Skip to main content

Table 6 Precision of the perimeter measurement

From: Wound perimeter, area, and volume measurement based on laser 3D and color acquisition

Seg. m.

CED

GrowCut

GrabCut

Wound A

143.8 ± 1.0

138.4 ± 2.4

136.8 ± 0.3

Wound B

110.6 ± 0.7

109.5 ± 0.5

105.4 ± 0.7

Wound C

139.0 ± 3.8

140.1 ± 5.8

135.1 ± 5.3

Wound D

81.2 ± 3.0

88.9 ± 5.1

78.4 ± 4.1

Average

2.1

3.5

2.6

  1. Precision of the perimeter measurement using different wound segmentation methods (Seg. m.) for all four test wounds. All values are in mm.