- Research
- Open Access
- Published:
PIV investigation of the flow fields in subject-specific vertebro-basilar (VA-BA) junction
BioMedical Engineering OnLine volume 18, Article number: 93 (2019)
Abstract
Background
As the only arterial structure of which two main arteries merged into one, the vertebro-basilar (VA-BA) system is one of the favorite sites of cerebral atherosclerotic plaques. The aim of this study was to investigate the detailed hemodynamics characteristics in the VA-BA system.
Methods
A scale-up subject-specific flow phantom of VA-BA system was fabricated based on the computed tomography angiography (CTA) scanning images of a healthy adult. Flow fields in eight axial planes and six radial planes were measured and analyzed by using particle image velocimetry (PIV) under steady flow conditions of \({Re}=300\), \({Re}=500\). A water–glycerin mixture was used as the working fluid.
Results
The flow in the current model exhibited highly three-dimensional characteristics. The confluence of VAs flow formed bimodal velocity distribution near the confluence apex. Due to the asymmetrical structural configuration, the bimodal velocity profile skewed towards left, and sharper peaks were observed under higher Reynolds condition. Secondary flow characterized by two vortices formed in the radial planes where 10 mm downstream the confluence apex and persists along the BA under both Reynolds numbers. The strength of secondary flow under \({Re}=500\) is around 8% higher than that under \({Re}=300\), and decayed nonlinearly along the flow direction. In addition, a low momentum recirculation region induced by boundary layer separation was observed near the confluence apex. The wall shear stress (WSS) in the recirculation area was found to be lower than 0.4 Pa. This region coincides well with the preferential site of vascular lesions in the VA-BA system.
Conclusions
This preliminary study verified that the subject-specific in-vitro experiment is capable of reflecting the detailed flow features in the VA-BA system. The findings from this study may help to expand the understanding of the hemodynamics in the VA-BA system, and further clarifying the mechanism that underlying the localization of vascular lesions.
Introduction
The vertebro-basilar (VA-BA) system is the only arterial structure in human that two large arteries merge into one, in which the VAs are arising from the subclavian arteries and join into BA. It provides a critical cerebral blood supply path that feeding the posterior circulation of the circle of Willis under normal conditions, and responsible for supplying compensational blood flow to anterior circulation when anatomical or pathological variations occurred [1,2,3].
Clinical observations have shown that the VA-BA region is a preferential site of vascular lesions. The prevalence of plaques in this region is around 50% in general population [4,5,6,7,8]. Moreover, approximately 25% of ischemic strokes were related to the lesions in VA-BA [9]. Compare with other causes of ischemic stroke, strokes caused by lesions in VA-BA would result in a much higher in-hospital mortality (8% vs. 20%) and worse functional outcomes [10].
Hemodynamics characteristics have long been associated with the initiation and the progression of the vascular diseases [4, 11,12,13]. Fry et al. [14] firstly suggested that the wall shear stress (WSS) excess 40 Pa could result in acute damages in the endothelial layer of vessels. In contrast, Caro et al. indicated that early lesions prefer to develop in low WSS regions [15]. This observation is supported by in-vivo and in-vitro studies that concerned hemodynamic at the carotid bifurcations [16, 17], coronary arteries [18,19,20,21], and descending thoracic aorta [22]. The results from the above studies have shown that the intimal thickening of vessels is strongly correlated with regions of low WSS. Further support of the atherogenic role of low WSS came from animal models [23, 24], computational fluid dynamics (CFD) simulations [10, 25], and studies at molecular and cellular scales [26, 27]. Based on the above evidence, low WSS theory is currently considered to be more convictive than high WSS hypotheses. In addition to low WSS, oscillatory shear stress [16, 17, 22, 28] and spatial wall shear stress gradient (SWSSG) [29] are also implicated as critical adverse factors in the disease process.
Most of the studies, however, are concerning the blood flow in bifurcations. Only limited attention has been paid to the flow characteristics in the arterial confluence. Despite the geometry of bifurcations is similar to junctions, the hemodynamic patterns in the arterial confluences are totally different from bifurcations due to the reversed flow direction, especially near the apex region.
One of the first studies concerning the flow in the VA-BA system was conducted by McDonald et al. [30]. In their in-vivo animal experiment, flows in the BA were visualized by injecting ink into the vessels. The study showed that the streams from symmetrical VAs would not mix in the BA. Limited by the method itself, however, no quantitative data were provided. Thereafter, multiple research methods were applied in the investigation of hemodynamic characteristics in VA-BA. Numerically, simulations were performed by applying two-dimensional models [31,32,33], three-dimensional symmetric models [31, 34, 35] and patient-specific models [10, 36]. Clinically, the using of MRA made the in-vivo measurement of flow in the confluence area possible [36,37,38].
