Biomechanical evaluation of three surgical scenarios of posterior lumbar interbody fusion by finite element analysis
https://doi.org/10.1186/1475-925X-11-31
© Xiao et al.; licensee BioMed Central Ltd. 2012
- Received: 23 March 2012
- Accepted: 18 June 2012
- Published: 18 June 2012
Abstract
Background
For the treatment of low back pain, the following three scenarios of posterior lumbar interbody fusion (PLIF) were usually used, i.e., PLIF procedure with autogenous iliac bone (PAIB model), PLIF with cages made of PEEK (PCP model) or titanium (Ti) (PCT model) materiel. But the benefits or adverse effects among the three surgical scenarios were still not fully understood.
Method
Finite element analysis (FEA), as an efficient tool for the analysis of lumbar diseases, was used to establish a three-dimensional nonlinear L1-S1 FE model (intact model) with the ligaments of solid elements. Then it was modified to simulate the three scenarios of PLIF. 10 Nm moments with 400 N preload were applied to the upper L1 vertebral body under the loading conditions of extension, flexion, lateral bending and torsion, respectively.
Results
Different mechanical parameters were calculated to evaluate the differences among the three surgical models. The lowest stresses on the bone grafts and the greatest stresses on endplate were found in the PCT model. The PCP model obtained considerable stresses on the bone grafts and less stresses on ligaments. But the changes of stresses on the adjacent discs and endplate were minimal in the PAIB model.
Conclusions
The PCT model was inferior to the other two models. Both the PCP and PAIB models had their own relative merits. The findings provide theoretical basis for the choice of a suitable surgical scenario for different patients.
Keywords
- Spine
- Cage
- PEEK
- Autogenous iliac bone
- Ligaments
Introduction
The aims of posterior lumbar interbody fusion (PLIF) procedure using cages or bone grafts are to provide stability of the motion segment and to facilitate the fusion process. After about 60 years of development and update, the surgical scenarios with cages or autogenous iliac bone (AIB) have been widely used.
The PLIF with AIB provided high fusion rate because the AIB was histo-compatible and non-immunogenic [1, 2]. However, several studies reported the major complications of this surgical method with a wide range of incidence varying between 1% and 39%, such as collapse, retropulsion of the grafted bone, and pseudoarthrosis [3–6]. To resolve such problems, the PLIF with cages was designed in 1991 [7]. The advantage of this surgical scenario was that the cages separated the mechanical and biologic functions of the PLIF. Many studies reported that the PLIF with cages could provide satisfactory clinical results [8–10]. However, this surgical scenario produced new problems such as adjacent segment degeneration (ASD), fine motion and mote of cages, and implants damage [11, 12].
Recently, with the development of material industry, polyetheretherketone (PEEK) aroused wide concern. Previous studies showed that PEEK was non-resorbable and elicited minimal cellular response, intracutaneous, and intramuscular toxicity [13, 14]. Both the in vitro and finite element (FE) studies showed that the implants made of PEEK material provided good experimental and clinical performances [15–20].
The finite element method, as an essential complement for the in vitro biomechanical studies, has been widely used for the study of lumbar spine [9–12, 20–22]. However, the major deficiency of FE model of lumbar spine was the simplification of both the anatomic structures and material properties of ligaments.
Although the influences of fusion rate, and ASD on the range of motion, stiffness, flexibility of lumbar spine following the PLIF procedure with AIB [21] and PLIF with cages made of PEEK [20] or Ti materiel [9, 10] have been investigated using FE method, respectively. To our knowledge, there are few studies evaluating the benefits or adverse effects among these three surgical scenarios using the model contained ligaments of three-dimensional (3D) solid elements. The aim of this study was to comparatively investigate the differences among three types of fusion construct, which may provide theoretical basis for the choice of a suitable surgical scenario for different patients.
Materials and methods
Establishment of the intact L1-S1 segment model
Bony boundary and the disc outlines from DICOM image based on different gray value threshold. (The region in the yellow outline was the disc and the region in the purple outline was bone structure.)
3D model of spinal segment L1-S1. ANN - annulus fibrosus, NUC - nucleus pulposus, ALL - anterior longitudinal ligament, PLL - posterior longitudinal ligament, IL - intertransverse ligament, LF - ligamenta flava, ISL - interspinal ligament, SSL - supraspinal ligament.
