Effect of glutaraldehyde based cross-linking on the viscoelasticity of mitral valve basal chordae tendineae

Abstract

Background

Mitral valve failure can require repair or replacement. Replacement bioprosthetic valves are treated with glutaraldehyde prior to implantation. The aim of this study was to determine the changes in mechanical properties following glutaraldehyde fixation of mitral valve chordae.

Methods

To investigate the impact of glutaraldehyde on mitral valve chordae, 24 basal chordae were dissected from four porcine hearts. Anterior and posterior basal (including strut) chordae were used. All 24 chordae were subjected to a sinusoidally varying load (mean level 2N, dynamic amplitude 2N) over a frequency range of 0.5–10 Hz before and after glutaraldehyde treatment.

Results

The storage and loss modulus of all chordal types decreased following glutaraldehyde fixation. The storage modulus ranged from: 108 to 119 MPa before fixation and 67.3–87.4 MPa following fixation for basal chordae; 52.3–58.4 MPa before fixation and 47.9–53.5 MPa following fixation for strut chordae. Similarly, the loss modulus ranged from: 5.47 to 6.25 MPa before fixation and 3.63–4.94 MPa following fixation for basal chordae; 2.60–2.97 MPa before fixation and 2.31–2.93 MPa following fixation for strut chordae.

Conclusion

The viscoelastic properties of mitral valve chordae are affected by glutaraldehyde fixation; in particular, the reduction in storage moduli decreased with an increase in chordal diameter.

Background

The mitral valve is an atrioventricular valve located in the left side of the heart [1, 2]. Its function is to enable the unidirectional flow of blood between the left atrium and left ventricle [3,4,5]. The subvalvular apparatus includes the anterior and posterior leaflets, two papillary muscles and chordae tendineae [4, 6,7,8]. Chordae connect the papillary muscles to the anterior and posterior leaflets [9]. Chordae can be categorised as marginal or basal. Marginal chordae insert into the free edge of the leaflet and are typically thinner and less extensible than other chordae [6]. Basal chordae are thicker in diameter and insert away from the free edge of the leaflet but between the free edge and the mitral annulus [10]. Strut chordae are thick basal chords which insert into the anterior leaflet only [6, 9]; whereas, commissural chordae branch radially and insert into both leaflets at the commissures between anterior and posterior leaflets [9, 11]. Distinct chordal categories enable specific mechanisms of mitral valve competence to be preserved [12].

The importance of the mechanical and material properties of chordae tendineae relates to its significance in improving the surgical procedure for the correction of mitral valve regurgitation. Chordal failure can lead to the need for surgical correction, of which two approaches are available: surgical repair [10, 13], or surgical replacement. Surgical repair allows for retention of native chordae, providing a less invasive procedure [6, 10, 14,15,16]. Alternatively, surgical replacement involves replacing the natural valve with an alternative including bioprosthetic or mechanical valves, and homografts [17]. Bioprosthetic valves may be recommended for older patients to avoid the need of anticoagulants associated with mechanical valves [18]. Due to the prevalence of heart failure in the aged population, there is interest in the functionality and durability of bioprosthetic replacement [19].

Clinically, mitral valves are replaced using porcine aortic valves where they are inserted into the mitral valve position. Due to this, studies have predominantly focused on aortic valve replacements where tissues are treated with glutaraldehyde as a fixative. Glutaraldehyde is used to prevent tissue degradation, reduce antigenicity, and improve durability [17, 20,21,22]. However, such treatments are associated with structural alterations, notably an increase in tissue stiffness [20, 22]. The effects of glutaraldehyde treatment on biological tissues are the focus of several studies due to its use in bioprostheses. However, studies investigating the effects of glutaraldehyde treatment on porcine aortic valves have led to seemingly contradictory findings. For instance, under tension, glutaraldehyde has little impact on tissue stiffness [23]. Conversely, under bending and shear tissue stiffness has been seen to increase [22, 24]. Uniaxial testing also highlighted a decrease in tissue extensibility following glutaraldehyde fixation [25]. Though aortic valves are commonly used for mitral valve replacement, studies investigating the use of bioprosthetic alternatives have concluded positive results; for instance, the use of bovine pericardium as stentless mitral valve replacements [26].

Understanding the effects of glutaraldehyde on the mechanical properties of chordae is crucial in determining the feasibility of using, say, porcine mitral valves for bioprosthetics. However, studies evaluating the effects of glutaraldehyde treatment of mitral valve chordae have been more limited, with one study finding that glutaraldehyde treatment reduced the ultimate tensile strength (UTS) of marginal chordae [17]. Another study demonstrated that glutaraldehyde fixation only affected a portion of the non-linear stress–strain relationship of chordae [27]. Further, chordae are viscoelastic and exposed to dynamic loading conditions [11, 28, 29] yet the viscoelastic properties of glutaraldehyde treated mitral valve chordae are unknown.