In addition to the numerical and in-vivo method, the in-vitro experiment is another powerful research tool to explore the hemodynamics in vascular and to validate the numerical results. Ravensbergen et al. experimentally investigated the impacts of confluence angle [39] and merging flows [34] on flow characteristics in a rectangular cross-section confluence phantom by using laser Doppler anemometry (LDA). The flow velocity profiles in the confluence area were measured, and the results validated the presence of secondary flow in the confluence area that reported in an earlier numerical simulation [35]. Lutz et al. [40] visualized the confluence flow patterns in the VA-BA system by using dye injection, and quantified the mixing effect by introducing the concept of mixing index that based on the measurement of dye concentration. Kobayashi et al. [41] investigated the velocity profiles in VA-BA segments excised from elderly cadavers under different steady flow rates by using high-speed-camera. These studies provided unique and valuable in-vitro perspectives of the flow patterns in the VA-BA region. However, constrained by the measurement tools, certain hemodynamic characteristics in the confluence region, such as shear stress and velocity field, are yet to be fully investigated.
This study aims to investigate the detailed hemodynamics characteristics in the VA-BA system. To achieve the goal, a subject-specific 3D VA-BA flow phantom was fabricated base on CTA scanning images, and the detailed flow features in the VA-BA system were investigated through in-vitro experiments under different Reynolds numbers. Hemodynamics parameters in the confluence area, including velocity fields, shear stress distribution, and secondary flow, were analyzed to provide a better understanding of the role of hemodynamics in the localized vascular lesions.
Methods
Image acquisition
The cerebral CTA data of a healthy adult were provided by the First Affiliated Hospital of Xi’an Jiaotong University (Xi’an, Shaanxi, China). The scanning was performed on a 64 detector spiral CT (Aquilion 64, Toshiba Medical Systems, California, USA). The field of view (FOV), number of slices, tube voltage, tube current, scan time, and slice thickness were \(265\,\text {mm}\times 265\,\text {mm}\), 967, 120 kV, 350 mA, 500 ms, and 0.5 mm, respectively (Fig. 1).
Fabrication of the flow phantom
First of all, a 3D digital model of the VA-BA region was reconstructed from the CTA images. The process of reconstruction was introduced in our previous paper in detail [1]. To enable the accurate measurement, the derived VA-BA model was scaled-up to 5 times of original size in Solidworks and exported to STL format (Fig. 2). Then the STL file was sent to the 3D printing system (SCPS350, Xi’an Jiaotong University, China), and a resin phantom of the vascular structure was printed in a resolution of 0.1 mm (Fig. 3a). Finally, the printed vascular model was fixed into a small tank fulfilled with silicone gel (Sylgard 184, Dow Corning Inc., USA) (Fig. 3b). After the curing of the silicone, the resin vascular model was melted by heating, and a transparent phantom with internal flow tunnel was prepared for the in-vitro PIV measurements (Fig. 3c).
Experimental setup
The experimental setup was composed of a flow loop and a measurement system (Fig. 4).
The steady flow was supplied by an upstream overflow tank that provided constant pressure. According to the empirical equation (Eq. 1) [6], the length of tube that connected the tank was set to 1.6 m and flow-regulating honeycombs were put inside the tubes to ensure the inlet flow fully developed.
The inlet flow rates of bilateral VAs were controlled by adjusting the valves placed at the upstream of the VA entrances and monitored by the electromagnetic flow meters. At the efferent of BA, an adjustable resistor was connected into the flow loop. The efferent flow of the phantom was collected in a downstream tank and pumped back to the overflow tank. An adjustable thermostat heater was placed in the overflow tank to control the temperature of the working fluid.
Working fluid and flow parameters
Blood is a multiphase fluid that behaves shear-thinning property. However, the viscosity of blood is relatively constant at shear rates above 100 s−1 [7]. Non-Newtonian nature of blood was neglected in the present study because of the shear rates in large arteries as ICA and VA are rarely lower than 100 s−1.
The working fluid that selected for this experiment was the mixture of glycerin and water with a density of 1157 kg/m3, a viscosity of 10.6 cP at \(37\,^{\circ }\text {C}\) and a refractive index of 1.41. The refractive index of the working fluid matched well with the silicon phantom, and no distortion was observed at the fluid-solid interface (Fig. 5). Hollow glass spheres with an average diameter of \(10\,\upmu \text {m}\) were seeded into the working fluid for PIV data acquisition.
Matching the Reynolds number in the experimental condition with the in-vivo arterial flow condition is the key to obtain meaningful results from a scaled flow phantom. As the Reynolds number ranges from 100 to 1000 in medium-size arteries, Reynolds number 300 and 500 were chosen as the experimental conditions. The Reynolds number was defined as follows:
Flow visualization
PIV system was used to capture the flow velocity in the phantom. The PIV system consisted of a CCD camera (LaVision Image Pro 4M CCD, \(2048\times 2048\) pix2), a 200 mJ Nd:Yag laser (Gemini 200, 532 nm), and an optic lens that produces light sheet about 1 mm thick.
To investigate the hemodynamics characteristics in the VA-BA system in detail, the flow fields in eight planes parallel to the flow direction (axial planes) (Fig. 5a) and six planes orthogonal to the mainstream (radial planes) (Fig. 5b) were captured. The axial planes were evenly spaced from top to bottom with an interval of 1 mm. The radial planes were distributed from upstream to downstream with an interval of 5 mm.
The two-frame cross-correlation method was used for image acquisitions. 100 image pairs were recorded in each acquisition to diminish the error caused by random events. The acquired raw images in the IM7 format were post-processed in the DaVis software to obtain the velocity vector files. The interrogation windows size was set to \(32\times 32\) pix with 50% overlap. A self-developed MATLAB algorithm was utilized to eliminate the noise outside the flow region.
