Modeling of the native knee with kinematic data derived from experiments using the VIVO™ joint simulator: a feasibility study
BioMedical Engineering OnLine volume 23, Article number: 85 (2024)
Abstract
Background
Despite advances in total knee arthroplasty, many patients are still unsatisfied with the functional outcome. Multibody simulations enable a more efficient exploration of independent variables compared to experimental studies. However, to what extent numerical models can fully reproduce knee joint kinematics is still unclear. Hence, models must be validated with different test scenarios before being applied to biomechanical questions.
Methods
In our feasibility study, we analyzed a human knee specimen on a six degree of freedom joint simulator, applying a passive flexion and different laxity tests with sequential states of ligament resection while recording the joint kinematics. Simultaneously, we generated a subject-specific multibody model of the native tibiofemoral joint considering ligaments and contact between articulating cartilage surfaces.
Results
Our experimental data on the sequential states of ligament resection aligned well with the literature. The model-based knee joint kinematics during passive flexion showed good agreement with the experiment, with root-mean-square errors of less than 1.61 mm for translations and 2.1° for knee joint rotations. During laxity tests, the experiment measured up to 8 mm of anteroposterior laxity, while the numerical model allowed less than 3 mm.
Conclusion
Although the multibody model showed good agreement to the experimental kinematics during passive flexion, the validation showed that ligament parameters used in this feasibility study are too stiff to replicate experimental laxity tests correctly. Hence, more precise subject-specific ligament parameters have to be identified in the future through model optimization.
Background
Despite tremendous advances in total knee arthroplasty (TKA), only 80%–85% of patients are satisfied with the postoperative outcome [1, 2]. Biomechanical studies are conducted to better understand knee joint kinematics before and after TKA [3,4,5]. Such studies are performed experimentally or numerically, while computer-based studies need experimental data to be validated [6,7,8].
On the other hand, studies on patients enable the investigation of the native knee joint and functionality of TKA, considering the soft tissue structures [9]. However, extensive parameter studies are not possible due to ethical aspects. In addition, investigating different motion sequences is technically challenging, and the measurement of kinematics on the patient is partially prone to errors [10]. Contrarily, experimental setups with human specimens allow for examining the knee joint at different complexity scales. For this purpose, mechanical knee rigs [11, 12] and industrial robots [13,14,15] have been used. Although these setups provide valuable insights into the knee joint dynamics, test rigs usually cannot apply defined loads and torques about all three spatial axes, and the usage of robots can be very complex and time-consuming. To overcome these limitations, the six degrees of freedom joint simulator VIVO™ (Advanced Mechanical Technology, Inc., Watertown, USA) is used to study the kinematics of the native human knee joint and the contribution of specific ligaments to overall joint stability [16, 17]. It can also be used for testing implant components with virtual ligaments represented by numerical strain-force laws [18, 19].
Based on the experimental data, multibody simulations are used to noninvasively analyze the knee joint kinematics and serve as a tool for virtual tests of different implant designs [20,21,22,23]. However, the quality of the simulation results highly depends on the boundary conditions used, and it is still unclear to what extent they fully reproduce the kinematics of the human knee joint [24]. Especially the parameters used to model the ligaments are described with a broad range in the literature and significantly contribute to joint kinematics [25].
Therefore, the aim of our present study is to generate a subject-specific multibody model of the right knee joint that mimics an experimental setup for investigating the human knee joint kinematics. For this setup, a knee specimen is experimentally examined on the VIVO™ joint simulator using different load cases, i.e., passive knee flexion and various laxity tests, to characterize knee joint kinematics. Additionally, tests are repeated with a sequential resection strategy of the knee ligaments. Based on medical images of the human specimen, a complex subject-specific multibody model comprising ligament structures and contact between articulating cartilage surfaces is generated to mimic the experimental tests. A more physiological simulation of the human knee joint kinematics will significantly contribute to several biomechanical questions, such as soft tissue management during TKA [26].
Results
Experimentally derived knee kinematics
Passive flexion
Figure 1 shows the kinematics over time during the passive knee flexion for the different resection states of the ligaments. Results were evaluated for the tibia coordinate system with respect to the femur coordinate system according to the Grood and Suntay convention [27].
Starting from initial values of 19–22 mm, the anterior translations for all resection states peaked approximately near the highest flexion value at 12.5 s with maximum values between 27 and 33 mm before returning to the original initial values with the shift back to extension. Resection of the patella slightly reduced anterior translation during the transition from flexion to extension, whereas the reverse was unaffected.
The varus/valgus curve showed a similar trend for all resection stages. Starting from a slight valgus position between − 2.4° and − 1.5°, the varus angle constantly decreased until midflexion (30–45° flexion) before it rose to its original value. While resection of the patella and lateral collateral ligament (LCL) decreased the varus/valgus rotation, removal of the medial collateral ligament (MCL) increased the rotation by up to 0.5°.
The course of external/internal rotation was similar for all resection states. It started at the same point at about 1° external rotation, then climbed to a maximum during midflexion before returning to its initial value. The rotation depends on whether it goes into or comes out of flexion. Patellar resection reduced external rotation by up to 3°. Resection of the anterior cruciate ligament (ACL) and especially the posterior cruciate ligament (PCL) further reduced the range of external/internal rotation.
