Semivariogram and Semimadogram functions as descriptors for AMD diagnosis on SD-OCT topographic maps using Support Vector Machine

A Correction to this article was published on 08 November 2018

This article has been updated

Abstract

Background

Age-related macular degeneration (AMD) is a degenerative ocular disease that develops by the formation of drusen in the macula region leading to blindness. This condition can be detected automatically by automated image processing techniques applied in spectral domain optical coherence tomography (SD-OCT) volumes. The most common approach is the individualized analysis of each slice (B-Scan) of the SD-OCT volumes. However, it ends up losing the correlation between pixels of neighboring slices. The retina representation by topographic maps reveals the similarity of these structures with geographic relief maps, which can be represented by geostatistical descriptors. In this paper, we present a methodology based on geostatistical functions for the automatic diagnosis of AMD in SD-OCT.

Methods

The proposed methodology is based on the construction of a topographic map of the macular region. Over the topographic map, we compute geostatistical features using semivariogram and semimadogram functions as texture descriptors. The extracted descriptors are then used as input for a Support Vector Machine classifier.

Results

For training of the classifier and tests, a database composed of 384 OCT exams (269 volumes of eyes exhibiting AMD and 115 control volumes) with layers segmented and validated by specialists were used. The best classification model, validated with cross-validation k-fold, achieved an accuracy of 95.2% and an AUROC of 0.989.

Conclusion

The presented methodology exclusively uses geostatistical descriptors for the diagnosis of AMD in SD-OCT images of the macular region. The results are promising and the methodology is competitive considering previous results published in literature.

Background

Age-related macular degeneration (AMD) is a progressive chronic disease of the central retina and a leading cause of vision loss worldwide [1]. Its development is characterized by the appearance of soft drusen (discrete deposits of residues) and abnormalities in pigmentation of the retinal sublayer known as retinal pigment epithelium (RPE) [2]. Damage in this region causes the inability to read, to recognize faces, etc. Its progression can lead to vision impairment, or even total blindness.

It is estimated that AMD is the cause of 8.7% of cases of blindness worldwide, with a higher incidence in developed countries and people over 60 years of age. For 2020, the global projected number of cases is of 196 million and this estimation reaches 288 million for the year 2040 [3].

Recent advances in retinal imaging technology have allowed the handling of age-related macular degeneration through the creation of spectral domain optical coherence tomography (SD-OCT). It is a noncontact, noninvasive, three-dimensional imaging technique that obtains images representing a sectional view of retina at micrometer resolution [4].

This examination modality reconstructs the tomographic image by measuring the optical backscattering of tissues. This technique has gained space as a diagnostic tool for AMD, mainly due to its quickness to acquire images and dispenses the use of contrasts and thus the risks involved.

The evaluation of AMD by SD-OCT occurs through the analysis of successive sectional cuts in search of morphological changes caused by the disease. SD-OCT acquisition generates a large number of sectional images, which requires that the expert visualize and evaluate a large number of volume slices seeking for changes, especially if they are minimal and sparse in the early stages of pathology [5].

A typical SD-OCT examination produces about 100 images with a resolution of \(512 \times 1000\) pixels (fast acquisition protocol). This demands considerable human effort for evaluation. If we consider that a physician is responsible for the care and follow-up of dozens of patients, there is a necessity for technologies that support the clinical practice of SD-OCT image analysis.

Accordingly, the development of computer aided diagnostic methods (CADx) can collaborate with the specialist for the detection of AMD by taking advantage of the ability to evaluate the large amount of data that the SD-OCT examination can provide. Even in its variations, a large amount of anatomical retinal information is available to characterize the disease [6].

Typically, we have two main approaches for automated AMD diagnosis based on SD-OCT. One of the approaches involves the following steps to be effectively addressed: (a) obtaining relevant biomedical characteristics for the differentiation between healthy and compromised retinas; (b) use of classifiers to accurately determine the presence or absence of the disease; and (c) validation of the method for generating reliable results from a properly classified image database [4, 5, 7,8,9,10,11,12,13]. This approach has the advantage of constructing robust classifiers, based on fixed mathematical models that can characterize well a specific pathology.

Another approach is based on the use of deep neural networks techniques [14] to identify pathologies that compromise the retina [15,16,17,18,19]. These methods present the main advantage of flexibility for training to discriminate various conditions, especially AMD and diabetic macular edema, and also do not require the search for discriminatory characteristics which are determined by the deep learning framework.

In general, these methods use the reflectance values of retinal pixels. They use simple segmentation techniques based on the elimination of the background [8], a warping of the image from flattening the retina [4, 13, 17], a cut of predefined dimensions in the region of greatest intensity of pixels [9] or even semi-automatic segmentation of retinal layers [5] to delimit the region of interest.

The main related works are compared in Table 1. We can notice that in some works the images are pre-processed for the normalization of intensity values of pixels [7] and noise reduction. This filtering is normally done with filters based on wavelet [20] or with the Total Variation algorithm [8, 10]. The main advantage of these techniques is providing better definition of anatomic alterations that distinguish images compromised by AMD.

