Table 4 Quantification of electric field strength for needle electrode models – calculated Ett min , Ett max and E max parameters.

From: Analytical and numerical quantification and comparison of the local electric field in the tissue for different electrode configurations

Needle electrode configuration

U/d = 1.15 V/cm

U/d = 1300 V/cm

 

Target tissue

Ettmax(V/cm)

Target tissue

Ettmin(V/cm)

Entire tissue

Emax(V/cm)

Target tissue

Ettmax(V/cm)

Target tissue

Ettmin(V/cm)

Entire tissue

Emax(V/cm)

2 (Fig. 4a)

0.804

0.527

6.618

905.4

591.7

7450.3

4 (Fig. 4b)

0.824

0.779

5.829

928.7

876.9

6562.5

6 (Fig. 4c)

1.049

0.911

5.166

1180.9

1025.2

5816.6

6 (Fig. 4d)

0.822

0.778

5.794

925.4

875.9

6523.1

6 (Fig. 4e)

1.064

0.835

7.443

1197.9

940.1

8379.6

7 (Fig. 4f)

0.21

~0

5.17

236.4

1.038

5820.8

7 (Fig. 4g)

8.1

0.84

8.1

9119.2

945.7

9119.2

  1. The results for models shown in Figs. 4a – 4e were calculated for applied voltage U = 1 V ( d = 5 3  mm MathType@MTEF@5@5@+=feaafiart1ev1aaatCvAUfKttLearuWrP9MDH5MBPbIqV92AaeXatLxBI9gBaebbnrfifHhDYfgasaacH8akY=wiFfYdH8Gipec8Eeeu0xXdbba9frFj0=OqFfea0dXdd9vqai=hGuQ8kuc9pgc9s8qqaq=dirpe0xb9q8qiLsFr0=vr0=vr0dc8meaabaqaciaacaGaaeqabaqabeGadaaakeaacqWGKbazcqGH9aqpcqaI1aqndaGcaaqaaiabiodaZaWcbeaakiabbccaGiabb2gaTjabb2gaTbaa@349D@ ) and for models shown in Figs. 4f – 4g for applied voltage U = 0.575 V (d = l = 5 mm) giving U/d = 1.15 V/cm. Furthermore, by scaling the results we also calculated the parameters Ett min , Ett max and E max for U/d = 1300 V/cm.