Skip to main content

Table 2 PEEP and compliance values before and after the automated PEEP titration

From: SOLVe: a closed-loop system focused on protective mechanical ventilation

Subject

Trial

Before titration

After PEEP titration

Difference

PEEP

\(C_\text {rs}\)

PEEP

\(C_\text {rs}\)

\(\Delta\)PEEP

\(\Delta\) \(C_\text {rs}\)

[mbar]

[ml/mbar]

[mbar]

[ml/mbar]

[mbar]

[%]

A

1

15

15

16

15

+1

0.0

A

2

18

13

14

16

– 4

23.1

B

1

12

13

14

16

+2

23.1

B

2

14

17

14

24

0

41.2

C

1

15

17

14

18

– 1

5.9