From: In silico evaluation of geometry variations with respect to the thermal spread during coagulation of egg white using bipolar vessel sealing instruments

Fig. 3

Simulated coagulation progress at \(35\,{V_p}\). The columns correspond to the wedge angles of 180°, 176°, 160°, 141° and 90° respectively. The first row shows the beginning of the coagulation, when temperatures reach locally at least 57 °C. The second row shows the moment when temperatures reach locally at least 60 °C. The third row shows a representative intermediate step of the coagulation. The last row shows the final shape of the coagulated egg white. The instrument’s shape is overlayed in white. The white lines separate the outer isolator from the electrode