Skip to main content

Table 7 Performance of FFRFAST in patients on a per-vessel level against invasive FFR (n = 402)

From: A novel physics-based model for fast computation of blood flow in coronary arteries

 

FFRFAST ≤ 0.80

FFRCFD ≤ 0.80

P value

True positive, no.

114

114

–

False positive, no.

23

20

–

True negative, no.

242

245

–

False negative, no.

23

23

–

Accuracy (%)

88.6 [84.9, 91.4]

89.3 [85.8, 92.1]

0.60

Sensitivity (%)

83.2 [75.7, 88.8]

83.2 [75.7, 88.8]

1.00

Specificity (%)

91.3 [87.1, 94.3]

92.5 [88.4, 95.2]

0.49

PPV (%)

83.2 [75.7, 88.8]

85.1 [77.6, 90.4]

0.50

NPV (%)

91.3 [87.1, 94.3]

91.4 [87.2, 94.4]