Each part of the model was imported into ABAQUS software and the surface mesh was converted to volumetric mesh. The minimum edge length of the TET elements was 0.7 mm and the maximum edge length was 0.9 mm. The intact model totally contained 56 parts, 5596653 TET elements and 1067522 nodes.
Establishment of the surgical models
A) simulation of the PLIF with cages
The implants of L4-5 surgical models, (a) M8 PSF, (b) cages, (c) bone grafts, (d) AIB.
The L4-L5 segment of the intact model was modified to simulate the PLIF procedure as shown in Figure 3. Two surgical models were established with Ti (E = 110 GPa) and PEEK (E = 3.5 GPa) material properties assigning to the cages of the PCT model and the PCP model, respectively.
B) simulation of the PLIF with autogenous iliac bone
All the modeling process of the PAIB model was the same as the modeling process of the PCT and PCP, except that the AIB was inserted between the vertebral bodies instead of the cages, as shown in Figure 3 (d).
The numbers of elements and nodes of the parts in the surgical models
M8 | L-Cage | R-Cage | Graft1 | Graft 2 | Graft 3 | AIB | Retained-ANN | |
---|---|---|---|---|---|---|---|---|
Elements | 339012 | 172203 | 171826 | 72966 | 80307 | 18562 | 74665 | 22157 |
Nodes | 67652 | 34284 | 34225 | 14013 | 15320 | 3787 | 14464 | 5009 |
Type | C3D4 | C3D4 | C3D4 | C3D4 | C3D4 | C3D4 | C3D4 | C3D4 |
Material properties
The material of the model can be divided into three types: bone, soft tissue, and surgical instrumentations. The material property of the bone structure was represented in MIMICS using some empirical expressions of the relationship among density, CT value, and Young’s modulus [23]. Since the gray values of thin cortical shell, inner trabecular core and endplates were different, the 200 kinds of different densities were calculated from the gray values based on the expressions. Then the 200 kinds of different elastic moduli were calculated from the above densities. This method would not only distinguish the material among the thin cortical shell, inner trabecular core and endplates, but also represent non-uniform material distribution.
The bone grafts were made from the mashed spinous process of L4, so the elastic modulus of bone grafts was low (E = 100 MPa) [10]. The iliac crest is mainly made up of cancellous bone, covered with a thin cortex. The material property of the iliac crest in this study referred to a previous study who defined the elastic modulus of the iliac crest as 1500 MPa [24].
Summary of the material properties used in our finite element models
Model | Material model | Material property | Literature | ||
---|---|---|---|---|---|
Bone | Linear-elastic | ρ = 1.067 * HU + 131 (g/cm3) | 23 | ||
E = 0.09882 ρ1.56 (MPa) | |||||
Intervertebral disc | Mooney-Rivlin formulation (MPa) | ||||
c10 | c01 | D 1 | |||
NUC | Hyper-elastic | 0.12 | 0.09 | 1 | 25 |
ANN | 0.56 | 0.14 | 1 | 26 | |
Ligaments | Ogden-3 formulation (MPa) | 27, 28 | |||
ALL | Hyper-elastic | (See Table 3) | |||
SSL | |||||
ISL | |||||
PLL | Hyper-elastic | (See Table 3) | |||
LF | |||||
IL | Hyper-elastic | (See Table 3) | |||
Facet Joint | Linear-elastic | Spring elements: E=250 (MPa) | 29 | ||
Modulus of Elasticity | Poisson’s Ratio (μ) | ||||
Graft bone | Linear-elastic | 100 (MPa) | 0.2 | 10 | |
AIB | Linear-elastic | 1500 (MPa) | 0.3 | 24 | |
PEEK | Linear-elastic | 3500 (MPa) | 0.3 | 20 | |
Titanium | Linear-elastic | 110 (GPa) | 0.3 | 11 |
The parameters of the fitting results for the different ligaments
μ 1 | α 1 | μ 2 | α 2 | μ 3 | α 3 | D 1 | D 2 | D 3 | |
---|---|---|---|---|---|---|---|---|---|
ALL, SSL, ISL | 0.177 | -3.080 | 0.627 | -13.860 | -0.357 | -6.800 | 1 | 1 | 1 |
PLL, LF | 0.159 | -1.126 | 0.770 | -18.540 | -0.390 | -9.600 | 1 | 1 | 1 |
IL | -6412.7 | -7.3 | 4159.0 | -9.0 | 2254.5 | 4.09 | 1 | 1 | 1 |
Contact, boundary and loading conditions
The interaction property “TIE” in ABAQUS was used to define all the “surface to surface” contacts. The FJs were simulated as spring elements. The intact model contained 95 TIE interactions and 300 spring elements [29].