A viscoelastic material can be characterised by its storage and loss modulus [11, 30, 31]. The elastic properties of the material are characterised by its ability to store energy, known as the storage modulus. The loss modulus of a material characterises the ability of the material to dissipate energy. Dynamic Mechanical Analysis (DMA) is a technique which enables the characterisation of viscoelastic materials, in terms of storage and loss moduli, under dynamic loading conditions [32,33,34,35].

The aim of this study is to assess the effects of glutaraldehyde fixation on the viscoelastic properties of mitral valve chordae tendineae. The viscoelastic properties of chordae have been characterised using DMA at frequencies relevant to physiological and patho-physiological heart rates [11]. For this study, porcine basal chordae have been assessed. However, given the potential for regional variation in mitral valve mechanical properties [36], anterior and posterior basal chordae have been compared, as well as basal and strut chordae, in terms of their dynamic viscoelasticity.

Methods

Tissue preparation

Chordae were obtained from four porcine hearts. Hearts were frozen upon extraction and delivered frozen and sealed by Fresh Tissue Supplies (Fresh Tissue Supplies, East Sussex, UK); no animals were sacrificed specifically for this study. They were stored at – 40 °C wrapped in tissue paper coated in Ringer’s solution following established procedures [11, 37]. Freezing biological tissues using this method does not adversely affect their mechanical properties [38, 39]. Six chordae (2 strut, 2 anterior basal, and 2 posterior basal) from each heart were obtained for testing (Fig. 1).

Fig. 1
figure 1

Porcine specimen. The different chordal types have been identified on the anterior mitral valve leaflet

DMA

Twenty-four chordae were subjected to DMA using a Bose ElectroForce 3200 testing machine (Bose Corporation, Electroforce Systems Group, Minnesota, USA) previously used for mechanical testing of mitral valve chordae [11, 37] and leaflets [36]. Length and chordal diameter were measured using ImageJ software (ImageJ 1.0, Maryland, USA); images were acquired of chordae when placed within the grips of the testing machine, from which measurements were taken using ImageJ. Chordae were gripped by using emery paper (P400 with two smaller rectangles of P240) to coat the grips of the testing machine, similar to that used previously [11, 33, 37].

Chordae viscoelastic properties are frequency dependent [11], they were sinusoidally loaded between 1 and 3N over a frequency sweep. Specimens were preloaded sinusoidally at 1 Hz for 100 cycles. Subsequently, a frequency sweep of 0.5, 1, 1.2, 3.5, 5, 7, and 10 Hz was used for testing. Chordae were tested at 1.2 Hz to correspond to the 70 beats/min used to represent in vivo conditions, whilst 3.5 and 7 Hz were estimated to match maximum in vivo strain rates [11, 40, 41]. Mitral valve components have been characterised up to 10 Hz of loading [36], hence, the highest loading frequency. Specimens were tested under the same loading protocols, and frequency sweep, before and after glutaraldehyde treatment (“Fixation of chordae” section). Specimens were maintained hydrated throughout testing using Ringer’s solution [11, 33, 37].

At each frequency of loading, an oscillating force (1–3 N) was applied and the out-of-phase displacement was measured and recorded using WinTest 4.1 DMA software (Bose Corporation, Electroforce Systems Group, Minnesota, USA). The WinTest DMA software conducted a Fast Fourier Transform (FFT) of the force and displacement waves at each frequency. The magnitude of the load (F*), displacement (d*), and the phase angle (δ) were then determined [31]. Subsequently, the complex stiffness is calculated using Eq. (1) and storage and loss stiffness are derived (Eqs. 2 and 3, respectively).

$${k^*} = \frac{F^*}{d^*}$$
(1)
$$k' = {k^*}\cos \delta \;$$
(2)
$$k'' = {k^*}\sin \delta$$
(3)

where \(k'\) and \(k''\) refer to the storage and loss stiffness respectively, and k* is the magnitude of the complex stiffness [11, 31]. The storage and loss moduli are then calculated by normalising for a given shape, using a shape factor (S); Eqs. 4 and 5, respectively.

$$E' = \frac{{{k^*}\cos \delta }}{S}\;$$
(4)
$$E'' = \frac{{{k^*}\sin \delta }}{S}\;$$
(5)

where \(E'\) and \(E''\) are the storage and loss moduli, respectively. Chordae have been approximated to be cylindrical [11, 37], and the shape factor was defined by:

$$S = \frac{{\pi {d^2}}}{4l}\;$$
(6)

where d is the average diameter of the chord and l is its length [11, 37].