Laxity tests
All three translational and rotational degrees of freedom during loading in anteroposterior, varus/valgus, and external/internal directions were recorded over 10 s and averaged to evaluate the laxity tests. The range of movement is shown via bars in Fig. 2. The upper end of the bars represents the position for 40 N posterior/5 Nm varus/2.5 Nm internal loading. A line highlights the position during passive flexion, while the lower end of the bars represents the position for 40 N anterior/5 Nm valgus/2.5 Nm external loading. Resection of the patella had no considerable influence on the knee laxity results. With resection of the ACL, laxity in the anterior direction increased by up to 5 mm for flexion angles of 30° and 60°. At 90° flexion, the load had to be reduced to 10 N due to the limitation of the working space of the joint simulator. These cases are marked with a triangle in Fig. 2. Laxity during anterior loading was not changed by the resection of the PCL and LCL at 30° and 60° flexion. At 90° flexion, the maximum load could not be applied during the working space's limitation. Laxity during posterior loading increased by approximately 5 mm after resecting the PCL, 5 mm after resecting the LCL, and 2.5 mm after resection of the MCL at a flexion angle of 30°. The varus/valgus rotation limits at 5 Nm load showed almost no influence on the patella, ACL, and PCL resection for 30° and 60° flexion angles. However, the resection of the LCL and MCL had a distinct influence on valgus/varus rotation. Some load cases could not be included due to the specimen's instability and are marked with a rhombus in Fig. 2. External and internal rotation limits were not reasonably changed by the patella, ACL, and PCL resection at a 30° flexion angle. For 60° and 90° flexion angles, resecting the PCL increased the internal laxity by about 5°. Resection of the LCL resulted in higher external rotation for all flexion angles, whereas resecting the MCL led to higher external and internal rotation at all flexion angles.
Numerically derived kinematics of passive flexion
The root mean square error (RMSE) was calculated for the kinematic parameters obtained from the multibody simulations of the passive flexion cycle to compare the numerical data with the experimental results. In the case of translational kinematics, the RMSE was 1.32 mm in the mediolateral direction, 1.61 mm in the anteroposterior direction, and 1.34 mm in the superoinferior direction. For the rotational kinematics, the RMSE was 0.84° for varus/valgus and 2.12° for external/internal rotation.
The data of anterior translation of the tibia with increasing flexion angles derived from the multibody modeling agreed with the experimental kinematics (Fig. 3). The varus/valgus rotation displayed the same progression with increasing flexion angle but slightly different start values and positions during 90° of flexion. For external and internal rotation, numerical and experimental kinematics showed the maximum external rotation at 60° flexion and an external rotation when reaching 90° extension and the end of the motion. The amplitude of the rotations was in the same range, whereby the numerical kinematics showed more deviations in the rotation angle.
A maximum contact force during passive flexion of 1000 N was calculated from the multibody model at a flexion angle of 90°. In general, numerical modeling resulted in lower laxity of the ligaments (Fig. 4). While the experiments showed a range of laxity between 6 and 8 mm in anteroposterior direction for different flexion angles when all ligaments were intact, the numerical model only allowed 2–3 mm of movement when applying the same loads. For varus/valgus laxity, the experiments showed a laxity of about 10°, while in the numerical model, the laxity was reduced to approximately 1.5° for all flexion angles. Laxity for external and internal rotation was also reduced in the numerical model from approximately 25° in the experiments to approximately 10° in the multibody simulation. The influence of sequential resection could not be further investigated in the numerical model, as the resection of individual ligaments led to significant instability.
Ligament forces calculated by the numerical model varied between 300 and 550 N during passive flexion for the ACL and between 0 and 420 N for the PCL. The progression of the forces in the cruciate ligaments is shown in Fig. 5. The force in the ACL decreased with flexion angle, while the PCL force increased with flexion angle. Results for the LCL and MCL forces are available as Additional file 1 in the Supplementary Information.
Discussion
This feasibility study aimed to investigate individual knee kinematics during different loading scenarios by generating a subject-specific multibody model that mimics the experimental data. The measured kinematics during passive flexion and different laxity measurements were in good agreement with the literature for both native and resected states. Nevertheless, it has to be noted that the direct comparison with other studies is only of limited value due to the different and often insufficiently documented positioning of the joint coordinate systems. In future work the coordinate systems should be defined as standardized as possible (e.g. using 'REFRAME' [28]) to ensure better comparability with studies of other groups.
While the kinematics of passive flexion of the experimental testing could be well reproduced with the multibody model, it showed a reduced range of laxity compared to the experimental tests. We conclude that the ligament parameters from the literature used in our study are too stiff to fully represent our experimental test results.
The test results considering tensile loading of the quadriceps femoris tendon showed increased anterior translation of the tibia during passive flexion. It has been demonstrated in a previous study that isolated muscle loading affects tibiofemoral kinematics [29]. Passive flexion induced an anterior displacement of the tibia, primarily caused by the tension of the PCL with increasing knee flexion [30]. This displacement could be seen in the experimentally recorded data and was comparable to data from other groups [25, 29, 31]. After PCL resection, the translation was less consistent with the literature [25, 32]. In contrast, removing the ACL increased anterior translation of the tibia, especially in early flexion.
The patella or the ligaments' resection did not considerably affect the varus/valgus rotation during passive flexion. There was a slight increase in valgus rotation after the resection of the LCL and a slight rise in varus rotation after the resection of the MCL, corresponding to the ligaments' physiological directions of action. The resections of the collateral ligaments resulted in increased valgus and varus rotation, as both the MCL [33] and the LCL [34] function as stabilisers for varus/valgus rotations in the human knee joint. This agrees with the literature, where the varus/valgus rotation does not depend on the flexion angle [35].
