Figure 6From: A Parametric Study Delineating Irreversible Electroporation from Thermal Damage Based on a Minimally Invasive Intracranial Procedure Electric field, conductivity, temperature and thermal damage distributions at the conclusion of an 80 s IRE treatment simulation. A model of eighty 50 μs pulses with an applied voltage of 1000 V at a repetition rate of 1 Hz is presented. The grid resolution in the distributions is 1.0 mm. For this specific simulation, the tissue was treated only with IRE since thermal damage occurs when Ω > 0.53.Back to article page