Table 2 Performance measures of tested heart beat detector on SCG signals m001–m020

From: Comparison of HRV indices obtained from ECG and SCG signals from CEBS database

Signal TP FP FN Beats Se PPV
m001 3794 129 113 3907 0.971 0.967
m002 3205 20 20 3225 0.994 0.994
m003 2283 339 836 3119 0.732 0.871
m004 3404 22 3 3407 0.999 0.994
m005 1949 1656 1650 3599 0.542 0.541
m006 3086 32 30 3116 0.990 0.990
m007 2596 23 18 2614 0.993 0.991
m008 5008 3 11 5019 0.998 0.999
m009 2949 234 211 3160 0.933 0.926
m010 2148 842 840 2988 0.719 0.718
m011 3450 148 146 3596 0.959 0.959
m012 3744 233 246 3990 0.938 0.941
m013 3707 5 4 3711 0.999 0.999
m014 3378 39 39 3417 0.989 0.989
m015 3204 2 1 3205 1.000 0.999
m016 3860 2 1 3861 1.000 0.999
m017 3574 11 12 3586 0.997 0.997
m018 4011 69 93 4104 0.977 0.983
m019 3178 14 17 3195 0.995 0.996
m020 3386 15 12 3398 0.996 0.996
Total 65,914 3838 4303 70,217 0.939 0.945