Table 4 Test data for AEDs

From: Adhoc electromagnetic compatibility testing of non-implantable medical devices and radio frequency identification

   Probably not clinically significant Device in operable Probably clinically significant All experiments
AED #1 (2008) Occurrences of EMI 3 0 0 3 (15%)
Max Distance of EMI (cm) 21 - - 21
RFID Reader Code causing EMI 15,17,18 N/A N/A 15,17,18
AED #2 (2002) Occurrences of EMI 1 3 2 3 (15%)
Max Distance of EMI (cm) 11 10 18 18
RFID Reader Code causing EMI 17 3,5,17 3,5 3,5,17
AED #3 (2009) Occurrences of EMI 2 4 4 8 (40%)
Max Distance of EMI (cm) 14 51 18 51
RFID Reader Code causing EMI 15,17 7,8,10,11 3,5,7,17 3,5,7,8,10,11,15,17
AED #4 (2004) Occurrences of EMI 2 1 1 4 (20%)
Max Distance of EMI (cm) 38 0 1 38
RFID Reader Code causing EMI 15,16 17 10 10,15-17
  1. The EMI Occurrences row represents the number of times EMI occurred during a test (n = 20). One experiment was defined as all testing performed on one medical device with one RFID system. Therefore the All Experiments column indicates whether any EMI occurred for a particular RFID system and does not indicate whether multiple effects were caused by the same RFID system. Production years for devices tested in parenthesis.