Respiratory modulation of oscillometric cuff pressure pulses and Korotkoff sounds during clinical blood pressure measurement in healthy adults

Abstract

Background

Accurate blood pressure (BP) measurement depends on the reliability of oscillometric cuff pressure pulses (OscP) and Korotkoff sounds (KorS) for automated oscillometric and manual techniques. It has been widely accepted that respiration is one of the main factors affecting BP measurement. However, little is known about how respiration affects the signals from which BP measurement is obtained. The aim was to quantify the modulation effect of respiration on oscillometric pulses and KorS during clinical BP measurement.

Methods

Systolic and diastolic BPs were measured manually from 40 healthy subjects (from 23 to 65 years old) under normal and regular deep breathing. The following signals were digitally recorded during linear cuff deflation: chest motion from a magnetometer to obtain reference respiration, cuff pressure from an electronic pressure sensor to derive OscP, and KorS from a digital stethoscope. The effects of respiration on both OscP and KorS were determined from changes in their amplitude associated with respiration between systole and diastole. These changes were normalized to the mean signal amplitude of OscP and KorS to derive the respiratory modulation depth. Reference respiration frequency, and the frequencies derived from the amplitude modulation of OscP and KorS were also calculated and compared.

Results

Respiratory modulation depth was 14 and 40 % for OscP and KorS respectively under normal breathing condition, with significant increases (both p < 0.05) to 16 and 49 % with deeper breathing. There was no statistically significant difference between the reference respiration frequency and those derived from the oscillometric and Korotkoff signals (both p > 0.05) during deep breathing, and for the oscillometric signal during normal breathing (p > 0.05).

Conclusions

Our study confirmed and quantified the respiratory modulation effect on the oscillometric pulses and KorS during clinical BP measurement, with increased modulation depth under regular deeper breathing.

Background

Manual auscultatory and automated oscillometric methods are non-invasive ways usually used for blood pressures (BP) measurement. The manual auscultatory method is based on the auscultation of Korotkoff sound (KorS), while the automated oscillometric method is based on the analysis of oscillometric cuff pressure pulses (OscP). The reliability of OscP and KorS signals are therefore important for accurate BP measurement for both manual and automated methods. It has been widely accepted that respiration is one of the main factors affecting BP measurement [1]. Published studies have shown that respiration influences both systolic and diastolic blood pressures (SBP and DBP) [15]. With regular deep breathing, both SBP and DBP decreased significantly by 4.4 and 4.8 mmHg respectively, in comparison with normal breathing [6]. According to a major review in the Journal of the American Medical Association (JAMA), a systematic error of 5 mmHg error of BP measurement would result in 27 million Americans being exposed to unnecessary treatment or 21 million being denied treatment [7]. Therefore, it is clinically important to understand the effect of respiration on OscP and KorS.

Until now, the underlying mechanisms of the effect of respiration on OscP and KorS and on BP measurement have not been fully understood. It is generally believed that respiration influences the central venous pressure through the chest expansion and compression. Inspiration decreases central venous pressure, leading to the increase of venous return and right atrial filling and the decrease of pulmonary venous flow to the left side of the heart, and finally reduces stroke volume [8]. The opposite occurs during expiration. The respiratory modulation of stroke volume will directly influence the amplitude of OscP and then affect the BP measured using the automated oscillometric method. On the other hand, although the genesis of KorS is still a matter of debate, one longstanding hypothesis is that KorS is generated by the distension of the arterial wall caused by the changing transmural pressure gradient [912]. The respiratory modulation of stroke volume will directly influence the force deployed in opening the artery and then influence the blood flow sound, which will be reflected on the amplitude of the KorS signal and then influence the BP measured through the manual auscultatory method.

