Table 1 Requirement specifications for the developed chronic wound monitoring system

From: A flexible and low power telemetric sensing and monitoring system for chronic wound diagnostics

Parameters

Type/ category

Specifications

Sensors

Temperature

At least 01 sensor, 0.5°C resolution, miniature and flexible

Moisture

At least 01 sensor, 5% RH resolution, miniature and flexible

Pressure

At least 01 sensor, 5 mmHg resolution, miniature and flexible

System size

NA

Small enough for placement within a compression bandage

Power consumption

NA

Must be kept as low as possible for long-term operation

Transmission

Protocol

Any suitable protocol e.g. Zigbee®, Bluetooth®, WiFi®

Range

Suitable to receive data in a clinical setup

Power

Within the safe exposure limits for human body

Frequency

2.4 GHz band to keep the components’ size smaller

Interval

Should be kept high to save battery power, e.g. 10 minutes

Ergonomics

NA

Must be flexible, bio-compatible, and non-invasive

Response time

NA

Less than or equal to a second