The relationship between compression garments and electrocardiogram signals during exercise and recovery phase

Abstract

Background

The direction of the current research was to investigate whether electrocardiogram (ECG) signals have been impacted by using compression garments during exercise and recovery phase. Each subject is non-athletes, conducted two running tests, wearing either non-compression garments (NCGs) or compression garments (CGs) throughout experiments and 2-h of the recovery phase. Experiment 1 (number of participants (n) = 8; 61.4 ± 13.7 kg, 25.1 ± 3.8 years, 165.9 ± 8.3 cm) focused on the exercising phase while Experiment 2 (n = 14; 60.9 ± 12.0 kg, 24.7 ± 4.5 years, 166.0 ± 7.6 cm) concentrated on the recovery phase. Electrocardiogram (ECG) data were collected through wearable biosensors.

Results

The results demonstrated a significant difference between compression garments and non-compression garments at the end of the tests and from 90 min onwards during the recovery phase (p < 0.05). Corrected QT (QTc), ST interval and heart rate (HR) indicated the significant difference between NCGs and CGs. Conclusion: Based on the findings, the utilization of compression garments showed a positive influence in non-athletes based on the quicker recovery in HR, ST, and QTc.

Introduction

Compression garments are becoming popular for sports activities in athletes and non-athletes. Many studies have claimed that compression garments may have positive effects and could increase performance. For example, during distance running, wearing CGs correlates with significantly lower muscle activation and longer median frequencies which might prevent muscle-fatigue and enhance running performance [1]. Other research investigating the use of upper body CGs indicates that performance and comfort can be increased as a result of enhanced proprioceptive cues during upper-body movements in high-level athletes [2]. Similarly, there was an improvement in performance resulting from altering the running technique and reducing perceived exertion [3]. However, another article has also identified non-affective major hemodynamic parameters, sensation, physiological factors and performance compared with loose-fit breeches [4]. Essentially, the effects of compression garments during exercise remain incomplete.

Compression garments have also been recommended as a recovery tool due to many advantages. For instance, CGs decreased muscle soreness post 24 h of 25 repetitions on a modified preacher curl machine exercise [5], post 48 h of three treadmill runs at a 6% elevation rate [6] or after 72 h of a vertical jump test [7]. CGs also represented a positive effect on delayed onset muscle fatigue and swelling [8]. While the studies as mentioned above showed the usefulness of CGs on recovery, there are still many other studies which indicated that the extent of the recovery rate or peripheral fatigue following activity was not affected by CGs [9]. Therefore, the effects of CGs on recovery need further investigation.

Overall, many studies have shown that a cardiovascular function has some correlation to exercising performance and recovery. Substantial evidence exists from previous research demonstrates that the increase in heart rate variability (HRV) indicates lowering risk profiles and a higher number of risk factors related to the decrease of HRV [10]. Additionally, heart rate (HR) post-exercise can be an effective parameter in designing the training strategy of both non-athletes and athletes and plays an important role in performance’s monitoring [11, 12]. For instance, slower heart rate recovery (HRR) correlated to the higher risk of autonomic dysfunction endothelial dysfunction diabetes, cardiovascular mortality, and metabolic syndrome [13].

The increase of hypertension also relates to higher HR [14]. Moreover, a higher value of resting HR can predict the risk of increasing hypertension [15]. However, dynamic mechanisms of HR underlying the influence of compression garments on wearers are not well understood [16]. In this study, effects of CGs were investigated and focused on non-athletes only.

Additional parameters comprised ST, QRS, TpTe, QT and corrected of QT (QTc) were provided via ECG signals. Many previous investigations identified the strong relationship between the improvement in cardiac arrhythmias and prolongation of QTc or QT intervals [17]. An adverse effect on the performance in training of athletes has associated with the higher values of QT and QTc dispersion [18]. Similarly, research of Mandyam (2012) suggested that the development of incident atrial fibrillation and stroke were also affected by a longer QT interval [19]. Moreover, there are many studies associated with ST-elevation such as a relationship between ST-elevation and predicting left ventricular [20]. However, the ST intervals are still lacking investigation. To the best of our knowledge, although some studies relating to a monitoring system and an automatic wearable ECG classification were published [21], there is no research related to the effects of using CGs on ECG signals on both during and post-exercise.

This present study examines the underlying mechanism of cardiovascular effects associated with compression garments through exercise and recovery based on ECG signals.