Shear stress conversion
Shear stress is defined as follow:
Thus, shear stress can be got from the velocity vector field. To compare the shear stress in this scaled phantom to that of the original size, a scale factor was applied in this study. The scale factor is derived from Buckingham Pi theorem [10, 42]:
Results
Geometrical structure of the VA-BA system
The current VA-BA model has an asymmetrical structural that the sagittal plane of the body did not pass through the center of junction apex. The original diameters of the left VA (LVA), right VA (RVA), and BA were 2.37 mm, 2.75 mm, and 2.83 mm, respectively. The diameter ratio of the bilateral VAs was 1.16 and the confluence angle was \(63^\circ\).
Axial flow patterns
Figure 6 showed the overall axial velocity distributions in the VA-BA region under different Reynolds numbers.
Detailed axial velocity fields (solid contour) and velocity profiles (vector arrows) in each plane were illustrated in Fig. 7.
As the figure showed, the velocity distributions in the bilateral VAs were quasi-parabolic. When the flows confronted, a bimodally distributed velocity profile with a trough at the arterial axis appeared in BA. Under the flow condition of \({Re}=300\), it takes 18 mm in the median plane for the bimodal velocity profile restore to the quasi-parabolic distribution (Fig. 7d, e). In contrast, the bimodal velocity distribution pattern existed all along the BA segment of flow phantom under the flow condition of \({Re}=500\). The peaks of bimodal velocity distribution are more bias to the left side under higher Reynolds number. In addition, a triangle shaped flow stagnation region was found at the confluence apex, where the flow velocity magnitude is below 0.05 m/s. The area of the flow stagnation decreased with the increasing of Reynolds number, which is 74.09 mm2 under \({Re}=300\) and 59.69 mm2 under \({Re}=500.\)
Streamlines over the whole VA-BA region under different Reynolds numbers were illustrated in (Fig. 8).
Detailed streamlines in each plane were illustrated in Fig. 9. The overall streamline distributions were similar between different Reynolds numbers. The flows from the bilateral VAs confronted each other at the beginning of the BA where 10 mm downstream the confluence apex. In the upper planes, stream from the left side fully dominated the flow in the BA (Fig. 9a, b). When comes to the lower planes of the flow phantom, the flow from RVA gradually took dominance in the BA (Fig. 9c–h), this phenomenon became stronger as the Reynolds number increases. However, no signs of flow disturbance and mixing were observed in the entire flow region. The two streams from the bilateral VAs remain parallel in BA, and a clear boundary between the bilateral streams could be observed. Moreover, it is interesting to note that stable recirculation zones formed at the bifurcation site while no shedding was observed under experimental conditions.
Shear stress obtained from the experimental data has been converted based on Eq. 4 (Fig. 10). Detailed shear stress distributions in each plane under different Reynolds numbers were plotted in Fig. 11. Low shear stress \((<0.4 \; \text {Pa})\) was observed near the wall of the confluence apex and the left side of the BA. High shear stress was observed along with the confluence interface and the vessel walls of VA and BA.
Radial flow patterns
Velocity fields in the radial planes were shown in Figs. 12 and 13. Flow from the left VA was wrapped by that from the right side and formed unique three-dimensional flow patterns downstream the confluence apex. An obvious boundary between the flow from the bilateral VAs was observed in the BA (Fig. 13c–f). This phenomenon agrees with the observations in the axial planes.
The velocity vectors in radial planes were illustrated in Figs. 14 and 15 further clarified the radial flow characteristic in the VA-BA region. It can be seen that the secondary flows in the BA were characterized by two vortices rotate in opposite directions (Fig. 15c–f). To quantify the strength of the secondary flow, a dimensionless term “mean secondary velocity” that introduced by Ravensbergen et al. [39] was utilized in this study. The mean secondary velocity is defined as the ratio of the cross-sectional mean of the secondary velocities to the cross-sectional mean of the axial velocities at each location. The impacts of Reynolds numbers and distance from the confluence apex on the strength of secondary flow were quantitatively analyzed and showed in Fig. 16, and the specific data were listed in Table 1.
Shear stress distributions in radial planes were calculated and illustrated in Figs. 17 and 18. It was noticed that a low shear stress zone appeared at the apex site in the radial flow (Fig. 18b) as well, with the value that lower than 0.4 Pa.
Discussion
In the present study, a subject-specific VA-BA phantom was reconstructed, and in-vitro investigations were conducted to elucidate the hemodynamics features of the VA-BA arterial system. Detailed flow patterns and shear stress distributions in the confluence region were studied and compared under different Reynolds numbers.
