Skip to main content

Table 2 The average threshold for different electrode diameter up to 200 μm is T (μA) ± standard error of the mean (SEM). The average current density is CD (mA/cm2) ± SEM

From: Electrochemical characteristics of microelectrode designed for electrical stimulation

Electrode diameter (μm)

Pulse width

60 μs

400 μs

1000 μs

T ± SEM

CD ± SEM

T ± SEM

CD ± SEM

T ± SEM

CD ± SEM

10

6.9 ± 0.6

87.90 ± 7.64

1.05 ± 0.04

13.38 ± 0.51

0.87 ± 0.03

11.09 ± 0.38

30

6.0 ± 0.4

8.49 ± 0.57

1.84 ± 0.09

2.60 ± 0.08

1.32 ± 0.04

1.87 ± 0.06

60

21.7 ± 0.5

7.68 ± 0.18

3.20 ± 0.20

1.13 ± 0.07

3.20 ± 0.10

1.13 ± 0.04

200

13.5 ± 0.4

0.43 ± 0.01

3.3 ± 0.10

0.11 ± 0.00

2.33 ± 0.07

0.07 ± 0.00