Methods

Participants

Eight healthy volunteers [age: 25.1 ± 3.8-year-old, weight: 61.4 ± 13.7 kg (kg), height: 165.9 ± 8.3 cm (cm)], including three women and five men, attended for Experiment 1. Fourteen healthy subjects (age: 24.7 ± 4.5-year-old, height: 166.0 ± 7.6 cm, weight: 60.9 ± 12.0 kg), including seven women and seven men, conducted in Experiment 2.

All participants did not have a smoking history and were non-athletes. Exclusion criteria comprise respiratory, cerebrovascular and cardiovascular disease. The consumption of alcohol and caffeine were not allowed for 24 h before the trials. The volunteers also were required to have a healthy night sleep. All subjects finished consent forms and the questionnaire of a basic medical situation before the experiments. The protocols were accepted by The Human Ethics Committee at the University of Technology Sydney (UTS). (Study 1: 2014000844; Study 2: ETH16-0696). The detailed parameters of the subjects of the two studies are indicated in Table 1.

Table 1 Subjects characteristics

Test conditions and compression garments

Compression garments with the whole-body were used in the experiments. SportSkins Classic CGs come from Skins, Campbelltown, New South Wales, Australia. The compression garments comprised a long-sleeved top (neck–waist) and full-leg pants (waist–ankle). These components included 24% of Roica Spandex and 76% of Meryl Microfiber and Nylon. Based on participants’ body stature and mass, the corrected size garments were chosen for CGs according to the manufacturer’s suggestion. The NCGs were short and T-shirt which provided 0% level of pressure during the experiments. The volunteers used the same shoes and socks in both tests. The types of garments are shown in Fig. 1.

Fig. 1
figure 1

Subject use CGs (a1) and NCGs (a2). Position of lead II (b). Intervals detection (c). RR intervals (d)

Each subject came to the laboratory two separate days throughout the experiments period. Two kinds of garments (NCGs and CGs) were used randomly through two exercises on a treadmill. The normal range of temperature was set from 20 to 22 °C in both testing days.

Experimental design

On two different days, subjects conducted in two running trials with compression garments and without compression garments. Before the experiments, participants attended a rest of about 10 min. ECG of rest-data was calculated during this step from Thought Technology Ltd, Canada using a Flexcomp Infiniti Monitor. ECG electrodes were applied following the position of lead II (Fig. 1).

Experiment 1: Subjects conducted a 10-min running on 0% grade at six km/h of a treadmill. Following 2 min of running, volunteers were requested for data collection about 1.5 min (90 s) by stopping on the treadmill. Subsequently, the velocity was increased by 1 km/h following each pre-determined speed. The protocol continued when participants reached a velocity of 11 km/h. The data were collected as soon as subjects completed each speed during the stop.

Experiment 2: Subjects conducted a 10-min running on 0% grade at 6 km/h of a treadmill. Following every 2 min of running, the treadmill was speeded up by 1 km/h. The tests remained until the velocity of 11 km/h was finished. After that, participants took 2 h rest for recovery. ECG data were collected during the 2 h following the tests.

Data collection

The electrical signals, caused by heart muscle, were detected by the ECG sensor. The electrodes captured this electrical signal. After that, the sensor amplified and filtered the signal. Digital signals were then converted by the encoder. Computer processed, recorded and displayed this digital signal. The BioGraph Infiniti software was used for this step. All the progress was showed in Fig. 2.

Fig. 2
figure 2

ECG detection

The sample rate is at 2048 samples per second. The ECG amplifier is the ECG-Flex/Pro (SA9306M). Heartbeat was detected by collecting the electrical activity of the heart. The measurement’s unit is in millivolts (1 µV).

Notch filters were applied to avoid line interference noise. The power line frequency was removed by the ECG notch filter (60 or 50 Hertz). The distance of 3 meters (10 feet) from any radio transmitting devices and 1 m (3 feet) from any electronic equipment was required for the device.

Matlab was applied for statistical analysis and calculating peaks. Mean values (mean) and standard deviation (std) were assessed for all statistical results. The significant level was chosen when the value of p was lower than 0.05. All data analyses were conducted according to a version of Matlab 2016b.

HRV parameters were calculated throughout the 90 s data collection period. The intervals were assessed during the first 10 s immediate stop for analysis in Experiment 1.