External and internal rotation showed the screw-home mechanism when reaching extension in experimental and numerical data, which is well documented in the literature [25, 31, 32, 36]. When full flexion of 90° was achieved, an internal rotation of approximately 4° was also evident in the experimental results. During the last 10° of extension, a final rotation of approximately 2° could be observed. Resection of the ACL and PCL reduced the changes in external/internal rotation angle during flexion due to missing wrapping of the cruciate ligaments. In general, rotation was more internally rotated after the resection of the ACL, which can be explained physiologically by the lack of wrapping of the cruciate ligaments around each other. Similar observations were documented by Markolf et al. [31]. Resection of the collateral ligaments led to a further reduction in tibial rotation during passive flexion, as the tibial torque was reduced since at least one of the ligaments is usually under tension throughout flexion movements [37].
Liu-Barba et al. [38] described a correlation between the applied compression force and the resulting translations and rotations. Different amounts of applied compression force complicate the comparability between the studies. Increased internal rotation during extension compared with flexion has been similarly documented by Markolf et al. [31]. Testing with the patella and loaded quadriceps femoris tendon also showed increased internal rotation, which may be explained by the individually acting lever arms of the muscle forces and cannot be evaluated with sufficient accuracy without considering a force effect in the hamstring muscles.
Considering the resection states, the results of the laxity tests agreed with the literature data [8]. An almost identical AP laxity during the tests with and without the patella supports the assumption that the differences in kinematics with and without the patella during passive flexion were due to the load applied to the tendon of the quadriceps femoris muscle. Nevertheless, further studies should be performed to verify this assumption. Resection of the ACL increased the possible translation in the anterior direction, and resection of the PCL showed a more posterior translation of the tibia as supported by Willinger et al. [39] and Kennedy et al. [40]. In addition, the possible internal rotation during laxity tests after resection of the ACL and PCL was higher, especially at large flexion angles, which is plausible since the wrapping of the cruciate ligaments no longer restricts internal rotation. Resection of the LCL essentially allowed more valgus rotation, while removal of the MCL increased varus rotation. In addition, more external rotation was possible after the resection of the LCL and more rotation in general after the resection of the MCL, which can explain the instability of the knee joint. The translation values during anterior and posterior loading of the intact preparation were lower than those of Moslemian et al. [36]. However, they agreed well with the data of Pedersen et al. [8]. On the other hand, the values of varus/valgus and external/internal rotations were comparable to the results of Moslemian et al. [36].
Nevertheless, the comparison is limited because the applied loads and the definition of the coordinate systems may differ in several studies. More precisely, the loads applied in this study were reduced by 50% compared to those documented in previous literature [8, 16, 17]. This modification was necessary as initial trials with increased loads exceeded the machine's working space after multiple resections due to rising laxity.
A multibody model of the tibiofemoral joint has been created to mimic the experimental kinematics during passive flexion and laxity tests. The RMSE was considerably low for all degrees of freedom. Hence, a good agreement was reached between experimental and numerical kinematics. The range of motion of the tibial anteroposterior translation in the simulation was slightly lower compared to the experiment. This could be due to stiffer cruciate ligaments in the multibody model restricting the range of motion.
The varus/valgus curve also showed a similar curve in the multibody simulation. However, it started with about 1.5° more valgus rotation. Contrary to the experimentally measured data, whether the flexion angle was approached from flexion or extension was also not essential in the multibody model, as the values do not differ. This is a general difference in comparing the kinematics between the experiment and the multibody model, which becomes even more apparent when considering the external/internal rotation. While the numerical curves show a symmetry axis in the middle of the load cycle (at 12.5 s), the experiment's results differ between the extension and flexion phases. This may be due to the neglected time dependency of the ligament properties. The reduced range of motion in rotation in comparison to the literature was also reflected by the model. The resection of a single ligament led to a considerable imbalance in the simulation, which the remaining ligaments could not compensate for.
While the same progression of change in the ACL and PCL force over passive flexion was shown in other models [25, 41,42,43] and experimental studies [43], the value of the forces in the simulation is considerably higher. The decreased range of movement in simulated laxity tests indicates that ligament forces in the model are much higher than in the experiment. Too high ligament stiffness is probably the reason for model instabilities in the simulation of the individual resection stages. This points out the necessity for further optimization of ligament modeling.
Some limitations of our feasibility study have to be taken into consideration. First, only one human specimen was tested, which does not provide a statistical indication of the reproducibility of the tests. This specimen was preconditioned before the start and rewetted at regular intervals between the experiments to ensure the same test conditions before each test. Nevertheless, it can not be excluded that the physiology changed over the testing period of eight hours. The friction properties of the contact surfaces on the preparation may also have changed over time. Physiological lubrication conditions were assumed in the tests on the intact specimen. After patellar resection, the contact surfaces were continuously wetted with sodium chloride to prevent them from drying out. However, this may also have an influence on the friction properties of the preparation due to different properties compared to human synovial fluid [44]. In the experimental setup, when testing the intact joint, only the quadriceps muscles were loaded over the patella and only with a constant force, while the posterior muscles were neglected. When performing the tests, only the joint reaction forces were controlled. However, we do not have any information about the actual contact forces acting in the joint, which would have been beneficial for the subsequent validation of the multibody model.
