Skip to main content

Table 3 Evaluation of the precision and application effect of AI

From: Application effect of an artificial intelligence-based fundus screening system: evaluation in a clinical setting and population screening

 

Accuracy(%)

Kappa value

Conformance strength

PPV (%)

NPV (%)

Normal

86.66

0.709*

High

85.09

90.05

ARMD

97.33

0.593*

Middle

44.83

99.84

DR

99.21

0.844*

Strongest

87.50

99.52

RVO

99.53

0.901*

Strongest

82.35

100.00

Referable glaucoma

91.05

0.510*

Middle

48.65

96.63

PM

97.80

0.836*

Strongest

81.25

99.15

Other abnormalities

86.34

0.234*

Reasonable

94.12

86.13

Any fundus abnormalities

86.66

0.709*

High

90.05

85.09

  1. PPV positive predictive value, NPV negative predictive value
  2. *P < 0.001