Structural characteristics
The geometrical structure of the VA-BA arterial system varies among populations. Previous clinical studies have reported that the confluence angle of the VAs ranges from \(10\,^{\circ }\) to \(160^{\circ }\,(60^{\circ }{\pm }30^{\circ })\) [10, 39]. Wake-Buck et al. [10] classified the structure of VA-BAs into three types according to the spatial configurations, including Walking type, Tuning fork type, and Lambda type. A recent retrospective clinical study showed that 64% of patients have a Lambda type VA-BA, and the prevalence of Walking type and Tuning fork type configuration is 17% and 19%, respectively [43]. In present study, the confluence angle between the VAs was \(63^{\circ }\) and the diameter ratio of RVA to LVA was 1.16, which are within the reported physiological range. The angle between the LVA and BA was \(176^{\circ }\) and RVA joints the BA in a pseudo-T-junction, which could be classified as the lambda type VA-BA.
Flow characteristics
Axial flow
As a unique arterial structure, the flow in the VA-BA system also exhibited distinctive characteristics. When confronted in the proximal end of the BA, the streams from the bilateral VAs flowed on their own side without mixing. Such patterns suggested the flow in the BA is laminar. In addition, bimodal velocity profiles appeared immediately after the confluence. Due to the asymmetrical structure, the left peak of the velocity profile is sharper than the right side, which represents the eccentric flow patterns in BA. These findings were supported by several CFD [10], in-vitro [41] and in-vivo [30, 38, 44] studies.
To make the data suitable for quantitative comparison, the normalized velocity that defined as the maximum magnitude in each velocity profile (Umax) divided by the averaged cross-sectional velocity (Umean) in BA were calculated, and data were listed in Table 2. According to the Poiseuille law, the ratio of Umax to Umax should be 2 in the fully developed laminar flow inside a tube, and the length required for the flow development was given in Eq. 1. Because of the limited length, it was impossible for the flow in BA fully developed. Thus, normalized velocity would between 1 and 2, and its magnitude would decrease with increasing of Reynolds number. The results from the current study agree well with the theoretical analysis. Meanwhile, due to the eccentric flow induced by the asymmetrical structure, the normalized velocity under \({Re}=300\) was close to 2 near the confluence apex. The similar phenomenon was reported by Kobayashi et al. [41]. The decreasing of normalized velocity along flow direction revealed that the flow development in BA was affected by the confluence of bilateral streams. Moreover, despite the flows from VAs are parallel along the BA in the median planes, (Fig. 9c–f), helical flow was observed in upper (Fig. 9a, b) and lower planes (Fig. 9g, h) near the distal end of BA segment. This observation further suggested the fluid downstream the confluence flows in a layered pattern along radials direction.
When focused on the confluence apex region, a zone of low shear stress that coincides with low momentum recirculation as the result of boundary layer separation was observed in this study. The area of this region decreased with the increasing of Reynolds number. The WSS in the VA-BA was derived from the shear stress field that one pixel from to the wall, and low WSS \((<0.4\,\text {Pa})\) region at the apex of VAs junction was predicted in this study. This finding agrees with the results from the idealize and patient-specific models [8, 10, 45]. Several studies have suggested low WSS yield atheroprotective range could trigger an inflammatory-cell-mediated pathway that associated with the growth of atherosclerotic plaques [14, 46,47,48]. The observation of the localized low WSS provided additional hemodynamic evidence of the preferential of vascular plaques in the confluence apex region. Besides, the low-velocity recirculation zone may accelerate the progression of atherosclerosis by gathering and depositing of blood components.
Radial flow
Highly three-dimensional flows in the VA-BA system have been reported in literature [8, 10, 34, 39]. Thus, the investigation of flow patterns in radial direction would be important for a more comprehensive understanding of the hemodynamics characteristics in the VA-BA system.
In previous studies, Ravensbergen’s team investigated the radial flow in a generalized tuning fork type VA-BA system by using CFD simulations [34, 35, 39]. Their results showed that secondary flow with a distinct four-vortex pattern appeared in the BA as the consequence of flow confluence. Furthermore, the same team quantitatively analyzed the impact of flow conditions on the secondary flow strength by using the non-dimensional term mean secondary velocity, which suggested that the secondary flow is stronger under higher Reynolds number and decayed along the flow direction.
The present study is the first in-vitro attempt to investigate the secondary flow in the subject-specific phantom of VA-BA system. Our results showed that the secondary flow was established around 10 mm downstream the confluence apex due to the boundary layer separation, and persists along the BA. The secondary flow strength at the plane 10 mm downstream apex was 42.1% under \({Re}=300\) and 45.5% under \({Re}=500\), and decreased along the flow direction in a non-linear pattern (Fig. 16). Though the directly quantitative comparison of secondary strength between studies is difficult due to the use of different model and flow conditions, its distribution patterns in this study agree well with the results from previous ones [34, 35]. However, we only observed two vortices rotate oppositely in the secondary flow. This phenomenon is mainly due to the skewed flow in the BA, which was induced by the lambda type structural configuration.
Though the average shear stress level in the radial planes is low, relatively high shear stress areas appeared at the interface of two streams and near the artery wall of BA. The secondary flow may bring some blood from a relatively low shear stress area to a high shear stress area repeatedly. Blood components in the high and low shear stress cycle may result in blood damage due to fatigue [49]. Meanwhile, secondary flow in the artery could also play a key role in the erosion and endothelial response at the early stage of atherosclerosis pathology [50].
Simplifications made in this study
Restricted by the viable experimental condition, it is difficult to fully mimic the physiological flow in the VA-BA system. Thus, several simplifications have been made in the current study.