Regarding the multibody model, limitations were mainly associated with ligament modeling and tibiofemoral contact. While being three-dimensional organic structures, ligaments were modeled as one-dimensional force spring elements. However, according to the literature [37], this assumption is legitimate to simulate the ligaments accurately. The multibody model does not use time-dependent force elements for the ligaments, whereas in-vivo, this property can be attributed to the viscoelasticity of the ligaments and the cartilage [45]. As this may be a reason for differences in the symmetry of the kinematic, the influence of time dependence when modeling the ligaments should be investigated in the future. Wrapping the ligament around bones or other soft tissues was omitted. In future work, the virtual ligament model of the joint simulator shall be used, which does not provide such wrapping around obstacles [46,47,48]. In addition, previous studies [37, 49] showed that the absence of wrapping is essential for non-physiological loading and movement scenarios, which were not investigated in this work. Moreover, identifying suitable ligament parameters is still challenging as the reported parameters in the literature are inconsistent [37]. The results of the numerical laxity tests indicate that the entire ligament apparatus is slightly too stiff to reproduce the experimental investigations more accurately. The computation of ligament force as Wismans' method, which does not consider viscoelasticity, is also predefined by the virtual ligament modeling used in the joint simulator. This factor could explain the observed asymmetry in the experimental studies, which was not replicated numerically. The current model is also limited by the contact implementation, where the menisci are neglected, and viscoelasticity is missing at the contact element between the femoral and tibial cartilage. In addition, fluid effects, which have an influence on contact mechanics in numerical studies, are not considered by the contact element used [50, 51]. Furthermore, the parametrization of contact modeling is always a challenge in biomechanical simulations [37].
Conclusions
In our feasibility study, a human fresh-frozen specimen was successfully investigated experimentally on a complex six-degree-of-freedom joint simulator with passive flexion and different laxity tests under sequential ligament resections. A subject-specific multibody model of the tibiofemoral joint was generated and showed good agreement with the experimental kinematics during passive flexion. Differences in the laxity tests between the multibody model and the experimental investigations suggest that ligament parameterization is still challenging and needs further improvement. Even with a good kinematic agreement between experiment and simulation for the passive flexion, it is necessary to validate multibody models with different load cases that especially consider the stiffness of specific ligaments, such as laxity tests. Currently, a multibody optimization algorithm is developed to determine subject-specific ligament parameters using the experimental results obtained in this work. By coupling the multibody model with an optimisation algorithm, the ligament parameters will be iterated until the error between the experimentally determined kinematics and the calculated model kinematics is minimized [14]. These ligament parameters can be used in future experimental studies using the joint simulator and in computational investigations of knee joint dynamics under different loading scenarios.
Methods
Experimental testing
Ethical approval was obtained from the Institutional Review Board (A 2023-0055) for our experimental testing. Before preceding the experiments on the VIVO™ joint simulator, a fresh-frozen human specimen (male, 76 years, 56.7 kg) of the right knee was analyzed by computed tomography (SOMATOM Perspective, Siemens Healthineers, Erlangen, Germany). Based on these data, bone structures of the fibula, tibia, femur as well as the femoral cartilage structures were segmented and reconstructed with the software Amira 5.4.1 (Thermo Fisher Scientific, Waltham, Massachusetts, USA and Zuse Institute Berlin, Berlin, Germany). The tibial cartilage was measured and digitised with an Artec Space Spider 3D scanner (Artec 3D, Senningerberg, Luxembourg), as it was not possible to clearly distinguish between tibial cartilage and menisci during the segmentation process using the CT scan of the frozen specimen. All cartilage surfaces were processed in Geomagic Studio 2013 (3D Systems, Rock Hill, South Carolina, USA). Joint coordinate systems were defined according to the literature [52,53,54]. Relative motion between the femur and tibia was defined in Grood and Suntay coordinates [27]. In Fig. 6, a brief overview of the convention of Grood and Suntay is depicted based on [27].
Two actuators realize the movement of the joint simulator. The upper actuator consists of an abduction arm coupled with a flexion arm, thus enabling two rotations (flexion/extension, varus/valgus-rotation). The lower actuator, the xyz table, can perform three translations and one rotation around the vertical axis (internal/external). During the test, a kinematic loop equation maps the movements of the machine's hydraulic actuators to the Grood and Suntay coordinates in real time [55]. The knee was aligned in the joint simulator to match the previously defined femoral and tibial joint coordinate systems. Thus, the knee joint flexion axis corresponded to the joint simulator's flexion axis. For fixation, the femoral and tibial bones were resected proximally and distally and embedded in fast-curing resin GP 010 (Gößl und Pfaff, Brautlach, Germany) to ensure the correct positioning (Fig. 7). VIVOTMs degrees of freedom, as well as the experimental setup, are depicted in Fig. 7.
Regarding the experimental protocol, the human specimen was thawed for twelve hours before testing. Afterward, the specimen was moved with 20 cycles of passive flexion to precondition the soft tissues [55]. Different resection states were performed, and simultaneous knee joint kinematics were recorded during the passive flexion and laxity tests. Before each recording, the specimen was moved with at least two cycles of passive flexion for preconditioning purposes. To simulate passive flexion, the flexion axis of the simulator was position-controlled with a prescribed flexion angle as a sine wave with a period 25 s for a cycle between full extension (0°) and flexion (90°). The two other rotational degrees of freedom, the varus/valgus rotation and the tibial internal/external rotation, were torque-controlled to a constant value of 0 Nm.
In comparison, the three translation axes were force-controlled to 0 N for the two horizontal axes and to 50 N for the vertical axis representing a compressive load. The same compressive load was applied for the laxity tests, while flexion angles of 30°, 60° and 90° were prescribed. Precisely, loads in anterior/posterior, varus/valgus, and external/internal directions were applied for 10 s, according to Table 1. At the same time, forces and torques of all other degrees of freedom were force- and torque-controlled to 0 N and 0 Nm, respectively. The data were generated with 100 Hz frequency and smoothed using a four-pole Butterworth filter. The applied loads were adapted from the literature [8, 16, 17].
