Free-living core body temperature monitoring using a wrist-worn sensor after COVID-19 booster vaccination: a pilot study
BioMedical Engineering OnLine volume 22, Article number: 25 (2023)
Abstract
Core body temperature (CBT) is a key vital sign and fever is an important indicator of disease. In the past decade, there has been growing interest for vital sign monitoring technology that may be embedded in wearable devices, and the COVID-19 pandemic has highlighted the need for remote patient monitoring systems. While wrist-worn sensors allow continuous assessment of heart rate and oxygen saturation, reliable measurement of CBT at the wrist remains challenging. In this study, CBT was measured continuously in a free-living setting using a novel technology worn at the wrist and compared to reference core body temperature measurements, i.e., CBT values acquired with an ingestible temperature-sensing pill. Fifty individuals who received the COVID-19 booster vaccination were included. The datasets of 33 individuals were used to develop the CBT prediction algorithm, and the algorithm was then validated on the datasets of 17 participants. Mean observation time was 26.4 h and CBT > 38.0 °C occurred in 66% of the participants. CBT predicted by the wrist-worn sensor showed good correlation to the reference CBT (r = 0.72). Bland–Altman statistics showed an average bias of 0.11 °C of CBT predicted by the wrist-worn device compared to reference CBT, and limits of agreement were − 0.67 to + 0.93 °C, which is comparable to the bias and limits of agreement of commonly used tympanic membrane thermometers. The small size of the components needed for this technology would allow its integration into a variety of wearable monitoring systems assessing other vital signs and at the same time allowing maximal freedom of movement to the user.
Introduction
Core body temperature (CBT) is an important vital sign (VS) and fever is an important indicator of disease [1]. In humans, CBT is closely controlled around its normal value of 36.5 °C with 24-h variations of only 1.5 °C. Fever represents a controlled deviation of the host from the otherwise precisely maintained temperature homeostasis and is a complex adaptive response to several immune challenges whether infectious or non-infectious [2]. “Fever” is also the single most frequently reported manifestation of coronavirus disease 2019 (COVID-19) and has been reported to be one of the earliest signs of symptomatic COVID-19 [3, 4]. Therefore, CBT monitoring plays an important role both in the surveillance of infected individuals and in screening programs.
The gold standard for the determination of CBT is pulmonary artery catheter measurement, but this method is invasive and time-consuming. Other techniques to determine CBT are measurement of rectal, bladder, and oesophageal temperature, which are also invasive and not practicable in many settings. Ingestible telemetric temperature sensors provide CBT measurements in good agreement with oesophageal or rectal thermometers and are often used for outpatient field-based studies [5]. However, these telemetric pills may be excreted as soon as 8 h after ingestion, limiting their use for CBT monitoring in clinical settings. Peripheral temperature measurements are less precise and reliable than CBT measurements but easier to assess. Many methods have been developed to approximate CBT as closely as possible, such as tympanic, axillary, and oral thermometers [6, 7]. Although these techniques allow obtaining an estimation of body temperature easily, the need of close contact of the operator to the potentially infectious individual still limits their use in the context of highly transmittable diseases, such as COVID-19. Commercially available dermal thermometers designed for home use are applied to the individual’s skin with the help of adhesive tape and allow remote and continuous monitoring of CBT. However, such devices often lack precision and thus are not considered to be suitable for clinical use [8]. Methods that continuously record CBT using heat flux provide a reliable measurement of CBT (3 M™ Bair Hugger™ temperature monitoring system [9] (formerly called 3 M SpotOn [10, 11]), Drägerwerk™ Tcore™ temperature monitoring system [12, 13], Medisim™ Temple Touch Pro™ [14]). However, in these systems, the sensors are applied on the subjects’ forehead and permanent electrical supply is needed, thus they are not practical for use in patients outside the operating room (OR) or the intensive care unit (ICU). Recently, a novel technology (CALERA®, greenTEG®, Switzerland [15]) was shown to accurately monitor CBT in hospitalized stroke patients while worn on the chest [16].