The nodes of the inferior surface of S1 were completely fixed in all directions.
To validate our intact model, pure unconstrained 10 Nm extension (e), 10 Nm flexion (f), 10 Nm lateral bending (l), and 10 Nm torsion (t) moments were applied to the superior surface of L1 vertebral body, respectively. Five load steps were applied to reach to the 10 Nm moments in each loading condition.
To validate the intradiscal pressure (IDP) of the intact model, the L4-L5 segment was modeled and calculated independently. The inferior endplate of the L5 vertebral body was rigidly fixed. Pure unconstrained moments of 10 Nm extension and flexion were applied to the superior endplate of the L4 vertebral body, respectively.
To compare the differences among the three surgical models under physiological loading condition, the surgical models were stressed with a 400 N of axial compression and 10 Nm moments to simulate extension, flexion, lateral bending and torsion. The intact model was also recalculated under the above loading conditions.
Results
Model validation
To validate our intact model, the FE results of range of motion (ROM) were compared with a previous in vitro experimental study under the same loading conditions [30, 31]. As Table 4 shown, a good agreement was obtained between our numerical results and the reported data.
The comparison of the ROM between our FE model and the previous in vitro experimental study
Segment | Moment (Nm) | Extension and flexion | Left and right lateral bending | Left and right torsion | |||
---|---|---|---|---|---|---|---|
In vitro | FEA | In vitro | FEA | In vitro | FEA | ||
2.5 | 7.5(1.35) | 6.64 | 6.5 (1.4) | 6.9 | 2.7(1.3) | 3.62 | |
L1-L2 | 5 | 8.3(1.3) | 8.69 | 7.6 (1.35) | 8.5 | 2.6(1.5) | 3.04 |
7.5 | 9.2(1.25) | 8.93 | 8.9(1.2) | 9.32 | 3.9(1.1) | 3.47 | |
10 | 10.4(1.45) | 10.2 9 | 10.2(1.5) | 9.78 | 4(1.7) | 4.02 | |
2.5 | 6.5(1.25) | 7.62 | 8.8(1.2) | 9.23 | 2.7(0.9) | 3.24 | |
L2-L3 | 5 | 10.1(1.2) | 9.06 | 10.7(1.3) | 11.2 | 3.15(1) | 3.78 |
7.5 | 11.1(1.2) | 10.94 | 12(1.55) | 13 | 3.9(0.85) | 4.47 | |
10 | 11.5(1.3) | 10.29 | 13(1.15) | 13.43 | 5(0.95) | 5.48 | |
2.5 | 8.3(1.75) | 7.65 | 8.3(1.85) | 8.43 | 2.5(1.1) | 3.51 | |
L3-L4 | 5 | 10.5(2.2) | 8.93 | 10.5(2.15) | 9.5 | 3.7(1.05) | 4.27 |
7.5 | 9.9(1.6) | 10.24 | 11.4(1.5) | 11.9 | 4.7(1.15) | 4.79 | |
10 | 11.3(1.7) | 1.78 | 12.2(1.7) | 13.2 | 5.1(1.38) | 5.76 | |
2.5 | 10(1.3) | 11.25 | 7.5(1.45) | 8.21 | 1.8(1.02) | 2.57 | |
L4-L5 | 5 | 12.5(2.25) | 12.95 | 10.2(2.25) | 10 | 2.4(0.85) | 3.27 |
7.5 | 14(2) | 14.6 | 11.1(2.1) | 10 | 2.8(1.2) | 3.94 | |
10 | 14.8(2.1) | 14.2 | 12.2(2.25) | 13.23 | 3.7(1.5) | 4.23 | |
2.5 | 13.5(3.3) | 13.68 | 7.7(2.3) | 8.2 | 1.5(0.55) | 2.09 | |
L5-S1 | 5 | 15.1(1.95) | 16.17 | 9.3(2.15) | 8.3 | 1.8(0.75) | 2.35 |
7.5 | 16.4(2) | 17.66 | 10.2(2.2) | 11 | 2(0.65) | 2.48 | |
10 | 16.9(2.05) | 17.29 | 11.3(2.35) | 12.56 | 2.5(0.75) | 2.7 |
The comparison of the IDP between our FE model and the previous in vitro experimental study
Moment (Nm) | Extension | Flexion | ||
---|---|---|---|---|
In vitro | FEA | In vitro | FEA | |
1 | 0.04 (0.03) | 0.025 | 0.055 (0.035) | 0.037 |
2.5 | 0.08 (0.07) | 0.066 | 0.108 (0.05) | 0.087 |
5 | 0.13 (0.11) | 0.1335 | 0.18 (0.06) | 0.198 |
7.5 | 0.16 (0.135) | 0.198 | 0.27 (0.068) | 0.312 |
10 | 0.17 (0.16) | 0.262 | 0.35 (0.05) | 0.384 |
Element size, number of elements and calculational time of the model with different mesh density
Model A | Model B | Model C | Model D | Model E | |
---|---|---|---|---|---|
The edge length of the TET elements (mm) | 0.4-0.5 | 0.5-0.7 | 0.7-0.9 | 0.9-1.1 | 1.1-1.3 |