Some earlier studies have preliminarily investigated the effect of respiration on OscP and KorS recordings. Di Marco et al. [13] showed the respiratory modulation on KorS, but they did not quantitatively analyze the different respiratory modulation effects between normal and deep breathing. Zheng et al. [8] extended those studies and demonstrated the modulation effect of respiration on both OscP and KorS, but their study was performed under only static cuff pressures. There were significant differences between conditions of static cuff pressure and normal clinical BP measurement with deflating cuff pressure. As the cuff pressure during clinical BP measurement significantly influences the amplitudes of OscP and KorS, this may make the derived respiratory amplitude modulation effect on those two signals different from that observed with static cuff pressures. Therefore, studies are needed to explore the modulation effect of respiration on OscP and KorS during normal clinical BP measurement with deflating cuff pressures.

In order to make the investigation close to actual clinical conditions, the present study aimed to investigate the respiratory modulation effect on OscP and KorS during normal BP measurement. Two respiratory patterns [normal breathing and regular deep breathing] were used in this study to quantitatively assess: (1) the presence of amplitude modulation of respiration on OscP and KorS recordings; and (2) the differences in respiratory modulation on OscP and KorS between the two respiratory patterns.

Methods

Subjects and data acquisition

40 healthy subjects were enrolled in the study and their general information is summarized in Table 1. The study was carried out according to the Declaration of Helsinki (1989) of the World Medical Association, and was approved by Newcastle and North Tyneside NHS Research Ethics Committee. Informed and written consent was obtained from all subjects.

Table 1 General data information for the subjects studied

Manual SBP and DBP were measured by a trained operator according to the recommendations of the British and European Hypertension Societies [14]. The cuff was inflated to 200 mmHg and then deflated linearly at 2–3 mmHg/s. For each subject, six BP measurements were performed in total, with three repeats under two breathing patterns (normal and regular deep breathing with subjects breathing at their own comfortable rate). The order of the two breathing patterns was randomized between subjects. Between every two consecutive measurements, a resting period of appropriately 1 min was given to allow cardiovascular stabilization.

During linear cuff deflation, KorS was recorded by a piezo-electric microphone, with the bell-shaped stethoscope terminal connected to the microphone and placed on the antecubital fossa of the forearm. OscP was derived from the cuff pressure signals, while the reference respiration signal (Resp) was obtained by a chest magnetometer detecting chest wall movement [15]. All signals were digitally recorded at 2 kHz and 16 bits, and stored to a computer for offline processing.

Signal processing

Figure 1 shows the signal processing procedure to analyze the amplitude modulation effect of respiration on OscP and KorS during cuff deflation between SBP and DBP (a). On panel (b), reference Resp (from magnetometer) and the amplitude modulation signals on OscP and KorS are displayed to show the amplitude modulation effect of respiration on OscP and KorS. On panel (c) the normalized power spectral density of Resp and the amplitude modulation signals on OscP and KorS are displayed to show the respiratory amplitude modulation effect in the frequency domain. In the following sub-sections, each step of the signal processing procedure is described. Analysis was performed on anonymised data.

Fig. 1
figure 1

Signal processing procedure to analyze the amplitude modulation effect of respiration on OscP and KorS. a Original OscP (dotted line) and KorS (solid line) acquired during cuff deflation between SBP and DBP. The OscP signal between 9 and 11 s (marked with rectangular) is zoomed in and shown on the top-right corner of panel; b Reference Resp signal (dashed line) and amplitude modulation signals on OscP (dotted line) and KorS (solid line); and c normalized power spectral density of Resp (dashed line) and amplitude modulation signals on OscP (dotted line) and KorS (solid line)

Reference respiration signal

Reference Resp signal was firstly down-sampled to 4 Hz (after low-pass filtering to prevent aliasing). This sampling rate satisfies the Nyquist condition, as the normal range of respiratory rates of healthy adults under resting conditions is between 0.1 and 0.5 Hz. Respiration depth was defined as the mean of peak-to-valley differences of Resp [8].

Amplitude modulation signals on OscP and KorS

The linearly deflating baseline cuff pressure was first removed to obtain OscP. The peak and valley of OscP and KorS pulses were then located. All pulse amplitudes of OscP and KorS were calculated as the amplitude difference between the peak and the valley. These amplitudes were resampled at 4 Hz by cubic spline interpolation to generate the amplitude modulation signals on OscP and KorS.

As shown in Fig. 1b, the cycles of Resp and amplitude modulation signals on OscP and KorS are nearly identical, indicating the amplitude modulation effect of respiration on OscP and KorS.