In the past decade, there has been a growing interest in the development of remote patient monitoring (RPM) systems and sensors of vital functions embedded in wearable devices [17,18,19,20]. The COVID-19 pandemic has further raised the interest for affordable and reliable health technology that allows RPM and that may be used for large-scale screening programs [21,22,23,24,25]. While wrist-worn sensors allow continuous assessment of VS such as heart rate and oxygen saturation with good accuracy, reliable measurement of CBT at the wrist remains challenging [12, 26].
The present study was conducted to evaluate CBT monitoring and detection of elevated CBT at the wrist using a novel device based on the CALERA® technology. Elevated CBT is reported to occur in 20–37% of individuals within two days after administration of the second or third dose of an mRNA-based COVID-19 vaccine [27, 28]. Thus, CBT was measured continuously with a wrist-worn sensor and compared to CBT values acquired with an ingestible telemetric pill in healthy volunteers during 24 h after administration of a COVID-19 booster vaccination.
Results
Data construction
Sixty-one participants were included in the study. Eleven participants were excluded because of technical issues (e.g., sensor failures) or human errors (e.g., forgetting to turn on the senor) that rendered the measurements invalid. Fifty participants completed the entire study protocol. The dataset was split into two groups with allocation of participants performed to achieve similar distribution of elevated CBT proportion in both groups. The datasets of group one (33 participants) were used for model development and training. The datasets of group two (17 participants) were used for model validation (Fig. 1).
All participants received the mRNA-1273 SARS-CoV-2 vaccine (Moderna™ Spikevax™). The participants reported low physical activity level during the observation time: nearly all participants reported staying at home and resting. Few participants reported low-intensity physical activity, such as household duties (e.g., cleaning or cooking), yoga or outdoor walking. A minority of participants undertook grocery shopping, commuting by bike or home-office working. No adverse event occurred during the study. Important participants’ characteristics are listed in Table 1.
Prediction performance
In group two, the prediction mean absolute error (MAE) was 0.34 °C (standard deviation, SD 0.12). The Bland–Altman analysis shows good agreement between predicted CBT and reference CBT, with a bias of 0.11 °C (SD 0.23) (Fig. 2). The upper limit of agreement (LoA) was 0.93 °C and the lower LoA was − 0.67 °C. In Fig. 3, predicted CBT was plotted against reference CBT for each measurement. The correlation coefficient of the two measurement methods was r = 0.72 (SD 0.2). Overall, fever detection performance resulted in a false positive rate (FPR) of 14.9% and false negative rate (FNR) of 23.6%. Figure 4 shows a participant’s predicted and reference CBT over the entire observation time.
Discussion
To the best of our knowledge, this is the first study to investigate continuous CBT monitoring and detection of elevated CBT with a wrist-worn device in free-living individuals. The technology had previously been shown to have high accuracy in detection of elevated CBT in the context of physical activity (integrated in the greenTEG® CORE® device) [29], as well as in a clinical context when placed on the chest [16]. There is no universally accepted definition of “fever” or for the upper limit of normal CBT [30]. For this study, we defined “CBT elevation” as elevation of CBT > 38.0 °C, a cut-off often used in clinical practice [1, 31]. Approximately 66% of the participants developed elevated CBT in the 24 h after the COVID-19 vaccination, which is slightly higher than previous observations [27, 28]. The precision of the wrist-worn CBT prediction system was comparable to that of other thermometers routinely used. A meta-analysis of 32 studies found similar mean bias and LoA compared to centrally measured CBT for tympanic membrane thermometers and axillary thermometers (− 0.08 °C [− 1.42 to 1.26 °C], and − 0.33 °C [− 0.94 to 0.27 °C], respectively) [7]. The false positive and false negative rates for the detection of elevated CBT (14.9% and 23.6%, respectively) are explained by the algorithm’s intrinsic tendency to average values and by the fact that it had previously been trained in normal CBT. Further training of the algorithm in the high CBT range is expected to further increase its accuracy.
