS:

Statistical Package for the Social Sciences

References:

  1. Cho J, Kim Y, Choi Y. Three-dimensional analysis of facial asymmetry after zygomaticomaxillary complex fracture reduction: a retrospective analysis of 101 East Asian patients. Arch Craniofac Surg. 2021;22(3):148.

    Article  Google Scholar 

  2. Chen T, Hsu Y, Li J, Hu J, Khadka A, Wang Q, Wang D. Correction of zygoma and zygomatic arch protrusion in East Asian individuals. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;112(3):307–14.

    Article  Google Scholar 

  3. Zou C, Niu F, Liu J, Chen Y, Wang M, Su R, Xu J, An G, Gui L. Midface contour change after reduction malarplasty with a modified L-shaped osteotomy: a surgical outcomes study. Aesthetic Plast Surg. 2014;38(1):177–83.

    Article  Google Scholar 

  4. Jirawatnotai S, Sriswadpong P. The difference in the location of the malar summit between genders in Southeast Asians with appropriate references. Arch Craniofac Surg. 2021;22(2):78.

    Article  Google Scholar 

  5. Onizuka T, Watanabe K, Takasu K, Keyama A. Reduction malar plasty. Aesthetic Plast Surg. 1983;7(2):121–5.

    Article  Google Scholar 

  6. Wang T, Gui L, Tang X, Liu J, Yu D, Peng Z, Song B, Song T, Niu F, Yu B. Reduction malarplasty with a new L-shaped osteotomy through an intraoral approach: retrospective study of 418 cases. Plast Reconstr Surg. 2009;124(4):1245–53.

    Article  Google Scholar 

  7. Tang K, Hsu Y, Hu J, Zhao Y, Wang Q, Li J. New horizontal v-shaped osteotomy for correction of protrusion of the zygoma and the zygomatic arch in East Asians: indication and results. Br J Oral Maxillofac Surg. 2014;52(7):636–40.

    Article  Google Scholar 

  8. Hong SE, Liu SY, Kim JT, Lee JH. Intraoral zygoma reduction using L-shaped osteotomy. J Craniofac Surg. 2014;25(3):758.

    Article  Google Scholar 

  9. Gao Z-W, Wang W-G, Zeng G, Lu H, Ma H-H. A modified reduction malarplasty utilizing 2 oblique osteotomies for prominent zygomatic body and arch. J Craniofac Surg. 2013;24(3):812–7.

    Article  Google Scholar 

  10. Nakanishi Y, Nagasao T, Shimizu Y, Miyamoto J, Kishi K, Fukuta K. The boomerang osteotomy–a new method of reduction malarplasty. J Plast Reconstr Aesthet Surg. 2012;65(5):e111–20.

    Article  Google Scholar 

  11. Kook M-S, Jung S, Park H-J, Ryu S-Y, Oh H-K. Reduction malarplasty using modified L-shaped osteotomy. J Oral Maxillofac Surg. 2012;70(1):e87–91.

    Article  Google Scholar 

  12. Yang X, Mu X, Yu Z, Gu Q, Cao D, Yu D, Wei M, Chang T. Compared study of Asian reduction malarplasty: wedge-section osteotomy versus conventional procedures. J Craniofac Surg. 2009;20(8):1856–61.

    Article  Google Scholar 

  13. Hwang YJ, Jeon JY, Lee MS. A simple method of reduction malarplasty. Plast Reconstr Surg. 1997;99(2):348–55.

    Article  Google Scholar 

  14. Kim Y-H, Seul J-H. Reduction malarplasty through an intraoral incision: a new method. Plast Reconstr Surg. 2000;106(7):1514–9.

    Article  Google Scholar 

  15. Kim JW, Hwang W. Optimal fixation location in intraoral reduction malarplasty using an l-shaped osteotomy. J Craniofac Surg. 2019;30(8):2490–2.