Number of elements | 3668172 | 1859589 | 977162 | 495715 | 362041 |
Calculational Time (h) | 36.98 | 8.43 | 3.84 | 1.3 | 0.45 |
The results of IDP of models A to E. (a) the results of IDP in extension, (b) the results of IDP in flexion. (Models A to E were five remeshed L4-L5 models with different mesh densities. The detailed parameters of the mesh densities of these models were shown in Table 6.)
The biomechanical behaviors of the three surgical models were compared with those of the intact model, respectively.
Stresses on the implants
The maximum stress on the M8 PSF was found to be 239.153 MPa in the PAIB model under lateral bending loading condition. The maximum stress on the M8 PSF was minimal in the PCT model, which was 93.3163 MPa under extension loading condition. The maximum stress on the M8 PSF in the PAIB model increased by 3.51% compared with the PCP model. The maximum stress on the M8 PSF in the PCP model increased by 3.76% compared with the PCT model.
There were obvious differences in the maximum stress and the average stresses on the cages of the PCT and PCP models. The maximum stress on cages of the PCT model was 3.6 times that of the PCP model, and the average stresses on cages of the PCT model was 3.36 times that of the PCP model, at the most.
Stresses and strains on ligaments
The increase/decrease rates or percentage changes in this study were described by the following equation: increase/decrease rate or percentage change = (Data of surgical model - Data of intact model)/Data of intact model × 100%.
The percentage change of the maximum Von Mises stress on different ligaments among three surgical models
ALL (L5-S1) | PLL (L3-L4) | LF (L1-L2) | IL (L3-L4) | ISL (L5-S1) | SSL (L5-S1) | |
---|---|---|---|---|---|---|
PCT | 44.75% | 22.16% | 1.78% | 31.45% | 8.26% | 14.35% |
PAIB | 44.08% | 16.63% | 1.93% | 39.04% | 67.31% | 72.31% |
PCP | 44.44% | 22.70% | 1.68% | 31.40% | 8.75% | 14.71% |
The percentage change of the average strains on different ligaments among three surgical models
ALL | PLL | LF | IL | ISL | SSL | |
---|---|---|---|---|---|---|
PCT | -19.02% | -7.46% | 0.81% | 23.72% | -4.01% | -5.15% |
PAIB | -17.09% | -9.48% | 5.93% | 30.89% | 2.27% | 1.54% |
PCP | -18.83% | -7.44% | 0.86% | 23.69% | -5.01% | -4.89% |
Stresses of the adjacent discs
The percentage change of the maximum Von Mises stress on the intervertebral discs among three surgical models
PCT | PAIB | PCP | ||
---|---|---|---|---|
Extension | L1-L2 | -27.92% | -26.11% | -27.92% |
L2-L3 | -34.31% | -31.34% | -34.20% | |
L3-L4 | -27.70% | -4.80% | -27.60% | |
L5-S1 | -35.70% | -5.65% | -35.70% | |
Flexion | L1-L2 | -3.65% | -3.48% | -3.60% |
L2-L3 | -20.79% | -19.68% | -20.59% | |
L3-L4 | -27.74% | -8.34% | -27.66% | |
L5-S1 | -41.67% | -11.27% | -41.41% | |
Lateral bending | L1-L2 | -7.89% | -6.60% | -7.72% |
L2-L3 | -9.15% | -7.76% | -9.03% | |
L3-L4 | -17.07% | -2.70% | -17.28% | |
L5-S1 | -62.97% | -62.39% | -63.14% | |
Torsion | L1-L2 | 13.50% | 13.45% | 13.49% |
L2-L3 | 6.64% | 7.62% | 6.67% | |
L3-L4 | 8.18% | 15.61% | 8.36% | |
L5-S1 | 60.89% | 61.58% | 62.72% |
Stresses on endplates of surgical segment
The maximum stress on L4 inferior endplate was 20.83 and 15.76 times that of the intact model at the most in the PCT and PCP models, respectively. The maximum stress on L5 superior endplate was 3.67 and 2.09 times that of intact model, at the most. And the maximum stress on endplate of the PCP model was smaller than the PCT model in each loading condition. The stresses on the endplates in the PAIB model were smaller than the intact model with the maximum decrease rate of 86%.