The values are represented by an assessment of RR interval variability. The RR interval was calculated as the interval from the peak of one QRS complex to the peak for the next, shown in Fig. 2. HR was calculated from RR intervals (second), following Formula (1). ECG intervals including ST, TpTe, QRS, QT intervals were assessed from positions of Tp, Te, S, Q during 20 s following each recovery phase of 10-min in Experiment 2. Although many various formulas can calculate QTc, the Bazett was recommended as the most popular description to reveal the considerable findings. The mentioned calculation is represented in Formula (2).

$$ HR = \frac{60}{RR} $$
(1)
$$ QTc = \frac{QT}{{\sqrt[{}]{RR}}} $$
(2)

Results

Heart rate variability

Experiment 1, HRV parameters were considered within 90 s of the collection period. The results demonstrated some statically significant differences between using CGs and NCGs in HR after completing velocity of 7 km h−1 (NCGs: 126.2 ± 19.9, CGs: 115.2 ± 17.5, p = 0.0033), of 8 km h−1 (NCGs: 134.0 ± 23.2, CGs: 127.2 ± 20.2, p = 0.0473), of 9 km h−1 (NCGs: 144.0 ± 22.2, CGs: 132.1 ± 18.3, p = 0.0251), of 10 km h−1 (NCGs: 149.2 ± 21.4, CGs: 141.9 ± 17.9, p = 0.0804) and of 11 km h−1 (NCGs: 155.9 ± 17.1, CGs: 148.6 ± 12.4, p = 0.0165), following Fig. 3. Other parameters demonstrated a non-significant difference between the two groups of garments.

Fig. 3
figure 3

Heart rate during exercise in both groups of CGs and NCGs.*Significant difference (*p < 0.05, **p < 0.01)

Experiment 2, CGs group indicated significantly lower HR compared with NCGs group after recovery of 30-min (CGs: 92.1 ± 11.8, NCGs: 96.0 ± 11.5, p = 0.0485). Similarly, the significant influence of the use of CGs was revealed from the recovery of 80 min onwards. For instance, the results were demonstrated to have significant differences at the recovery of 80-min (NCGs: 85.4 ± 13.9, CGs: 80.1 ± 9.6, p = 0.0376), at the recovery of 90-min (NCGs: 85.2 ± 10.8, CGs: 79.5 ± 10.5, p = 0.0066), at the recovery of 100-min (NCGs: 83.1 ± 10.5, CGs: 79.1 ± 9.6, p = 0.0072), at the recovery of 110-min (NCGs: 82.8 ± 9.5, CGs: 76.9 ± 8.3, p = 0.001) and at the recovery of 120-min (NCGs: 82.3 ± 9.2, CGs: 76.4 ± 8.5, p = 0.0001). There was no significant change between using CGs and using non-CGs in other HRV values. The results are shown in Fig. 4.

Fig. 4
figure 4

Heart rate in both groups of CGs and NCGs post-exercise.*Significant difference (*p < 0.05, **p < 0.01, ***p < 0.001)

Intervals

Experiment 2, the parameters concluded a significantly lower corrected QT intervals in the group using CGs at 90-min recovery (NCGs: 403.8 ± 21.6 ms, CGs: 391.3 ± 18.0 ms, p = 0.0035). The QTc was also influenced by the utilization of CGs with p = 0.0369 (NCGs: 405.3 ± 29.8 ms, CGs: 391.0 ± 13.1 ms) at 100-min recovery. Consistent with this, the value was affected by using CGs at the recovery of 110-min (NCGs: 399.1 ± 22.5 ms, CGs: 388.9 ± 16.5 ms, p = 0.0073). At the recovery of 120-min, there was a significant difference between two groups of garments (NCGs: 398.3 ± 18.1 ms, CGs: 387.6 ± 16.7 ms, p = 0.0128). The findings are shown in Fig. 5.

Fig. 5
figure 5

QTc in both groups of CGs and NCGs post-exercise.*Significant difference (*p < 0.05, **p < 0.01)

After 60-min recovery, ST intervals reported a significant difference with p = 0.0269 (NCGs: 274.3 ± 28.1 ms, CGs: 282.6 ± 26.6 ms). Additionally, the application of CGs at 90-min recovery also determined a substantial difference in p = 0.0452 compared with NCGs with CGs: 289.2 ± 25.5 ms and NCGs: 279.4 ± 30.9 ms. At the recovery time after 100 min and 110 min, the finding demonstrated p < 0.001 with NCGs: 278.2 ± 28.3, CGs: 287.5 ± 28.0 ms (p = 0.0008) and NCGs: 279.5 ± 27.6 ms, CGs: 292.2 ± 28.1 ms (p = 0.0005), respectively. After 120 min of the recovering phase, the results reported a significant change in ST intervals between NCGs: 281.7 ± 27.2 ms and CGs: 290.8 ± 27.2 ms with p = 0.0216. The findings are shown in Fig. 6.