For all tests of the knee specimen before the resection states, the patella was loaded via the quadriceps tendon with a constant tensile force of 20 N to stabilize the patella during testing [56]. The first resection state included resecting the patella and, thus, the quadriceps load. After repeated testing, including passive flexion and laxity tests according to Table 1, further resection states include the anterior cruciate ligament (ACL), the posterior cruciate ligament (PCL), the lateral collateral ligament (LCL) and the medial collateral ligament (MCL) (Fig. 8).
Multibody model
The multibody model of the tibiofemoral joint was generated in the software Simpack (v2022, Dassault Systèmes, Vélizy-Villacoublay, France) to mimic the previously described experimental setup. The reconstructions of the CT dataset were used to set up a multibody topology with the femur and tibia. The multibody simulation did not include the patellofemoral joint, as the additional contact considerably increases the complexity. The femoral and tibial coordinate systems were implemented according to the experiment.
Regarding the multibody topology, the femur was fixed in the world coordinate system. A five degrees of freedom joint with contact between femoral and tibial cartilage was defined to connect the tibia to the femur so that the flexion could be specified at any time. The five degrees of freedom were set according to their state of equilibrium depending on internal (contact) and external (ligaments) forces. Ligaments were modeled as non-linear spring force elements according to Wismans et al. [57], and Blankevoort et al. [58], with the exerting force defined by
where \(\varepsilon\) is the strain, \(k\) the stiffness and \({\varepsilon }_{l}\) a constant set to 0.03 [59]. The strain can be calculated with
where l is the actual length of the ligament and \(l_{0}\) is the slack length of the ligament. Accordingly, the slack length can be calculated from the reference length \(l_{r}\) and the reference strain \(\varepsilon_{r} ,\)
Regarding the multibody model, the ACL and PCL, the LCL, and the MCL are divided into a superficial and deep medial collateral ligament (sMCL, dMCL) and the oblique popliteal ligament (OPL) consisting of two bundles each (anterior and posterior bundle). Additionally, one bundle each was added for the posterior oblique ligament (POL), the arcuate popliteal ligament (APL), and medial and lateral posterior capsular structures (mCAP, lCAP). Ligament insertion points were identified using the CT data, literature [60,61,62,63] and bony landmarks and were approved by an experienced board-certified orthopedic surgeon. The mechanical parameters of the ligaments used in the simulation are based on a previous study [20] and are summarized in Table 2.
To consider the interaction of the articulating surfaces, a contact force element between femoral and tibial cartilage was defined using an elastic foundation model according to Hippmann et al. [64]. The compressive force of 50 N in the experiment and the loads required for the laxity tests were applied via external force elements in the multibody model. Gravity was neglected to mimic the experimental setup. Figure 9 shows the multibody model of the right knee with the prescribed flexion angle for passive flexion.
Availability of data and materials
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Abbreviations
- \(l\) :
-
Actual ligament length
- \({l}_{0}\) :
-
Slack length of the ligament
- \({l}_{r}\) :
-
Reference length of the ligament
- \(k\) :
-
Ligament stiffness
- \(\varepsilon\) :
-
Actual strain of the ligament
- \({\varepsilon }_{l}\) :
-
Transition strain—transition from quadratic region to linear region for 2εl
- \({\varepsilon }_{r}\) :
-
Reference strain
- TKA:
-
Total knee arthroplasty
- ACL:
-
Anterior cruciate ligament
- PCL:
-
Posterior cruciate ligament
- LCL:
-
Lateral collateral ligament
- MCL:
-
Medial collateral ligament
- sMCL:
-
Superficial medial collateral ligament
- dMCL:
-
Deep medial collateral ligament
- POL:
-
Posterior oblique ligament
- APL:
-
Arcuate popliteal ligament
- mCAP:
-
Medial posterior capsular structures
- lCAP:
-
Lateral posterior capsular structures
- RMSE:
-
Root mean square error
References
Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468(1):57–63. https://doi.org/10.1007/s11999-009-1119-9.
Gibon E, Amanatullah DF, Loi F, Pajarinen J, Nabeshima A, Yao Z, Hamadouche M, Goodman SB. The biological response to orthopaedic implants for joint replacement: part i: metals. J Biomed Mater Res B Appl Biomater. 2017;105(7):2162–73. https://doi.org/10.1002/jbm.b.33734.
Sharma GB, Kuntze G, Kukulski D, Ronsky JL. Validating dual fluoroscopy system capabilities for determining in-vivo knee joint soft tissue deformation: a strategy for registration error management. J Biomech. 2015;48(10):2181–5. https://doi.org/10.1016/j.jbiomech.2015.04.045.
Lee TQ. Biomechanics of hyperflexion and kneeling before and after total knee arthroplasty. Clin Orthop Surg. 2014;6(2):117–26. https://doi.org/10.4055/cios.2014.6.2.117.
Koh IJ, Lin CC, Patel NA, Chalmers CE, Maniglio M, Han SB, McGarry MH, Lee TQ. Kinematically aligned total knee arthroplasty reproduces more native rollback and laxity than mechanically aligned total knee arthroplasty: a matched pair cadaveric study. Orthop Traumatol Surg Res. 2019;105(4):605–11. https://doi.org/10.1016/j.otsr.2019.03.011.
Uzuner S, Kuntze G, Li LP, Ronsky JL, Kucuk S. Creep behavior of human knee joint determined with high-speed biplanar video-radiography and finite element simulation. J Mech Behav Biomed Mater. 2022;125: 104905. https://doi.org/10.1016/j.jmbbm.2021.104905.
Blankevoort L, Huiskes R. Validation of a three-dimensional model of the knee. J Biomech. 1996;29(7):955–61. https://doi.org/10.1016/0021-9290(95)00149-2.
