Semi-automated ultrasound guidance applied to nasogastrojejunal tube replacement for enteral nutrition in critically ill adults

Abstract

Background and objective

At present, the enteral nutrition approaches via nose and duodenum (or nose and jejunum) are the preferred method of nutritional support in the medical engineering field, given the superiority of in line with physiological processes and no serious complication. In this study, the authors adopted saline as the acoustic window, and gave enteral nutrition support to critically ill patients, via the nasogastrojejunal approach guided by semi-automated ultrasound. These above patients benefited a lot from this kind of nutrition support treatment, and we aimed to report the detailed information.

Methods

Critically ill patients (n = 41) who had been treated with hospitalized intestine nutrition were identified. The Apogee 1200 ultrasonic diagnostic apparatus, and nasogastrojejunal tubes were adopted to carry out intestine nutrition treatment guided by semi-automated ultrasound. In order to confirm the specific positions of cardia, gastric body, antrum of stomach, and pylorus, the semi-automated ultrasound was utilized to probe the stomach cavity. And then, the ultrasonic probe was placed in the cardia location, and the nasogastrojejunal tube was slowly inserted through the metal thread. After operation, the nursing service satisfaction of patients and mean operation time were calculated, respectively.

Results

All the patients were treated with enteral nutrition via nasogastrojejunal tube, and the whole procedure was under the guidance of semi-automated ultrasonography. The end of the feeding tube is attached to the surface of the stomach with a greater curvature, which can be bent on account of a no gastric peristalsis squeeze function, and thereby were prevented from entering into the antrum and pylorus locations. After this procedure, the mental thread was taken out, and the tube was pushed forward by a “drift” approach in order to allow it to enter into the intestine. The total nursing service satisfaction of patients was 90.24%, and the total incidence of adverse reactions was 17.07%.

Conclusions

In summary, the application of saline can be taken as sound window, and the metal wire as the tracking target, the bedside nasogastrojejunal tube guided by semi-automated ultrasound is an effective feeding tube placement method, with relatively good clinical application value in medical engineering.

Background

Nutrition support is one major development of clinical medicine in the twentieth century, and has become an indispensable constituent part in the treatment of critically ill patients, in order to alleviate the nutritional deficit [1, 2]. Enteral nutrition has achieved significant advances in decades, and is beneficial for the patients who have functional guts but can not meet their nutritional requirements via normal diet, on account of cancer, HIV, stroke, multiple sclerosis, dementia, etc. [3,4,5]. This kind of enteral feeding can be delivered by means of various approaches, including nasogastric tube, percutaneous endoscopic gastrostomy, jejunostomy. In general, during the period of enteral nutrition, the providers also need to assess the nutritional status, and evaluate the nutritional requirements of patients [6]. Besides, the development of enteral nutrition also requires multidisciplinary teams, such as the extended roles for dietitians and nurses, etc. [7]. In addition, more and more serious aging society, various diseases mentioned above, the swallowing difficulties and malnutrition resulted from various complications, are all the main reasons why rapidly increasing enteral nutrition is needed [8, 9].

At present, the enteral nutrition approaches via nose and duodenum (or nose and jejunum) are the preferred method of nutritional support in medical engineering, given the superiority of in line with physiological processes and no serious complication [10, 11]. In this study, the authors adopted saline as the acoustic window, and gave enteral nutrition support treatment to critically ill patients, via the nasogastrojejunal approach guided by semi-automated ultrasound. These above patients benefited a lot from this kind of nutrition support, and we reported the detailed information as followed.

Methods

Clinical data of the identified patients

This research was approved beforehand by the institution ethics committee in our hospital. According to the relevant regulations of ethics, the informed consent of patients had been obtained before investigation. 41 critically ill patients who would be treated with hospitalized intestine nutrition were identified in our department, from February 2015 to January 2017. In detail, 34 males and 7 females, the age ranged from 21 to 86 years old. 15 cases suffered from primary acute severe pancreatitis, 21 cases suffered from chronic obstructive pulmonary disease and respiratory failure, 3 cases suffered from cerebral hemorrhage, 1 case suffered from erosive gastritis, 1 case suffered from multiple organ failure, as shown in Table 1 and Fig. 1. And, Fig. 2 illustrated the case number per year from 2015 to 2017.