First of all, the inlet flow was assumed as fully developed at the entrance in the current study. Although this assumption is not physiologically correct, it has been widely accepted in in-vitro experimental studies [51, 52] and provided a basis for comparison between studies.
Secondly, due to the deformation of the vessels is small and blood behaves as Newtonian fluid in arteries of the size of the VA-BA system, the arterial wall elasticity and the non-Newtonian properties of blood were neglected.
In addition, despite the blood flow in the vascular system was pulsatile under physiological conditions, several numerical [16, 53] and in-vitro [41, 54, 55] studies have suggested that steady boundary conditions are able to predict the non-temporal related cerebral flow characteristics at the corresponding point of the pulsatile flow profile. Among which, a comparison between pulsatile results [33] and steady results [31] in two-dimensional VA-BA models showed that the most important flow patterns are the same in both cases. Kobayashi et al. [41] further suggested that the VA-BA flow phenomena occurred in pulsatile flow are essentially the same as those found in steady flow. Thus, the steady flow boundary conditions adopted in this study is capable of revealing the flow characteristics in the region of interesting as no temporal hemodynamics terms were involved.
Moreover, the flow phantom utilized in this study was scaled-up. To overcome the spatial restrictions, the scaling of vascular replicas based on the principle of dynamic similarity, also known as dynamic scaling, has been widely used in the in-vitro experiments [56,57,58,59]. The dynamic scaling is a well-established concept that ensures the development of flow in scaled phantom identical to the original object. In the incompressible steady flow within rigid domains, the dynamic similarity could be ensured by matching of the Reynolds number. In this study, a scaled-up VA-BA phantom was deployed due to the difficulties of flow visualization in the original size. As described in the method section of this manuscript, the dynamic scaling has been achieved by matching the Reynolds number between the scaled phantom and in-vivo conditions. Thus, the measurements from the scaled phantom are capable of reflecting the flow characteristics in the non-scaled situations.
Limitations and future works
It is important to emphasize that the current study carries inherent limitations associated with the in-vitro modeling techniques. Firstly, the arteries that branch out from BA were removed. To investigate the flow in VA-BA system in a more detailed manner, these branch arteries should be included in future work. Secondly, the elasticity of the arterial wall was neglected. As discussed before, though the rigid wall assumption is acceptable in the current study, it still may lead to a slight overestimation of WSS [60]. In addition, more samples should be included in the future investigations to systemically evaluate the impact of various spatial characteristics [21]. Finally, the three-dimensional flow field should be measured in future work to assess the complex spatial flow in the junction site.
Conclusions
In this study, in-vitro experiments were conducted to investigate the detailed hemodynamics characteristics in a subject-specific VA-BA flow phantom. Flow characteristics in the axial and radial planes were visualized and analyzed by using PIV. The preliminary results showed that the flow in VA-BA system behaves a highly three-dimensional feature, and the flow patterns were affected by the spatial structures and inflow Reynolds numbers. Further, a low WSS region coincides well with the preferential region of atherosclerotic plaques was found near the confluence apex. The findings from this study could help to expand the understanding of the hemodynamics in the VA-BA system, and further clarifying the mechanism that underlying the localization of vascular lesions.
Availability of data and materials
Not applicable.
Abbreviations
- BA:
-
basilar artery
- CFD:
-
computational fluid dynamics
- CoW:
-
circle of Willis
- CTA:
-
computed tomography angiography
- LDL:
-
low density lipoprotein
- NURBS:
-
non-uniform rational B-splines
- PIV::
-
particle image velocimetry
- VA::
-
vertebral artery
References
Zhu GY, Yuan Q, Yang J, Yeo JH. The role of the circle of Willis in internal carotid artery stenosis and anatomical variations: a computational study based on a patient-specific three-dimensional model. Biomed Eng Online. 2015;14(1):107. https://doi.org/10.1186/s12938-015-0105-6.
Zhu GY, Yuan Q, Yang J, Yeo JH. Experimental study of hemodynamics in the circle of Willis. Biomed Eng Online. 2015;14(Suppl 1):10. https://doi.org/10.1186/1475-925X-14-S1-S10.
Liu X, Gao Z, Xiong H, Ghista D, Ren L, Zhang H, Wu W, Huang W, Hau WK. Three-dimensional hemodynamics analysis of the circle of Willis in the patient-specific nonintegral arterial structures. Biomech Model Mechanobiol. 2016;15(6):1439–56. https://doi.org/10.1007/s10237-016-0773-6.
Cecchi E, Giglioli C, Valente S, Lazzeri C, Gensini GF, Abbate R, Mannini L. Role of hemodynamic shear stress in cardiovascular disease. Atherosclerosis. 2011;214(2):249–56. https://doi.org/10.1016/j.atherosclerosis.2010.09.008.
Schaffer S, Schwartz C, Wagner W. A definition of the intima of human arteries and of its atherosclerosis-prone regions. Circulation. 1992;85(1):391–405. https://doi.org/10.1161/01.CIR.85.1.391.
Bamford J, Sandercock P, Dennis M, Burn J, Warlow C. Classification and natural history of clinical identifiable subtypes of cerebral infarction. Lancet. 1991;337(8756):1521–6. https://doi.org/10.1016/0140-6736(91)93206-O.