Fixation of chordae

Chordae were fixed using 0.6% glutaraldehyde (Fluka Analytical, Sigma Aldrich, St Louis, MO, USA) solution diluted with 0.2 M PBS (Phosphate Buffered Solution; Sigma Aldrich, Darmstadt, Germany); as previously used with biological tissues [22, 27, 42]. Bioprosthetic replacements are typically fixed at 0.6% glutaraldehyde concentration [17, 25, 27]. Samples were submerged within 5 ml of the glutaraldehyde solution for 1 h as recommended elsewhere [43,44,45]. Following fixation, specimens were washed thrice in 0.2 M PBS for 10 min to remove excess glutaraldehyde [46].

Data analysis

Statistical analysis was performed using SigmaPlot v12.0 (SYSTAT, San Jose, CA, USA). Linear regression analysis was used to determine the frequency dependent nature of the storage and loss moduli of each chord (p < 0.05). Significant differences (p < 0.05) in the storage and loss moduli pre-and post-fixation were determined using the Wilcoxon Signed Rank test for paired non-parametric data. Non-parametric tests were used due to the small sample size (four hearts; i.e. four independent measurements).

A Kruskal–Wallis one-way analysis of variance (ANOVA) on ranks was used to determine any differences between basal chordal types (i.e. anterior, posterior and strut chordae). Significant differences (p < 0.05) were evaluated using Tukey’s multiple comparisons test.

Results

Storage modulus

The storage modulus for all specimens was frequency dependent, where the storage modulus increased with increasing frequency. Regression analysis showed that the storage modulus, \(E'\), was proportional to the natural logarithm of frequency, f (Eq. 7). The storage modulus is represented by the function below where A and B are coefficients (Table 1).

Table 1 Linear regression results for the storage and loss modulus of normal chordae
$$E' = A\ln \left( f \right) + B\;$$
(7)

The mean storage modulus increased: from 52.3 MPa at 0.5 Hz to 58.4 MPa at 10 Hz (Fig. 2a) for anterior strut chordae; from 108 MPa at 0.5 Hz to 118 MPa at 10 Hz for anterior basal chordae (Fig. 2b); from 109 MPa at 0.5 Hz to 119 MPa at 10 Hz for posterior basal chordae (Fig. 2c).

Fig. 2
figure 2

The effects of glutaraldehyde fixation on the storage modulus of a Anterior strut chordae, b Anterior basal chordae, and c Posterior basal chordae. Points represent the sample means with 95% confidence intervals included. Means were calculated based on 8 samples from 4 different hearts for each chordal type. Black dots refer to unfixed chordae, while white dots represent glutaraldehyde treated chordae

Fixation decreased the storage modulus for all chordal types across the frequency range (Tables 2 and 3), but storage moduli still increased with frequency. Following glutaraldehyde fixation, the mean storage modulus for anterior strut chordae increased from 47.9 MPa at 0.5 Hz to 53.5 MPa at 10 Hz. The mean storage modulus of fixed anterior and posterior basal chordae increased from 67.3 MPa and 77.7 MPa at 0.5 Hz, to 75.7 MPa and 87.4 MPa at 10 Hz, respectively. The mean storage modulus of anterior strut chordae decreased by 4.2–4.9 MPa over the frequency range following glutaraldehyde fixation.

Table 2 Viscoelastic properties of normal chordae under sinusoidally varying load across a range of frequencies
Table 3 Viscoelastic properties of chordae treated with glutaraldehyde under a sinusoidally varying load across a range of frequencies

Fixation with glutaraldehyde resulted in a significant (p < 0.05) decrease in the mean storage modulus of anterior basal chordae of 40.5–42.3 MPa over the frequency range. A mean storage modulus decrease of 30.8–31.9 MPa over the frequency range was present because of glutaraldehyde fixation for posterior basal chordae. The influence of chordal diameter on the decrease in storage modulus was evaluated at 1 Hz (Fig. 3). As chordal diameter increases, the change in storage modulus (\(\Delta {E^{\prime}}\)) due to fixation decreases.

Fig. 3
figure 3

Regression of chordal diameter against reduction in storage modulus (percentage,  %) for 24 specimens at 1 Hz. Linear regression analysis determined the relationship to be significant (p = 0.018) with an R2 value of 0.2293. Note, any percentage reduction below 0% refers to an increase in storage modulus

Prior to fixation there were significant (p < 0.05) differences in the storage modulus of anterior strut chordae and both anterior and posterior basal chordae. The storage modulus of basal chordae was 56.2 MPa at 0.5 Hz, and 60.9 MPa at 10 Hz which was greater than that of anterior strut chordae (Table 2). Following fixation, there were no statistically significant differences between each of the chordal types (p > 0.05).

