Skip to main content

Table 4 Comparison of the experimental and calculated clearance data for urea and creatinine

From: An experimental and numerical study of the flow and mass transfer in a model of the wearable artificial kidney dialyzer

  Urea Clearance (ml/min) Creatinine Clearance (ml/min)
Case # Exp. Num. Err (%) Exp. Num. Err (%)
21 37 40 -8 24 26 -8
22 31 34 -9 21 22 -5
23 30 30 0* 20 20 0*
24 24 24 0 16 16 0
25 13 12 8 9 8 11
31 44 48 -8 28 32 -13
32 43 37 14 28 25 11
35 40 34 15 28 22 21
41 25 29 -14 16 18 -11
44 32 35 -9 21 22 -5
47 39 40 -3 24 25 -4
50 52 51 2 31 30 3
53 28 32 -13 19 21 -10
56 35 33 6 22 20 9
59 40 41 -2 25 24 4
62 52 50 4 31 30 3
  1. *The mass transfer coefficients have been determined from this case