Amplitude modulation depth (AMD) between two breathing conditions

Over the time period between manual SBP and DBP, AMD was defined as the mean change of amplitude modulation signals normalized to mean amplitude of OscP and KorS signals. AMD was then used to investigate the difference in amplitude modulation of respiration on OscP and KorS between the two breathing patterns.

Respiratory frequencies from Resp and amplitude modulation signals on OscP and KorS

The normalized power spectral density of Resp and amplitude modulation signals on OscP and KorS were estimated by the Welch periodogram using a Hamming window (between 0.1 and 0.5 Hz and with frequency resolution of 0.001 Hz). Respiratory frequencies from Resp (fR), OscP (fO) and KorS (fK) were calculated as the peak frequency of the corresponding power spectral density. As shown in the example in Fig. 1c, the peak frequencies of the power spectral density of Resp and amplitude modulation signals on OscP and KorS are almost identical, indicating the amplitude modulation effect of respiration on OscP and KorS.

Statistical analysis

One-way analysis of variance (ANOVA) was used to analyze the repeatability between the three repeated measurements. The two-tail t test (with a significance level of α = 0.05) was used to assess BP, respiratory frequency, respiratory depth and AMD between the two respiratory patterns. Bland–Altman analysis with limits of agreement was used to investigate the agreement between fR and fO or fK [16].

Results

Repeatability between measurements

There was no statistically significant difference (all p > 0.05) for all the parameters derived in this study for the three repeat recordings, indicating that subjects kept relatively still and breathed at a relatively stable rate for both conditions. This was important for enabling changes to be quantified for the different measurements.

Blood pressure changes with regular deep breathing

Table 2 presents the BPs (SBP, MAP and DBP) measured manually under normal and regular deep breathing conditions. With deep breathing, there were statistically significant decreases for all BPs (SBP, MAP and DBP) (all p < 0.001), confirming that regular deep breathing influenced BP.

Table 2 BP parameters in normal and deep breathing conditions

Reference respiratory parameters under the two respiratory patterns

As shown in Table 3, regular deep breathing significantly (all p < 0.001) decreased respiratory rate, and increased respiratory depth in comparison with the normal breathing condition.

Table 3 Respiratory parameters in normal and deep breathing conditions

Amplitude modulation depth difference under normal and deep breathing conditions

As shown in Table 4, AMD increased significantly (both p < 0.05) from 14 to 40 % for OscP and KorS respectively under normal breathing condition to 16 and 49 % with deeper breathing, indicating a significantly stronger amplitude modulation effect on both OscP and KorS signals with deep breathing.

Table 4 Amplitude modulation depth of OscP and KorS in normal and deep breathing conditions

Derived respiratory frequency difference under normal and deep breathing conditions

Figures 2 and 3 show respiratory frequency estimation errors for OscP and KorS in comparison with fR under both normal and deep breathing conditions. In comparison with fR, over 60 % measurements of fO and fK agreed with fR within 0.05 Hz or less. This was increased to nearly 80 % with deeper breathing. There was no statistically significant difference between the reference respiration frequency and these derived from the OscP and KorS signals (both p > 0.05) during deep breathing, and for the OscP signal during normal breathing (p > 0.05). As given in Table 5, the minimum frequency difference was observed from OscP in deep breathing with the smallest limit of agreement of [−0.109 to 0.098 Hz].

Fig. 2
figure 2

Bland–Altman difference plot of respiratory frequency estimation from OscP and KorS in normal and deep breathing conditions. Dotted black line indicates bias and solid black lines indicate 1.96-SD limits of agreement

Fig. 3
figure 3

Histogram of respiratory frequency estimation error from OscP and KorS under normal and deep breathing conditions. Frequency bin width is 0.1 Hz

Table 5 Comparison of fR with fO and fK

Discussion

In this study, normal and regular deep breathing were adopted at each subject’s comfortable respiratory rate to study the influence of respiration on OscP and KorS recordings. As expected and agreed with published studies [13, 17], regular deep breathing significantly decreased respiratory rate, and increased respiratory depth in comparison with the normal breathing condition. The statistically significant difference for respiratory parameters between regular deep breathing and normal breathing was a precondition for studying the different modulation effects of respiration on OscP and KorS between breathing conditions.