Pedersen D, Vanheule V, Wirix-Speetjens R, Taylan O, Delport HP, Scheys L, Andersen MS. A novel non-invasive method for measuring knee joint laxity in four dof: In vitro proof-of-concept and validation. J Biomech. 2019;82:62–9. https://doi.org/10.1016/j.jbiomech.2018.10.016.
Ewing JA, Kaufman MK, Hutter EE, Granger JF, Beal MD, Piazza SJ, Siston RA. Estimating patient-specific soft-tissue properties in a TKA knee. J Orthop Res. 2016;34(3):435–43. https://doi.org/10.1002/jor.23032.
Andersen MS, Dzialo CM, Marra MA, Pedersen D. A methodology to evaluate the effects of kinematic measurement uncertainties on knee ligament properties estimated from laxity measurements. J Biomech Eng. 2021;143(6):061003. https://doi.org/10.1115/1.4050027
Bloemker KH, Guess TM, Maletsky L, Dodd K. Computational knee ligament modeling using experimentally determined zero-load lengths. Open Biomed Eng J. 2012;6:33–41. https://doi.org/10.2174/1874230001206010033.
Steinbrück A, Schröder C, Woiczinski M, Fottner A, Müller PE, Jansson V. Patellofemoral contact patterns before and after total knee arthroplasty: an in vitro measurement. Biomed Eng Online. 2013;12:58. https://doi.org/10.1186/1475-925X-12-58.
Vap AR, Schon JM, Moatshe G, Cruz RS, Brady AW, Dornan GJ, Turnbull TL, Laprade RF. The role of the peripheral passive rotation stabilizers of the knee with intact collateral and cruciate ligaments: a biomechanical study. Orthop J Sports Med. 2017;5(5):2325967117708190. https://doi.org/10.1177/2325967117708190.
Winter E, Schröder I, Bader R, Woernle C. Influence of ligament modelling on knee joint kinematics with respect to multibody optimisation. In the 5th joint international conference on multibody system dynamics. 2018.
Kebbach M, Grawe R, Geier A, Winter E, Bergschmidt P, Kluess D, D’Lima D, Woernle C, Bader R. Effect of surgical parameters on the biomechanical behaviour of bicondylar total knee endoprostheses—a robot-assisted test method based on a musculoskeletal model. Sci Rep. 2019;9(1):14504. https://doi.org/10.1038/s41598-019-50399-3.
Sarpong NO, Sonnenfeld JJ, LiArno S, Rajaravivarma R, Donde S, Sneddon E, Kaverina T, Cooper HJ, Shah RP, Geller JA. Virtual reconstruction of the posterior cruciate ligament for mechanical testing of total knee arthroplasty implants. Knee. 2020;27(1):151–6. https://doi.org/10.1016/j.knee.2019.10.023.
Moslemian A, Arakgi ME, Roessler PP, Sidhu RS, Degen RM, Willing R, Getgood AMJ. The Medial structures of the knee have a significant contribution to posteromedial rotational laxity control in the PCL-deficient knee. Knee Surg Sports Traumatol Arthrosc. 2021;29(12):4172–81. https://doi.org/10.1007/s00167-021-06483-1.
Vakili S, Lanting B, Getgood A, Willing R. Development of multibundle virtual ligaments to simulate knee mechanics after total knee arthroplasty. J Biomech Eng. 2023;145(9):091003. https://doi.org/10.1115/1.4062421
Sekeitto AR, McGale JG, Montgomery LA, Vasarhelyi EM, Willing R, Lanting BA. Posterior-stabilized total knee arthroplasty kinematics and joint laxity: a hybrid biomechanical study. Arthroplasty (London, England). 2022;4(1):53. https://doi.org/10.1186/s42836-022-00153-4.
Kebbach M, Darowski M, Krueger S, Schilling C, Grupp TM, Bader R, Geier A. Musculoskeletal multibody simulation analysis on the impact of patellar component design and positioning on joint dynamics after unconstrained total knee arthroplasty. Materials. 2020;13(10):2365. https://doi.org/10.3390/ma13102365.
Marra MA, Vanheule V, Fluit R, Koopman BH, Rasmussen J, Verdonschot N, Andersen MS. A subject-specific musculoskeletal modeling framework to predict in vivo mechanics of total knee arthroplasty. J Biomech Eng. 2015;137(2):020904. https://doi.org/10.1115/1.4029258
Kebbach M, Geier A, Darowski M, Krueger S, Schilling C, Grupp TM, Bader R. Computer-based analysis of different component positions and insert thicknesses on tibio-femoral and patello-femoral joint dynamics after cruciate-retaining total knee replacement. Knee. 2023;40:152–65. https://doi.org/10.1016/j.knee.2022.11.010.
Tzanetis P, Fluit R, de Souza K, Robertson S, Koopman B, Verdonschot N. Pre-planning the surgical target for optimal implant positioning in robotic-assisted total knee arthroplasty. Bioengineering. 2023;10(5):543. https://doi.org/10.3390/bioengineering10050543
Wang L, Wang CJ, Berryman F. Modeling of the knee joint. In Computational modelling of biomechanics and biotribology in the musculoskeletal system. Woodhead Publishing; 2021. p. 437–72.
Kia M, Schafer K, Lipman J, Cross M, Mayman D, Pearle A, Wickiewicz T, Imhauser C. A multibody knee model corroborates subject-specific experimental measurements of low ligament forces and kinematic coupling during passive flexion. J Biomech Eng. 2016;138(5):051010. https://doi.org/10.1115/1.4032850
Johnston H, Abdelgaied A, Pandit H, Fisher J, Jennings LM. The effect of surgical alignment and soft tissue conditions on the kinematics and wear of a fixed bearing total knee replacement. J Mech Behav Biomed Mater. 2019;100: 103386. https://doi.org/10.1016/j.jmbbm.2019.103386.