Table 1 Related work comparison

Most described methods are based on the individual analysis of the SD-OCT volumes’ B-Scans, not considering the correlation between successive slices as seen in [12, 13, 15, 21]. This type of application is common mainly due to being more accessible to define a region of interest from a B-Scan than from any other volume representation. Normally, the successive slices suffer a misalignment caused by eye movement during capture. Thus, it becomes more practical that each slice be processed individually.

The main disadvantage of methods that analyze individual B-scans is to disregard the correlation between pixels of successive slices. Moreover, after the step of detecting characteristics that indicate the presence of AMD per slice, a new classification is necessary to indicate the presence of the disease by volume.

The approach presented in [5] differs from others by representing the SD-OCT volume in thickness maps of total retina (TR), neurosensory retina (NSR), and especially of the retinal pigmented epithelium and drusen complex (RPEDC), which region has a higher incidence of AMD-induced alterations. With this, it is possible to generate a representation of a whole SD-OCT volume for the detection of AMD. Following, the topographic maps generated are redimensioned from \(1000 \times 100\) to \(1001 \times 1001\) pixels, reconstructing the original appearance of the retina, however at the expense of inserting interpolated data in the image.

From this point, they generate mean thickness maps and determine superior and inferior limits for the classification of retinas compromised or not by AMD. Despite this work presenting good results (AUROC = 0.991), we believe it is very dependent on the mean thickness measures and does not analyze how the local variation of thickness occurs along the RPEDC surface. So, we realize it is possible to propose a new approach for describing RPEDC conditions with local thickness variation analysis. Submitting these features to a classification method allows us to develop a methodology for the detection of AMD in human eyes.

Given the similarity of the generated retinal topographic maps with geological investigation maps like soil analysis and remote sensing [22], we propose using the geostatistical functions of semivariogram and semimadogram as descriptors for characterization of the macular region of the retina.

These geostatistical functions have already been used with success in biomedical contexts such as lung nodule detection [23] and breast cancer classification [24]. In these works, geostatistical functions are applied directly on the intensity values of pixels in regions of the image to differentiate between healthy tissue or tissue compromised by cancer. In [25], these functions are used to automatically detect eyes, differentiating them from other facial features. We can see that these methods well describe the texture of tissue or a topographic map, like proposed in this work.

The main advantage of using geostatistical functions over topographic maps is that they allow a representation of the volume as a whole, not limiting the analysis by individual slices (not being subject to the misalignment of slices problem) and providing a local analysis of thickness variation along the RPEDC in macular region.

In this work, we used the topographic map proposed by [5] but, innovate on not use the topographic map to describe only the thickness of the layers. Otherwise, we try to describe the global texture presented in the topographic map. The texture representation of the topographic map is the central contribution/innovation point of the proposed work.

To describe the macular region texture we propose the use of the semivariogram and semimadogram functions that we believe can describe the layers thickness variations of the total retina, NSR, e RPEDC. Thus, the complete methodology is innovative in itself, as it is the use of geostatistical functions to discriminate against the texture of the topographic map. This is a way to represent the retina globally.

Our approach is innovative in the sense that it uses the analysis of local variation of the thickness as a texture map, based on the assumption that druse formation modifies these measures. This gives us a way to evaluate thickness information with their spatial location building a robust method for AMD detection.

The occurrence of druses interferes in the retinal layers morphology, deforming them which reflects on generated topographic maps. Thus, our method of AMD diagnosis is based on the evaluation of the retinal layers thickness alterations. The topographic maps describe how their thickness is varying along the macular region. So it is an appropriate tool to be used in automatic methods. We developed a way to describe these variations based on the assumption that the texture representation of this maps will be a strong indication of healthy and unhealthy patients.

So, we believe this methodology may be useful to more robustly distinguish impairment cases and let us develop a computer-based method for automatic diagnosis of AMD impairment. Such a system can offer a second opinion, to provide more information to the specialist, supporting their clinical decision. Processing that data through machine learning methods can allow rapid studies, facilitating population screening, remote diagnosis and even early detection of the disease. The problem addressed in this research will be research on a methodology for the automatic diagnosis of AMD by using SD-OCT exams.

Methods

The methodology proposed in this work for the development of a CADx architecture is based on the following stages: (1) image acquisition, (2) image representation, (3) features extraction and (4) classification. The flow chart of the proposed methodology is shown in Fig. 1.

Fig. 1
figure 1

Flow chart of the proposed methodology. This picture shows a view of the necessary stages to perform AMD diagnosis over SD-OCT images

In the first step, the images are acquired from the database provided by [5] which contain precise markings of the borders of the total retina, RPEDC, and neurosensory retina. In the second step, we generate topographic maps and reconstruction en face of the SD-OCT volumes. The third step is the central part of the proposed methodology where we extract geostatistical features on the topographic maps obtained from SD-OCT volumes. Finally, in the fourth stage, we submit the extracted features to a classifier method to determine if an SD-OCT volume is whether or not affected by AMD. We detail each step below.

Image acquisition

The image set used for the validation of the proposed methodology belongs to the database provided by [5]. It is a subset of the AREDS2 [26] database containing 269 eye volumes with AMD and 115 control volumes (eyes considered normal). This data was collected for the purpose of studying quantitative indicators on the presence of AMD in adults.