Real-time CBT monitoring with a device allowing users to follow their daily routine is promising regarding several aspects. During the COVID-19 pandemic, RPM programs with VS and symptoms assessed several times a day and reported to health care professional in telemedicine consultations have shown to avoid short-term hospital admissions [23] and to allow early hospital discharge of infected patients [24, 32]. However, patients were provided with a large amount of technical equipment (blood pressure monitors, pulse oximeter, thermometer), and VS measurements as well as reporting of the results had to be performed manually. Gruwez et al. describe poor patient compliance after day 5 of the RPM program [25]. Furthermore, the repeated measurements and website or webcall-based reporting reduced the access to these programs for people with poor digital literacy. Thus, typically the elderly, who most likely would have benefited from monitoring programs, often were excluded from the RPM programs. The development of automated “all-in-one” devices that include a CBT sensor, may help to facilitate such initiatives by making VS assessment easier and improving compliance.
Global warming has increased attention about the limits of human adaptability to high environmental temperatures and has raised concern about vulnerable individuals’ health during the increasingly frequent heat waves [33,34,35,36]. Wearable CBT monitoring technology could be helpful to monitor persons at risk for heat-related illness during such episodes and thus to support the development of early detection programs.
Traditionally, CBT values have been interpreted dichotomously: patients either have a fever or are afebrile. Although used as indicator of disease since antiquity, “fever” has no universally accepted definition today [30]. The International Society of Physiological Studies (IUPS) defines fever as “a state of elevated core temperature” [37]. However, the meaning of “a state of elevated core temperature” is still debated in the clinical context [30]. Since the seminal works of Wunderlich in the nineteenth century [38], many efforts have been undertaken to define “normal” human body temperature, its deviations and the relationship between disease and temperature [1, 2, 31], and it has been demonstrated that body temperature is influenced by many factors, most importantly age and site of measurement [1]. Attempts have also been made to distinguish fever patterns and their significance, but none of these approaches was accurate enough to support clinical decisions [39]. However, recent evidence suggests that CBT pattern analysis can provide valuable clinical information, regardless of whether the patients meet fever threshold criteria, such as prediction of sepsis development [40], adverse events in immunocompromised hosts [41] and discrimination of bacterial vs. other cause of fever in patients hospitalized for suspected bacterial infection [42]. Continuous temperature monitoring is routinely used for perioperative temperature management [43, 44], and non-invasive zero-heat-flux and double sensor technologies have been studied for application in this context [9, 45, 46]. However, the placement of these devices’ probes on patients’ foreheads makes them unsuitable for use on general care wards. Ward monitor systems should use small, wireless and wearable sensors and be easy to use, giving hospitalized patients the freedom to move within their rooms and the health-care facility [47,48,49]. The technology tested in the present study has been shown to be a reliable alternative to tympanic membrane thermometers in hospitalized patients [16]. Evidence exists that automated monitoring of VS such as heart rate, oxygen saturation and blood pressure in patients hospitalized on the general care ward can improve patient outcome compared to intermittently measured VS [50,51,52]. Thus, there is potential utility of a small, wireless, and power-autonomous CBT sensor in the clinical context, especially if integrated into monitoring systems that also assess other VS.
We identified some limitations in our studies. All data collected included temperature patterns below 39 °C. The relatively mild CBT elevation induced by the second or third COVID-19 vaccine doses did not allow training and validation of the algorithm in higher CBT ranges. Second, the behavior and activity of most free-living individuals lacked variety. Future investigations should consider a variety of active daily life scenarios. An in-depth analysis of effects from environmental and contextual confounders is needed to better clarify the sensor’s ability of tracking fever patterns in outdoor free-living. Third, the relatively young age and high motivation of the study population recruited in a special study setting hinders extrapolation to the adherence to the device of other populations, such as elderly or ill patients.