Table 1 General information of patients identified in this study
Fig. 1
figure 1

Distribution of primary diseases involved in this research

Fig. 2
figure 2

Case number per year from 2015 to 2017

Instrument and methods

The Apogee 1200 portable ultrasonic diagnostic apparatus (Japan), and Ch 12 nasogastrojejunal tubes (Manufacturer: Bai Tong in China) were adopted to carry out intestine nutrition treatment guided by semi-automated ultrasound. These included patients fasted for 6–8 h, and then received gastrointestinal decompression treatment for 1 h in order to vent the gas and food residues in the stomach, so that the images could be clearly displayed. The patients kept sober and took semi-recumbent, in order to make them coordinate work during treatment. The routine abdomen ultrasound was utilized to probe the stomach cavity, in order to confirm the specific positions of cardia, gastric body, antrum of stomach, and pylorus. And then, the ultrasonic probe was placed in the cardia location, and the nasogastrojejunal tube was slowly inserted through the metal thread. The ultrasonic diagnostic apparatus could capture the image of metal thread, and the intubationist slowly injected pre-heated physiological saline 300–400 ml. Afterwards, the ultrasonic probe was used to perform fan-shaped scan, with the purpose of knowing the specific location of nasogastrojejunal tube and the status of gastric motility. The intubationist slowly rotated the guide wire to make it follow the gastric motility, and at the same time slowly pushed the tube. The nasogastrojejunal tube straightly went into the pylorus, and was forwarded smoothly. When the intubationist felt empty, the tube entered into duodenum. The location of ultrasonic probe was adjusted, the duodenal bulb could be observed, and the ultrasonic echo of metal thread was stronger. And then the thread was pushed forward to another 110–120 cm. The bedside abdomen radiographs were obtained to confirm whether the tube position was proper, and then the guide wire was took out. And then, the operator made sure that the patient was in a good condition, and then another 20 ml saline was injected again to assure the tube was unobstructed, the gas over the water tone could be heard by umbilical auscultation. At last, the enteral nutrition pump was connected with the tube. After operation, the nursing service satisfaction of patients and mean operation time were calculated, respectively.

Results

General information

All the patients were treated by using nasogastrojejunal tubes via guidance of semi-automated ultrasonography, and the status of the patients during operations were all good. 40 cases were successfully implanted, 1 case failed. The failed case suffered from multiple organ failure, and the gastric motility disappeared. The end of the feeding tube is attached to the surface of the stomach with a greater curvature, which can be bent on account of a no gastric peristalsis squeeze function, and thereby prevented from entering into the antrum and pylorus locations. After this procedure, the mental thread was taken out, and the tube was pushed forward by a “drift” approach in order to allow it to enter into the intestine. Nevertheless, the operation time lasted longer, and the nasogastrojejunal tube was placed into the intestine after 2 h. And, Fig. 3 illustrated the captured image when the nasogastrojejunal tube entered into pylorus and duodenal bulb.

Fig. 3
figure 3

Captured image when the nasogastrojejunal tube entered into pylorus and duodenal bulb

Average operation time of intubation

The average operation time of intubation in this study was displayed in Table 2. 15 cases suffered from primary acute severe pancreatitis, and the average operation time was 48.20 ± 12.50 min. 21 cases suffered from chronic obstructive pulmonary disease, and the average operation time was 26.80 ± 6.52 min. 3 cases suffered from Cerebral hemorrhage, and the average operation time was 44.21 ± 4.59 min. 1 patients with Erosive gastritis, and 1 case failed.

Table 2 Average operation time of intubation in this study

Patients’ satisfaction degree

Besides, we made another questionnaire survey to investigate the patients’ satisfaction degree of nasogastrojejunal tube, and the result was illustrated in Table 3 and Fig. 4. The total nursing service satisfaction of patients was 90.24%. The satisfactory degree of patients with primary acute severe pancreatitis was 93.33%, the satisfactory degree of patients with chronic obstructive pulmonary disease was 90.48%, the satisfactory degree of patients with cerebral hemorrhage and erosive gastritis both were 100%. In general, the total nursing service satisfaction appeared high, more than 90%, which suggested that this kind of technique could be acceptable very well. Nevertheless, several serious complications occurred, on account of the diseases or the operators, and the nursing service satisfaction of these patients decreased.