Graziano F, Ganau M, Iacopino DG, Boccardi E. Vertebro-basilar junction aneurysms: a single centre experience and meta-analysis of endovascular treatments. Neuroradiol J. 2014;27(6):732–41. https://doi.org/10.15274/NRJ-2014-10100.
Ravensbergen J, Ravensbergen JW, Krijger JK, Hillen B, Hoogstraten HW. Localizing role of hemodynamics in atherosclerosis in several human vertebrobasilar junction geometries. Arterioscler Thromb Vasc Biol. 1998;18(5):708–16. https://doi.org/10.1161/01.ATV.18.5.708.
Bogousslavsky J, Van Melle G, Regli F. The Lausanne stroke registry: analysis of 1,000 consecutive patients with first stroke. Stroke. 1988;19(9):1083–92. https://doi.org/10.1161/01.STR.19.9.1083.
Wake-Buck AK, Gatenby JC, Gore JC. Hemodynamic characteristics of the vertebrobasilar system analyzed using MRI-based models. PLoS ONE. 2012;7(12):1354–7. https://doi.org/10.1371/journal.pone.0051346.
Cebral JR, Mut F, Weir J, Putman CM. Association of hemodynamic characteristics and cerebral aneurysm rupture. Am J Neuroradiol. 2011;32(2):264–70. https://doi.org/10.3174/ajnr.A2274.
Kulcsár Z, Ugron Á, Marosfoi M, Berentei Z, Paál G, Szikora I. Hemodynamics of cerebral aneurysm initiation: the role of wall shear stress and spatial wall shear stress gradient. Am J Neuroradiol. 2011;32(3):587–94. https://doi.org/10.3174/ajnr.A2339.
Sforza DM, Kono K, Tateshima S, Vinuela F, Putman C, Cebral JR. Hemodynamics in growing and stable cerebral aneurysms. J Neurointerv Surg. 2015;. https://doi.org/10.1136/neurintsurg-2014-011339.
Fry DL. Acute vascular endothelial changes associated with increased blood velocity gradients. Circ Res. 1968;22(2):165–97. https://doi.org/10.1161/01.RES.22.2.165.
Caro CG, Fitz-Gerald JM, Schroter RC. Arterial wall shear and distribution of early atheroma in man. Nature. 1969;223(5211):1159–60. https://doi.org/10.1038/2231159a0.
Ku DN, Giddens DP, Zarins CK, Glagov S. Pulsatile flow and atherosclerosis in the human carotid bifurcation. Positive correlation between plaque location and low oscillating shear stress. Arterioscler Thromb Vasc Biol. 1985;5(3):293–302. https://doi.org/10.1161/01.ATV.5.3.293.
Friedman MH, Deters OJ, Bargeron CB, Hutchins GM, Mark FF. Shear-dependent thickening of the human arterial intima. Atherosclerosis. 1986;60(2):161–71. https://doi.org/10.1016/0021-9150(86)90008-0.
Perktold K, Hofer M, Rappitsch G, Loew M, Kuban BD, Friedman MH. Validated computation of physiologic flow in a realistic coronary artery branch. J Biomech. 1998;31(3):217–28. https://doi.org/10.1016/S0021-9290(97)00118-8.
Friedman MH, Deters OJ. Correlation among shear rate measures in vascular flows. J Biomech Eng. 1987;109(1):25–6. https://doi.org/10.1115/1.3138637.
Chatzizisis YS, Coskun AU, Jonas M, Edelman ER, Feldman CL, Stone PH. Role of endothelial shear stress in the natural history of coronary atherosclerosis and vascular remodeling: molecular, cellular, and vascular behavior. J Am Coll Cardiol. 2007;49(25):2379–93. https://doi.org/10.1016/j.jacc.2007.02.059.
Yang Y, Liu X, Xia Y, Liu X, Wu W, Xiong H, Zhang H, Xu L, Wong KKL, Ouyang H, Huang W. Impact of spatial characteristics in the left stenotic coronary artery on the hemodynamics and visualization of 3D replica models. Sci Rep. 2017;7(1):15452. https://doi.org/10.1038/s41598-017-15620-1.
Moore JE, Xu C, Glagov S, Zarins CK, Ku DN. Fluid wall shear stress measurements in a model of the human abdominal aorta: oscillatory behavior and relationship to atherosclerosis. Atherosclerosis. 1994;110(2):225–40. https://doi.org/10.1016/0021-9150(94)90207-0.
Gambillara V, Chambaz C, Montorzi G, Roy S, Stergiopulos N, Silacci P. Plaque-prone hemodynamics impair endothelial function in pig carotid arteries. Am J Physiol Heart Circ Physiol. 2006;290(6):2320–8. https://doi.org/10.1152/ajpheart.00486.2005.
Buchanan JR, Kleinstreuer C, Truskey GA, Lei M. Relation between non-uniform hemodynamics and sites of altered permeability and lesion growth at the rabbit aorto-celiac junction. Atherosclerosis. 1999;143(1):27–40. https://doi.org/10.1016/S0021-9150(98)00264-0.