Conclusion
Continuous CBT monitoring and detection of elevated CBT with a wrist-worn sensor is reliable and accurate in free-living individuals. Continuous CBT monitoring has the potential to improve the surveillance of both patients included in remote monitoring programs as well as patients hospitalized on general care wards.
Methods
Participants and experimental protocol
The study was conducted over a 7-month period between September 2021 and March 2022 at the University Hospital Basel, Switzerland. Information about the study was provided to individuals presenting to the hospital’s employees’ COVID-19 vaccination center. Interested individuals were assessed for eligibility. Eligible participants were defined as those aged between 18 and 60 years and who received their second or third COVID-19 vaccination at the day of inclusion. Exclusion criteria were: inability to sign consent, inability to swallow pills, history of major gastro-intestinal surgery, ≤ 40 kg body weight, scheduled magnetic resonance imaging (MRI) examination in the period from the start of the measurements until seven days after ingestion of the telemetric pill, pregnancy, impairment or disability of the upper extremity likely to have a negative impact on the quality of measurements (e.g., wounds, active venous access, amputation, dialysis shunt, edema, axillary dissection, continuous long-term monitoring of blood pressure, tattoos), known allergy to plastic or latex, and language problems.
The participants were equipped with a prototype sensor containing the CALERA® CBT technology worn at the left wrist, and an optical heart rate monitor (Wahoo® Tickr fit®) placed on the participants’ left upper arm. All participants also ingested a telemetric temperature-sensing pill (eCelsius®, BodyCap, Caen, France) and received a corresponding readout monitor. A bracelet informing about the ingested telemetric pill and its limitation regarding MRI scans (no-MRI) was placed on the right wrist (Fig. 5).
After administration of the vaccine, participants were free to leave the vaccination center and to resume their usual activity. They were asked to record activities (e.g., meals, sleep, physical activity, showering, etc.) in a paper diary time sheet with hour-minute resolution. Continuous data were collected for a minimum of 12 h. The participants were instructed to perform a daily control of the data sent by the ingested telemetric pill with the corresponding readout monitor.
All procedures performed in this study were in accordance with the Declaration of Helsinki. The local ethics committee (Ethikkommission Nordwest- und Zentralschweiz, EKNZ) approved the study protocol and procedures (EKNZ 2021–00,690). All participants signed an informed consent before their participation in the study.
CALERA sensor technology
The CALERA® technology (greenTEG®, Zurich, Switzerland) consists of a miniaturized heat flux sensor combined with skin temperature and heart rate sensors to continuously monitor CBT. A combination of physiological sensing, classical statistical modelling and embedded machine learning provides a CBT estimation at each one-minute sample. The heat flux signal is used to derive thermal resistance changes of the skin by compensating for skin temperature fluctuations caused by variations of the environmental conditions. CBT is estimated on-device and transmitted to a receiver by Bluetooth and ANT communication protocols. The small size of the heat flux sensor (2 × 2 mm) and its low power requirements facilitate the integration of the system into a wrist-worn device.
Evaluation metrics
The CBT prediction performance was computed by comparing the model output with the reference CBT measured by the ingested telemetric pill. The following evaluation metrics were employed: (1) the bias, (2) the mean absolute error (MAE), and (3) the Pearson correlation coefficient:
-
(1)
Bias \(\frac{{\Sigma }_{i=1}^{n}{y}_{i} -{x}_{i}}{n}\)
-
(2)
MAE \(\frac{{\Sigma }_{i=1}^{n}{|y}_{i} -{x}_{i}|}{n}\)
where i indicates a one-minute sample, x_i is the CBT reference, y_i is the CBT prediction and n is the total number of samples; and
-
(3)
Pearson correlation: \({{r}_{yx}=\frac{{\Sigma }_{i}{x}_{i}{y}_{i}-n\overline{x}\overline{y}}{\sqrt{{\Sigma }_{i}{{x}^{2}}_{i}-n{\overline{x}}^{2}}\sqrt{{\Sigma }_{i}{{y}^{2}}_{i}-n{\overline{y}}^{2}}}}\)where \(\overline{x}\) is the CBT reference sample mean, \(\overline{y}\) is the CBT prediction sample mean.
