Skip to main content

Table 3 Molybdenum/Rhodium and Rhodium/Rhodium regression distribution quantities (using the same format as in Table 1).

From: Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications

Mo/Rh 28 kVp

μ

σ

Δ

l

σl

R2

FatFL

0.503

0.002

0.005

5.343

0.011

0.999

 

0.490–0.516 (1.9%)

  

5.28–5.40 (0.1%)

  

FatKA

0.511

0.001

0.005

5.34

0.010

0.999

 

0.498–0.525

  

5.27–5.40

  

GlandFL

0.718

0.002

0.013

5.155

0.013

0.998

 

0.683–0.753 (1.0%)

  

4.98–5.32 (1.0%)

  

GlandKS

0.711

0.010

0.012

5.118

0.020

0.999

 

0.677–0.745

  

4.95–5.28

  

Mo/Rh 29 kVp

      

FatFL

0.495

0.002

0.004

5.47

0.011

0.999

 

0.482–0.507 (1.2%)

  

5.41–5.53 (0.20%)

  

FatKA

0.501

0.001

0.004

5.48

0.010

0.999

 

0.489–0.513

  

5.42–5.54

  

GlandFL

0.700

0.002

0.014

5.274

0.014

0.998

 

0.661–0.739 (1.2%)

  

5.08–5.46 (0.20%)

  

GlandKS

0.692

0.010

0.013

5.263

0.020

0.998

 

0.655–0.729

  

5.08–5.44

  

Rh/Rh 30 kVp

      

FatFL

0.442

0.001

0.002

5.648

0.010

0.999

 

0.435–0.449 (0.21%)

  

5.61–5.68 (1.1%)

  

FatKA

0.443

0.001

0.002

5.712

0.010

0.999

 

0.434–0.451

  

5.66–5.75

  

GlandFL

0.644

0.002

0.010

5.557

0.010

0.999

 

0.621–0.666 (2.4%)

  

5.45–5.66 (1.0%)

  

GlandKS

0.629

0.005

0.010

5.593

0.014

0.999

 

0.609–0.648

  

5.49–5.69

 Â