Skip to main content

Table 5 Differences in mean wave rise time (MWRT) between simultaneous ICP signals

From: Simultaneous monitoring of static and dynamic intracranial pressure parameters from two separate sensors in patients with cerebral bleeds: comparison of findings

PatID MWRT (sec; average + std) Percentage of 6-s time windows with difference in MWRT:
  Signal 1 Signal 2 > 0.1 s > 0.2 s
Category A
1 0.27 + 0.04 0.27 + 0.04 - -
2 0.21 + 0.02 0.21 + 0.02 - -
3 0.14 + 0.04 0.14 + 0.03 1 -
4 0.18 + 0.05 0.18 + 0.05 - -
5 0.28 + 0.06 0.28 + 0.05 - -
Category B
6 0.29 + 0.04 0.24 + 0.04 13 -
7 0.25 + 0.02 0.25 + 0.04 2 -
8 0.23 + 0.02 0.21 + 0.03 2 -
9 0.14 + 0.04 0.14 + 0.04 3 1
10 0.28 + 0.03 0.24 + 0.06 14 -
Category C
11 0.26 + 0.05 0.28 + 0.02 9 -
12 0.33 + 0.13 0.23 + 0.09 45 17
13 0.36 + 0.06 0.45 + 0.12 50 20
14 0.33 + 0.09 0.39 + 0.10 25 7
15 0.22 + 14.3 0.22 + 14.3 8 1
16 0.12 + 0.01 0.12 + 0.01 - -
17 0.24 + 0.05 0.25 + 0.03 7 -
  1. MWRT: mean wave rise time.