Table 4 Lesion Volume with No Tissue Perfusion Values represent the total volume of tissue necroses calculated over the course of the simulated ablation using various cell damage thresholds (D) and isothermal temperatures (IT).

From: Thermal modeling of lesion growth with radiofrequency ablation devices

Source Voltage (Volts) D = 63% (mm3) D = 100% (mm3) IT = 42°C (mm3)1 IT = 47°C (mm3)1 IT = 60°C (mm3) IT = 90°C (mm3)
2.5 0 0 0 0 0 0
5.0 0 0 0 0 0 0
7.5 0 0 0 0 0 0
10.0 0 0 89 0 0 0
12.5 0 0 444 13 0 0
15.0 0 0 942 216 0 0
17.5 9 3 1649 526 6 0
20.0 121 67 2508 915 87 0
22.5 314 242 3549 1414 296 0
25.0 577 495 4860 2070 547 6
27.5 923 809 6452 2866 870 55
30.0 1388 1228 8386 3830 1243 202
  1. 1 – The 42°C and 47°C isothermal volumes were chosen specifically because they are frequently used to establish damage thresholds in hyperthermia and radiofrequency ablation, respectively.
\