Table 4 Percent Error between Constant Conductivity and Temperature Dependent Data without Perfusion

From: Finite Element Analysis of Hepatic Radiofrequency Ablation Probes using Temperature-Dependent Electrical Conductivity

Applied Volts (V) Maximum Temperature Achieved (°C) Change Electric Field (%) Change Current Density (%) Change Temperature (%) Change Heat Flux (%) Change Electrical Conductivity (%) Change SAR (%)
0.0 37 0.0 0.0 0.0 0.0 0.0 0.0
2.5 38 -0.1 1.4 0.0 0.4 1.5 1.4
5.0 41 -0.2 5.8 0.1 0.8 6.1 5.6
7.5 45 -0.5 13.5 0.3 1.7 14.0 12.9
10.0 51 -0.8 24.6 0.7 2.8 25.7 23.6
12.5 59 -1.3 40.0 1.4 4.3 41.8 38.2
15.0 69 -1.8 60.3 2.5 6.1 63.3 57.3
17.5 80 -2.5 87.1 3.9 8.2 91.9 82.5
20.0 94 -3.2 120.2 5.7 10.7 127.5 113.2
  1. Percent error for all cases is calculated as (<Temperature Dependent Parameter> – <Constant Parameter>) / <Constant Parameter>. All percentage changes refer to the maximum change in the parameter.