Stresses on bone grafts
The average Von Mises stresses on bone grafts of three surgical models in each loading condition
Model | Extension | Flexion | Lateral bending | Torsion | |
---|---|---|---|---|---|
graft-1 | PCT | 0.122157 | 0.194958 | 0.016597 | 0.128982 |
PCP | 0.228447 | 0.409267 | 0.04922 | 0.288651 | |
graft-2 | PCT | 0.147298 | 0.306561 | 0.144857 | 0.170196 |
PCP | 0.249845 | 0.571044 | 0.28286 | 0.328011 | |
graft-3 | PCT | 0.20448 | 0.302199 | 0.105593 | 0.202865 |
PCP | 0.27222 | 0.507136 | 0.18066 | 0.337146 | |
AIB | PAIB | 0.292492 | 0.737651 | 0.296522 | 0.499107 |
Discussion
Finite element analysis (FEA) is a sophisticated simulation method, and also an effective tool for elucidating biomechanics in the spine. In the biomechanical evaluations based on FEA, it is important to establish a model that can accurately reproduce the mechanical property of each part. Establishing such a model requires accurate data on anatomic structures and material properties [13]. However, since ligaments show complicated material properties and large deformation, it is difficult to establish an accurate model of ligaments in FEA. Many researchers used two-dimensional tension-only truss or cable elements to describe the function of ligaments [11, 13, 22, 26, 29]. In the present study, the surrounding ligaments were modeled with three-dimensional solid elements. The material properties of ligaments were simulated by hyper-elastic Ogden-3 formulation based on the experimental data. The validated results indicated that the model established in this study could effectively reproduce the mechanical behaviors of L1-S1 lumbar segment. In addition, another advantage of the model established in this study was that it could directly obtain the stresses and strains of the ligaments. The results may be useful to predict the chronic degeneration and disease of ligaments.
The intensive discussions among the three surgical scenarios were shown below.
Stresses of M8 PSF
The largest maximum stress on the M8 PSF was found in the PAIB model, with the PCP model following and the PCT model being the least in each loading condition. The increase in the stresses on the M8 PSF may induce the increase in the risk of the breakage of PSF. The maximum stress on the M8 PSF was significant larger in lateral bending and extension than in flexion and torsion. Therefore, clinically the patients were recommended to avoid excessive lateral bending and extension movements in the process of treatment and recuperation.
Stresses of cages
As the Ti material was stiffer than the PEEK material, both the maximum stress and the average stresses on the cages of PCT model were larger than those on the PCP model, which indicated that the Ti material cages in the PCT models suffered more stresses concentration than the PEEK cages. The greater stresses on cages may increase the risk of fine motion and mote on cages. The fine motion and mote of cages would cause inflammation of the fused segment and have adverse effect on the fusion process. So the PCT model was obviously inferior to the PCP model in this respect.
Stresses and strains on ligaments
Compared with the intact model, the stresses on ligaments of the three surgical models increased significantly and the maximum increase of the stresses located at the segments that proximally adjacent to fusion segment (See Table 7). The greater stresses on ligaments were found in the PAIB model than the other two models. The maximum stress on PAIB model was about 8 times that of the PCT model and 5 times that of the PCP model, at the most.
