Table 4 Detailed results for each patient from part 1 of the study

From: Interference between surgical magnetic drapes and pacemakers: an observational study comparing commercially available devices and a new magnetically isolated drape

Subject

Date

Age (years)

Gender

Weight (kg)

Height (m)

BMI

Pacemaker model

Asynchronous rhythm with CVD drape

Displacementa

Asynchronous rhythm with prototype drape (with and without displacement)

1

20/10/2011

82

F

68

1.68

24.02

Boston Scientific

No

No

2

20/10/2011

73

M

82

1.78

25.75

Boston Scientific

Yes

Yes

No

3

20/10/2011

81

F

60

1.60

23.44

Boston Scientific

No

No

4

20/10/2011

65

M

84

1.73

28.20

Boston Scientific

No

No

5

27/10/2011

82

M

66

1.75

21.49

Boston Scientific

No

No

6

27/10/2011

52

M

102

1.78

32.19

Boston Scientific

No

No

7

27/10/2011

78

F

54

1.63

20.47

Boston Scientific

Yes

Yes

No

8

27/10/2011

75

F

88

1.60

34.38

Boston Scientific

No

No

9

27/10/2011

84

F

59

1.60

23.05

Boston Scientific

No

No

10

27/10/2011

86

M

77

1.70

26.68

Boston Scientific

No

No

11

27/10/2011

82

M

95

1.73

31.84

Boston Scientific

Yes

No

No

12

27/10/2011

81

F

61

1.60

23.91

Boston Scientific

Yes

No

No

13

27/10/2011

70

F

55

1.52

23.81

Boston Scientific

No

No

14

27/10/2011

76

M

85

1.74

28.17

Boston Scientific

Yes

Yes

No

15

27/10/2011

76

M

56

1.64

20.82

Boston Scientific

Yes

No

No

16

27/10/2011

80

F

77

1.65

28.17

Boston Scientific

Yes

Yes

No

17

27/10/2011

73

F

102

1.65

37.50

Boston Scientific

No

No

18

3/11/2011

81

M

64

1.68

22.50

Medtronic

Yes

Yes

No

19

3/11/2011

92

M

57

1.70

19.62

Medtronic

Yes

Yes

No

20

3/11/2011

85

F

69

1.63

25.78

Medtronic

Yes

Yes

No

21

3/11/2011

75

M

94

1.75

30.79

Medtronic

Yes

Yes

No

22

3/11/2011

85

M

66

1.65

24.17

Medtronic

Yes

Yes

No

23

3/11/2011

77

M

93

1.65

34.31

Medtronic

Yes

Yes

No

24

3/11/2011

76

M

108

1.83

32.25

Medtronic

Yes

No

No

25

3/11/2011

90

M

61

1.68

21.68

Medtronic

Yes

Yes

No

26

3/11/2011

87

F

45

1.55

18.90

Medtronic

Yes

Yes

No

27

3/11/2011

77

F

93

1.65

34.16

Medtronic

Yes

No

No

  1. aDisplacement was necessary in order to obtain asynchronous rhythm