Table 1 Molybdenum/Molybdenum regression distribution quantities. Quantities are summaries of the field of view: μ is the mean effective x-ray attenuation coefficient, σ is the standard deviation of the μ distribution, Δ is the mean variation in μ derived from averaging the individual regression variation, l is the mean logarithmic intercept, σ l is the standard deviation of the l distribution, and R2 is the mean of the associated distribution. Two sided 95% mean confidence interval is cited under the respective μ and l quantities. Location is indicated by the subscripts of the left hand column entries. Parenthetical FL entries indicate the absolute value percent difference between respective location quantities.

From: Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications

Mo/Mo 26 kVp

μ

σ

Δ

l

σl

R2

FatFL

0.572

0.001

0.005

5.093

0.010

0.998

 

0.557–0.586 (1.9%)

  

5.03–5.15 (2.6%)

  

FatKS

0.583

0.001

0.005

4.964

0.010

0.998

 

0.568–0.599

  

4.89–5.03

  

GlandFL

0.833

0.002

0.015

4.893

0.001

0.998

 

0.784–0.882 (0.60%)

  

4.68–5.10 (2.6%)

  

GlandKS

0.838

0.010

0.012

4.767

0.020

0.998

 

0.799–0.877

  

4.60–4.93

  

Mo/Mo 27 kVp

μ

σ

Δ

l

σl

R2

FatFL

0.559

0.001

0.005

5.262

0.010

0.999

 

0.544–0.575 (2.1%)

  

5.19–5.32 (1.8%)

  

FatKS

0.571

0.001

0.005

5.165

0.010

0.999

 

0.554–0.586

  

5.09–5.16

  

GlandFL

0.805

0.003

0.020

5.043

0.010

0.998

 

0.750–0.860 (0.40%)

  

4.81–5.28 (1.9%)

  

GlandKS

0.808

0.005

0.014

4.950

0.020

0.999

 

0.763–0.854

  

4.75–5.14

 Â