Skip to main content

Table 1 Percent relative computational times of the NVCA and moving average (MA) filters calculated with respect to the computational time of VBM4D, taken as reference

From: Real-time algorithm for Poissonian noise reduction in low-dose fluoroscopy: performance evaluation

Spatial mask dimension

3 × 3

5 × 5

7 × 7

Temporal mask dimension

1

2

3

5

7

9

1

2

3

5

7

9

1

2

3

5

7

9

NVCA-to-VBM4D computation time (%)

1.5

2.1

2.4

3.1

3.7

4.3

2.7

3.9

4.7

6.6

8.5

10.0

4.7

6.4

8.1

11.7

15.2

18.3

MA-to-VBM4D computation time (%)

1.1

1.2

1.2

1.3

1.4

1.5

1.1

1.4

1.7

2.1

2.3

2.7

1.6

2.0

2.5

3.1

3.7

4.3