From: Integrated EIT system for functional lung ventilation imaging

Fig. 1

TV images and TV values extracted from the images. a TV images of a mechanically ventilated pig with 6 different tidal volume settings. b Linear relation between TV from the reconstructed TV images and the air volume supplied from the mechanical ventilator. c Bland–Altman plot with ± 2 standard deviation (SD)