Loss modulus

Regression analysis for the loss modulus demonstrated that 10 chordae were frequency independent (p > 0.05) whilst the remaining 14 were frequency dependent (p < 0.05). From the 14 frequency dependent results, 12 increased with increasing frequency, whilst two decreased. The loss modulus is represented by Eq. 8 below, where C and D are coefficients (Table 1).

$$E'' = C\ln ( f) + D$$
(8)

The mean loss modulus ranged from: 2.60 MPa at 0.5 Hz, to 2.97 MPa at 7 Hz (Fig. 4a) for normal anterior strut chordae; 5.78 MPa at 0.5 Hz, to 6.25 MPa at 7 Hz (Fig. 4b) for normal anterior basal chordae; 5.47 MPa at 0.5 Hz, to 5.77 MPa at 7 Hz (Fig. 4c) for normal posterior basal chordae.

Fig. 4
figure 4

The effects of glutaraldehyde fixation on the loss modulus of a Anterior strut chordae, b Anterior basal chordae, and c Posterior basal chordae. Points represent the sample means with the 95% confidence intervals included. Means were calculated based on 8 samples from 4 different hearts for each chordal type. Black dots refer to unfixed chordae, while white dots represent glutaraldehyde treated chordae

Fixation decreased the loss modulus of all chordae (Fig. 4) whilst affecting the frequency dependency. Following fixation, 22 specimens were frequency dependent (Eq. 8) while 2 remained frequency independent. The mean loss modulus of fixed anterior strut chordae ranged from 2.31 MPa at 0.5 Hz, to 2.93 MPa at 7 Hz. Following fixation, the mean loss modulus of anterior and posterior basal chordae ranged from 3.63 MPa and 4.29 MPa at 0.5 Hz, to 4.08 MPa and 4.94 MPa at 7 Hz (Fig. 4).

Fixation resulted in a significant (p < 0.05) decrease in the mean loss modulus of 2.15–2.17 MPa over the frequency range for anterior basal chordae. Fixation also resulted in a significant (p < 0.05) decrease in the mean loss modulus for anterior strut chordae of 0.29 MPa at 0.5 Hz. The decrease in loss modulus following fixation for the remaining results was not significant (p > 0.05).

For unfixed chordae, the minimum difference in loss modulus (\(\Delta {E^{''}}\)) between chordal types was 3.18 MPa at 0.5 Hz and the maximum difference 3.28 MPa at 7 Hz. This was reduced to a minimum difference of 1.97 MPa at 1 Hz and a maximum difference of 2.02 MPa at 1.2 Hz following glutaraldehyde fixation. Although the difference in loss modulus was not significant (Tables 2 and 3) before or after glutaraldehyde fixation, it can be observed from Fig. 4 that the loss modulus for anterior strut chordae was lower over the frequency range than that of other basal chordae.

Discussion

Overview

The effects of glutaraldehyde fixation on the viscoelastic properties of mitral valve chordae tendineae have been characterised. Both the storage and loss modulus decreased following glutaraldehyde fixation. The storage modulus of normal and fixed chordae were frequency dependent. Prior to fixation, the loss modulus of chordae varied across the range of specimens, showing a mix of frequency dependent and frequency independent results. Over a wider range of frequencies, it is possible that a correlation between loss modulus and frequency might be more evident. Following fixation, the overall trend showed chordae as frequency dependent, with the loss modulus increasing with frequency. The storage and loss modulus of basal chordae was larger than that of the anterior strut chordae before and after fixation. As diameter increased, the change in storage modulus due to fixation decreased.

Effect of cross-linking on the dynamic viscoelasticity of chordae

The viscoelastic properties of mitral valve chordae have previously been characterised by their storage and loss moduli [11, 28]. Wilcox et al. [11] found that the storage modulus increased with frequency, whilst Lim et al. [28] found it to be frequency independent. The results from our current study are consistent with the findings that the storage modulus increases with frequency. However, the results are somewhat contradictory with respect to the loss modulus. Lim et al. [28] described the loss modulus to decrease with frequency, explaining that this would allow the valve to ensure complete closure at high heart rates. The results from our study are inconsistent with Wilcox et al. [11] and Lim et al. [28], such that the loss modulus of 12 specimens increased with increasing frequency. Alternatively, 10 specimens were frequency independent, consistent with Wilcox et al. [11], whilst the remaining 2 decreased with increasing frequency. Though reasons for this inconsistency are not obvious, the discrepancy obtained by Lim et al. [28] could be the result of human specimens being used as opposed to porcine specimens used by ourselves and Wilcox et al. [11].