    Article  Google Scholar 

  16. Baek R-M, Kim J, Lee SW. Revision reduction malarplasty with coronal approach. J Plast Reconstr Aesthet Surg. 2010;63(12):2018–24.

    Article  Google Scholar 

  17. Lee K-C, Ha S-U, Park J-M, Kim S-K, Park S-H, Kim J-H. Reduction malarplasty by 3-mm percutaneous osteotomy. Aesthetic Plast Surg. 2006;30(3):333–41.

    Article  Google Scholar 

  18. Yang DB, Chung JY. Infracture technique for reduction malarplasty with a short preauricular incision. Plast Reconstr Surg. 2004;113(4):1253–61.

    Article  Google Scholar 

  19. Hwang CH, Lee MC. Reduction malarplasty using a zygomatic arch-lifting technique. J Plast Reconstr Aesthet Surg. 2016;69(6):809–18.

    Article  MathSciNet  Google Scholar 

  20. Kim JJ, Lee EY, Seok H, Kang JY. An improved technique for zygoma reduction malarplasty. J Craniomaxillofac Surg. 2018;46(4):654–9.

    Article  Google Scholar 

  21. Wang Y, Hsu Y, Chen G, Bi D, Wang Y, Li J. The importance of the mortice and tenon joint on the zygomatic arch in reduction malarplasty-a retrospective study. J Craniomaxillofac Surg. 2022;50(4):316–21.

    Article  Google Scholar 

  22. Lee YH, Lee SW. Zygomatic nonunion after reduction malarplasty. J Craniofac Surg. 2009;20(3):849–52.

    Article  Google Scholar 

  23. Baek R-M, Lee SW. Face lift with reposition malarplasty. Plast Reconstr Surg. 2009;123(2):701–8.

    Article  Google Scholar 

  24. Baek R-M, Kim J. Kim B-k: Three-dimensional assessment of zygomatic malunion using computed tomography in patients with cheek ptosis caused by reduction malarplasty. J Plast Reconstr Aesthet Surg. 2012;65(4):448–55.

    Article  Google Scholar 

  25. Jin H. Reduction malarplasty using an L-shaped osteotomy through intraoral and sideburns incisions. Aesthetic Plast Surg. 2011;35(2):242–4.

    Article  Google Scholar 

  26. Zhang Y, Tang M, Jin R, Zhang Y, Zhang Y, Wei M, Qi Z. Comparison of three techniques of reduction malarplasty in zygomaticus and massateric biomechanical changes and relevant complications. Ann Plast Surg. 2014;73(2):131–6.

    Article  Google Scholar 

  27. Shao Z, Xie Y, Yu B, Liu L, Du T. A new assisted fixation technique to prevent zygoma displacement in malar reduction. Aesthetic Plast Surg. 2013;37(4):692–6.

    Article  Google Scholar 

  28. Rohner D, Tay A, Meng CS, Hutmacher DW, Hammer B. The sphenozygomatic suture as a key site for osteosynthesis of the orbitozygomatic complex in panfacial fractures: a biomechanical study in human cadavers based on clinical practice. Plast Reconstr Surg. 2002;110(6):1463–71 (discussion 1472).

    Google Scholar 

  29. Deveci M, Eski M, Gurses S, Yucesoy CA, Selmanpakoglu N, Akkas N. Biomechanical analysis of the rigid fixation of zygoma fractures: an experimental study. J Craniofac Surg. 2004;15(4):595–602.

    Article  Google Scholar 

  30. Hanemann M Jr, Simmons O, Jain S, Baratta R, Guerra AB, Metzinger SE. A comparison of combinations of titanium and resorbable plating systems for repair of isolated zygomatic fractures in the adult: a quantitative biomechanical study. Ann Plast Surg. 2005;54(4):402–8.

    Article  Google Scholar 

  31. Shyam Sundar S, Nandlal B, Saikrishna D, Mallesh G. Finite element analysis: a maxillofacial surgeon’s perspective. J Maxillofac Oral Surg. 2012;11(2):206–11.

