Table 1 Identification performance compared to clinical inspection for each patient and overall without subdividing by asynchrony type, which is in Table 3

From: Automated evaluation of typical patient–ventilator asynchronies based on lung hysteretic responses

Patient

Total breaths

# Asynchronous

# No asynchrony

Sensitivity

Specificity

Accuracy

1

500

154

346

99.4%

99.4%

99.4%

2

500

22

478

95.5%

99.8%

99.6%

3

500

89

411

92.1%

99.3%

98.0%

4

500

496

4

99.0%

N/A

98.6%

5

500

274

226

93.8%

99.6%

96.4%

6

500

231

269

94.8%

93.7%

94.2%

7

500

218

282

99.5%

96.1%

97.6%

8

500

433

67

94.9%

92.5%

94.6%

9

500

216

284

97.2%

97.5%

97.4%

10

500

10

490

90.0%

97.1%

97.0%

11

500

0

500

N/A

100.0%

100.0%

Overall

5500

2143

2857

96.6%

97.8%

97.3%