The PAIB model also produced larger strains on the majority of ligaments. The ligaments were pre-stressed due to the increase of strains on ligaments, which reduced the ability of ligaments to resist stretching. The ligaments would be injured or fragmented more easily, when there is external load applied on the spine. The increase of stresses and strains on the ligaments also changed the normal physiological and mechanical environments of ligaments. These changes were likely relevant to the invocation of early pain and prone to cause chronic soft tissue injury and degeneration. In this respect, the PCP and PCT models were better than PAIB model. To our knowledge, there were few reports describing the stresses on ligaments in the PLIF procedure.
Stresses of the adjacent intervertebral discs
Both the postoperative following-up and biomechanical studies showed that the PLIF accelerated degeneration of adjacent segment and segmental instability [11]. The FE results showed that great changes were found in the stresses on the discs proximally adjacent to the fusion segment. These great changes in discs could be used to interpret the clinical findings of early degeneration of adjacent disc [12]. The increase of the maximum Von Mises stresses on adjacent discs during torsion was probably due to the following reasons: the M8 PSF restrained more ROM than other loading conditions. Thus, the ROM of the adjacent segment increased a lot. And the Von Mises stresses on adjacent discs also increased. The decreases of the stresses on the discs in the PAIB model were smaller than the other two surgical models, especially at fusion adjacent segments under extension, flexion and torsion. This result showed that the surgical method using AIB could decrease the risk of degeneration of fusion adjacent discs.
Maximum stress on the endplate of surgical segment
Of all the structures, the most significant changes in the maximum stress occurred on the L4 inferior endplate and L5 superior endplate. There were two main reasons that caused the tremendous increase in the stress on the endplates. Firstly, although the jagged edges of the cages avoided the relevant moments between endplates and cages, this design resulted in stress concentration. Secondly, the materials of PEEK and Ti were much stiffer than the bone grafts. So the phenomenon of stress shielding on cages was serious. The majority of the load was transferred onto the cages instead of the bone grafts. So the PLIF with cages caused tremendous increase in the stress on the endplates. Excessive stresses on the endplate may cause osteolysis of the endplate and subsidence of the fused segment. Compared with the surgical model using cages, the surgical model using AIB could reduce the stresses concentration on the endplates obviously, thus protect the endplate of the surgical segment.
From the results of stresses on the adjacent discs and the endplates, it was shown that the PAIB model was better than the other two models. Therefore, this surgical method was recommended for the elderly patients who had already suffered from the ASD and osteoporosis. This was because that the two major risks faced in the PLIF procedure were the further degenerative diseases of surgery adjacent segments and the subsidence or damage of endplate, which would eventually result in the failure of fusion surgery. Compared to the two models using cages, the PAIB model could effectively abate such phenomenon.
Average stresses on bone grafts
Contour plots of Von Mises stresses on bone grafts for each surgical model in torsion. (a) the graft-1 in the PCT model, (b) the graft-2 in the PCT model, (c) the graft-3 in the PCT model, (d) the graft-1 in the PCP model, (e) the graft-2 in the PCP model, (f) the graft-3 in the PCP model, (g) the AIB in the PAIB model.
To conclude, the comparative results showed that the greatest stresses on cages and endplates were found in the PCT model, but the stresses on the bone grafts were found lowest in this model, so it may be inferior to the other two surgical models. The PCP model and PAIB model showed a similar considerable stresses on the bone grafts. However, the PCP model showed a decrease in the percentage change of ligaments and less stresses on PLIF. The PAIB model showed a decrease in the percentage change of adjacent discs and lower stresses on endplates.
There are certain limitations in this FE study. Firstly, The ANN was modeled with an isotropic hyper-elastic material model without the fiber-reinforced structure. Secondly, the facet joints and capsular ligaments were simplified to 30 spring elements. Under the actual condition, the structures are more complex. Thirdly, muscle contractions were not simulated in the current study. The muscle contractions may bring complicated external forces that have significant influences on the biomechanical perspective [9]. The above factors will be considered in our further study. Although there were certain simplifications in our FE model, the FE model was well validated by the previous in vitro study. Therefore, the model established in this study is reasonable and can be used as an efficient tool to evaluate the effects of three surgical scenarios on the lumbar spine.
Conclusions
The PCT model may be inferior to the other two surgical models. Therefore it was not recommended to use cages made of Ti material in an instrumented PLIF. Both the PCP model and PAIB model had their own relative merits, so the doctor could choose the most suitable surgical method based on the finding in this research for the different clinical circumstances.