Fig. 6
figure 6

ST in both groups of CGs and NCGs post-exercise.*Significant difference (*p < 0.05, **p < 0.01)

Discussion

At the end of the running protocol, Experiment 1 indicated a significantly lower HR (p < 0.05). The present result is consistent with previous studies. For instance, in a study of well-trained runners, at a low intensity from 8 to 10 km/h, HR showed no significant difference between CGs and NCGs. However, at the high intensity from 12 to 18 km/h, HR showed a significantly lower value in CGs group than NCGs group [22]. In the current research, the significant difference in HR was found at a velocity of 7 km/h afterward. The level of intensity was lower than the abovementioned study because all volunteers are non-trained runners (non-athletes). Based on the observation, the significantly lower heart rate could be found when participants reached the high intensity and were exhausted. Consistent with this, earlier research of incremental cycling bouts demonstrated a significantly lower value of HR when wearing CGs compared with the application of regular above-knee cycling shorts [23].

Based on the results of higher performance in a study of an archery-competition, the lower HR may have a relationship to lower stress and more concentration [16]. This finding is also consistent with research involving ten golfers and eleven pitchers relating to the enhancement of performance in driving accuracy, fastball accuracy, chipping accuracy and shot accuracy by the use of CGs [24].

In Experiment 2, both NCGs and CGs indicated a decrease in HR throughout the recovering time. Through the first 80-min recovery, the application of CGs demonstrated non-significant difference compared with NCGs. This investigation is consistent with some previous research which reported no alteration between CGs and NCGs in heart rate of 6 steps incremental experiment after 20-min recovery on a kayak ergometer [25] and 3 running power trials of post 20-min intermittent maximal anaerobic [26]. Similarly, research of 40 km cycling time trial performance [27] and a match simulation exercise [28] determined a non-significant difference after half-hour recovery in HR in a comparison between wearing lower-body CGs (full long-leg pants) and using NCGs. Additionally, after 60-min recovery of a maximal cycling ergometer protocol, the findings also suggested no effects by using compression stocking compared with NCGs [23]. However, the results showed the significant lower HR (p < 0.05) after 80-min recovery in the participants with compression clothing compared with without compression garments.

Enhancement of redistribution of blood flow and blood flow was associated with a lower mean heart rate [24]. Additionally, glycogen depletion can be caused by high-intensity exercise relating to inadequate muscle glycogen synthesis. However, the issue can be solved by the development of blood flow via the utilization of CGs [27]. The quicker recovery of HR after the recovery of 80-min using CGs supports many previous studies associated with the suggestion of CGs as a recovering strategy [29], [30]. This investigation of the current research is that compression garments may have benefits after the running exercise in HR.

The present study investigated a significant lower QTc when using compression garments compared with non-compression garments after a 90-min of recovering time (p < 0.05). Similarly, an earlier study indicated a shorter value of QTc at the end of a running experiment in CGs group [31]. The previous finding proved that the longer QTc related to the development of atrial fibrillation and incidence of stroke [19]. Moreover, a significant increase in cardiac arrhythmias also associated with the prolongation of QTc [18]. The current results are consistent with our previous observation suggesting that the pressure of CGs may effect on QTc parameters [31]. Based on the findings of lower QTc, CGs may have a positive influence on the recovering phase.

In this present study, ST intervals indicated the decrease at the end of the tests and the increase of normal parameters throughout the recovering time. The application of CGs showed a higher value of ST intervals during the recovery. At 60-min of recovering time and after 90-min recovery, CGs demonstrated significantly longer ST intervals compared with NCGs (p < 0.05). Specifically, participants wearing CGs reached the normal parameters of ST intervals quicker than using non-CGs. Evidence of previous findings concluded that ST intervals correlated with low diastolic blood pressure (DBP) and high systolic blood pressure (SBP) [32]. The observation of the current result supported these previous results associated with the influence of CGs and indicated significantly higher SBP parameters during the standing phases and the supine [33].

