Table 1 Time parameters describing the pupillary light reflex

From: A high-resolution binocular video-oculography system: assessment of pupillary light reflex and detection of an early incomplete blink and an upward eye movement

 

All

Direct

Consensual

p-value

T1 (s)

0.199 ± 0.031

0.203 ± 0.021

0.195 ± 0.039

>0.05

T2 (s)

0.638 ± 0.083

0.642 ± 0.082

0.633 ± 0.086

>0.05

T3 (s)

1.562 ± 0.234

1.609 ± 0.215

1.516 ± 0.249

>0.05

  1. Mean and standard deviation of time parameters of the PLR obtained from the sequences. A comparison between direct and consensual reflexes revealed no significant differences.