Table 2 Summary of the mean operating conditions of the experimental cases (P(mmHg), Q(ml/min), UF(ml/min), KUF(ml/hr/mmHg))

From: An experimental and numerical study of the flow and mass transfer in a model of the wearable artificial kidney dialyzer

Case P b,in P b,out P d,in P d,out Q b,in Q b,out Q d,in Q d,ou TMP UF K UF
Roller pump for blood, centrifugal pump for dialysate, with afterload
1 108 100 166 172 83 100 45 28 -65 -17 16.0
2 71 63 158 162 49 76 54 26 -93 -28 18.0
3 58 50 112 116 46 65 38 18 -60 -20 20.0
4 40 33 88 89 28 49 33 14 -52 -20 23.0
5 136. 127 167 173 100 118 45 26 -39. -18 28.0
6 162. 152 171 177 121 132 38 27 -17 -10 36.0
7 188 180 177 183 142 144 31 28 4 -3 -40.0
Roller pump for blood, centrifugal pump for dialysate, without afterload
15 62 57 17 9 82 71 42 53 47 11 14.0
16 27 23 8 2 40 39 43 44 20 2 4.5
17 17 14 -1 -9 28 26 31 33 21 2 4.9
18 13 10 -4 -12 20 19 24 26 20 2 5.3
21 62 55 22 20 82 72 50 59 38 10 15.0
22 44 37 9 9 65 58 40 47 32 7 13.0
23 32 27 3 2 52 45 34 40 27 6 14.0
24 22 19 -6 -8 42 36 20 26 28 6 14.0
25 11 8 -13 -14 23 19 9 13 23 4 10.0
WAK pump for both blood and dialysate, with afterload
31 39 33 -25 -32 77 51 42 65 64 24 23
32 24 21 -24 -31 50 34 38 56 50 17 20
33 19 16 -21 -29 37 25 33 44 42 12 16
34 6 5 -19 -25 10 4 30 35 27 5.5 12
35 22 18 -22 -30 43 29 36 50 46 14 18
WAK pump for both blood and dialysate, without afterload, porcine blood
41 31 30 13 10 37 39 40 38 19 -2 -6
44 33 34 10 5 46 48 60 58 26 -2 -5
47 43 43 12 7 57 56 56 57 34 1 2
50 51 45 7 -1 81 81 112 112 45 0 0
Roller pump for blood, centrifugal pump for dialysate, without afterload, porcine blood
53 34 32 24 20 44 46 51 49 11 -2 -13
56 36 35 26 24 51 54 67 64 11 -3 -15
59 39 36 27 25 60 62 79 77 12 -2 -12
62 55 48 34 27 82 84 112 110 21 -2 -7