For all chordae, the storage modulus was greater than that of the loss modulus across all frequencies, a trend observed in previous investigations [11, 28]. However, the magnitude of the results differs. Wilcox et al. [11] noted that small increases in load result in large increases in the storage modulus. The results from the investigation are within the range of 52.3–119 MPa for the storage modulus and 1.24–4.31 MPa for the loss modulus. The storage moduli are within the range identified in literature [11, 28] and align closely with the range identified by Lim et al. [28] between 20 and 140 MPa.

Glutaraldehyde fixation decreased the storage and loss modulus of chordae. Since knowledge of the effects of glutaraldehyde treatment on the dynamic properties of mitral valve chordae is limited, the following hypothesis is proposed. A link between the fibril crimp of collagen and their extensibility has been characterised [6]. It was shown that thicker chordae had a smaller crimp period than thinner chordae, and thus were more highly crimped. This has been hypothesised to characterise differences between the storage modulus of thick and thin chordae, whereby thicker chordae have a lower storage modulus [11]. Glutaraldehyde fixation is assumed to cause a chemical reaction between the aldehyde groups of glutaraldehyde and the ε-amine groups of lysine and hydroxylysine present in collagen [47, 48]. The result of this reaction is the formation of cross-links between the collagen fibres, which is thought to increase the fibril crimp, providing an explanation for the results (i.e. decrease in storage modulus). Further, since fixation is thought to increase the fibril crimp [23], we can extrapolate from these inferences to hypothesise that as thinner chordae become more predisposed to additional cross-linking, they experience a greater change in storage modulus; consistent with our current results. Thus, assuming fixation causes the chordae to become more crimped, we would expect the storage modulus to decrease due to fixation, this is reflected in the results for unfixed and fixed chordae.

Our results have shown that all three chordal types tested exhibited a decrease in storage modulus. However, the magnitude of the decrease varied. Although strut chordae are categorised as a type of basal chord, their diameter is larger, with strut diameter ranging from 0.7 to 1.1 mm as opposed to the diameter for other basal chordae ranging from 0.41 to 0.84 mm [11, 37]. Liao and Vesely [6] used a simple model to detail the effects of chordal diameter on possible linkages between fibrils. They concluded that the smaller the diameter, the greater number of fibril linkages that can occur. Consequently, the magnitude of fibril linkages that could occur within basal chordae would be larger than strut chordae. Thus, we hypothesise that increased fibril linkages would be expected to increase the storage moduli of chordae; the caveat being that if fibril linkages are increased via a chemical fixation process, then there may be a trade-off with changes in crimp (as described above).

The results from this study clearly show that the loss modulus of chordae decreased following fixation with glutaraldehyde. To explain these changes, we can consider the impact on the fibre matrix. It has previously been suggested that the loss modulus of chordae is related to the fibre matrix interactions [11], though this has not been investigated. Previously, we have only considered the impact of glutaraldehyde on the collagen molecules, however, since research into the effects of fixation of chordae is limited, evidence to support this is not currently conclusive. Since chordae have a complex structure made up of fibres (collagen; 65% dry weight) and elastin (9.7% dry weight) as well as proteoglycans (glycosaminoglycans; GAGs), the effects of fixation on more than just the collagen [8, 11, 49] should potentially be considered. Consequently, it is possible that these changes in the loss modulus because of glutaraldehyde fixation are due to changes occurring during potential fibre–GAGs interaction, not just the collagen molecules [11], though evidence to support this is currently limited.

Clinical implications

Glutaraldehyde fixation alters the viscoelastic properties of chordae. Clinically, this would affect the functionality of a putative bioprosthetic replacement. Though marginal chordae were not considered during this investigation, it is likely that fixation would affect the tissues similarly. However, the crimp period of marginal chordae is larger than that of basal and strut chordae [6]. Consequently, fixation could cause marginal chordae to become more highly crimped. This is further supported by the hypothesis that chordae with smaller diameters are susceptible to larger amounts of fibril linkages. Thus, it is possible the reduction in stiffness of marginal chordae would exceed that of basal and strut chordae. This is of importance due to the function of marginal chordae in ensuring valve closure [11, 40, 50]. Though a decrease in stiffness for basal and strut chordae will be less critical, a decrease in marginal chordae stiffness could result in insufficient closure, consequently leading to valve failure [12].

