Fig. 9 | BioMedical Engineering OnLine

Fig. 9

From: A new radiofrequency balloon angioplasty device for atherosclerosis treatment

Fig. 9

Setup for phantom gel experiment: a diagram of the catheter and thermocouples holder; b the dimensions of the holes and the relative position for the catheter and thermocouples: the larger hole with radius of 1.5 mm was for the catheter and the TC0 (the blue circle) located on the surface of the catheter, the small holes were for the thermocouples; c, d a picture of the phantom into which thermocouples and catheter were inserted: c side view, the thermocouple 3 and 4 were on the back of the tube; d top view, the thermocouple 1 was shielded by the thermocouple 2. The positive and negative electrodes were outlined with a red and blue rectangle, respectively