The main population characteristics of the AREDS2 trials are: Mean age, 74 years; 57% female; 97% white; 7% current smokers; 19% with prior cardiovascular disease; and 44% and 50% taking statin-class cholesterol-lowering drugs and aspirin, respectively. Ocular characteristics include 59% with large bilateral drusen, 32% with advanced AMD in 1 eye and mean visual acuity of 20/32 in eyes without advanced AMD [26].

The SD-OCT examinations were captured at four clinics using SD-OCT scanners manufactured by Bioptigen, Inc (Research Triangle Park, NC). The volumes were acquired in the rectangular region of \(6.7 \,{\text {mm}} \times 6.7\) mm centered on the fovea with rapid scan protocol, resulting in volumes of size \(1000 \times 512 \times 100\) voxels.

The subset provided by [5] (also used in this work) has subjects aged between 50 and 85 years exhibiting intermediate AMD with druses greater than \(125\,\upmu\)m in both eyes or large drusen in one eligible eye and advanced AMD in the fellow eye, with no history of vitreoretinal surgery or ophthalmologic disease that might affect acuity in either eye. The control group selection used the same inclusion criteria in the AREDS2 study with the exception that the subject did not present any evidence of the presence of drusen.

This image set provided segmented retinal layer borders that limit the total retina (inner limiting membrane and Bruch’s membrane) that were segmented altogether with the border of the retinal pigment epithelium. This segmentation was initially performed automatically by the software DOCTRAP, developed by Duke University, and subjected to careful review and adjustments by certified specialists from Duke Advanced Research in Spectral Domain OCT Imaging Laboratory. We present a single B-Scan of an SD-OCT volume in Fig. 2.

Fig. 2
figure 2

Representation of SD-OCT layer’s boundaries marks. The image on the left represents the surfaces that determine the marking of the borders of the retina divisions. The right image, in turn, is a B-Scan of the same volume extracted from the image base provided by [5] with emphasis on the demarcation of the borders. In both images the borders that delimit the neurosensorial retina (NSR) and the retinal pigmented epithelium and drusen complex (RPEDC) are represented. The NSR is delineared by internal layer membrane (IML, colored red) and the pigmented epithelium border (PEB, colored blue). In turn, the retinal pigmented epithelium and drusen complex (RPEDC) is delineared by PEB and Bruch’s membrane (BM, colored green) including the drusenoids alterations. The total retina (TR) comprehend the whole region between IML and BM

After segmentation, the specialists manually marked the center point of the fovea. From this marking, they delimited a cylinder centered at this point as the area of interest of each volume. For tests and validation of this step, we utilized all 384 provided volumes and its layers boundaries marks.

In spite of the existence of automated segmentation techniques, at the moment, we use the segmentation provided with the base due to it being reviewed by specialists. This gives us greater security for validation of the proposed method since segmentation errors may impact its accuracy.

Image representation

In this step, we generated two-dimensional representations based on topographic maps and reconstruction en face for each SD-OCT 3D volume of the image set. Topographic maps represent the thickness of the retina (as well as the neurosensory layer and the RPEDC) for each point on its surface. The reconstruction en face generates images that correspond to the reflectance values of the SD-OCT on the border of the RPE. The image set used represents borders reviewed by specialists as surfaces \(B_i=(x, y)\), \(i\in \{1, 2, 3\}\). These three areas correspond respectively to the i limiting membrane (ILM), border of the pigment epithelium and the border of Bruch’s membrane (BM). These surfaces are used for the representations as detailed below.

Generation of topographic maps

From the surfaces defining the borders of the layers, we generated three two-dimensional surfaces \(L_i(x, y)\), where \(i\in \{1, 2, 3\}\), to represent each SD-OCT 3D volume. The coordinate \(z_i = L_i(x, y)\) corresponds to the axial depth of a point located in the coordinates (xy) in a substructure of the retina. This is done by summing the amount of voxels existing between the two borders, or simply calculating \(L_1 = B_3 - B_1\), \(L_2 = B_2 - B_1\) and \(L_3 = B_3 - B_2\).

The first topographic map generated (surface \(L_1\)) represents the total retina thickness, corresponding to the region between the border of the internal limiting membrane (ILM) and the border of Bruch’s membrane (BM). The second map (surface \(L_2\)) uses the neurosensorial retina (NSR), region between the ILM and the border of the pigment epithelium. The last topographic map (\(L_3\)) corresponds to the retinal pigment epithelium and drusen complex (RPEDC), which is the region between the BM and the border of the RPE. In order to better illustrate, Fig. 3 presents the generation of the topographic map for the surface \(L_1\).

Fig. 3
figure 3

Total retinal topographic map generation from a SD-OCT volume. Complete volume is represented in A. B and C presents successive elimination of unnecessary regions. B also demonstrates the delimitation of a ROI with radius \(r = 5\) mm centered in the marking of fovea marked by specialists. Finally, D presents the topographic map that represents the thickness of the retina for each point

En face reconstruction of RPE

The reconstruction en face of the border of the pigment epithelium generates a two-dimensional surface \(F_{RPE}\) representing the SD-OCT volume in which the reflectance values existing in the inner border of the pigment epithelium are associated to each pixel. Thus, an image is extracted bearing resemblance to a depiction of an eye fundus. However, each pixel presented corresponds to the voxels located on the border of the RPE, in accordance to:

$$\begin{aligned} F_{RPE} = R(L_2,x,y,z) \end{aligned}$$
(1)

where, R is the \(L_2\) surface reflectance for position x, y and z.