Future studies considering the effect of fixation on chordal types would allow for complete characterisation of the effects of glutaraldehyde fixation on mitral valve chordae. Since alternative fixatives are under investigation, considering the effect of these alternatives on the dynamic properties would be of importance for bioprosthetic functionality [17].

Conclusion

Fixation with glutaraldehyde decreased the magnitude of storage and loss properties of mitral valve basal chordae tendineae, which characterise its viscoelasticity. The storage and loss moduli of chordae decreased following glutaraldehyde fixation for all basal chordal types. The storage modulus of chordae before and after fixation remained frequency dependent. However, following fixation the frequency-dependency of the loss moduli of chordae was altered to a, typically, increased loss modulus with frequency.

References

  1. Guyton AC, Hall JE. Textbook of medical physiology. Philadelphia: Elsevier Saunders; 1995.

    Google Scholar 

  2. Rausch MK, Tibayan FA, Ingels NB, Miller DC, Kuhl E. Mechanics of the mitral annulus in chronic ischemic cardiomyopathy. Ann Biomed Eng. 2013;41:2171–80.

    Article  Google Scholar 

  3. Alavi SH, Sinha A, Steward E, Milliken JC, Kheradvar A. Load dependent extracellular matrix organization in atrioventricular heart valves: differences and similarities. Am J Physiol Heart Circ Physiol. 2015;309:H276–84.

    Article  Google Scholar 

  4. Rabbah J-PM, Saikrishnan N, Siefert AW, Santhanakrishnan A, Yoganathan AP. Mechanics of healthy and functionally diseased mitral valves: a critical review. J Biomech Eng. 2013;135:21007.

    Article  Google Scholar 

  5. Sacks MS, Yoganathan AP. Heart valve function: a biomechanical perspective. Philos Trans R Soc Lond B Biol Sci. 2007;362:1369–91.

    Article  Google Scholar 

  6. Liao J, Vesely I. A structural basis for the size-related mechanical properties of mitral valve chordae tendineae. J Biomech. 2003;36:1125–33.

    Article  Google Scholar 

  7. Ranganathan N, Lam JHC, Wigle ED, Silver MD. Morphology of the human mitral valve: II. Valve Leaflets. Circ. 1970;41:459–67.

    Article  Google Scholar 

  8. Ritchie J, Warnock JN, Yoganathan AP. Structural characterization of the chordae tendineae in native porcine mitral valves. Ann Thorac Surg. 2005;80:189–97.

    Article  Google Scholar 

  9. Lam JH, Ranganathan N, Wigle ED, Silver MD. Morphology of the human mitral valve. I. Chordae tendineae: a new classification. Circulation. 1970;41:449–58.

    Article  Google Scholar 

  10. Al-Atabi M, Espino DM, Hukins DW, Buchan KG. Biomechanical assessment of surgical repair of the mitral valve. Proc Inst Mech Eng Part H J. Eng Med. 2012;226:275–87.

    Article  Google Scholar 

  11. Wilcox AG, Buchan KG, Espino DM. Frequency and diameter dependent viscoelastic properties of mitral valve chordae tendineae. J Mech Behav Biomed Mater. 2014;30:186–95.

    Article  Google Scholar 

  12. Espino DM, Shepherd DET, Hukins DWL, Buchan KG. The role of chordae tendineae in mitral valve competence. J Heart Valve Dis. 2005;14:603–9.

    Google Scholar 

  13. Espino DM, Hukins DWL, Shepherd DET, Buchan KG. Mitral valve repair: an in vitro comparison of the effect of surgical repair on the pressure required to cause mitral valve regurgitation. J Heart Valve Dis. 2006;15:375–81.

    Google Scholar 

  14. Akins CW, Hilgenberg D, Buckley MJ, Vlahakes GJ, Torchiana DF, Daggett WM, Austen WG. Mitral valve reconstruction versus replacement for degenerative or ischemic mitral regurgitation. Ann Thorac Surg. 1994;58:668–76.

    Article  Google Scholar 

  15. Barber JE, Kasper FK, Ratliff NB, Cosgrove DM, Griffin BP, Vesely I. Mechanical properties of myxomatous mitral valves. J Thorac Cardiovasc Surg. 2001;122:955–62.

    Article  Google Scholar 

  16. Gillinov AM, Wierup PN, Blackstone EH, Bishay ES, Cosgrove DM, White J, Lytle BW, McCarthy PM, Miller DC, Wierup RAD, Carpentier A. Is repair preferable to replacement for ischemic mitral regurgitation? J Thorac Cardiovasc Surg. 2001;122:1125–41.

