Table 2 Identification performance compared to clinical inspection by type of asynchrony and for non-asynchrony breaths

From: Automated evaluation of typical patient–ventilator asynchronies based on lung hysteretic responses

Asynchrony types

TP

FN

TN

FP

Sensitivity (%)

Specificity (%)

Accuracy (%)

fa

68

8

4912

12

89.5

99.8

99.6

rt

595

13

4384

8

97.9

99.8

99.6

pc

383

22

4581

14

94.6

99.7

99.3

dt

241

5

4748

6

98.0

99.9

99.8

dc

202

15

4741

42

93.1

99.1

98.9

ie

133

9

4848

10

93.7

99.8

99.6

at

415

34

4543

8

92.4

99.8

99.2

Non-asynchrony breaths

2542

52

2328

78

98.0

96.8

97.4