Grood ES, Suntay WJ. A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng. 1983;105(2):136–44. https://doi.org/10.1115/1.3138397.
Vasquez AO, Taylor WR, Postolka B, Schütz P, Maas A, Woiczinski M, Sauer A. A reproducible and robust representation of tibiofemoral kinematics of the healthy knee joint during stair descent using REFRAME—Part I: REFRAME foundations and validation. 2024.
Victor J, Labey L, Wong P, Innocenti B, Bellemans J. The influence of muscle load on tibiofemoral knee kinematics. J Orthop Res. 2010;28(4):419–28. https://doi.org/10.1002/jor.21019.
Aumüller G, Aust G, Engele J, Maio G, Kirsch J. Anatomie. Stuttgart: Thieme; 2020.
Markolf KL, Jackson SR, Foster B, McAllister DR. ACL forces and knee kinematics produced by axial tibial compression during a passive flexion-extension cycle. J Orthop Res. 2014;32(1):89–95. https://doi.org/10.1002/jor.22476.
Akbari Shandiz M, Boulos P, Saevarsson SK, Yoo S, Miller S, Anglin C. Changes in knee kinematics following total knee arthroplasty. Proc Inst Mech Eng Part H J Eng Med. 2016;230(4):265–78. https://doi.org/10.1177/0954411916632491
Matsumoto H, Suda Y, Otani T, Niki Y, Seedhom BB, Fujikawa K. Roles of the anterior cruciate ligament and the medial collateral ligament in preventing valgus instability. J Orthop Sci. 2001;6(1):28–32. https://doi.org/10.1007/s007760170021.
Yaras RJ, O'Neill N, Yaish AM. Lateral collateral ligament knee injury. In: StatPearls [Internet] 2022 May 20. StatPearls Publishing.
Wilson D, Feikes J, Zavatsky A, O’Connor J. The components of passive knee movement are coupled to flexion angle. J Biomech. 2000;33(4):465–73. https://doi.org/10.1016/S0021-9290(99)00206-7.
Moslemian A, Sidhu R, Roessler P, Wood R, Degen R, Getgood A, Willing R. Influence of the posterior cruciate ligament on kinematics of the knee during experimentally simulated clinical tests and activities of daily living. J Biomech. 2021;115:110133. https://doi.org/10.1016/j.jbiomech.2020.110133.
Farshidfar SS, Cadman J, Deng D, Appleyard R, Dabirrahmani D. The effect of modelling parameters in the development and validation of knee joint models on ligament mechanics: a systematic review. PLoS ONE. 2022;17(1): e0262684. https://doi.org/10.1371/journal.pone.0262684.
Liu-Barba D, Hull ML, Howell SM. Coupled motions under compressive load in intact and ACL-deficient knees: a cadaveric study. J Biomech Eng. 2007;129(6):818–24. https://doi.org/10.1115/1.2800762.
Willinger L, Athwal KK, Williams A, Amis AA. An anterior cruciate ligament in vitro rupture model based on clinical imaging. Am J Sports Med. 2021;49(9):2387–95. https://doi.org/10.1177/03635465211017145.
Kennedy NI, Wijdicks CA, Goldsmith MT, Michalski MP, Devitt BM, Årøen A, Engebretsen L, Laprade RF. Kinematic analysis of the posterior cruciate ligament, part 1: the individual and collective function of the anterolateral and posteromedial bundles. Am J Sports Med. 2013;41(12):2828–38. https://doi.org/10.1177/0363546513504287.
Ali AA, Harris MD, Shalhoub S, Maletsky LP, Rullkoetter PJ, Shelburne KB. Combined measurement and modeling of specimen-specific knee mechanics for healthy and ACL-deficient conditions. J Biomech. 2017;57:117–24. https://doi.org/10.1016/j.jbiomech.2017.04.008.
Shelburne KB, Pandy MG, Anderson FC, Torry MR. Pattern of anterior cruciate ligament force in normal walking. J Biomech. 2004;37(6):797–805. https://doi.org/10.1016/j.jbiomech.2003.10.010.
Hosseini Nasab SH, List R, Oberhofer K, Fucentese SF, Snedeker JG, Taylor WR. Loading patterns of the posterior cruciate ligament in the healthy knee: a systematic review. PloS One. 2016;11(11):e0167106. https://doi.org/10.1371/journal.pone.0167106
Bortel E, Charbonnier B, Heuberger R. Development of a synthetic synovial fluid for tribological testing. Lubricants. 2015;3(4):664–86. https://doi.org/10.3390/lubricants3040664.
Robi K, Jakob N, Matevz K, Matjaz V. The physiology of sports injuries and repair processes. Curr Issues Sports Exerc Med. 2013.
Charlton IW, Johnson GR. Application of spherical and cylindrical wrapping algorithms in a musculoskeletal model of the upper limb. J Biomech. 2001;34(9):1209–16. https://doi.org/10.1016/S0021-9290(01)00074-4.
Zarifi O, Stavness I. Muscle wrapping on arbitrary meshes with the heat method. Comput Methods Biomech Biomed Engin. 2017;20(2):119–29. https://doi.org/10.1080/10255842.2016.1205043.
Lloyd JE, Roewer-Despres F, Stavness I. Muscle path wrapping on arbitrary surfaces. IEEE Trans Biomed Eng. 2021;68(2):628–38. https://doi.org/10.1109/TBME.2020.3009922.
