Table 3 Distribution of observations wherein sensors 1 and 2 displayed values above/below certain thresholds

From: Baseline pressure errors (BPEs) extensively influence intracranial pressure scores: results of a prospective observational study

   Percentage of combinations
PatID Number of 6-sec TS observations Sensor 1 ≥10 mmHg/Sensor 2 <10 mmHg or Sensor 1 <10 mmHg/Sensor 2 ≥10 mmHg Sensor 1 ≥15 mmHg/Sensor 2 <15 mmHg or Sensor 1 <15 mmHg/ Sensor 2 ≥15 mmHg Sensor 1 ≥20 mmHg/ Sensor 2 <20 mmHg or Sensor 1 <20 mmHg/ Sensor 2 ≥20 mmHg
1 77,900 31% 2% -
2 66,612 37% 19% 12%
3 133,387 3% 1% -
4 84,355 8% 1% -
5 41,708 4% 2% -
6 47,972 33% 3% -
7 127,305 9% 2% 1%
8 147,216 35% 12% 2%
9 170,327 20% 7% 2%
10 139,019 14% 2% -
11 58,886 3% 5% 2%
12 123,334 5% 3% -
13 61,780 8% 2% -
14 83,510 6% 1% -
15 218,383 10% 1% -
16 112,501 57% 15% 3%