Table 3 Nursing service satisfaction of patients identified in this study
Fig. 4
figure 4

Nursing service satisfaction of patients suffering from different diseases

Incidence of adverse drug reactions

In addition, in terms of incidence of adverse drug reactions among the identified patients in this study, we calculated them and listed in Table 4 and Fig. 5. The main adverse drug reactions in this research were stomachache, headache, nausea and vomiting. The incidence of adverse reactions of primary acute severe pancreatitis was 6.67%, the incidence of chronic obstructive pulmonary disease was 14.29%, the incidence of cerebral hemorrhage was 66.67%, the incidence of erosive gastritis was 0%. At last, the total incidence of adverse reactions was 17.07%.

Table 4 Incidence of adverse reactions among the identified patients in this study
Fig. 5
figure 5

Adverse reactions resulting from different diseases

Discussion

With the rapid development of medical engineering, the medical care of patients with critical illness has become increasingly complicated [12,13,14]. And the nutrition support has long been recognized as supportive therapy in critical care. In recent decades, the clinical nutrition has been evolving all the time, and it has been deemed as a form of therapy, rather than just a way of delivering nutrition [15,16,17]. Because several nutrition might exert therapeutic effects via immunomodulation or liver protection, etc. For instance, among cancer patients, the enteral nutrition contributes to decrease the postoperative complications and prolong survival time [18]. At present, the application of immune-enhancing enteral nutrition is an integral part of surgical guidelines. On the basis of latest ESICM clinical practice guidelines, the initiating enteral nutrition has been recommended to be provided within the first 24–48 h after intensive care unit admission, if the patients can not eat by themselves [19, 20].

The main approaches of enteral nutrition are various, but the commonly used and noninvasive method is nasogastrojejunal tube. The traditional method of intubation tube always depends on X ray or gastroscope, or only relies on the operator experience. When the intubationist just operates by his own experience, the success rate of intubation appears lower [21, 22]. When the intubationist operates under the guidance of X ray, the patients have to be moved, but the majority of patients are critically ill in the intensive care unit and are hard to be carried, especially for the patients who need mechanical ventilation [23, 24]. Nasogastrojejunal tube insertion is based on minimally invasive catheterization procedure that is combined with ultrasound guidance. This semi-automated device pertains to a class of medical imaging assisted equipment that can help patients in terms of enteral nutrition [25].

On the other hand, the superiorities of intubation guided by semi-automated ultrasound are as follows. (1) The operation by the bed is permitted, without moving the patient. And the operator can observe the location of nasogastrojejunal tube. (2) The operation guided by semi-automated ultrasound do not bring about radioactive injury, and the painful feeling of patients was smaller when receiving this noninvasive method. (3) It is cost-effective method, when compared with digital subtraction angiography and gastroscope. In addition, there are several points need to be addressed during implementation. (1) Before operation, the abrosia for 6–8 h is needed, as well as gastrointestinal decompression for 1 h, in order to exclude interference resulted from food residue and gas in the stomach, and thereby to improve the image quality, to clearly display the tube position, and to reduce the failure rate of operation. (2) When the tube enters into gastral cavity, another 300–400 ml saline should be injected into tube. The saline should be placed in the incubator and be heated to 37 °C, to avoid unnecessary stimulation. The injected saline forms an acoustic window, so that the nasogastrojejunal can be clearly displayed. (3) The tube should be carried forward, with the help of gastric peristalsis, and this is also the pivotal point. The stomach cavity has two physiological curves, namely, stomach angle and pyloric region. The metal wire always remains extended position, and it is difficult to go through the two bending angles. When the peristaltic wave gets through the two physiological bending, the bending angles disappear, and the extrusion force pushes forward the tube, until entering into duodenum. The only one failure case attributed to the disappearance of gastric motility in the patient suffering from multiple organ failure.

Conclusion

To sum up, the application of saline can be taken as sound window, and the metal wire as the tracking target, the bedside nasogastrojejunal tube guided by semi-automated ultrasound is an effective feeding tube placement method, with relatively good clinical application value in medical engineering.