Conclusion

Based on the mentioned positive influence of cardiovascular function, the application of whole body compression garments may support both exercise and recovery phase. The current findings included a quicker recovery in heart rate, ST, and QTc. Between activity episodes, compression garments should be suggested to athletes and coaches for improving the recovery. The above observations can increase the training’s quality and lead to enhancement in subsequent performance. Further observations are required to investigate the underlying mechanisms of compression garments on recovery strategy and performance in athletes.

References

  1. Hsu W-C, Tseng L-W, Chen F-C, et al. Effects of compression garments on surface EMG and physiological responses during and after distance running. J Sport Health Sci. 2017. https://doi.org/10.1016/j.jshs.2017.01.001.

    Article  Google Scholar 

  2. Hooper DR, Dulkis LL, Secola PJ, et al. Roles of an upper-body compression garment on athletic performances. J Strength Condition Res. 2015;29(9):2655–60.

    Article  Google Scholar 

  3. Born D-P, Holmberg H-C, Goernert F, et al. A novel compression garment with adhesive silicone stripes improves repeated sprint performance-a multi-experimental approach on the underlying mechanisms. BMC Sports Sci Med Rehabil. 2014;6(1):21.

    Article  Google Scholar 

  4. Venckūnas T, Trinkūnas E, Kamandulis S, et al. Effect of lower body compression garments on hemodynamics in response to running session. Sci World J. 2014;2014:353040.

    Article  Google Scholar 

  5. Kim J, Kim J, Lee J. Effect of compression garments on delayed-onset muscle soreness and blood inflammatory markers after eccentric exercise: a randomized controlled trial. J Exerc Rehabil. 2017;13(5):541.

    Article  Google Scholar 

  6. Struhár I, Kumstát M, Králová DM. Effect of compression garments on physiological responses after uphill running. J Human Kinetics. 2018;61(1):119–29.

    Article  Google Scholar 

  7. Maloon GE. Self-myofascial release foam rolling vs. compression garments: which is more effective at reducing DOMS? San Francisco State University, 2017.

  8. Marqués-Jiménez D, Calleja-González J, Arratibel-Imaz I, et al. Influence of different types of compression garments on exercise-induced muscle damage markers after a soccer match. Res Sports Med. 2017;1:27–42.

    Google Scholar 

  9. Goto K, Mizuno S, Mori A. Efficacy of wearing compression garments during post-exercise period after two repeated bouts of strenuous exercise: a randomized crossover design in healthy, active males. Sports Med Open. 2017;3(1):25.

    Article  Google Scholar 

  10. Thayer JF, Yamamoto SS, Brosschot JF. The relationship of autonomic imbalance, heart rate variability and cardiovascular disease risk factors. Int J Cardiol. 2010;141(2):122–31.

    Article  Google Scholar 

  11. Lamberts R, Swart J, Capostagno B, et al. Heart rate recovery as a guide to monitor fatigue and predict changes in performance parameters. Scand J Med Sci Sports. 2010;20(3):449–57.

    Article  Google Scholar 

  12. Aubry A, Hausswirth C, Louis J, et al. The development of functional overreaching is associated with a faster heart rate recovery in endurance athletes. PLoS ONE. 2015;10(10):e0139754.

    Article  Google Scholar 

  13. Dimkpa U. Post-exercise heart rate recovery: an index of cardiovascular fitness. J Exe Physiol. 2009;12:19–22.

    Google Scholar 

  14. Jae SY, Carnethon MR, Heffernan KS, et al. Slow heart rate recovery after exercise is associated with carotid atherosclerosis. Atherosclerosis. 2008;196(1):256–61.

    Article  Google Scholar 

  15. Inoue T, Iseki K, Iseki C, et al. Higher heart rate predicts the risk of developing hypertension in a normotensive screened cohort. Circ J. 2007;71(11):1755–60.

    Article  Google Scholar 

  16. Kaikkonen P, Rusko H, Martinmäki K. Post-exercise heart rate variability of endurance athletes after different high-intensity exercise interventions. Scand J Med Sci Sports. 2008;18(4):511–9.

    Article  Google Scholar 

  17. Felix M, Dimitriu AG. Original study the effects of physical training over QT dispersion and QTc dispersion to the beginner sportsmen. Fascicula. 2011;2:105–10.

    Google Scholar 

  18. Mandyam MC, Soliman EZ, Alonso A, et al. A longer Qt interval is associated with increased risk of atrial fibrillation. Circulation. 2012;126(21 Supplement):A16639.

    Google Scholar 

  19. Benn M, Nordestgaard B, Nielsen S, et al. Short and long QTc interval, cardiovascular disease and mortality. Eur Heart J. 2013;34(suppl 1):P4957.

