Table 5 AED4 test results

From: Radiated radiofrequency immunity testing of automated external defibrillators - modifications of applicable standards are needed

Device designation

Type of interference observed

Field strength for worst-case failure (V/m)

Modulation

(Hz)

Frequency range of failure (MHz)

Device's behavior (Observations from user standpoint)

Notes

AED4

None

20

1 Hz pulsed (square wave)

No failures

No verbal warnings heard during testing at any frequency. Device recommended shock as expected.

V-Fib applied - No interference or waveform distortion in ECG.