Buchanan J. Hemodynamics simulation and identification of susceptible sites of atherosclerotic lesion formation in a model abdominal aorta. J Biomech. 2003;36(8):1185–96. https://doi.org/10.1016/S0021-9290(03)00088-5.
Cheng C, Van Haperen R, De Waard M, Van Damme LCa, Tempel D, Hanemaaijer L, Van Cappellen GWa, Bos J, Slager CJ, Duncker DJ, Van Der Steen AFW, De Crom R, Krams R. Shear stress affects the intracellular distribution of eNOS: direct demonstration by a novel in vivo technique. Blood. 2005;106(12):3691–8. https://doi.org/10.1182/blood-2005-06-2326.
Lutgens E, Gijbels M, Smook M, Heeringa P, Gotwals P, Koteliansky VE, Daemen MJaP. Transforming growth factor-beta mediates balance between inflammation and fibrosis during plaque progression. Arterioscler Thromb Vasc Biol. 2002;22(6):975–82. https://doi.org/10.1161/01.ATV.0000019729.39500.2F.
Ku DN. Blood flow in arteries. Annu Rev Fluid Mech. 1997;29(1):399–434. https://doi.org/10.1146/annurev.fluid.29.1.399.
Lei M, Kleinstreuer C, Truskey GA. Numerical investigation and prediction of atherogenic sites in branching arteries. J Biomech Eng. 1995;117(3):350–7. https://doi.org/10.1115/1.2794191.
McDonald DA, Potter JM. The distribution of blood to the brain. J Physiol. 1951;114(3):356–71. https://doi.org/10.1113/jphysiol.1951.sp004627.
Krijger JK, Hillen B, Hoogstraten HW. Mathematical models of the flow in the basilar artery. J Biomech. 1989;22(11–12):1193–202. https://doi.org/10.1016/0021-9290(89)90221-2.
Krijger JKB, Hillen B, Hoogstraten HW, van den Raadt MPMG. Steady two-dimensional merging flow from two channels into a single channel. Appl Sci Res. 1990;47(3):233–46. https://doi.org/10.1007/BF00418053.
Krijger JKB, Hillen B, Hoogstraten HW. A two-dimensional model of pulsating flow in the basilar artery. J Appl Math Phys. 1991;42(5):649–62. https://doi.org/10.1007/BF00944764.
Ravensbergen J, Krijger JKB, Hillen B, Hoogstraten HW. Merging flows in an arterial confluence: the vertebro-basilar junction. J Fluid Mech. 1995;304(11):119–41. https://doi.org/10.1017/S0022112095004368.
Krijger JKB, Heethaar RM, Ravensbergen J. Computation of steady three-dimensional flow in a model of the basilar artery. J Biomech. 1992;25(12):1451–65. https://doi.org/10.1016/0021-9290(92)90058-9.
Zhao X, Zhao M, Amin-Hanjani S, Du X, Ruland S, Charbel FT. Wall shear stress in major cerebral arteries as a function of age and gender-a study of 301 healthy volunteers. J Neuroimaging. 2015;25(3):403–7. https://doi.org/10.1111/jon.12133.
Bockman MD, Kansagra AP, Shadden SC, Wong EC, Marsden AL. Fluid mechanics of mixing in the vertebrobasilar system: comparison of simulation and MRI. Cardiovasc Eng Technol. 2012;3(4):450–61. https://doi.org/10.1007/s13239-012-0112-8.
Smith AS, Bellon JR. Parallel and spiral flow patterns of vertebral artery contributions to the basilar artery. Am J Neuroradiol. 1995;16(8):1587–91.
Ravensbergen J, Krijger JKB, Hillen B, Hoogstraten HW. The influence of the angle of confluence on the flow in a vertebro-basilar junction model. J Biomech. 1996;29(3):281–99. https://doi.org/10.1016/0021-9290(95)00064-X.
Lutz RJ, Warren K, Balis F, Patronas N, Dedrick RL. Mixing during intravertebral arterial infusions in an in vitro model. J Neurooncol. 2002;58(2):95–106. https://doi.org/10.1023/A:1016034910875.
Kobayashi N, Karino T. Flow patterns and velocity distributions in the human vertebrobasilar arterial system. Laboratory investigation. J Neurosurg. 2010;113(4):810–9. https://doi.org/10.3171/2010.1.JNS09575.
Bale-Glickman JM. Experimental studies of physiological flows in replicated diseased carotid bifurcations. Ph.D. thesis, University of California, Berkeley; 2005. https://books.google.co.jp/books?id=KELFNwAACAAJ.
Yu J, Zhang S, Li M-L, Ma Y, Dong Y-R, Lou M, Feng F, Gao S, Wu S-W, Xu W-H. Relationship between the geometry patterns of vertebrobasilar artery and atherosclerosis. BMC Neurol. 2018;18(1):83. https://doi.org/10.1186/s12883-018-1084-6.
Taveras JM, Wood EH. Diagnostic neuroradiology, vol. 1. Baltimore: Williams & Wilkins; 1976.
Chong BW, Kerber CW, Buxton RB, Frank LR, Hesselink JR. Blood flow dynamics in the vertebrobasilar system: correlation of a transparent elastic model and MR angiography. Am J Neuroradiol. 1994;15(4):733–45.