Our study showed that the amplitude modulation depths of respiration on OscP and KorS were significantly higher under deep breathing condition. This might be attributed to the significantly higher respiratory depth with deep breathing which increases the amplitude modulation effect of respiration on OscP and KorS. This result was in accordance with the findings of Saul et al. [18], where a direct mechanical coupling between respiration and the vasculature has been shown. The results also agreed with what we have reported on the respiratory effect on OscP recorded under static pressures [8], but the difference in amplitude modulation effect on KorS had not been observed in the previous study. This could be caused by the signals recorded from different measurement conditions. The KorS signal in our previous study was recorded from static pressures between 0.2 and 0.3 Hz, while the current one was from deflating cuff pressures between 0.19 and 0.25 Hz. In the future, it would be worth investigating the relationship between respiratory rate and the modulation effect.

This study also extended the study of Di Marco et al. [13] and Zheng et al. [8], and showed the amplitude modulation effect of respiration on OscP and KorS during normal clinical BP measurement. Although the amplitudes of OscP and KorS change with deflating cuff pressure, a good agreement between fR and fO or fK has been observed. The derived respiration frequency from OscP and KorS signals during deep breathing, and from the OscP signal during normal breathing had no statistically significant difference from the reference frequency, indicating the respiratory modulation effect on OscP and KorS during normal clinical BP measurement. This also reveals the potential to derive respiration rate from OscP during BP measurement.

The agreement between the reference respiratory frequency and that derived from KorS under normal breathing condition was not as good as that under the deep breathing condition. This could also be caused by lower amplitude modulation depth under normal breathing condition, which makes the respiratory amplitude modulation signal on KorS more sensitive to noise interference.

Note that a limitation of this work is that only healthy subjects with normal blood pressure were recruited to participate in this study. There may be differences in the magnitude of the effect of respiration on BP measurements in patients with cardiovascular disease due to factors such as differences in central arterial stiffness. A further clinical study is warranted to compare the amplitude modulation effect of respiration on OscP and KorS between patients with known cardiovascular disease (e.g., hypotensive and hypertensive) and age-matched healthy subjects.

Conclusions

In conclusion, we have provided scientific evidence for (1) the presence of an amplitude modulation effect of respiration on OscP and KorS during normal clinical BP measurement, and (2) the significantly stronger amplitude modulation effect of respiration on OscP and KorS for regular deep breathing than for the normal breathing condition.

Abbreviations

AMD:

amplitude modulation depth

ANOVA:

analysis of variance

BP:

blood pressure

DBP:

diastolic blood pressure

fK :

frequency of the amplitude modulation signal on Korotkoff sound

fO :

frequency of the amplitude modulation signal on oscillometric cuff pressure pulses

fR :

frequency of respiration signal

JAMA:

Journal of the American Medical Association

KorS:

Korotkoff sound

MAP:

mean arterial pressure

OscP:

oscillometric cuff pressure pulses

Resp:

respiration signal

SBP:

systolic blood pressure

ΔfK :

fR − fK

ΔfO :

fR − fO

References

  1. Novak V, Novak P, De Champlain J, Le Blanc AR, Martin R, Nadeau R. Influence of respiration on heart rate and blood pressure fluctuations. J Appl Physiol. 1993;74(2):617–26.

    Google Scholar 

  2. Radaelli A, Raco R, Perfetti P, Viola A, Azzellino A, Signorini MG, et al. Effects of slow, controlled breathing on baroreceptor control of heart rate and blood pressure in healthy men. J Hypertens. 2004;22(7):1361–70.

    Article  Google Scholar 

  3. Jagomagi K, Raamat R, Talts J, Lansimies E, Jurvelin J. Effect of deep breathing test on finger blood pressure. Blood Press Monit. 2003;8(5):211–4.