A distance parameter d can be specified to take into account 2D neighboring voxels located at the maximum distance of d up or down. Therefore, the representation en face is generated by the average values obtained from this set of voxels. The purpose of this representation by average is to render it more robust in relation to noise.

Figure 4 presents the reconstruction en face of the border of REP for a random volume with AMD and another considered normal. It is possible to visually perceive variations in the texture of the two images, where the darker regions correspond to points of lesser reflectance such as blood vessels and drusen formation.

Fig. 4
figure 4

Reconstruction en face of the border of the RPE. The figure shows the surface of the pigmented epithelium for a retina of the control group (a) and a retina with a diagnosis of AMD (b). The clearer values represent points of greater reflectance

Feature extraction

A key concept in geostatistical image analysis is the notion of spatial continuity, which represents the likelihood of a particular pixel value at a determined location given neighboring or regional data values [27]. Thus, it allows the study of the randomness of the sampled data in order to identify possible spatial structures. Events occurring in a particular location tend to have similar behavior in neighboring locations. This characterizes what we call spatial dependence.

The degree of spatial dependence (or variablity of the data) between samples (pixels, in this case) can be measured by semivariance. It analyzes the variance between pairs of data points over a range of distances. Its magnitude is directly proportional to the distance between the points, so the larger the distance, the larger the semivariance value.

Semivariogram

The semivariance values as a function of distance between pair of points is called semivariogram and can be written as [28]:

$$\begin{aligned} \gamma (h)= \frac{1}{2N(h)} \sum _{(i,j)|h_{ij}=h} (DN_i - DN_j)^2 \end{aligned}$$
(2)

where the parameter \(h_{i,j}\) called lag is the distance between the pairs of points located in i and j where \(DN_i\) and \(DN_j\) are respectively the values of pixels i and j. N(h) is the number of points separated by h.

The semivariogram usually displays a characteristic curve (Fig. 5) in which semivariance grows from small lags to large lags. At a given distance the semivariance becomes approximately constant. The plateau where \(\gamma\) stabilizes is called the sill.

Fig. 5
figure 5

Semivariogram parameters (left) and the characteristic curve (right)

The distance at which the sill is reached is called range. At this point, the semivariance is equal to the variance of the data of the sample considered, meaning that there is no relation between the pairs of observations considered at this distance [23]. On the other hand, if the distance h is very small, the compared points are very similar to each other, which implies very low values of semivariance for these initial distances, hence the value of the semivariogram will start with very low values \(\gamma > 0\). This value, as well as the range and the sill, also represents a characteristic of the semivariogram called the nugget effect, or simply nugget.

Some parameters are used for the calculation of the semivariogram: the direction, the measurement (or lag distance), the maximum bandwidth, the lag tolerance and the angular tolerance. Setting these values limits which pixels will be considered for the calculation. Figure 5 show an illustration of the boundaries applied by these parameters to the pairs of pixels.

Semimadograma

Similar to the semivariogram, the semimadogram is another function also used to measure spatial dependence [23]. It is given by the mean of the absolute difference measured in the sample pairs as a function of distance and direction. The function is defined by:

$$\begin{aligned} m(h)= \frac{1}{2N(h)} \sum _{i=1}^{N(h)}|x_i - y_i| \end{aligned}$$
(3)

where h is the lag, \(x_i\) and \(y_i\) are the values of pixels of an image separated by h. N(h) is the number of pairs from the distance h.

To represent the instances in feature vectors, we calculated semivariogram and semimadogram over the topographic maps of RT(\(L_1\)), NSR(\(L_2\)) and RPEDC(\(L_3\)). These functions were also calculated on the reconstruction en face of the border of the RPE. The calculations were performed in four directions (\(0^\circ , 45^\circ , 90^\circ\) and \(135^\circ\)) that are commonly used in computational image processing. The values of lag = \(h \in \{1, 2, 3, ..., 15\}\), the tolerance for lag \(\Delta h = \,\pm 3\), tolerance for the direction angle was equal to \(10^\circ\) and the maximum bandwidth of 3 pixels. The feature vectors are used as inputs for the classification step.

Classification

To classify between AMD and non-AMD cases, we have used the Support Vector Machines (SVM) [29] which is a supervised learning method for data classification and pattern recognition. Its principle is based on separating data in a sample space from training on a set of previously labeled data. The separation of these data is done through hyperplanes that segment the sample space into partitions, thus separating the instances into distinct classes [30].

SVM is a proven effective method for general classification tasks. In spite of other classifiers that require large amounts of data, such as deep learning [14], SVM is a very robust classifier for tests with a few amount of samples (384 in our case).