    Article  Google Scholar 

  20. Xu L, Huang X, Ma J, et al. Value of three-dimensional strain parameters for predicting left ventricular remodeling after ST-elevation myocardial infarction. Int J Cardiovasc Imaging. 2017;33(5):663–73.

    Article  Google Scholar 

  21. Xia Y, Zhang H, Xu L, et al. An automatic cardiac arrhythmia classification system with wearable electrocardiogram. IEEE Access. 2018;6:16529–38.

    Article  Google Scholar 

  22. Duffield R, Cannon J, King M. The effects of compression garments on recovery of muscle performance following high-intensity sprint and plyometric exercise. J Sci Med Sport. 2010;13(1):136–40.

    Article  Google Scholar 

  23. Rimaud D, Messonnier L, Castells J, et al. Effects of compression stockings during exercise and recovery on blood lactate kinetics. Eur J Appl Physiol. 2010;110(2):425–33.

    Article  Google Scholar 

  24. Driller MW, Halson SL. The effects of wearing lower body compression garments during a cycling performance test. Int J Sports Physiol Perform. 2013;8:300–6.

    Article  Google Scholar 

  25. Dascombe B, Laursen P, Nosaka K, et al. No effect of upper body compression garments in elite flat-water kayakers. Eur J Sport Sci. 2013;13(4):341–9.

    Article  Google Scholar 

  26. Leicht AS, Crowe MJ, Sinclair WH, et al. Influence of compression garments on physiological and perceptual responses during intermittent anaerobic running and recovery. In: Proceedings of ACSM conference on integrative physiology of exercise, Miami Beach, FL, USA, 22–25 September 2010.

  27. de Glanville KM, Hamlin MJ. Positive effect of lower body compression garments on subsequent 40-kM cycling time trial performance. J Strength Condition Res. 2012;26(2):480–6.

    Article  Google Scholar 

  28. Pruscino CL, Halson S, Hargreaves M. Effects of compression garments on recovery following intermittent exercise. Eur J Appl Physiol. 2013;113(6):1585–96.

    Article  Google Scholar 

  29. Davies V, Thompson KG, Cooper S-M. The effects of compression garments on recovery. J Strength Condition Res. 2009;23(6):1786–94.

    Article  Google Scholar 

  30. Jakeman JR, Byrne C, Eston RG. Lower limb compression garment improves recovery from exercise-induced muscle damage in young, active females. Eur J Appl Physiol. 2010;109(6):1137–44.

    Article  Google Scholar 

  31. Nguyen LTN, Eager D, Nguyen HT. Effect of wearing whole body compression garments on cardiovascular function using ECG signals. Telehealth Assist Technol. 2016;26(3.1):24–9.

    Google Scholar 

  32. Yeom H, Kim HC. MPS 08-04 The association between ST interval and blood pressure in a general population of Korea. J Hypertens. 2016;34:e261.

    Article  Google Scholar 

  33. Privett SE, George KP, Whyte GP, et al. The effectiveness of compression garments and lower limb exercise on post-exercise blood pressure regulation in orthostatically intolerant athletes. Clin J Sport Med. 2010;20(5):362–7.

    Article  Google Scholar 

Download references

Authors’ contributions

LTNN carried out the experimental studies, participated in the analysis of the results, and drafted the manuscript. DE participated in the analysis and confirmed the validity of the results. As the senior researcher, HN planned out the studies, coordinated the ethics application, provided guidance in the analysis of the results, and confirmed the validity of the results. All authors read and approved the final manuscript.

Acknowledgements

The authors would like to thank the participants in this current research and Mr. Raymond Clout for his technical support.

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

Please contact the author for data requests.

Consent for publication

Written informed consent for publication of the data used in the current research was collected from non-athletes before the examination. All personal data were removed from the data files.

Funding

Not applicable.

Ethics approval and consent participate

The studies have been approved by the Human Ethics Committee at University of Technology Sydney, Australia. Study 1: UTS HREC REF No. 2014000844; Study 2: UTS HREC REF No. ETH16-0696.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hung Nguyen.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nguyen, L.T.N., Eager, D. & Nguyen, H. The relationship between compression garments and electrocardiogram signals during exercise and recovery phase. BioMed Eng OnLine 18, 27 (2019). https://doi.org/10.1186/s12938-019-0645-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12938-019-0645-2

Keywords