From: SOLVe: a closed-loop system focused on protective mechanical ventilation
Target | Unit | Target low | Target high | |
---|---|---|---|---|
SpO\(_2\) | % | 91 | 93 | |
PETCO\(_2\) | mmHg | 30 | 60 | |
Protective target region | V\(_\text {T,pbw}\) | ml/kg | 4 | 6 |
\(\Delta P\) | mbar | 5 | 14 | |
Protective acceptable region | V\(_\text {T,pbw}\) | ml/kg | 6 | 8 |
\(\Delta P\) | mbar | 5 | 14 | |
Protective compromise region | V\(_\text {T,pbw}\) | ml/kg | 4 | 6 |
\(\Delta P\) | mbar | 14 | 20 |