Skip to main content

Table 6 Default targets for the SOLVe algorithm as used during the in vivo experiments

From: SOLVe: a closed-loop system focused on protective mechanical ventilation

Target

 

Unit

Target low

Target high

SpO\(_2\)

 

%

91

93

PETCO\(_2\)

 

mmHg

30

60

Protective target region

V\(_\text {T,pbw}\)

ml/kg

4

6

\(\Delta P\)

mbar

5

14

Protective acceptable region

V\(_\text {T,pbw}\)

ml/kg

6

8

\(\Delta P\)

mbar

5

14

Protective compromise region

V\(_\text {T,pbw}\)

ml/kg

4

6

\(\Delta P\)

mbar

14

20