Lusis A. Atherosclerosis. Nature. 2000;407(6801):233–41. https://doi.org/10.1038/35025203.Atherosclerosis.
Dejana E, Valiron O, Navarro P, Lampugnani MG. Intercellular junctions in the endothelium and the control of vascular permeability. Ann N Y Acad Sci. 1997;811:36–44. https://doi.org/10.1111/j.1749-6632.1997.tb51986.x.
Meng H, Tutino VM, Xiang J, Siddiqui a. High WSS or low WSS? Complex interactions of hemodynamics with intracranial aneurysm initiation, growth, and rupture: toward a unifying hypothesis. Am J Neuroradiol. 2014;35(7):1254–62. https://doi.org/10.3174/ajnr.A3558.
Zhang JN, Bergeron AL, Yu Q, McBride L, Bray PF, Dong JF. Duration of exposure to high fluid shear stress is critical in shear-induced platelet activation-aggregation. Thromb Haemost. 2003;90(4):672–8. https://doi.org/10.1160/TH03-03-0145.
Mohamied Y, Rowland EM, Bailey EL, Sherwin SJ, Schwartz Ma, Weinberg PD. Change of direction in the biomechanics of atherosclerosis. Ann Biomed Eng. 2014;43(1):16–25. https://doi.org/10.1007/s10439-014-1095-4.
Katritsis D, Kaiktsis L, Chaniotis A, Pantos J, Efstathopoulos EP, Marmarelis V. Wall shear stress: theoretical considerations and methods of measurement. Architecture. 2007;49(5):307–29. https://doi.org/10.1016/j.pcad.2006.11.001.
Brunette J, Mongrain R, Laurier J, Galaz R, Tardif JC. 3D flow study in a mildly stenotic coronary artery phantom using a whole volume PIV method. Med Eng Phys. 2008;30(9):1193–200. https://doi.org/10.1016/j.medengphy.2008.02.012.
Hillen B, Drinkenburg BA, Hoogstraten HW, Post L. Analysis of flow and vascular resistance in a model of the circle of Willis. J Biomech. 1988;21(10):807–14. https://doi.org/10.1016/0021-9290(90)90307-O.
Chen L. Hemodynamics in the cerebral circulation: numerical studies and experimental investigation. Doctor of philosophy. Nanyang: Nanyang Technological University; 2005.
Fahy P, McCarthy P, Sultan S, Hynes N, Delassus P, Morris L. An experimental investigation of the hemodynamic variations due to aplastic vessels within three-dimensional phantom models of the circle of Willis. Ann Biomed Eng. 2014;42(1):123–38. https://doi.org/10.1007/s10439-013-0905-4.
Beier S, Ormiston J, Webster M, Cater J, Norris S, Medrano-Gracia P, Young A, Cowan B. Vascular hemodynamics with computational modeling and experimental studies. In: Computing and visualization for intravascular imaging and computer-assisted stenting. Amsterdam: Elsevier; 2017. p. 227–51. https://doi.org/10.1016/B978-0-12-811018-8.00009-6. https://linkinghub.elsevier.com/retrieve/pii/B9780128110188000096.
Beier S, Ormiston J, Webster M, Cater J, Norris S, Medrano-Gracia P, Young A, Cowan B. Impact of bifurcation angle and other anatomical characteristics on blood flow–a computational study of non-stented and stented coronary arteries. J Biomech. 2016;49(9):1570–82. https://doi.org/10.1016/j.jbiomech.2016.03.038.
Hasler D, Obrist D. Three-dimensional flow structures past a bio-prosthetic valve in an in-vitro model of the aortic root. PLoS ONE. 2018;13(3):0194384. https://doi.org/10.1371/journal.pone.0194384.
Friedman MH, Kuban BD, Schmalbrock P, Smith K, Altan T. Fabrication of vascular replicas from magnetic resonance images. J Biomech Eng. 1995;117(3):364. https://doi.org/10.1115/1.2794193.
Torii R, Oshima M, Kobayashi T, Takagi K, Tezduyar TE. Influence of wall thickness on fluid–structure interaction computations of cerebral aneurysms. Int J Numer Methods Biomed Eng. 2010. https://doi.org/10.1002/cnm.1289.
Acknowledgements
We would like to acknowledge the help from our technician Mr. Zhenwei Ji during the setting of test rig.
Funding
This project is supported by grants from the National Natural Science Foundation of China (NSFC) (11802227), China postdoctoral science foundation grant (2016M600781), and the Fundamental Research Funds for the Central Universities (XJJ2017032).
Author information
Authors and Affiliations
Contributions
GYZ carried out the experimental study, drafted the manuscript, and acquired the funding. YW participated in the manuscript drafting. QY participated in the design of this study. JHY participated in the design of the study and helped to revise the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Zhu, G., Wei, Y., Yuan, Q. et al. PIV investigation of the flow fields in subject-specific vertebro-basilar (VA-BA) junction. BioMed Eng OnLine 18, 93 (2019). https://doi.org/10.1186/s12938-019-0711-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12938-019-0711-9
Keywords
- PIV
- In-vitro
- Hemodynamics
- VA-BA