All metrics were calculated sample-wise (i.e., on a minute-to-minute basis) on individual participant data. A correlation scatter plot was used to visualize the sample-wise correlation between prediction and ground truth, and a Bland–Altman plot was used to identify any patterns or biases in the residuals across the range of ground truth and to calculate limits of agreement (LoA). The bias was calculated by averaging all the one-minute errors, where ground truth was subtracted from the predicted signal. A positive mean bias indicates that the prediction overestimates the ground truth and a negative mean bias indicates that the prediction underestimates the ground truth.
Considering an elevated CBT event any sample higher than 38.0 °C, we also calculated elevated CBT detection performance by employing false positive rate (FPR) and false negative rate (FNR), as follows:
FPR = \(\frac{\mathrm{FP}}{\mathrm{FP }+\mathrm{ TN}}\)
and
FNR = \(1- \frac{\mathrm{TP}}{\mathrm{TP }+\mathrm{ FN}}\),where FP is the number of samples wrongly categorized as elevated CBT events, TN is the number of samples correctly categorized as non-elevated CBT state, TP is the number of samples correctly categorized as elevated CBT events, and FN is the number of samples wrongly categorized as non-elevated CBT state.
Availability of data and materials
Restrictions apply to the availability of these data, which were used under license for this study. Data are available SE, with the permission of greenTEG AG.
Abbreviations
- CBT:
-
Core body temperature
- COVID-19:
-
Coronavirus disease 2019
- RPM:
-
Remote patient monitoring
- SARS-CoV-2:
-
Severe acute respiratory syndrome corona virus 2
- VS:
-
Vital signs
References
Geneva II, et al. Normal body temperature: a systematic review. Open Forum Infect Dis. 2019;6(4):ofz032.
Bartfai T, Conti B. Fever. ScientificWorldJournal. 2010;10:490–503.
Chams N, et al. COVID-19: a multidisciplinary review. Front Public Health. 2020;8:383.
Atzrodt CL, et al. A Guide to COVID-19: a global pandemic caused by the novel coronavirus SARS-CoV-2. Febs j. 2020;287(17):3633–50.
Byrne C, Lim CL. The ingestible telemetric body core temperature sensor: a review of validity and exercise applications. Br J Sports Med. 2007;41(3):126–33.
Hart D, et al. Non-invasive zero-heat-flux technology compared with traditional core temperature measurements in the emergency department. Am J Emerg Med. 2020;38(11):2383–6.
Niven DJ, et al. Accuracy of peripheral thermometers for estimating temperature: a systematic review and meta-analysis. Ann Intern Med. 2015;163(10):768–77.
Zsuzsanna Balla H, Theodorsson E, Ström JO. Evaluation of commercial, wireless dermal thermometers for surrogate measurements of core temperature. Scand J Clin Lab Invest. 2019;79(1–2):1–6.
Conway A, et al. Accuracy and precision of zero-heat-flux temperature measurements with the 3M™ Bair Hugger™ temperature monitoring system: a systematic review and meta-analysis. J Clin Monit Comput. 2021;35(1):39–49.
Brandes IF, et al. Evaluation of a novel noninvasive continuous core temperature measurement system with a zero heat flux sensor using a manikin of the human body. Biomed Tech (Berl). 2015;60(1):1–9.
Bräuer A, et al. Accuracy of zero-heat-flux thermometry and bladder temperature measurement in critically ill patients. Sci Rep. 2020;10(1):21746.
Tsadok I, et al. Assessing rectal temperature with a novel non-invasive sensor. J Therm Biol. 2021;95: 102788.
Mazgaoker S, et al. Measuring core body temperature with a non-invasive sensor. J Therm Biol. 2017;66:17–20.
Bräuer A, et al. Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature. J Clin Monit Comput. 2022. https://doi.org/10.1007/s10877-022-00851-z.
