Table 2 Comparison of PO and vPPG triggers during MRA of a healthy volunteer

From: Initial evaluation of prospective cardiac triggering using photoplethysmography signals recorded with a video camera compared to pulse oximetry and electrocardiography at 7T MRI

Scan

Cardiac cycles

Missed cardiac cycles

Cardiac cycle durations (ms)

Trigger delay and jitter (ms)

PO PPG

vPPG

PO PPG

vPPG

vPPG

1.

73

0

0

931 ± 48

923 ± 42

401 ± 157

2.

74

0

0

975 ± 76

970 ± 79

312 ± 129

  1. Scan (1) is associated with the upper MIP and scan (2) with the lower MIP in Fig. 12. The trigger delay was measured as the mean difference in absolute trigger times between both methods and the trigger jitter as the sd