The modeling method using the ligaments with 3D solid elements can be extended to other body parts such as knee joint, ankle joint and shoulder joint, in which the ligaments play an important role. The model can also be used as the basis for our further study, in which surgical models having cages with different shapes and grafts will be developed. Besides, bone remodeling theory will be introduced to predict the long-term bone graft fusion, which could provide theoretical basis for clinical postoperative rehabilitation.
Declarations
Acknowledgement
This work was supported by National Natural Science Foundation of China (Nos.10832012, 10972090), and The National Key Basic Research Development Program (973 Program) Prior Special Study (No. 2012CB723802).
Authors’ Affiliations
References
- Arrington ED, Smith WJ, Chambers HG, et al.: Complications of iliac crest bone graft harvesting. Clinorthop Relat R 1996, 329: 300–309.View ArticleGoogle Scholar
- Asazuma T, Masuoka K, Motosuneya T, et al.: Posterior lumbar interbody fusionusing dense hydroxyapatite blocks and autogenous iliac bone. Clinical and radiographic examinations. J Spinal Disord Tech 2005, 18: S41–47. 10.1097/01.bsd.0000112043.70321.1aView ArticleGoogle Scholar
- Sengupta DK, Truumees E, Patel CK, et al.: Outcome of local bone versus autogenous iliac crest bone graft in the instrumented posterolateral fusion of the lumbar spine. Spine 2006, 31(9):985–991. 10.1097/01.brs.0000215048.51237.3cView ArticleGoogle Scholar
- Arai Y, Takahashi M, Kurosawa H, et al.: Comparative study of iliac bone graft and carbon cage with local bone graft in posterior lumbar interbody fusion. J Orthop Surg 2002, 10: 1–7.Google Scholar
- Hutter CG: Posterior intervertebral body fusion-a 25 year study. Clin Orthop 1983, 179: 86–96.View ArticleGoogle Scholar
- Brantigan JW: Pseudoarthrosis rate after allograft posterior lumbar interbody fusion with pedicle screw and plate fixation. Spine 1994, 19: 1271–1280.View ArticleGoogle Scholar
- Brantigan JW, Steffee AD, Geiger JM: A carbon fiber implant to aid interbody lumbar fusion. Mechanical testing. Spine 1991, 16(6):277–282.View ArticleGoogle Scholar
- Kuslich SD, Danielson G, Dowdle JD, et al.: Four-year follow-up results of lumbar spine arthrodesis using the bagby and kuslich lumbar fusion cage. Spine 2000, 25(20):2656–2662. 10.1097/00007632-200010150-00018View ArticleGoogle Scholar
- Kim HJ, Chun HJ, Moon SH, et al.: Analysis of biomechanical changes after removal of instrumentation in lumbar arthrodesis by finite element analysis. Med Biol Eng Comput 2010, 48: 703–709. 10.1007/s11517-010-0621-2View ArticleGoogle Scholar
- Kim Y: Finite element analysis of anterior lumbar interbody fusion threaded cylindrical cage and pedicle screw fixation. Spine 2007, 32(23):2558–2568. 10.1097/BRS.0b013e318158cdd8View ArticleGoogle Scholar
- Chosa E, Goto K, totoribe K, et al.: Analysis of the effect of lumbar spine fusion on the superior adjacent intervertebral disk in the presence of disk degeneration, using the three-dimensional finite element method. J Spinal Disord Tech 2004, 17(2):134–139. 10.1097/00024720-200404000-00010View ArticleGoogle Scholar
- Chiang MF, Zhong ZH, Chen CS, et al.: Biomechanical comparison of instrumented posterior lumbar interbody fusion with one or two cages by finite element analysis. Spine 2006, 31(19):682–689. 10.1097/01.brs.0000232714.72699.8eView ArticleGoogle Scholar
- Kurtz SM, Devine JN: PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials 2007, 28(32):4845–4869. 10.1016/j.biomaterials.2007.07.013View ArticleGoogle Scholar
- Toth JM, Wang M, Estes BT, et al.: Polyetheretherketone as a biomaterial for spinal applications. Biomaterials 2006, 27: 324–334. 10.1016/j.biomaterials.2005.07.011View ArticleGoogle Scholar