For evaluating the performance of the classifier, we used the measures of sensitivity, specificity, accuracy, AUROC and Cohen’s kappa [31]. Sensitivity is defined by \(TP/(TP + FN)\), specificity is defined by \(TN/(TN + FP)\), and accuracy is defined by \((TP + TN)/(TP + TN + FP + FN)\), where TN, is true-negative, FN is false-negative, FP is false-positive, and TP is true-positive.

Results

To evaluate the performance of the proposed methodology, we run an experiment with the 383 SD-OCT volumes provided by [5]. The missing volume was not used due to a smaller number of B-Scans it contains in comparison with the others (80 B-Scans versus 100 in the other volumes). Testing was performed on a computer with an Intel Core i7 processor with 8 GB of RAM. The implementation was fully developed in the Python 3.5 language in a Debian Linux environment.

The semivariogram and semimadogram for different topographic maps were calculated by generating eight different arrangements of maps/geostatistical functions to be evaluated for the ability to distinguish examination affected by AMD from those considered healthy. The arrangements are: (1) total retina (semivariogram), (2) total retina (semimadogram), (3) neurosensorial retina (semivariogram), (4) neurosensorial retina (semimadogram), (5) RPEDC (semivariogram), (6) RPEDC (semivariogram), (7) RPEDC en face (semivariogram) and (8) RPEDC en face (semimadogram).

For each arrangement, the geostatistical functions provide values that show local spatial dependency of the chosen maps. In Fig. 6, we show the output given by the geostatistical functions for two cases diagnosed with AMD and two control cases. This data is then submitted to the SVM classifier permitting the realization of tests and validation.

Fig. 6
figure 6

Semivariogram and semimadogram response for RPEDC Maps. Rows A and B correspond to exams afflicted by AMD, while C and D present control volumes. The first image of each row corresponds to the volume’s representation through RPEDC’s topographic map and the last two columns respectively show the semivariogram and semimadogram functions in the \(45^{\circ }\) direction

The validation tests were performed using the k-fold technique with repetitions. For training, we divided the data into 5 folds with 100 repetitions, setting a ratio of 80% for training and 20% for tests in each interaction. Using a large number of repetitions increases the reliability of the results given the fact that it decreases the contribution of folds that present overfitting to the overall result. We calculated the mean values, standard deviations for each test (Table 2).

As can be seen in Table 2, we describe that both the semivariogram and semimadogram when calculated on the topographic map of the total retina have low differentiation capacity between the two classes. The average AUROC found for these tests are, respectively, 0.576 and 0.756.

When using the topographic map of the neurosensory retina, we obtained better results in terms of sensitivity, specificity and AUROC. The best results for this map were obtained by the semimadogram function with average sensitivity of 90.0%, specificity mean of 84.3%, accuracy mean of 88.3%, and AUROC mean of 0.950. The kappa index of 0.724 indicates a substantial agreement.

We obtained the best results with topographic maps of the RPEDC, especially with semivariogram function. The average accuracy of \(95.2\%\) shows that this model has an excellent diagnostic differentiation capability for AMD, presenting an average AUROC of 0.989.

On the other hand, the semivariogram an semimadogram functions applied on RPE’s reconstruction en face, which considered the reflectance of the points in the upper border of this layer, did not demonstrate expressive ability to differentiate the control tests from examinations affected by AMD (average AUROC of 0.505 and 0.778, respectively).

For a visual comparison, in Fig. 7 we present the semivariograms and semimadograms responses calculated for the topographic maps. These were calculated for direction of \(45^\circ\) and a volume with AMD and a control one are directly compared.

Table 2 Results of AMD classification based on geostatistical features
Fig. 7
figure 7

Semivariogram and semimadogram functions plots for \(45^\circ\) of the retinal layers. The vertical scale is different for each layer

We further compare the performance of our best method (semivariogram over RPEDC) to the classical texture descriptors (Haralick) [32, 33] and with the Local Binary Patterns (LBP) [33] to evaluate the performance of our methods in comparison to the commonly used descriptors that were previously mentioned (Fig. 8).

The results of these experiments (Table 3) show that the commonly used descriptors possess lower values in comparison to the semivariogram descriptor. Specifically, the LBP and Haralick descriptor methods achieved AUROCs of 0.968 and 0.862 respectively. These values are lower than the best result presented in this work (arrangement 5: AUROC = 0.989). Table 3 presents the comparison between these results and Fig. 9 shows a graphical comparison of the different AUROC values.

Fig. 8
figure 8

ROC plots of each generated SVM model. The best obtained value was 0.989

Fig. 9
figure 9

ROC plot for comparison of semivariogram peformance with another methods. The best obtained value was 0.989

Table 3 Comparison with others techniques over same data

Discussion

Table 2 describes that both the semivariogram and the semimadogram when calculated on the topographic map of the total retina have low differentiation capacity for AMD detection. This was also demonstrated in [5] by using thickness average maps. Specifically in our work, this result indicates that there is no significant difference in terms of spatial dependence between classes when considering the total retina.

When using the topographic map of the neurosensory retina, the kappa index of 0.724 indicates a substantial agreement. Notwithstanding, these values are still slightly below the results found in the literature. The main difference between the results obtained by the maps of the total retina and the neurosensory retina is that the last map is calculated from the upper border of the RPE, a layer that is directly deformed from the accumulation of drusen.

