Table 2 Ablation size acquired by different methods

From: Ablation outcome of irreversible electroporation on potato monitored by impedance spectrum under multi-electrode system

Group

Experiment measurement (cm)

Eth for simulation (V/cm)

Size by simulation (cm)

Errors (%)

R rel

Size by functions (cm)

Errors (%)

10

X2 = 0.67 ± 0.05

180

0.608

8.57

12.12

0.72

7.49

10

X3 = 0.90 ± 0.11

180

1.126

25.2

2.71

1.02

13.90

X4 = 0.96 ± 0.23

1.126

16.9

2.43

0.94

1.57

10

X5 = 1.49 ± 0.23

180

1.517

1.65

1.42

1.38

7.58

11

X2 = 0.70 ± 0.12

180

0.608

12.52

12.13

0.71

2.87

11

X3 = 1.23 ± 0.16

180

1.126

8.38

2.91

1.08

12.08

X4 = 1.14 ± 0.11

1.126

0.94

2.72

1.03

9.47

11

X5 = 1.73 ± 0.10

180

1.517

12.46

1.50

1.43

17.4

12

X2 = 0.74 ± 0.15

140

0.734

0.94

13.63

0.77

4.50

12

X3 = 1.37 ± 0.02

140

1.359

1.09

3.92

1.36

0.98

X4 = 1.35 ± 0.11

1.359

0.37

3.70

1.30

3.95

12

X5 = 1.74 ± 0.04

140

1.901

9.48

2.04

1.76

1.42