    Article  Google Scholar 

  17. Gunning GM, Murphy BP. The effects of decellularization and cross-linking techniques on the fatigue life and calcification of mitral valve chordae tendineae. J Mech Behav Biomed Mater. 2016;57:321–33.

    Article  Google Scholar 

  18. Tillquist MN, Maddox TM. Cardiac crossroads: deciding between mechanical or bioprosthetic heart valve replacement. Patient Prefer Adherence. 2011;5:91–9.

    Article  Google Scholar 

  19. Nkomo VT, Gardin JM, Skelton TN, Gottdiener JS, Scott CG, Enriquez-Sarano M. Burden of valvular heart diseases: a population-based study. Lancet. 2006;368:1005–11.

    Article  Google Scholar 

  20. Hukins DWL, Leahy JC, Mathias KJ. Selection of replacement materials. J Mater Chem. 1999;3:629–36.

    Article  Google Scholar 

  21. Singhal P, Luk A, Butany J. Bioprosthetic heart valves: impact of implantation on biomaterials. ISRN Biomater. 2013;2013:728791.

    Article  Google Scholar 

  22. Talman EA, Boughner DR. Glutaraldehyde fixation alters the internal shear properties of porcine aortic heart valve tissue. Ann Thorac Surg. 1995;60:S369–73.

    Article  Google Scholar 

  23. Vesely I. A mechanism for the decrease in stiffness of bioprosthetic heart valve tissues after cross-linking. ASAIO J. 1996;42:993–9.

    Article  Google Scholar 

  24. Vesely I, Boughner D. Analysis of the bending behaviour of porcine xenograft leaflets and of natural aortic valve material: bending stiffness, neutral axis and shear measurements. J Biomech. 1989;22:655–9.

    Article  Google Scholar 

  25. Billiar KL, Sacks MS. Biaxial mechanical properties of the natural and glutaraldehyde treated aortic valve cusp—part I: experimental results. J Biomech Eng. 2000;122:23–30.

    Article  Google Scholar 

  26. Walther T, Lehmann S, Falk V, Walther C, Doll N, Rastan A, Metz S, Schneider J, Gummert J, Mohr FW. Midterm results after stentless mitral valve replacement. Circulation. 2003;108:85–9.

    Article  Google Scholar 

  27. Broom ND. The stress/strain and fatigue behaviour of glutaraldehyde preserved heart-valve tissue. J Biomech. 1977;10:707–24.

    Article  Google Scholar 

  28. Lim KO, Boughner DR, Smith CA. Dynamic elasticity of human mitral valve chordae tendieae. Can J Physiol Pharmacol. 1977;55:413–8.

    Article  Google Scholar 

  29. Lim KO, Boughner DR. Mechanical properties of human mitral valve chordae tendineae: variation with size and strain rate. Can J Physiol Pharmacol. 1975;53:330–9.

    Article  Google Scholar 

  30. Fulcher GR, Hukins DWL, Shepherd DET. Viscoelastic properties of bovine articular cartilage attached to subchondral bone at high frequencies. BMC Musculoskelet Disord. 2009;10:61.

    Article  Google Scholar 

  31. Lawless BM, Barnes SC, Espino DM, Shepherd DET. Viscoelastic properties of a spinal posterior dynamic stabilisation device. J Mech Behav Biomed Mater. 2016;59:519–26.

    Article  Google Scholar 

  32. Barnes SC, Lawless BM, Shepherd DET, Espino DM, Bicknell GR, Bryan RT. Viscoelastic properties of human bladder tumours. J Mech Behav Biomed Mater. 2016;61:250–7.

    Article  Google Scholar 

  33. Burton HE, Freij JM, Espino DM. Dynamic viscoelasticity and surface properties of porcine left anterior descending coronary arteries. Cardiovasc Eng Technol. 2017;8:41–56.

    Article  Google Scholar 

  34. Espino DM, Shepherd DET, Hukins DWL. Viscoelastic properties of bovine knee joint articular cartilage: dependency on thickness and loading frequency. BMC Musculoskelet Disord. 2014;15:205.

    Article  Google Scholar 

  35. Lawless BM, Sadeghi H, Temple DK, Dhaliwal H, Espino DM, Hukins DWL. Viscoelasticity of articular cartilage: analysing the effect of induced stress and the restraint of bone in a dynamic environment. J Mech Behav Biomed Mater. 2017;75:293–301.

    Article  Google Scholar 

  36. Baxter J, Buchan KG, Espino DM. Viscoelastic properties of mitral valve leaflets: an analysis of regional variation and frequency-dependency. Proc Inst Mech Eng Part H J Eng Med. 2017;231:938–44.