The best results (semivariogram on the RPEDC) achieved an AUROC of 0.989, a result very close to that achieved by [5]. The value of \(kappa = 0.886\) also indicates that despite being validated with an unbalanced sample, the model’s performance possesses a high concordance between classes.

On the other hand, the semivariogram applied on the reconstruction en face of the RPE, which solely considered the reflectance of points in the upper border of this layer, did not demonstrate expressive ability to differentiate the control tests of the examinations affected by AMD. Especially the semivariogram, which resulted in an average AUROC of 0.505, which is a result too similar to a random classifier. A slightly better AUROC (0.778) was achieved when using the semimadogram for this reconstruction. However, a lower kappa value (0.347) indicates that the generated model is biased and tends to present a high number of mismatched classifications.

The poorest performance of this last test against the promising results obtained with the topographic map of the RPEDC is due to the fact that the AMD does not cause significant changes in the reflectance of points on the border of the RPE as with the map of thickness. These variations are best observed when evaluating the deformations caused in this layer by the appearance of drusen and possibly choroidal neovascularization. In future work, we can take into consideration not only the border, but also an axial integration of the pixels of the RPE and thus identify whether reflectance changes in this layer are significant for the detection of AMD.

In Fig. 7 we present a visual comparison between the plotting of semivariogram and semimadogram functions relating to various tests performed. Therefrom, we can perceive that the semivariogram function produces very close values for the classes when applied on the map of the total retina and reconstruction en face of the RPE. The values, in turn, become well differentiated when applied on the topographic map of RPEDC, congruent with the best generated model.

Considering the results achieved, we verified that the best model generated for the diagnosis of AMD using topographic maps is arrangement 5 (RPEDC/Semivariogram). We noticed that the performance of the proposed model is greater than various works that are based in reflectance values. A comparison with related works is presented in Table 4.

Table 4 Comparison of performance of the main methodologies presented in related works

Some works have presented slightly higher accuracy, however they use smaller image sets for validation, and other ones have used individualized B-Scans. They do not perform an overall evaluation of the retina, which makes it difficult to compare the work directly. It is also important to note that the performance information is not presented in some papers.

Among the works that make use of the same SD-OCT volumes base [5, 11] along with the tests that were made with the Haralick and LBP techniques, we show through the results achieved with the semivariogram over RPEDC’s topographic map that the proposed method is robust for the detection of AMD.

With the direct comparison between AUROC values, we can see that the results are really promising. The work of [5] (which presents better AUROC) is based on the mean thickness of the retinal layers, which may vary slightly depending on the AMD stage and it does not consider the local variation of these values. We believe that our method has the advantage of acting on the pattern of the distribution of anatomical changes caused by AMD.

Haralick descriptors also do not make use of a spatial distribution parameter in their analysis. Lastly, although LBP does consider the data’s spatial localization, it was not able to adequately discriminate volumes with compromised retinas, which brings us to consider the use of a some type of pre-processing for this technique.

Also, we intend to investigate this in the future with other databases that are available, or with a base generated in conjunction with a local clinic.

We believe that the proposed method’s results, which is based on the local thickness variation, are promising. Although the AUROC found is slightly smaller than [5], it must be emphasized that the used base is small and there also exists the necessity of testing our method with bases that present cases of early AMD or that contain smaller differences in RPEDC thickness.

Conclusions

In this article, we proposed a methodology for automated diagnostic of AMD in SD-OCT images using geoestatistical semivariogram and semimadogram functions as descriptors. These descriptors were applied to the topographic map of the total retina as well as sublayers (neurosensory and retinal pigment epithelium) of the macular region together with the reconstruction en face of the pigment epithelium surface. The data was submitted to a Suport Vector Machine classifier after the feature extraction.

We exclusively applied geostatistical texture descriptors for the detection of AMD in optical coherence tomography images base. This is done by a texture classification of the topographic map of human retina layers and this method can be used as a disease screening tool or follow-up in studies of new treatments and prevention methods.

Considering that the main objective of this work was to demonstrate that these descriptors are effective for the diagnosis of AMD, we believe that we have achieved this goal with assuredly promising results. The best performed model presented an AUROC of 0.989. The validation of this model demonstrates that the proposed methodology can distinguish with excellent accuracy AMD from unafflicted eyes. Nevertheless, it is advisable to test with a larger database with larger varieties of thickness and even try the method with other pathologies that may affect the retina.

Change history

  • 08 November 2018

    After publication, it was highlighted that the original publication [1] contained a spelling mistake in the first name of Marcelo Gattas. This was incorrectly captured as Marelo Gattass in the original article which has since been updated.

References

  1. Lim LS, Mitchell P, Seddon JM, Holz FG, Wong TY. Age-related macular degeneration. Lancet. 2012;379(9827):1728–38. https://doi.org/10.1016/S0140-6736(12)60282-7.

    Article  Google Scholar 

  2. Bordon AF, Oshima A, Guia TA, Calucci D, Sallum JMF, Farah ME. Avaliação da tomografia de coerência óptica em pacientes portadores de degeneração macular relacionada à idade tratada com terapia fotodinâmica com verteporfina. Arquivos Brasileiros de Oftalmologia. 2008;71(6):799–804. https://doi.org/10.1590/S0004-27492008000600006.