Montgomery L. Biomechanical analysis of ligament modelling techniques and femoral component malrotation following TKA. The University of Western Ontario. https://ir.lib.uwo.ca/etd/8331
Uzuner S, Li LP. Alteration in ACL loading after total and partial medial meniscectomy. BMC Musculoskelet Disord. 2024;25(1):94. https://doi.org/10.1186/s12891-024-07201-x.
Uzuner S. Numerical analysis of a poroelastic cartilage model: Investigating the influence of changing material properties in osteoarthritis. Proc Inst Mech Eng Part E J Proc Mech Eng. 2024. https://doi.org/10.1177/09544089241248147.
Yin L, Chen K, Guo L, Cheng L, Wang F, Yang L. Identifying the functional flexion-extension axis of the knee: an in-vivo kinematics study. PLoS ONE. 2015;10(6): e0128877. https://doi.org/10.1371/journal.pone.0128877.
Victor J, Van Doninck D, Labey L, Van Glabbeek F, Parizel P, Bellemans J. A common reference frame for describing rotation of the distal femur: a ct-based kinematic study using cadavers. J Bone Jt Surg Br Vol. 2009;91(5):683–90. https://doi.org/10.1302/0301-620X.91B5.21827
Kahn TL, Soheili A, Schwarzkopf R. Outcomes of total knee arthroplasty in relation to preoperative patient-reported and radiographic measures: data from the osteoarthritis initiative. Geriatr Orthop Surg Rehabil. 2013;4(4):117–26. https://doi.org/10.1177/2151458514520634.
Henke P, Ruehrmund L, Bader R, Kebbach M. Exploration of the Advanced VIVOTM Joint Simulator: An In-Depth Analysis of Opportunities and Limitations Demonstrated by the Artificial Knee Joint. Bioengineering (Basel). 2024;11(2):178. https://doi.org/10.3390/bioengineering11020178.
Naghibi H, Mazzoli V, Gijsbertse K, Hannink G, Sprengers A, Janssen D, van den Boogaard T, Verdonschot N. A noninvasive MRI based approach to estimate the mechanical properties of human knee ligaments. J Mech Behav Biomed Mater. 2019;93:43–51. https://doi.org/10.1016/j.jmbbm.2019.01.022.
Wismans J, Veldpaus F, Janssen J, Huson A, Struben P. A three-dimensional mathematical model of the knee-joint. J Biomech. 1980;13(8):677–85. https://doi.org/10.1016/0021-9290(80)90354-1.
Blankevoort L, Kuiper JH, Huiskes R, Grootenboer HJ. Articular contact in a three-dimensional model of the knee. J Biomech. 1991;24(11):1019–31. https://doi.org/10.1016/0021-9290(91)90019-J.
Butler DL, Kay MD, Stouffer DC. Comparison of material properties in fascicle-bone units from human patellar tendon and knee ligaments. J Biomech. 1986;19(6):425–32. https://doi.org/10.1016/0021-9290(86)90019-9.
Tillmann B. Atlas der Anatomie des Menschen: Mit Muskeltabellen. Berlin: Springer; 2017.
Laprade RF, Ly TV, Wentorf FA, Engebretsen L. The posterolateral attachments of the knee: a qualitative and quantitative morphologic analysis of the fibular collateral ligament, popliteus tendon, popliteofibular ligament, and lateral gastrocnemius tendon. Am J Sports Med. 2003;31(6):854–60. https://doi.org/10.1177/03635465030310062101.
LaPrade RF, Morgan PM, Wentorf FA, Johansen S, Engebretsen L. The anatomy of the posterior aspect of the knee: an anatomic study. Jbjs. 2007;89(4):758–64. https://doi.org/10.2106/JBJS.F.00120
Athwal KK, Willinger L, Shinohara S, Ball S, Williams A, Amis AA. The bone attachments of the medial collateral and posterior oblique ligaments are defined anatomically and radiographically. Knee Surg Sports Traumatol Arthrosc. 2020;28(12):3709–19. https://doi.org/10.1007/s00167-020-06139-6.
Hippmann G. An algorithm for compliant contact between complexly shaped bodies. Multibody SysDyn. 2004;12(4):345–62. https://doi.org/10.1007/s11044-004-2513-4.
Acknowledgements
We thank Karl Lippert for providing the 3D scanner and helping scan the cartilage surface of the human specimen.
Funding
Open Access funding enabled and organized by Projekt DEAL. This work was supported by the German Research Foundation DFG (grant INST 2268/17-1 FUGG) and carried out as part of a project supported by Aesculap AG, Research and Development.
Author information
Authors and Affiliations
Contributions
Conceptualization, PH, JM, LR, SB, SK, TMG, CL, CW, RB and MK; methodology, PH, JM, LR, SB, SK, RB and MK; computer simulation, JM; data curation, JM; writing—original draft preparation, PH, JM; writing—review and editing PH, JM, LR, SB, SK, TMG, CL, CW, RB and MK; supervision, TMG, RB and MK; visualization, PH, JM, LR; project administration, RB and MK; funding acquisition, RB and MK; All authors have read and agreed to the published version of the manuscript.
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
Ethical approval was granted by the Ethics Committee of the University Medical Center Rostock according to A 2023-0055.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Additional file 1.
Predicted LCL and MCL forces of the multibody model.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Henke, P., Meier, J., Ruehrmund, L. et al. Modeling of the native knee with kinematic data derived from experiments using the VIVO™ joint simulator: a feasibility study. BioMed Eng OnLine 23, 85 (2024). https://doi.org/10.1186/s12938-024-01279-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12938-024-01279-z