    Article  Google Scholar 

  37. Millard L, Espino DM, Shepherd DET, Hukins DWL, Buchan KG. Mechanical properties of chordae tendineae of the mitral heart valve: young’s modulus, structural stiffness, and effects of aging. J Mech Med Biol. 2011;11:221–30.

    Article  Google Scholar 

  38. Chan RW, Titze IR. Effect of postmortem changes and freezing on the viscoelastic properties of vocal fold tissues. Ann Biomed Eng. 2003;31:482–91.

    Article  Google Scholar 

  39. Szarko M, Muldrew K, Bertram JE. Freeze-thaw treatment effects on the dynamic mechanical properties of articular cartilage. BMC Musculoskelet Disord. 2010;11:231.

    Article  Google Scholar 

  40. Jimenez JH, Soerensen DD, He Z, Ritchie J, Yoganathan AP. Mitral valve function and chordal force distribution using a flexible annulus model: an in vitro study. Ann Biomed Eng. 2005;33:557–66.

    Article  Google Scholar 

  41. Ritchie J, Jimenez J, He Z, Sacks MS, Yoganathan AP. The material properties of the native porcine mitral valve chordae tendineae: an in vitro investigation. J Biomech. 2006;39:1129–35.

    Article  Google Scholar 

  42. Broom ND, Thomson FJ. Influence of fixation conditions on the performance of glutaraldehyde-treated porcine aortic valves: towards a more scientific basis. Thorax. 1979;34:166–76.

    Article  Google Scholar 

  43. Beck SJ. Electron microscopy: a handbook of techniques for the biologist. New York: Nassau Community College; 1996.

    Google Scholar 

  44. Diagnostics, n.d. Factors affecting fixation [WWW Document]. https://www.nationaldiagnostics.com/histology/article/factors-affecting-fixation. Accessed 23 Oct 2017.

  45. Education, n.d. Histotechniques [WWW Document] http://library.med.utah.edu/WebPath/HISTHTML/HISTOTCH/HISTOTCH.html. Accessed 23 Oct 2017.

  46. Burton HE, Williams RL, Espino DM. Effects of freezing, fixation and dehydration on surface roughness properties of porcine left anterior descending coronary arteries. Micron. 2017;101:78–86.

    Article  Google Scholar 

  47. Olde Damink LHH, Dijkstra PJ, Van Luyn MJA, Van Wachem PB, Nieuwenhuis P, Feijen J. Glutaraldehyde as a crosslinking agent for collagen-based biomaterials. J Mater Sci Mater Med. 1995;6:460–72.

    Article  Google Scholar 

  48. Zeeman R. Cross-linking of collagen based materials. University of Twente, Enschede, The Netherlands, PhD thesis. 1998.

  49. Lis Y, Burleigh MC, Parker DJ, Child AH, Hogg J, Davies MJ. Biochemical characterization of individual normal, floppy and rheumatic human mitral valves. Biochem J. 1987;244:597–603.

    Article  Google Scholar 

  50. Kunzelman KS, Cochran RP, Verrier ED, Eberhart RC. Anatomic basis for mitral valve modelling. J Heart Valve Dis. 1994;3:491–6.

    Google Scholar 

Download references

Authors’ contributions

MC participated in the study’s design, performed mechanical testing and produced the first draft of the manuscript. HEB conceived the study, and participated in its design. BML conceived the study, and participated in its design. VG conceived the study, and participated in its design. KGB conceived the study, and participated in its design. DME conceived the study, participated in its design and helped to draft the manuscript. All authors read and approved the final manuscript.

Acknowledgements

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

The datasets during and/or analysed during the current study available from the corresponding author on reasonable request.

Consent for publication

Not applicable.

Ethics approval and consent to participate

Ethical approval was granted for this study by the University of Birmingham Research Support Group, [ERN_15-0032]. This study did not involve any participants, and no animals were sacrificed specifically for this study.

Funding

H.E.B. was funded by an Engineering and Physical Sciences Research Council scholarship [M114612B] for the duration of this study. The authors would like to thank Arthritis Research UK for the funding of the testing equipment used during this study (Grant Number: H0671). Funding bodies were not involved in the design of the study, collection, analysis, or interpretation of data, nor in writing of the manuscript.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to D. M. Espino.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Constable, M., Burton, H.E., Lawless, B.M. et al. Effect of glutaraldehyde based cross-linking on the viscoelasticity of mitral valve basal chordae tendineae. BioMed Eng OnLine 17, 93 (2018). https://doi.org/10.1186/s12938-018-0524-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12938-018-0524-2

Keywords