    Article  Google Scholar 

  3. Wong WL, Su X, Li X, Cheung CMG, Klein R, Cheng C-Y, Wong TY. Global prevalence of age-related macular degeneration and disease burden projection for 2020 and 2040: a systematic review and meta-analysis. Lancet Glob Health. 2014;2(2):106–16. https://doi.org/10.1016/S2214-109X(13)70145-1.

    Article  Google Scholar 

  4. Liu Y-Y, Ishikawa H, Chen M, Wollstein G, Duker JS, Fujimoto JG, Schuman JS, Rehg JM. Computerized macular pathology diagnosis in spectral domain optical coherence tomography scans based on multiscale texture and shape features. Invest Ophthalmol Vis Sci. 2011;52(11):8316–22.

    Article  Google Scholar 

  5. Farsiu S, Chiu SJ, O’Connell RV, Folgar FA, Yuan E, Izatt JA, Toth CA. Quantitative classification of eyes with and without intermediate age-related macular degeneration using optical coherence tomography. Ophthalmology. 2014;121(1):162–72. https://doi.org/10.1016/j.ophtha.2013.07.013.

    Article  Google Scholar 

  6. Fleckenstein M, Issa PC, Helb H-M, Schmitz-Valckenberg S, Finger RP, Scholl HP, Loeffler KU, Holz FG. High-resolution spectral domain-oct imaging in geographic atrophy associated with age-related macular degeneration. Invest Ophthalmol Vis Sci. 2008;49(9):4137–44. https://doi.org/10.1167/iovs.08-1967.

    Article  Google Scholar 

  7. Serrano-Aguilar P, Abreu R, Antón-Canalís L, Guerra-Artal C, Ramallo-Fariña Y, Gómez-Ulla F, Nadal J. Development and validation of a computer-aided diagnostic tool to screen for age-related macular degeneration by optical coherence tomography. Br J Ophthalmol. 2012;96(4):503–7. https://doi.org/10.1136/bjophthalmol-2011-300660.

    Article  Google Scholar 

  8. Albarrak A, Coenen F, Zheng Y, et al. Age-related macular degeneration identification in volumetric optical coherence tomography using decomposition and local feature extraction. In: Proceedings of 2013 international conference on medical image, understanding and analysis, 2013. p. 59–64.

  9. Srinivasan PP, Kim LA, Mettu PS, Cousins SW, Comer GM, Izatt JA, Farsiu S. Fully automated detection of diabetic macular edema and dry age-related macular degeneration from optical coherence tomography images. Biomed Opt Expr. 2014;5(10):3568–77. https://doi.org/10.1364/BOE.5.003568.

    Article  Google Scholar 

  10. Zhang Y, Zhang B, Coenen F, Xiao J, Lu W. One-class kernel subspace ensemble for medical image classification. EURASIP J Adv Sign Process. 2014;2014(1):17. https://doi.org/10.1186/1687-6180-2014-17.

    Article  Google Scholar 

  11. Venhuizen FG, van Ginneken B, Bloemen B, van Grinsven MJ, Philipsen R, Hoyng C, Theelen T, Sánchez, CI. Automated age-related macular degeneration classification in oct using unsupervised feature learning. In: Medical Imaging 2015: computer-aided diagnosis, 2015; 9414: p. 94141 . https://doi.org/10.1117/12.2081521. International Society for Optics and Photonics

  12. Wang Y, Zhang Y, Yao Z, Zhao R, Zhou F. Machine learning based detection of age-related macular degeneration (amd) and diabetic macular edema (dme) from optical coherence tomography (oct) images. Biomed Opt Expr. 2016;7(12):4928–40. https://doi.org/10.1364/BOE.7.004928.

    Article  Google Scholar 

  13. Sun Y, Li S, Sun Z. Fully automated macular pathology detection in retina optical coherence tomography images using sparse coding and dictionary learning. J Biomed Opt. 2017;22(1):016012. https://doi.org/10.1117/1.JBO.22.1.016012.

    Article  MathSciNet  Google Scholar 

  14. LeCun Y, Bengio Y, Hinton G. Deep learning. Nature. 2015;521(7553):436–44. https://doi.org/10.1038/nature14539.

    Article  Google Scholar 

  15. Ravenscroft D, Deng J, Xie X, Terry L, Margrain T.H, North R.V, Wood A. Learning feature extractors for amd classification in oct using convolutional neural networks. In: Signal Processing Conference (EUSIPCO), 2017 25th European. https://doi.org/10.23919/EUSIPCO.2017.8081167. New York: IEEE; 2017. p. 51–5.

  16. Fang L, Wang C, Li S, Yan J, Chen X, Rabbani H. Automatic classification of retinal three-dimensional optical coherence tomography images using principal component analysis network with composite kernels. J Biomed Opt. 2017;22(11):116011. https://doi.org/10.1117/1.JBO.22.11.116011.

    Article  Google Scholar 

  17. Karri SPK, Chakraborty D, Chatterjee J. Transfer learning based classification of optical coherence tomography images with diabetic macular edema and dry age-related macular degeneration. Biomed Opt Expr. 2017;8(2):579–92. https://doi.org/10.1364/BOE.8.000579.

