From: Prototype development of an electrical impedance based simultaneous respiratory and cardiac monitoring system for gated radiotherapy

Figure 5

Output traces from the bioimpedance-based respiratory monitoring unit developed and the RPM system for three healthy human volunteer. Each subject was instructed to exercise 25 seconds of normal breathing followed by 25 seconds of fast breathing followed by 25 seconds of slow breathing and finally 25 seconds of breath holding. Subject 1: female, 168 cm, 58 kg; Subject 2: male, 183 cm, 77 kg; Subject 3: male, 177 cm, 75 kg. Correlation coefficients: Subject 1 – 0.6623; Subject 2 – 0.9502; Subject 3 – 0.9599.