Table 1 Results of pressure measurement on the back area

From: Evaluation of the effect of different sitting assistive devices in reclining wheelchair on interface pressure

Modes

BCA (cm2)

BAP (kPa)

BPP (kPa)

BRM

152.44 ± 88.02

2.59 ± 0.31

4.71 ± 1.35

 P (BRM vs. LBM)

0.518

0.063

0.162

 P (BRM vs. FBM)

0.796

0.550

0.305

 P (BRM vs. LFBM)

0.918

0.132

0.023

LBM

166.19 ± 93.59

2.39 ± 0.20

4.10 ± 0.80

 P (LBM vs. FBM)

0.642

0.213

0.576

 P (LBM vs. LFBM)

0.938

0.569

0.407

FBM

153.69 ± 70.73

2.53 ± 0.35

4.44 ± 1.50

 P (FBM vs. LFBM)

0.959

0.300

0.255

LFBM

163.06 ± 83.18

2.42 ± 0.30

3.87 ± 1.26

 P (BRM vs. LBM vs. FBM vs. LFBM)

0.913

0.178

0.019

  1. Comparison of mean contact area (CA), average pressure (AP), and peak pressure (PP) on the back area (BCA, BAP, and BPP) across four different modes, which include back reclined mode (BRM), lumbar support with back reclined mode (LBM), femur upward with back reclined mode (FBM), and lumbar support with femur upward with back reclined mode (LFBM). Values are mean ± standard deviation (N = 16)
  2. P is given for statistical significance