Skip to main content

Table 4 Parkes consensus error grid analysis interval zones

From: A concept for human use of real-time and remote monitoring of diabetic subjects using intermittent scanned continuous glucose measurement

Data interval/lapse time (h)

None

200/50

100/25

64/16

32/8

16/4

4/1

Zone A (%)

93.06

96.02

96.62

95.77

96.46

97.54

98.15

Zone B (%)

6.76

3.83

3.19

4.03

3.41

2.31

1.68

Zone C (%)

0.18

0.15

0.19

0.19

0.13

0.15

0.04

Zone D (%)

0.00

0.00

0.00

0.00

0.00

0.00

0.00

Zone E (%)

0.00

0.00

0.00

0.00

0.00

0.00

0.00

  1. Values are reported as a percentage, h: hours