References

  1. Bloomer MJ, Clarke AB, Morphet J. Nurses’ prioritization of enteral nutrition in intensive care units: a national survey. Nurs Crit Care. 2017. https://doi.org/10.1111/nicc.12284.

    Google Scholar 

  2. Boland K, Maher N, O’Hanlon C, O’Sullivan M, Rice N, Smyth M, Reynolds JV. Home enteral nutrition recipients: patient perspectives on training, complications and satisfaction. Frontline Gastroenterol. 2017;8:79–84.

    Article  Google Scholar 

  3. Buyukcoban S, Akan M, Koca U, Eglen MY, Ciceklioglu M, Mavioglu O. Comparison of two different enteral nutrition protocol in critically ill patients. Turk J Anaesthesiol Reanim. 2016;44:265–9.

    Article  Google Scholar 

  4. Chen ZH, Lin SY, Dai QB, Hua J, Chen SQ. The effects of pre-operative enteral nutrition from nasal feeding tubes on gastric outlet obstruction. Nutrients. 2017;9:E373.

    Article  Google Scholar 

  5. Heerasing N, Thompson B, Hendy P, Heap GA, Walker G, Bethune R, Mansfield S, Calvert C, Kennedy NA, Ahmad T, Goodhand JR. Exclusive enteral nutrition provides an effective bridge to safer interval elective surgery for adults with Crohn’s disease. Aliment Pharmacol Ther. 2017;45:660–9.

    Article  Google Scholar 

  6. Klek S, Scislo L, Walewska E, Choruz R, Galas A. Enriched enteral nutrition may improve short-term survival in stage IV gastric cancer patients: a randomized, controlled trial. Nutrition. 2017;36:46–53.

    Article  Google Scholar 

  7. Ma L, Luo GY, Ren YF, Qiu B, Yang H, Xie CX, Liu SR, Liu SL, Chen ZL, Li Q, Fu JH, Liu MZ, Hu YH, Ye WF, Liu H. Concurrent chemoradiotherapy combined with enteral nutrition support: a radical treatment strategy for esophageal squamous cell carcinoma patients with malignant fistulae. Chin J Cancer. 2017;36:8.

    Article  Google Scholar 

  8. Ojo O, Brooke J. Recent advances in enteral nutrition. Nutrients. 2016;8:E709.

    Article  Google Scholar 

  9. O’Leary-Kelley C, Bawel-Brinkley K. Nutrition support protocols: enhancing delivery of enteral nutrition. Crit Care Nurse. 2017;37:e15–23.

    Article  Google Scholar 

  10. Reintam Blaser A, Starkopf J, Alhazzani W, Berger MM, Casaer MP, Deane AM, Fruhwald S, Hiesmayr M, Ichai C, Jakob SM, Loudet CI, Malbrain ML, Montejo Gonzalez JC, Paugam-Burtz C, Poeze M, Preiser JC, Singer P, van Zanten AR, De Waele J, Wendon J, Wernerman J, Whitehouse T, Wilmer A, Oudemans-van Straaten HM. Early enteral nutrition in critically ill patients: ESICM clinical practice guidelines. Intensive Care Med. 2017;43:380–98.

    Article  Google Scholar 

  11. Shu XL, Kang K, Gu LJ, Zhang YS. Effect of early enteral nutrition on patients with digestive tract surgery: a meta-analysis of randomized controlled trials. Exp Ther Med. 2016;12:2136–44.

    Article  Google Scholar 

  12. Fletcher J. Giving nutrition support to critically ill adults. Nurs Times. 2015;111:12–6.

    Google Scholar 

  13. Kiss N. Nutrition support and dietary interventions for patients with lung cancer: current insights. Lung Cancer (Auckl). 2016;7:1–9.

    Google Scholar 

  14. Lheureux O, Preiser JC. Role of nutrition support in inflammatory conditions. Nutr Clin Pract. 2017;32:310–7.

    Article  Google Scholar 

  15. Marshall S, Agarwal E, Young A, Isenring E. Role of domiciliary and family carers in individualised nutrition support for older adults living in the community. Maturitas. 2017;98:20–9.