Zahner M., H., Durrer L., Schwyter E. SENSOR UNIT FOR A PORTABLE COMPUTER SYSTEM AND INTEGRATION OF THE SENSOR UNIT. 2017. https://patentscope.wipo.int/search/en/detail.jsf?docId=WO2018114653. Accessed 11 Aug 2022.
Ajčević M, et al. A novel non-invasive thermometer for continuous core body temperature: comparison with tympanic temperature in an acute stroke clinical setting. Sensors. 2022;22(13):4760.
Loncar-Turukalo T, et al. Literature on wearable technology for connected health: scoping review of research trends, advances, and barriers. J Med Internet Res. 2019;21(9): e14017.
Farias FAC, et al. Remote patient monitoring: a systematic review. Telemed J E Health. 2020;26(5):576–83.
Parmar P, et al. Use of telemedicine technologies in the management of infectious diseases: a review. Clin Infect Dis. 2014;60(7):1084–94.
Seshadri DR, et al. Wearable sensors for COVID-19: a call to action to harness our digital infrastructure for remote patient monitoring and virtual assessments. Front Digit Health. 2020;2:8.
Awad A, et al. Connected healthcare: improving patient care using digital health technologies. Adv Drug Deliv Rev. 2021;178: 113958.
Mecklai K, et al. Remote patient monitoring—overdue or overused? N Engl J Med. 2021;384(15):1384–6.
Dirikgil E, et al. Home monitoring reduced short stay admissions in suspected COVID-19 patients: COVID-box project. Eur Respir J. 2021. https://doi.org/10.1183/13993003.00636-2021.
Grutters LA, et al. Home-monitoring reduces hospital stay for COVID-19 patients. Eur Respir J. 2021. https://doi.org/10.1183/13993003.01871-2021.
Gruwez H, et al. Remote patient monitoring in COVID-19: a critical appraisal. Eur Respir J. 2022. https://doi.org/10.1183/13993003.02697-2021.
Eggenberger P, et al. Prediction of core body temperature based on skin temperature, heat flux, and heart rate under different exercise and clothing conditions in the heat in young adult males. Front Physiol. 2018;9:1780.
Chapin-Bardales J, Gee J, Myers T. Reactogenicity following receipt of mRNA-based COVID-19 vaccines. JAMA. 2021;325(21):2201–2.
Baden LR, et al. Efficacy and safety of the mRNA-1273 SARS-CoV-2 vaccine. N Engl J Med. 2020;384(5):403–16.
Verdel N, et al. Reliability and validity of the CORE sensor to assess core body temperature during cycling exercise. Sensors (Basel). 2021;21(17):5932.
Mackowiak PA, Chervenak FA, Grünebaum A. Defining fever. Open Forum Infect Dis. 2021. https://doi.org/10.1093/ofid/ofab161.
Speaker SL, et al. Oral temperature of noninfected hospitalized patients. JAMA. 2021;325(18):1899–901.
Suárez-Gil R, et al. Efficacy of telemedicine and at-home telemonitoring following hospital discharge in patients with COVID-19. J Pers Med. 2022;12(4):609.
Kendrovski V, et al. Quantifying projected heat mortality impacts under 21st-century warming conditions for selected European countries. Int J Environ Res Public Health. 2017;14(7):729.
Watts N, et al. The Lancet Countdown: tracking progress on health and climate change. The Lancet. 2017;389(10074):1151–64.
Cottle RM, et al. Core temperature responses to compensable versus uncompensable heat stress in young adults (PSU HEAT Project). J Appl Physiol. 2022;133(4):1011–8.
Sorensen C, Hess J. Treatment and prevention of heat-related illness. N Engl J Med. 2022;387(15):1404–13.
IUPS. IUPS Thermal Comission Glossary of terms for thermal physiology, Third edition. Jpn J Physiol, 2001;51:245–280.
Wunderlich CRA. Das Verhalten der Eigenwärme in Krankheiten. Leipzig: Verlag von Otto Wigand; 1868.