- Boakye M, Mummaneni PV, Garrett M, et al.: Anterior cervical discectomy and fusion involving polyetheretherketone spacer and bone morphogenetic protein. J Neurosurg Spine 2005, 2(5):521–525. 10.3171/spi.2005.2.5.0521View ArticleGoogle Scholar
- Rivard CH, Rhalmi S, Coillard C, et al.: In vivo biocompatibility testing of peek polymer for a spinal implant system: a study in rabbits. J Biomed Mater Res 2002, 62(4):488–498. 10.1002/jbm.10159View ArticleGoogle Scholar
- Ponnappan KR, Serhan H, Zarda B, et al.: Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation. Spine 2009, 9: 263–267. 10.1016/j.spinee.2008.08.002View ArticleGoogle Scholar
- Rousseau MA, Lazennec JY, Saillant G, et al.: Circumferential arthrodesis using PEEK cages at the lumbar spine. J Spinal Disord Tech 2007, 20(4):278–281. 10.1097/01.bsd.0000211284.14143.63View ArticleGoogle Scholar
- Vaidya R, Sethi A, Bartol S, et al.: Complications in the use of rhBMP-2 in PEEK cages for interbody spinal fusions. Spinal Disord Tech 2008, 21: 557–562. 10.1097/BSD.0b013e31815ea897View ArticleGoogle Scholar
- Vadapalli S, Sairyo K, Goel VK, et al.: Biomechanical rationale for using polyetheretherketone (PEEK) spacers for lumbar interbody fusion finite element study. Spine 2006, 31(26):992–998. 10.1097/01.brs.0000250177.84168.baView ArticleGoogle Scholar
- Lee KK, Teo EC, Fuss FK, et al.: Finite-element analysis for lumbar interbody fusion under axial loading. IEEE T Bio-med Eng 2004, 51(3):393–400. 10.1109/TBME.2003.820994View ArticleGoogle Scholar
- Xiao ZT, Wang LY, Gong H, et al.: Establishment and verification of a non-linear finite element model for human L4-L5 lumbar segment. BMEI 2010 , 3: 1171–1175.Google Scholar
- Elise FN, Harun HB, Tony MK, et al.: Trabecular bone modulus–density relationships depend on anatomic site. J Biomech 2003, 36: 897–904. 10.1016/S0021-9290(03)00071-XView ArticleGoogle Scholar
- Tie Y, Wang DM, Ji T, et al.: Three-dimensional finite-element analysis investigating the biomechanical effects of human mandibular reconstruction with autogenous bone grafts. J Cranio Maxill Surg 2006, 34: 290–298. 10.1016/j.jcms.2006.03.004View ArticleGoogle Scholar
- Smit TH: The mechanical significance of the trabecular bone architecture in a human vertebra. Technische University Hamburg-Harburg 1996, 49–53.Google Scholar
- Goel VK, Monroe BT, Gilbertson LG, et al.: Interlaminar shear stresses and laminae separation in a disc. Finite element analysis of the L3–L4 motion segment subjected to axial compressive loads. Spine 1995, 20(6):689–698. 10.1097/00007632-199503150-00010View ArticleGoogle Scholar
- Sharma M, Langrana NA, Rodriguez J: Role of ligaments and facets in lumbar spinal sility. Spine 1995, 20: 887–900. 10.1097/00007632-199504150-00003View ArticleGoogle Scholar
- Xiao ZT, Wang LY, Gong H, et al.: A nonlinear finite element model of human L4-L5 lumbar spinal segment with three-dimensional solid element ligaments. Theor Appl Mech Lett 2011, 1(064001):1–6.Google Scholar
- Wang XS: Establishment of a simulation model and a finite element model, related biomechanical effects analysis of AIS of PUMCIIDII, different fusion level. Thesis, Peking Union Medical Colleg: Ph. D, ; 2003.Google Scholar
- Panjabi MM, Oxland TR, Yamamoto I, et al.: Mechanical behavior of the human lumbar and lumbosacral spine as shown by three-dimensional load–displacement curves. J Bone Joint Surg Am 1994, 76(3):413–424.Google Scholar
- Yamamoto I, Panjabi MM, Crisco JJ, et al.: Three-dimensional movements of the whole lumbar spine and lumbosacral joint. Spine 1989, 14: 1256–1260. 10.1097/00007632-198911000-00020View ArticleGoogle Scholar
- Heuer F, Schmidt H, Claes L, et al.: Stepwise reduction of functional spinal structures increase vertebral translation and intradiscal pressure. J Biomech 2007, 40: 795–803. 10.1016/j.jbiomech.2006.03.016View ArticleGoogle Scholar
Copyright
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.