    Article  Google Scholar 

  16. Mulherin DW, Cogle SV. Updates in nutrition support for critically Ill adult patients. Hosp Pharm. 2017;52:17–26.

    Article  Google Scholar 

  17. Sajeev M, Cohen J, Wakefield CE, Fardell JE, Cohn RJ. Decision aid for nutrition support in pediatric oncology: a pilot study. JPEN. 2017;41:1336–47.

    Article  Google Scholar 

  18. Wei J, Chen W, Zhu M, Cao W, Wang X, Shi H, Dong B, Sun J, Chen H, Zhou Y, Zhou S, Xu J. Guidelines for parenteral and enteral nutrition support in geriatric patients in China. Asia Pac J Clin Nutr. 2015;24:336–46.

    Google Scholar 

  19. Zhang H, Wang Y, Jiang ZM, Kondrup J, Fang H, Andrews M, Nolan MT, Mu SY, Zhang J, Yu K, Lu Q, Kang WM. Impact of nutrition support on clinical outcome and cost-effectiveness analysis in patients at nutritional risk: a prospective cohort study with propensity score matching. Nutrition. 2017;37:53–9.

    Article  Google Scholar 

  20. Holzinger U, Brunner R, Miehsler W, Herkner H, Kitzberger R, Fuhrmann V, Metnitz PG, Kamolz LP, Madl C. Jejunal tube placement in critically ill patients: a prospective, randomized trial comparing the endoscopic technique with the electromagnetically visualized method. Crit Care Med. 2011;39:73–7.

    Article  Google Scholar 

  21. Li J, Gu Y, Zhou R. Rhubarb to facilitate placement of nasojejunal feeding tubes in patients in the intensive care unit. Nutr Clin Pract. 2016;31:105–10.

    Article  Google Scholar 

  22. Li RH, Li DC, Lv XP, Quan B. Radiophotographically controlled nasointestinal intubation to treat recurrent postoperative adhesive ileus. Cell Biochem Biophys. 2015;72:509–14.

    Article  Google Scholar 

  23. Li RH, Li DC, Tian QZ, Wu P, Zhang XH. Diagnosis of small bowel obstruction using targeted enterography during nasointestinal decompression. Cell Biochem Biophys. 2015;72:833–7.

    Article  Google Scholar 

  24. Stimac D, Poropat G, Hauser G, Licul V, Franjic N, Valkovic Zujic P, Milic S. Early nasojejunal tube feeding versus nil-by-mouth in acute pancreatitis: a randomized clinical trial. Pancreatology. 2016;16:523–8.

    Article  Google Scholar 

  25. Wong KKL, Tu JY, Sun Z, Dissanayake DW. Methods in research and development of biomedical devices. Singapore: World Scientific Publishing Co.; 2013.

    Book  Google Scholar 

  26. Wong KKL, Hui SCN. Ethical principles and standards for the conduct of biomedical research and publication. Australas Phys Eng Sci Med. 2015;38(3):377–80.

    Article  Google Scholar 

  27. Wong KKL, Chu WCW. Ethics policies and procedures in imaging and interventional radiology. Australas Phys Eng Sci Med. 2015;38(2):375–6.

    Article  Google Scholar 

Download references

Authors’ contributions

Performed the literature review: YL; Carried out research: YL, YM, HR, and YY; Gave advice for setup: YY; Checked the validity of data: YL, YY. All authors read and approved the final manuscript.

Acknowledgements

The ICU research team leaded by Prof. Yongming Yao in the First Affiliated Hospital of PLA General Hospital gave the financial support for this research (Shenzhen’s Sanming Project, Grant No. SZSM201612011).

Competing interests

The authors declare that they have no competing interests.

Ethics approval and consent to participate

All the clinical information were obtained with informed consent (or a formal waiver of consent) with approval by the Research Ethics Boards in our hospital. And, the informed consents of patients were obtained beforehand [26, 27].

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yu Ye.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Y., Ye, Y., Mei, Y. et al. Semi-automated ultrasound guidance applied to nasogastrojejunal tube replacement for enteral nutrition in critically ill adults. BioMed Eng OnLine 17, 21 (2018). https://doi.org/10.1186/s12938-018-0452-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12938-018-0452-1

Keywords