Skip to main content

Table 3 Number of ECG time segments for which the standardized difference between Actiheart and Holter HRV estimates (in units of repeatability standard deviations, z) exceeds 1 and 2, respectively

From: Comparison of two systems for long-term heart rate variability monitoring in free-living conditions - a pilot study

HRV metric

s(repeatability)

Number of segments z>1

Percentage of totala

Number of segments z>2

Percentage of totala

ln(TP)

0.41

19

3.9%

4

0.8%

ln(LFP)

0.39

28

5.7%

10

2.0%

ln(HFP)

0.50

37

7.6%

12

2.5%

ln(LF/HF)

0.41

47

9.6%

12

2.5%

LFn.u.

0.078

39

8.0%

12

2.5%

HFn.u.

0.078

32

6.5%

8

1.6%

  1. a Total n = 489 ECG segments