Musher DM, et al. Fever patterns. Their lack of clinical significance. Arch Intern Med. 1979;139(11):1225–8.
Drewry AM, et al. Body temperature patterns as a predictor of hospital-acquired sepsis in afebrile adult intensive care unit patients: a case-control study. Crit Care. 2013;17(5):R200.
Flora C, et al. High-frequency temperature monitoring for early detection of febrile adverse events in patients with cancer. Cancer Cell. 2021;39(9):1167–8.
Vargas B, et al. Discriminating bacterial infection from other causes of fever using body temperature entropy analysis. Entropy (Basel). 2022;24(4):510.
Rauch S, et al. Perioperative hypothermia-a narrative review. Int J Environ Res Public Health. 2021;18(16):8749.
Rüffert H, et al. Consensus guidelines on perioperative management of malignant hyperthermia suspected or susceptible patients from the European Malignant Hyperthermia Group. Br J Anaesth. 2021;126(1):120–30.
Lauronen SL, et al. Comparison of zero heat flux and double sensor thermometers during spinal anaesthesia: a prospective observational study. J Clin Monit Comput. 2022. https://doi.org/10.1007/s10877-021-00799-6.
Janke D, et al. Measuring core body temperature using a non-invasive, disposable double-sensor during targeted temperature management in post-cardiac arrest patients. Front Med (Lausanne). 2021;8: 666908.
Saugel B, Hoppe P, Khanna AK. Automated continuous noninvasive ward monitoring: validation of measurement systems is the real challenge. Anesthesiology. 2020;132(3):407–10.
Khanna AK, Hoppe P, Saugel B. Automated continuous noninvasive ward monitoring: future directions and challenges. Crit Care. 2019;23(1):194.
Breteler MJM, et al. Vital signs monitoring with wearable sensors in high-risk surgical patients: a clinical validation study. Anesthesiology. 2020;132(3):424–39.
Subbe CP, Duller B, Bellomo R. Effect of an automated notification system for deteriorating ward patients on clinical outcomes. Crit Care. 2017;21(1):52.
Taenzer AH, et al. Impact of pulse oximetry surveillance on rescue events and intensive care unit transfers: a before-and-after concurrence study. Anesthesiology. 2010;112(2):282–7.
Lam T, et al. Continuous pulse oximetry and capnography monitoring for postoperative respiratory depression and adverse events: a systematic review and meta-analysis. Anesth Analg. 2017;125(6):2019–29.
Funding
Open access funding provided by University of Basel. The present study received financial support by Innosuisse—Swiss Innovation Agency.
Author information
Authors and Affiliations
Contributions
RO, GS, LD and JE had full access to the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis. All authors contributed substantially to the study design and interpretation. RO and GS collected and analyzed the data. SE, RO and GS prepared a first manuscript draft. RO and GS prepared Figs. 5, 2, 3 and 4. SE prepared Fig. 1. All authors contributed substantially to the writing of the manuscript, critically revised the manuscript for important intellectual content and approved the final version. All authors read and approved the final manuscript.
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
The local ethics committee (Ethikkommission Nordwest- und Zentralschweiz, EKNZ) approved the study protocol and procedures (EKNZ 2021-00690). All participants signed an informed consent before their participation in the study.
Consent for publication
Not applicable.
Competing interests
LD, RO, GS, FR, PE and MZ are employees of greenTEG AG and receive salary for their work. LD, RO, PE and MZ own stocks of the company. LD is the chief technology officer of the company and member of the executive management. MZ is part of the executive management. LD and MZ are owners of the patent for the CALERA technology. SE, MO, SB and JE declare that they have no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Etienne, S., Oliveras, R., Schiboni, G. et al. Free-living core body temperature monitoring using a wrist-worn sensor after COVID-19 booster vaccination: a pilot study. BioMed Eng OnLine 22, 25 (2023). https://doi.org/10.1186/s12938-023-01081-3
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12938-023-01081-3