    Article  Google Scholar 

  18. Lee CS, Baughman DM, Lee AY. Deep learning is effective for classifying normal versus age-related macular degeneration oct images. Ophthalmol Retina. 2017;1(4):322–7. https://doi.org/10.1016/j.oret.2016.12.009.

    Article  Google Scholar 

  19. Kermany DS, Goldbaum M, Cai W, Valentim CC, Liang H, Baxter SL, McKeown A, Yang G, Wu X, Yan F, et al. Identifying medical diagnoses and treatable diseases by image-based deep learning. Cell. 2018;172(5):1122–31. https://doi.org/10.1016/j.cell.2018.02.010.

    Article  Google Scholar 

  20. Srinivasan PP, Heflin SJ, Izatt JA, Arshavsky VY, Farsiu S. Automatic segmentation of up to ten layer boundaries in sd-oct images of the mouse retina with and without missing layers due to pathology. Biomed Opt Expr. 2014;5(2):348–65. https://doi.org/10.1364/BOE.5.000348.

    Article  Google Scholar 

  21. Ravenscroft D, Deng J, Xie X, Terry L, Margrain T.H, North R.V, Wood A. Amd classification in choroidal oct using hierarchical texton mining. In: International conference on advanced concepts for intelligent vision systems. https://doi.org/10.1007/978-3-319-70353-4_21. New York: Springer; 2017. p. 237–48.

    Chapter  Google Scholar 

  22. Sen Z. Cumulative semivariogram models of regionalized variables. Mathemat Geol. 1989;21(8):891–903. https://doi.org/10.1007/BF00894454.

    Article  Google Scholar 

  23. Silva AC, Carvalho PCP, Gattass M. Diagnosis of lung nodule using semivariogram and geometric measures in computerized tomography images. Comput Methods Programs Biomed. 2005;79(1):31–8. https://doi.org/10.1016/j.cmpb.2004.12.008.

    Article  Google Scholar 

  24. da Silva V.R, de Paiva A.C, Silva A.C, de Oliveira A.C.M. Semivariogram applied for classification of benign and malignant tissues in mammography. In: International conference image analysis and recognition. https://doi.org/10.1007/11867661_51. New York: Springer; 2006. p. 570–9.

    Google Scholar 

  25. Almeida JDS, Paiva AC, Silva AC. Automatic eye detection using semivariogram function and support vector machine. In: 17th International conference on systems, signals and image processing —IWSSIP, Rio de Janeiro - RJ, 2010. p. 174–7.

  26. Chew EY, Clemons T, SanGiovanni JP, Danis R, Domalpally A, McBee W, Sperduto R, Ferris FL. The age-related eye disease study 2 (AREDS2): study design and baseline characteristics (AREDS2 report number 1). Ophthalmology. 2012;119(11):2282–9. https://doi.org/10.1016/j.ophtha.2012.05.027.

    Article  Google Scholar 

  27. Schowengerdt RA. Remote sensing: models and methods for image processing. Burlington: Elsevier Science; 2006.

    Google Scholar 

  28. Isaaks EH, Srivastava RM. Applied geostatistics vol. 551.72 ISA. New York: Oxford University Press; 1989.

    Google Scholar 

  29. Vapnik VN, Vapnik V. Statistical learning theory, vol. 1. New York: Wiley; 1998.

    MATH  Google Scholar 

  30. Wang L. Support vector machines: theory and applications (Studies in Fuzziness and Soft Computing). Berlin: Springer; 2005. https://doi.org/10.1007/b95439.

    MATH  Google Scholar 

  31. Blackman NJ-M, Koval JJ. Interval estimation for cohen’s kappa as a measure of agreement. Stat Med. 2000;19(5):723–41.

    Article  Google Scholar 

  32. Gonzalez RC, Woods RE. Digital image processing. 3rd ed. Upper Saddle River: Prentice-Hall Inc; 2006.

    Google Scholar 

  33. Coelho LP. Mahotas: open source software for scriptable computer vision. J Open Res Softw. 2013. https://doi.org/10.5334/jors.ac.

    Article  Google Scholar 

Download references

Authors' contributions

AMS: responsible for the design and implementation ofresearch. APMS: responsible medical background, JDSA and ACS: geostatistical feature extraction. DLG and GBJr: responsible for data classification. AMS and SATM wrote the main body of the manuscript. ACP and MG supervised the work. All authors read and approved the final manuscript.

Acknowledgements

The authors thank IFMA, UFMA, CAPES, CNPQ and FAPEMA for the support to carry out this research.

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

The dataset used for the validation of the proposed methodology was made publicy available by Farsiu [5] and can be downloaded at the address http://people.duke.edu/~sf59/. Programming scripts used in this study are available from the corresponding author on reasonable request.

Consent for publication

All the authors have given consent for publication of this paper.

Ethics approval and consent to participate

Not applicable.

Funding

Not applicable.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Alex M. Santos.

Additional information

The original version of this article was revised as it was highlighted that the original article contained a spelling mistake in the name of Marcelo Gattass.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article