Table 2 Performance measures of tested heart beat detector on SCG signals m001–m020

From: Comparison of HRV indices obtained from ECG and SCG signals from CEBS database

Signal

TP

FP

FN

Beats

Se

PPV

m001

3794

129

113

3907

0.971

0.967

m002

3205

20

20

3225

0.994

0.994

m003

2283

339

836

3119

0.732

0.871

m004

3404

22

3

3407

0.999

0.994

m005

1949

1656

1650

3599

0.542

0.541

m006

3086

32

30

3116

0.990

0.990

m007

2596

23

18

2614

0.993

0.991

m008

5008

3

11

5019

0.998

0.999

m009

2949

234

211

3160

0.933

0.926

m010

2148

842

840

2988

0.719

0.718

m011

3450

148

146

3596

0.959

0.959

m012

3744

233

246

3990

0.938

0.941

m013

3707

5

4

3711

0.999

0.999

m014

3378

39

39

3417

0.989

0.989

m015

3204

2

1

3205

1.000

0.999

m016

3860

2

1

3861

1.000

0.999

m017

3574

11

12

3586

0.997

0.997

m018

4011

69

93

4104

0.977

0.983

m019

3178

14

17

3195

0.995

0.996

m020

3386

15

12

3398

0.996

0.996

Total

65,914

3838

4303

70,217

0.939

0.945