Skip to main content

Table 4 Root mean squared error of the stimulation based on the Kinect v2’s COM and the pressure mat’s COP compared against that based on the force plate’s filtered COP across the VFT balance exercises for the plantarflexors

From: Feasibility of using a depth camera or pressure mat for visual feedback balance training with functional electrical stimulation

 

Left plantarflexors (LPF)

Right plantarflexors (RPF)

Depth camera

(Mean ± SD mA)

Pressure mat

(Mean ± SD mA)

Depth camera

(Mean ± SD mA)

Pressure mat

(Mean ± SD mA)

Color

3.11 ± 0.99

3.32 ± 1.19

3.45 ± 1.02

2.16 ± 1.49

Hunt

2.25 ± 0.70

2.57 ± 1.21

2.30 ± 0.95

1.44 ± 0.96

Ellipse

1.81 ± 1.04

3.58 ± 1.36

1.98 ± 0.52

2.32 ± 1.37

Bullseye

0.54 ± 0.18

0.78 ± 0.45

0.62 ± 0.16

0.43 ± 0.13