    Article  Google Scholar 

  4. Mori H, Yamamoto H, Kuwashima M, Saito S, Ukai H, Hirao K, et al. How does deep breathing affect office blood pressure and pulse rate? Hypertens Res. 2005;28(6):499–504.

    Article  Google Scholar 

  5. Laude D, Goldman M, Escourrou P, Elghozi JL. Effect of breathing pattern on blood pressure and heart rate oscillations in humans. Clin Exp Pharmacol Physiol. 1993;20(10):619–26.

    Article  Google Scholar 

  6. Zheng D, Giovannini R, Murray A. Effect of respiration, talking and small body movements on blood pressure measurement. J Hum Hypertens. 2012;26(7):458–62.

    Article  Google Scholar 

  7. Jones DW, Appel LJ, Sheps SG, Roccella EJ, Lenfant C. Measuring blood pressure accurately: new and persistent challenges. JAMA. 2003;289(8):1027–30.

    Article  Google Scholar 

  8. Zheng D, Di Marco LY, Murray A. Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement. Med Biol Eng Comput. 2014;52(5):467–73.

    Google Scholar 

  9. Venet R, Miric D, Pavie A, Lacheheb D. Korotkoff sound: the cavitation hypothesis. Med Hypotheses. 2000;55(2):141–6.

    Article  Google Scholar 

  10. Tavel ME, Faris J, Nasser WK, Feigebaum H, Fisch C. Korotkoff sounds. Observations on pressure-pulse changes underlying their formation. Circulation. 1969;39(4):465–74.

    Article  Google Scholar 

  11. McCutcheon EP, Rushmer RF. Korotkoff sounds. An experimental critique. Circ Res. 1967;20(2):149–61.

    Article  Google Scholar 

  12. Ur A, Gordon M. Origin of Korotkoff sounds. Am J Physiol. 1970;218(2):524–9.

    Google Scholar 

  13. Di Marco LY, Zheng D, Murray A. Effects of deep breathing on blood pressure measurement in healthy subjects. Computing Cardiol (CinC). 2012;2012(2012):9–12.

    Google Scholar 

  14. O’Brien E, Asmar R, Beilin L, Imai Y, Mallion JM, Mancia G, et al. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. J Hypertens. 2003;21(5):821–48.

    Article  Google Scholar 

  15. Griffiths CJ, Gilmartin JJ, Gibson GJ, Murray A. Measurement of chest wall movement; design, performance and clinical use of a four-channel magnetometer instrument. Clin Phys Physiol Meas. 1983;4(4):363–71.

    Article  Google Scholar 

  16. Yousefi R, Nourani M, Ostadabbas S, Panahi I. A motion-tolerant adaptive algorithm for wearable photoplethysmographic biosensors. Biomed Health Inform, IEEE J. 2014;18(2):670–81.

    Article  Google Scholar 

  17. Clark F, Von Euler Cv. On the regulation of depth and rate of breathing. J Physiol. 1972;222(2):267–95.

    Article  Google Scholar 

  18. Saul JP, Berger R, Albrecht P, Stein S, Chen MH, Cohen R. Transfer function analysis of the circulation: unique insights into cardiovascular regulation. Am J Physiol Heart Circ Physiol. 1991;261(4):H1231–45.

    Google Scholar 

Download references

Authors’ contributions

DZ, AM and FC conceived and designed the study; DZ performed the experiments; DC, DZ, FC and AM jointly analyzed the data and prepared the manuscript. All authors read and approved the final manuscript.

Acknowledgements

This study was supported by the Engineering and Physical Sciences Research Council (EPSRC) Healthcare Partnership Award (reference number EP/I027270/1), and EPSRC standard grant (reference number EP/F012764/1). DC and FC are supported by the National Natural Science Foundation of China (Grant No. 61571213).

Competing interests

The authors declare that they have no competing interests.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Fei Chen.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chen, D., Chen, F., Murray, A. et al. Respiratory modulation of oscillometric cuff pressure pulses and Korotkoff sounds during clinical blood pressure measurement in healthy adults. BioMed Eng OnLine 15, 53 (2016). https://doi.org/10.1186/s12938-016-0169-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12938-016-0169-y

Keywords