    Article  Google Scholar 

  32. Szwedowski TD, Whyne CM, Fialkov JA. Toward characterization of craniofacial biomechanics. J Craniofac Surg. 2010;21(1):202–7.

    Article  Google Scholar 

  33. Meira JBC, Jikihara AN, Capetillo P, Roscoe MG, Cattaneo PM, Ballester RY. Chapter 3: Finite element analysis in dentistry. In: Dental Biomaterials. 2018; 67–89. https://doi.org/10.1142/9789813225688_0003.

    Chapter  Google Scholar 

  34. Prado FB, Freire AR, Cláudia Rossi A, Ledogar JA, Smith AL, Dechow PC, Strait DS, Voigt T, Ross CF. Review of in vivo bone strain studies and finite element models of the zygomatic complex in humans and nonhuman primates: implications for clinical research and practice. Anat Rec. 2016;299(12):1753–78.

    Article  Google Scholar 

  35. Hart RT, Hennebel VV, Thongpreda N, Van Buskirk WC, Anderson RC. Modeling the biomechanics of the mandible: a three-dimensional finite element study. J Biomech. 1992;25(3):261–86.

    Article  Google Scholar 

  36. Vollmer D, Meyer U, Joos U, Vegh A, Piffkò J. Experimental and finite element study of a human mandible. J Craniomaxillofac Surg. 2000;28(2):91–6.

    Article  Google Scholar 

  37. Lee TS, Park S. Advantages of a beveled osteotomy on the zygomatic arch during reduction malarplasty. J Craniofac Surg. 2017;28(7):1847–8.

    Article  Google Scholar 

  38. Lotz J, Cheal E, Hayes WC: Fracture prediction for the proximal femur using finite element models: part I—linear analysis. 1991. https://doi.org/10.1115/1.2895412

  39. Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin J-C, Pujol S, Bauer C, Jennings D, Fennessy F, Sonka M. 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging. 2012;30(9):1323–41.

    Article  Google Scholar 

  40. Harada K, Watanabe M, Ohkura K, Enomoto S. Measure of bite force and occlusal contact area before and after bilateral sagittal split ramus osteotomy of the mandible using a new pressure-sensitive device: a preliminary report. J Oral Maxillofac Surg. 2000;58(4):370–3.

    Article  Google Scholar 

  41. Ellis E III, Throckmorton GS, Sinn DP. Bite forces before and after surgical correction of mandibular prognathism. J Oral Maxillofac Surg. 1996;54(2):176–81.

    Article  Google Scholar 

  42. Okiyama S, Ikebe K, Nokubi T. Association between masticatory performance and maximal occlusal force in young men. J Oral Rehabil. 2003;30(3):278–82.

    Article  Google Scholar 

  43. Sarkarat F, Ebrahimi S, Kahali R, Pirhadi Rad A, Khosravi M, Rakhshan V. Finite element simulation of displaced ZMC fracture after fixation with resorbable and non-resorbable one-point mini-plates and applying normal to severe occlusal loads. Trauma Monthly. 2019;24(3):1–7.

    Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

MQA: conceptualisation, data collection, data analysis, provision of study materials, experimentation, draft preparation, and writing and reviewing. HG: conceptualisation, data collection, data analysis, provision of study materials, experimentation, draft preparation, and writing and reviewing. LS: data analysis, reviewing and editing. YH: data analysis, reviewing and editing. WY: data analysis, reviewing and editing. JL: conceptualisation, project administration, reviewing and editing, supervision. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Jihua Li.

Ethics declarations

Ethics approval and consent to participate

The research committees at West China Hospital of Stomatology and Sichuan University approved this study (WCHS-CRSE-2022-103).

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Al-Watary, M.Q., Gao, H., Song, L. et al. Stability of different fixation methods after reduction malarplasty under average and maximum masticatory forces: a finite element analysis. BioMed Eng OnLine 22, 37 (2023). https://doi.org/10.1186/s12938-023-01098-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12938-023-01098-8

Keywords