Skip to main content

Table 4 Results of the patient measurements

From: Novel wireless electroencephalography system with a minimal preparation time for use in emergencies and prehospital care

Patient 1 2 3 4 5
Age 23 25 55 58 75
Gender Female Male Female Male Female
Recording location Clin.Depart. Clin.Depart. ICU ICU Clin.Depart.
General estimation (clinical EEG) Normal Normal Highly Abnormal Highly Abnormal Highly Abnormal
Status Epilepticus No No No No Yes
Experienced clinical neurophysiologist 1      
General estimation (emEEG) Normal Normal Normal Highly Abnormal Highly Abnormal
Status Epilepticus No No No No Yes
Experienced clinical neurophysiologist 2      
General estimation (emEEG) Normal Normal Normal Highly Abnormal Highly Abnormal
Status Epilepticus No No No No Yes
  1. The results from the patient measurements were evaluated by two experienced clinical neurophysiologists. Three grades were used in the general estimation, normal, abnormal and highly abnormal based on the clinical practice. The recording location at the Department of Clinical Neurophysiology is denoted as Clin.Depart. and the intensive care unit as ICU.